:SENSe{1-16}:CORRection:COLLect:LRL:BAND{1-5} Subsystem
The :SENSe{1-16}:CORRection:COLLect:LRL:BAND{1-5} subsystem commands provide control of Line-Reflect-Line device parameters, configuration, execution, and output reporting. Some commands require a 4-port VNA instrument.
LRL Calibration Subsystems
The LRL-related calibration commands are organized into five (5) subsystems in the following sections:
:SENSe{1-16}:CORRection:COLLect:LRL:BAND{1-5}:PORT12:FREQuency <NRf>
:SENSe{1-16}:CORRection:COLLect:LRL:BAND{1-5}:PORT12:FREQuency?
Description
The command can be used by 2-port and 4-port VNA instruments.
Sets the reference frequency for loss of the given band on the indicated port-pair of the LRL calibration on the indicated channel.
The available port pair is 12 where it is fixed as Port 1 and Port 2.
Returns the reference frequency for loss of the given band on the indicated port-pair of the LRL calibration on the indicated channel.
Cmd Parameters
<NRf> The input parameter is in Hertz.
Query Parameters
NA
Query Output
<NR3> The output is expressed in Hertz.
Range
7E4 to 7E10
Default Value
0.00000000000E+000
Syntax Example
:SENS1:CORR:COLL:LRL:BAND1:PORT12:FREQ 1.0E9
:SENS1:CORR:COLL:LRL:BAND1:PORT12:FREQ?
:SENSe{1-16}:CORRection:COLLect:LRL:BAND{1-5}:PORT12:LOSS <NRf>
:SENSe{1-16}:CORRection:COLLect:LRL:BAND{1-5}:PORT12:LOSS?
Description
Sets the line loss of the given band on the indicated port-pair of the LRL calibration on the indicated channel.
The available port pair is 12.
Returns the line loss of the given band on the indicated channel.
Cmd Parameters
<NRf> The input parameter is in dB/mm.
Query Parameters
NA
Query Output
<NR3> The output is expressed in dB/mm.
Range
NA
Default Value
0.00000000000E+000
Syntax Example
:SENS1:CORR:COLL:LRL:BAND1:PORT12:LOSS 3.0E0
:SENS1:CORR:COLL:LRL:BAND1:PORT12:LOSS?