Part 1—AutoCal Adapter Removal Procedure 1. First setup for the Adapter Removal procedure. 2. VNA Test Port 1 3. 36583K (f-f) Adapter 4. 36585K-2MF AutoCal Module—For the duration of the calibration, the K (f-f) Adapter and the AutoCal Module must be connected as an assembly. Do not disassemble or disconnect from its Power and Signal cables. 5. Resultant calibration reference planes. 6. K (m-f) Test Cable 7. VNA Test Port 2 | Part 2—AutoCal Adapter Removal Procedure 8. Second setup for the Adapter Removal procedure. 9. After the first calibration, rotate the complete assembly so that the AutoCal K (f) connector is connected to VNA Test Port 1. Do not disconnect the Adapter from the AutoCal module. 10. Do not disconnect the AutoCal Module from its Power or Signal cables. 11. Connect the 36583 K (f-f) adapter to the Test Cable K (m-f). The reference planes remain in place. The user is guided through the remaining steps of the procedure. |
Note | For the 36581x and 36582x AutoCal modules, which only come in M-F versions, only the first type of AutoCal adapter removal will be available. |
MODIFY 2-PORT AUTOCAL SETUP Dialog Box—Auto Sense ON—Adapter Removal—Internal Thru |