:SENSe{1-16}:CORRection:COLLect:LRL[:CALa] Subsystem
The :SENSe{1-16}:CORRection:COLLect:LRL[:CALa] subsystem commands provide control of Line-Reflect-Line first calibration (or CALa) configuration parameters, execution, and output reporting.
LRL Calibration Subsystems
The LRL-related calibration commands are organized into five (5) subsystems in the following sections:
:SENSe{1-16}:CORRection:COLLect:LRL[:CALa]:BAND:COUNt <NRf>
:SENSe{1-16}:CORRection:COLLect:LRL[:CALa]:BAND:COUNt?
Description
Sets the number of bands to use in the LRL CALa calibration on the indicated channel.
Returns the number of bands to use in the LRL CALa calibration on the indicated channel.
Cmd Parameters
<NRf> The input parameter is an integer.
Query Parameters
NA
Query Output
<NR1> The output parameter is an integer.
Range
1 to 5
Default
1
Syntax Example
:SENS1:CORR:COLL:LRL:BAND:COUN 2
:SENS1:CORR:COLL:LRL:BAND:COUN?
:SENSe{1-16}:CORRection:COLLect:LRL[:CALa]:BAND{1-5}:FREQuency <NRf>
:SENSe{1-16}:CORRection:COLLect:LRL[:CALa]:BAND{1-5}:FREQuency?
Description
Sets the reference frequency for loss of the given band of the LRL CALa calibration on the indicated channel.
Returns the reference frequency for loss of the given band of the LRL CALa calibration on the indicated channel.
Cmd Parameters
<NRf> The input parameter is in Hertz.
Query Parameters
NA
Query Output
<NR3> The input parameter is in Hertz
Range
7E4 to 7E10
Default
0.00000000000E+000
Syntax Example
:SENS1:CORR:COLL:LRL:BAND1:FREQ 1.0E9
:SENS1:CORR:COLL:LRL:BAND1:FREQ?
:SENSe{1-16}:CORRection:COLLect:LRL[:CALa]:BAND{1-5}:LOSS <NRf>
:SENSe{1-16}:CORRection:COLLect:LRL[:CALa]:BAND{1-5}:LOSS?
Description
Sets the line loss of the given band of the LRL CALa calibration on the indicated channel.
Returns the line loss of the given band of the LRL CALa calibration on the indicated channel.
Cmd Parameters
<NRf> The input parameter is in dB/mm.
Query Parameters
NA
Query Output
<NR3>The input parameter is in dB/mm.
Range
NA
Default
0.00000000000E+000
Syntax Example
:SENS1:CORR:COLL:LRL:BAND1:LOSS 3.0E0
:SENS1:CORR:COLL:LRL:BAND1:LOSS?
:SENSe{1-16}:CORRection:COLLect:LRL[:CALa]:BAND{1-5}:REFLection:TYPe <char>
:SENSe{1-16}:CORRection:COLLect:LRL[:CALa]:BAND{1-5}:REFLection:TYPe?
Description
Sets the Band reflection type of the LRL CALa calibration on the indicated channel. where:
• OPENlike = The reflection type is more like an open.
• SHORTlike = The reflection type is more like a short.
• BOTH = The reflection type is like both an open and a short.
Returns the reflection type of the LRL CALa calibration on the indicated channel for a given band.
Cmd Parameters
<char> OPENlike | SHORTlike | BOTH
Query Output
<char> OPEN | SHORT | BOTH
Range
NA
Default
OPEN
Syntax Example
:SENS1:CORR:COLL:LRL:BAND1:REFL:TYP OPEN
:SENS1:CORR:COLL:LRL:BAND1:REFL:TYP?
:SENSe{1-16}:CORRection:COLLect:LRL[:CALa]:CKIT:LOAD <string>
Description
Loads an LRL cal kit file into CALa LRL/TRL calibration on the indicated channel.
No query.
Cmd Parameters
<string> Filename and path to the file in the form: 'x:\directory\filename.lcf' , where x:\directory\filename.lcf exists and filename.lcf is a valid LRL/TRL cal kit file.
Query Parameters
NA
Range
NA
Default
NA
Syntax Example
:SENS1:CORR:COLL:LRL:CKIT:LOAD ' c:\filename.lcf'
:SENSe{1-16}:CORRection:COLLect:LRL[:CALa]:CKIT:NAMe <string>
:SENSe{1-16}:CORRection:COLLect:LRL[:CALa]:CKIT:NAMe?
Description
Sets the name that will be stored in the LRL/TRL cal kit file.
Returns the name that was last set for the currently loaded file.
Cmd Parameters
<string> Filename and path to the file in the form: 'x:\directory\filename.lcf' , where x:\directory\filename.lcf exists and filename.lcf is a valid LRL/TRL cal kit file.
Query Parameters
NA
Query Output
<string> The name that is currently associated with the current cal kit file.
Range
NA
Default
NA
Syntax Example
:SENS1:CORR:COLL:LRL:CKIT:NAM ' C:\AnritsuVNA\mylrlcalkit.lcf'
:SENS1:CORR:COLL:LRL:CKIT:NAM?
:SENSe{1-16}:CORRection:COLLect:LRL[:CALa]:CKIT:SAVe <string>
Description
Saves an LRL cal kit file into CALa LRL calibration on the indicated channel.
No query.
Cmd Parameters
<string> Filename and path to the file in the form: 'x:\directory\filename.lcf' .
Query Parameters
NA
Range
NA
Default
NA
Syntax Example
:SENS1:CORR:COLL:LRL:CKIT:SAV ' c:\filename.lcf'
:SENSe{1-16}:CORRection:COLLect:LRL[:CALa]:DEVice{1-10}:LINE
Description
Initiates collection of the indicated Device Line Standard data for the CALa calibration on the indicated channel.
No query.
Cmd Parameters
NA
Range
NA
Default
NA
Syntax Example
:SENS1:CORR:COLL:LRL:DEV1:LINE
:SENSe{1-16}:CORRection:COLLect:LRL[:CALa]:DEVice{1-10}:LINE:DELay <NRf>
:SENSe{1-16}:CORRection:COLLect:LRL[:CALa]:DEVice{1-10}:LINE:DELay?
Description
Sets the delay of the given device for CALa of the LRL calibration of the indicated channel.
Returns the delay of the given device for CALa of the LRL calibration of the indicated channel.
Cmd Parameters
<NRf> The input parameter is in seconds.
Query Parameters
NA
Query Output
<NR3> The output parameter is in seconds.
Range
MPND
Default
Default value is dependent on the default dielectric of the substrate and the effective length default value.
Syntax Example
:SENS1:CORR:COLL:LRL:DEV2:LINE:DEL 20E-3
:SENS1:CORR:COLL:LRL:DEV2:LINE:DEL?
:SENSe{1-16}:CORRection:COLLect:LRL[:CALa]:DEVice{1-10}:LINE:PLENgth <NRf>
:SENSe{1-16}:CORRection:COLLect:LRL[:CALa]:DEVice{1-10}:LINE:PLENgth?
Description
Sets the physical length of the given device for CALa of the LRL calibration of the indicated channel.
Returns the physical length of the given device for CALa of the LRL calibration of the indicated channel.
Cmd Parameters
<NRf> The input parameter is in meters.
Query Parameters
NA
Query Output
<NR3> The output parameter is in meters.
Range
MPND
Default
Default value is dependent on the effective length default value.
Syntax Example
:SENS1:CORR:COLL:LRL:DEV2:LINE:PLEN 20E-3
:SENS1:CORR:COLL:LRL:DEV2:LINE:PLEN?
:SENSe{1-16}:CORRection:COLLect:LRL[:CALa]:DEVice{1-10}:LINE:FREQuency <NRf>
:SENSe{1-16}:CORRection:COLLect:LRL[:CALa]:DEVice{1-10}:LINE:FREQuency?
Description
The command can be used by 2-port and 4-port VNA instruments.
This command has effect only when an odd-numbered device is entered.
Sets the reference frequency for loss of the given device on the indicated port-pair of the LRL calibration on the indicated channel.
The available port pair is 12, where it is fixed as Port 1 and Port 2.
Returns the reference frequency for loss of the given device on the indicated port-pair of the LRL calibration on the indicated channel.
Recommend using the following command that sets the Reference Frequency per Band:
Cmd Parameters
<NRf> The input parameter is in Hertz.
Query Parameters
NA
Query Output
<NR3> The output parameter is in Hertz.
Range
7E4 to 7E10
Default
0.00000000000E+000
Syntax Example
:SENS1:CORR:COLL:LRL:DEV1:LINE:FREQ 1.0E7
:SENS1:CORR:COLL:LRL:DEV1:LINE:FREQ?
:SENSe{1-16}:CORRection:COLLect:LRL[:CALa]:DEVice{1-10}:LINE:LENGth <NRf>
:SENSe{1-16}:CORRection:COLLect:LRL[:CALa]:DEVice{1-10}:LINE:LENGth?
Description
Sets the line length of the given device of the LRL CALa calibration on the indicated channel.
Returns the line length of the given device of the LRL CALa calibration on the indicated channel.
Cmd Parameters
<NRf> The input parameter is in Meters.
Query Parameters
<NR3> The output parameter is in Meters.
Range
MPND
Default
0.00000000000E+000
Syntax Example
:SENS1:CORR:COLL:LRL:DEV1:LINE:LENG 1.0E0
:SENS1:CORR:COLL:LRL:DEV1:LINE:LENG?
:SENSe{1-16}:CORRection:COLLect:LRL[:CALa]:DEVice{1-10}:LINE:LOSS <NRf>
:SENSe{1-16}:CORRection:COLLect:LRL[:CALa]:DEVice{1-10}:LINE:LOSS?
Description
This command has effect only when an odd-numbered device is entered.
Sets the line loss of the given device on the indicated port-pair of the LRL calibration on the indicated channel.
The available port pair is 12.
Returns the line loss of the given device on the indicated port.
Recommend using the following command that sets the Loss per Band.
Cmd Parameters
<NRf> The input parameter is in dB/mm.
Query Parameters
NA
Query Output
<NR3> The output parameter is in dB/mm.
Range
NA
Default
0.00000000000E+000
Syntax Example
:SENS1:CORR:COLL:LRL:DEV1:LINE:LOSS 3.0E0
:SENS1:CORR:COLL:LRL:DEV1:LINE:LOSS?
:SENSe{1-16}:CORRection:COLLect:LRL[:CALa]:DEVice{1-10}:PORT{1-4}:MATCH:C0 <NRf>
:SENSe{1-16}:CORRection:COLLect:LRL[:CALa]:DEVice{1-10}:PORT{1-4}:MATCH:C0?
Description
The use of Port 3 and/or Port 4 requires a 4-port VNA instrument.
This command has effect only when an even-numbered device is entered.
Sets capacitance of the match device on the indicated port of the LRL CALa calibration on the indicated channel.
The C0 (C zero) coefficient is in Farads.
Returns capacitance of the match device on the indicated port of the LRL CALa calibration on the indicated channel.
Cmd Parameters
<NRf> The input parameter is in Farads.
Query Parameters
<NR3> The output parameter is in Farads.
Range
MPND
Default
0.00000000000E+000
Syntax Example
:SENS1:CORR:COLL:LRL:DEV2:PORT1:MATCH:C0 3.01E-12
:SENS1:CORR:COLL:LRL:DEV2:PORT1:MATCH:C0?
:SENSe{1-16}:CORRection:COLLect:LRL[:CALa]:DEVice{1-10}:PORT{1-4}:MATCH:C1 <NRf>
:SENSe{1-16}:CORRection:COLLect:LRL[:CALa]:DEVice{1-10}:PORT{1-4}:MATCH:C1?
Description
The use of Port 3 and/or Port 4 requires a 4-port VNA instrument.
This command has effect only when an even-numbered device is entered.
Sets the C1 (C one) coefficient of the match device on the indicated port of the LRL CALa calibration on the indicated channel.
The C1 coefficient is in Farads/Hertz.
Returns the C1 coefficient of the match device on the indicated port of the LRL CALa calibration on the indicated channel.
Cmd Parameters
<NRf> The input parameter is in Farads/Hertz.
Query Parameters
<NR3> The output parameter is in Farads/Hertz.
Range
MPND
Default Value
Syntax Example
:SENS1:CORR:COLL:LRL:DEV2:PORT1:MATCH:C1 2.0E0
:SENS1:CORR:COLL:LRL:DEV2:PORT1:MATCH:C1?
:SENSe{1-16}:CORRection:COLLect:LRL[:CALa]:DEVice{1-10}:PORT{1-4}:MATCH:C2 <NRf>
:SENSe{1-16}:CORRection:COLLect:LRL[:CALa]:DEVice{1-10}:PORT{1-4}:MATCH:C2?
Description
The use of Port 3 and/or Port 4 requires a 4-port VNA instrument.
This command has effect only when an even-numbered device is entered.
Sets the C2 coefficient of the match device on the indicated port of the LRL CALa calibration on the indicated channel.
The C2 coefficient is in Farads/Hertz^2.
Returns the C2 coefficient of the match device on the indicated port of the LRL CALa calibration on the indicated channel.
Cmd Parameters
<NRf> The input parameter is in Farads/Hertz^2.
Query Parameters
<NR3> The output parameter is in Farads/Hertz^2.
Range
MPND
Default Value
Syntax Example
:SENS1:CORR:COLL:LRL:DEV2:PORT1:MATCH:C2 2.0E0
:SENS1:CORR:COLL:LRL:DEV2:PORT1:MATCH:C2?
:SENSe{1-16}:CORRection:COLLect:LRL[:CALa]:DEVice{1-10}:PORT{1-4}:MATCH:C3 <NRf>
:SENSe{1-16}:CORRection:COLLect:LRL[:CALa]:DEVice{1-10}:PORT{1-4}:MATCH:C3?
Description
The use of Port 3 and/or Port 4 requires a 4-port VNA instrument.
This command has effect only when an even-numbered device is entered.
Sets the C3 coefficient of the match device on the indicated port of the LRL CALa calibration on the indicated channel.
The C3 coefficient is in Farads/Hertz^3.
Returns the C3 coefficient of the match device on the indicated port of the LRL CALa calibration on the indicated channel.
Cmd Parameters
<NRf> The input parameter is in Farads/Hertz^3.
Query Parameters
<NR3> The output parameter is in Farads/Hertz^3.
Range
MPND
Default Value
Syntax Example
:SENS1:CORR:COLL:LRL:DEV2:PORT1:MATCH:C3 2.0E0
:SENS1:CORR:COLL:LRL:DEV2:PORT1:MATCH:C3?
:SENSe{1-16}:CORRection:COLLect:LRL[:CALa]:DEVice{1-10}:PORT{1-4}:MATCH:L0 <NRf>
:SENSe{1-16}:CORRection:COLLect:LRL[:CALa]:DEVice{1-10}:PORT{1-4}:MATCH:L0?
Description
The use of Port 3 and/or Port 4 requires a 4-port VNA instrument.
This command has effect only when an even-numbered device is entered.
Sets inductance of the match device on the indicated port of the LRL CALa calibration on the indicated channel.
The L0 (zero) coefficient is in Henrys.
Returns inductance of the match device on the indicated port of the LRL CALa calibration on the indicated channel.
Cmd Parameters
<NRf> The input parameter is in Henrys.
Query Parameters
<NR3> The output parameter is in Henrys.
Range
MPND
Default
0.00000000000E+000
Syntax Example
:SENS1:CORR:COLL:LRL:DEV2:PORT1:MATCH:L0 2.0E-6
:SENS1:CORR:COLL:LRL:DEV2:PORT1:MATCH:L0?
:SENSe{1-16}:CORRection:COLLect:LRL[:CALa]:DEVice{1-10}:PORT{1-4}:MATCH:L1 <NRf>
:SENSe{1-16}:CORRection:COLLect:LRL[:CALa]:DEVice{1-10}:PORT{1-4}:MATCH:L1?
Description
The use of Port 3 and/or Port 4 requires a 4-port VNA instrument.
This command has effect only when an even-numbered device is entered.
Sets the L1 (L one) coefficient of the match device on the indicated port of the LRL CALa calibration on the indicated channel.
The L1 coefficient is in Henrys/Hertz.
Returns the L1 coefficient of the match device on the indicated port of the LRL CALa calibration on the indicated channel.
Cmd Parameters
<NRf> The input parameter is in Henrys/Hertz.
Query Parameters
<NR3> The output parameter is in Henrys/Hertz.
Range
MPND
Default
0.00000000000E+000
Syntax Example
:SENS1:CORR:COLL:LRL:DEV2:PORT1:MATCH:L1 1.4
:SENS1:CORR:COLL:LRL:DEV2:PORT1:MATCH:L1?
:SENSe{1-16}:CORRection:COLLect:LRL[:CALa]:DEVice{1-10}:PORT{1-4}:MATCH:L2 <NRf>
:SENSe{1-16}:CORRection:COLLect:LRL[:CALa]:DEVice{1-10}:PORT{1-4}:MATCH:L2?
Description
The use of Port 3 and/or Port 4 requires a 4-port VNA instrument.
This command has effect only when an even-numbered device is entered.
Sets the L2 coefficient of the match device on the indicated port of the LRL CALa calibration on the indicated channel.
The L2 coefficient is in Henrys/Hertz^2.
Returns the L2 coefficient of the match device on the indicated port of the LRL CALa calibration on the indicated channel.
Cmd Parameters
<NRf> The input parameter is in Henrys/Hertz^2.
Query Parameters
<NR3> The output parameter is in Henrys/Hertz^2.
Range
MPND
Default
0.00000000000E+000
Syntax Example
:SENS1:CORR:COLL:LRL:DEV2:PORT1:MATCH:L2 2.0E0
:SENS1:CORR:COLL:LRL:DEV2:PORT1:MATCH:L2?
:SENSe{1-16}:CORRection:COLLect:LRL[:CALa]:DEVice{1-10}:PORT{1-4}:MATCH:L3 <NRf>
:SENSe{1-16}:CORRection:COLLect:LRL[:CALa]:DEVice{1-10}:PORT{1-4}:MATCH:L3?
Description
The use of Port 3 and/or Port 4 requires a 4-port VNA instrument.
This command has effect only when an even-numbered device is entered.
Sets the L3 coefficient of the match device on the indicated port of the LRL CALa calibration on the indicated channel.
The L3 coefficient is in Henrys/Hertz^3.
Returns the L3 coefficient of the match device on the indicated port of the LRL CALa calibration on the indicated channel.
Cmd Parameters
<NRf> The input parameter is in Henrys/Hertz^3.
Query Parameters
<NR3> The output parameter is in Henrys/Hertz^3.
Range
MPND
Default
0.00000000000E+000
Syntax Example
:SENS1:CORR:COLL:LRL:DEV2:PORT1:MATCH:L3 2.0E0
:SENS1:CORR:COLL:LRL:DEV2:PORT1:MATCH:L3?
:SENSe{1-16}:CORRection:COLLect:LRL[:CALa]:DEVice{1-10}:PORT{1-4}:MATCH:OFF1 <NRf>
:SENSe{1-16}:CORRection:COLLect:LRL[:CALa]:DEVice{1-10}:PORT{1-4}:MATCH:OFF1?
Description
The use of Port 3 and/or Port 4 requires a 4-port VNA instrument.
This command has effect only when an even-numbered device is entered.
Sets the offset Length1 coefficient of the match device on the indicated port of the LRL CALa calibration on the indicated channel.
The OFF1 coefficient is measured in Meters/Hertz.
Returns the offset Length1 coefficient of the match device on the indicated port of the LRL CALa calibration on the indicated channel.
Cmd Parameters
<NRf> The input parameter is in Meters/Hertz.
Query Parameters
<NR3> The output parameter is in Meters/Hertz.
Range
MPND
Default
0.00000000000E+000
Syntax Example
:SENS1:CORR:COLL:LRL:DEV2:PORT1:MATCH:OFF1 2.0E0
:SENS1:CORR:COLL:LRL:DEV2:PORT1:MATCH:OFF1?
:SENSe{1-16}:CORRection:COLLect:LRL[:CALa]:DEVice{1-10}:PORT{1-4}:MATCH:OFF2 <NRf>
:SENSe{1-16}:CORRection:COLLect:LRL[:CALa]:DEVice{1-10}:PORT{1-4}:MATCH:OFF2?
Description
The use of Port 3 and/or Port 4 requires a 4-port VNA instrument.
This command has effect only when an even-numbered device is entered.
Sets the offset Length2 coefficient of the match device on the indicated port of the LRL CALa calibration on the indicated channel.
The OFF2 coefficient is measured in Meters/Hertz^2.
Returns the offset Length2 coefficient of the match device on the indicated port of the LRL CALa calibration on the indicated channel.
Cmd Parameters
<NRf> The input parameter is in Meters/Hertz^2.
Query Parameters
<NR3> The output parameter is in Meters/Hertz^2.
Range
MPND
Default
0.00000000000E+000
Syntax Example
:SENS1:CORR:COLL:LRL:DEV2:PORT1:MATCH:OFF2 2.0E0
:SENS1:CORR:COLL:LRL:DEV2:PORT1:MATCH:OFF2?
:SENSe{1-16}:CORRection:COLLect:LRL[:CALa]:DEVice{1-10}:PORT{1-4}:MATCH:OFF3 <NRf>
:SENSe{1-16}:CORRection:COLLect:LRL[:CALa]:DEVice{1-10}:PORT{1-4}:MATCH:OFF3?
Description
The use of Port 3 and/or Port 4 requires a 4-port VNA instrument.
This command has effect only when an even-numbered device is entered.
Sets the offset Length3 coefficient of the match device on the indicated port of the LRL CALa calibration on the indicated channel.
The OFF3 coefficient is measured in Meters/Hertz^3.
Returns the offset Length3 coefficient of the match device on the indicated port of the LRL CALa calibration on the indicated channel.
Cmd Parameters
<NRf> The input parameter is in Meters/Hertz^3.
Query Parameters
<NR3> The output parameter is in Meters/Hertz^3.
Range
MPND
Default
0.00000000000E+000
Syntax Example
:SENS1:CORR:COLL:LRL:DEV2:PORT1:MATCH:OFF3 2.0E0
:SENS1:CORR:COLL:LRL:DEV2:PORT1:MATCH:OFF3?
:SENSe{1-16}:CORRection:COLLect:LRL[:CALa]:DEVice{1-10}:PORT{1-4}:MATCH:OFFS <NRf>
:SENSe{1-16}:CORRection:COLLect:LRL[:CALa]:DEVice{1-10}:PORT{1-4}:MATCH:OFFS?
Description
The use of Port 3 and/or Port 4 requires a 4-port VNA instrument.
This command has effect only when an even-numbered device is entered.
Sets offset length of the match device on the indicated port of the LRL CALa calibration on the indicated channel.
Returns offset length of the match device on the indicated port of the LRL CALa calibration on the indicated channel.
Cmd Parameters
<NRf> The input parameter is in Meters.
Query Parameters
<NR3> The output parameter is in Meters.
Range
MPND
Default
0.00000000000E+000
Syntax Example
:SENS1:CORR:COLL:LRL:DEV2:PORT1:MATCH:OFFS 1.0E0
:SENS1:CORR:COLL:LRL:DEV2:PORT1:MATCH:OFFS?
:SENSe{1-16}:CORRection:COLLect:LRL[:CALa]:DEVice{1-10}:PORT{1-4}:MATCH:R <NRf>
:SENSe{1-16}:CORRection:COLLect:LRL[:CALa]:DEVice{1-10}:PORT{1-4}:MATCH:R?
Description
The use of Port 3 and/or Port 4 requires a 4-port VNA instrument.
This command has effect only when an even-numbered device is entered.
Sets resistance of the match device on the indicated port of the LRL CALa calibration on the indicated channel.
Returns resistance of the match device on the indicated port of the LRL CALa calibration on the indicated channel.
Cmd Parameters
<NRf> The input parameter is in Ohms.
Query Parameters
<NR3> The output parameter is in Ohms.
Range
MPND
Default
5.00000000000E+001
Syntax Example
:SENS1:CORR:COLL:LRL:DEV2:PORT1:MATCH:R 7.5E1
:SENS1:CORR:COLL:LRL:DEV2:PORT1:MATCH:R?
:SENSe{1-16}:CORRection:COLLect:LRL[:CALa]:DEVice{1-10}:PORT{1-4}:MATCH:S1P:FILE <string>
:SENSe{1-16}:CORRection:COLLect:LRL[:CALa]:DEVice{1-10}:PORT{1-4}:MATCH:S1P:FILE?
Description
This command has effect only when an even-numbered device is entered.
The command sets the MATCH S1P file on the selected port of the selected device for LRL CALa calibration for the indicated channel.
Returns the MATCH S1P file on the selected port of the selected device for LRL CALa calibration for the indicated channel.
Cmd Parameters
<string> s1p file path
Query Parameters
NA
Query Output
<char1> s1p file path
Range
NA
Default
NA
Syntax Example
:SENS1:CORR:COLL:LRL:DEV2:PORT1:MATCH:S1P:FILE ' x:\directory\filename.s1p'
:SENS1:CORR:COLL:LRL:DEV2:PORT1:MATCH:S1P:FILE?
:SENSe{1-16}:CORRection:COLLect:LRL[:CALa]:DEVice{1-10}:PORT{1-4}:MATCH:S1P[:STATe] <char>
:SENSe{1-16}:CORRection:COLLect:LRL[:CALa]:DEVice{1-10}:PORT{1-4}:MATCH:S1P[:STATe]?
Description
The use of Port 3 and/or Port 4 requires a 4-port VNA instrument.
This command has effect only when an even-numbered device is entered.
Sets the MATCH S1P file settings ON/OFF on the selected port of the selected device for LRL CALa calibration for the indicated channel.
ON means that the S1P file data will be used, OFF means the Circuit Model data will be used.
Returns the MATCH S1P file settings on the selected port of the selected device for LRL CALa calibration for the indicated channel.
Cmd Parameters
<char> 1 | 0 | ON | OFF
Query Parameters
NA
Query Output
<char1>1 | 0
Range
NA
Default
NA
Syntax Example
:SENS1:CORR:COLL:LRL:DEV2:PORT1:MATCH:S1P 1
:SENS1:CORR:COLL:LRL:DEV2:PORT1:MATCH:S1P?
:SENSe{1-16}:CORRection:COLLect:LRL[:CALa]:DEVice{1-10}:PORT{1-4}:MATCH:Z0 <NRf>
:SENSe{1-16}:CORRection:COLLect:LRL[:CALa]:DEVice{1-10}:PORT{1-4}:MATCH:Z0?
Description
The use of Port 3 and/or Port 4 requires a 4-port VNA instrument.
This command has effect only when an even-numbered device is entered.
Sets impedance of the match device on the indicated port of the LRL CALa calibration on the indicated channel.
Returns impedance of the match device on the indicated port of the LRL CALa calibration on the indicated channel.
Cmd Parameters
<NRf> The input parameter is in Ohms.
Query Parameters
<NR3> The output parameter is in Ohms.
Range
MPND
Default
5.00000000000E+001
Syntax Example
:SENS1:CORR:COLL:LRL:DEV2:PORT1:MATCH:Z0 7.5E1
:SENS1:CORR:COLL:LRL:DEV2:PORT1:MATCH:Z0?
:SENSe{1-16}:CORRection:COLLect:LRL[:CALa]:DEVice{1-10}:TYPe <char>
:SENSe{1-16}:CORRection:COLLect:LRL[:CALa]:DEVice{1-10}:TYPe?
Description
Sets the type of the given device in the LRL CALa calibration on the indicated channel.
Returns the type of the given device in the LRL CALa calibration on the indicated channel.
Cmd Parameters
<char> LINE | MATCH | DEVICE1 | DEVICE2
Query Parameters
<char> LINE | MATCH | DEVICE1 | DEVICE2
Range
NA
Default
LINE
Syntax Example
:SENS1:CORR:COLL:LRL:DEV1:TYP LINE
:SENS1:CORR:COLL:LRL:DEV1:TYP?
:SENSe{1-16}:CORRection:COLLect:LRL[:CALa]:FREQuency:BREakpoint <NRf>
:SENSe{1-16}:CORRection:COLLect:LRL[:CALa]:FREQuency:BREakpoint?
Description
Sets the breakpoint frequency of the LRL CALa calibration on the indicated channel.
Returns the breakpoint frequency of the LRL CALa calibration on the indicated channel.
Note: This is a legacy command. Use the command:
Cmd Parameters
<NRf> The input parameter is in Hertz.
Query Parameters
<NR3> The output parameter is in Hertz.
Range
0 to MPND
Default
3.00000000000E+009
Syntax Example
:SENS1:CORR:COLL:LRL:FREQ:BRE 1.0E7
:SENS1:CORR:COLL:LRL:FREQ:BRE?
:SENSe{1-16}:CORRection:COLLect:LRL[:CALa]:FREQuency:BREakpoint{1-4} <NRf>
:SENSe{1-16}:CORRection:COLLect:LRL[:CALa]:FREQuency:BREakpoint{1-4}?
Description
Sets the breakpoint frequency between bands of LRL calibration on a indicated channel.
Returns the breakpoint frequency between bands of LRL calibration on a indicated channel. For a 4-port system, the command is meant for CAL A of LRL.
In the command, the number next to BREakpoint has the following correspondence:
• BREakpoint1: Breakpoint frequency between bands 2 and 1
• BREakpoint2: Breakpoint frequency between bands 3 and 2
• BREakpoint3: Breakpoint frequency between bands 4 and 3
• BREakpoint4: Breakpoint frequency between bands 5 and 4
Cmd Parameters
<NRf> The input parameter is in Hertz.
Query Parameters
<NR3> The output parameter is in Hertz.
Range
MPND
Default
3E09
Syntax Example
:SENS1:CORR:COLL:LRL:FREQ:BRE1 4E09
:SENS1:CORR:COLL:LRL:FREQ:BRE1?
:SENSe{1-16}:CORRection:COLLect:LRL[:CALa]:PASSivity:ENForce[:STATe] <char>
:SENSe{1-16}:CORRection:COLLect:LRL[:CALa]:PASSivity:ENForce[:STATe]?
Description
Sets the state of whether to use the enforced passivity in the LRL CALa calibration on the indicated channel.
Returns whether enforced passivity will be used in the LRL CALa calibration on the indicated channel.
Cmd Parameters
<char> 1 | 0 | ON | OFF
Query Parameters
NA
Query Output
<char> 1 | 0
Range
NA
Default Value
0
Syntax Example
:SENS1:CORR:COLL:LRL:PASS:ENF 1
:SENS1:CORR:COLL:LRL:PASS:ENF?
:SENSe{1-16}:CORRection:COLLect:LRL[:CALa]:OPEN:OFFSet <NRf>
:SENSe{1-16}:CORRection:COLLect:LRL[:CALa]:OPEN:OFFSet?
Description
Sets the electrical offset length of the open like reflection of the LRL CALa calibration on the indicated channel.
Returns the electrical offset length of the open like reflection of the LRL CALa calibration on the indicated channel.
Cmd Parameters
<NRf> The input parameter is in Meters.
Query Parameters
<NR3> The output parameter is in Meters.
Range
MPND
Default
0.00000000000E+000
Syntax Example
:SENS1:CORR:COLL:LRL:OPEN:OFFS 1.0E0
:SENS1:CORR:COLL:LRL:OPEN:OFFS?
:SENSe{1-16}:CORRection:COLLect:LRL[:CALa]:REFPlane <char>
:SENSe{1-16}:CORRection:COLLect:LRL[:CALa]:REFPlane?
Description
Sets the location of the reference plane in the LRL CALa calibration on the indicated channel where:
• MIDdle = the middle of the transmission line
• END = the end of the transmission line
Returns the location of the reference plane in the LRL CALa calibration on the indicated channel.
Cmd Parameters
<char> MIDdle | END
Query Parameters
<char> MID | END
Range
NA
Default
END
Syntax Example
:SENS1:CORR:COLL:LRL:REFP MID
:SENS1:CORR:COLL:LRL:REFP?
:SENSe{1-16}:CORRection:COLLect:LRL[:CALa]:SHORT:OFFSet <NRf>
:SENSe{1-16}:CORRection:COLLect:LRL[:CALa]:SHORT:OFFSet?
Description
Sets the offset length of the short like reflection of the LRL CALa calibration on the indicated channel.
Returns the offset length of the short like reflection of the LRL CALa calibration on the indicated channel.
Cmd Parameters
<NRf> The input parameter is in Meters.
Query Parameters
<NR3> The output parameter is in Meters.
Range
MPND
Default
0.00000000000E+000
Syntax Example
:SENS1:CORR:COLL:LRL:SHORT:OFFS 1.0E0
:SENS1:CORR:COLL:LRL:SHORT:OFFS?
1-800-ANRITSU
10450-00040 Rev. AD