VectorStar™ MS464xB Series Microwave Vector Network Analyzer Programming Manual : SCPI Commands : :SENSe{1-16}:SEGMent Subsystem
 
:SENSe{1-16}:SEGMent Subsystem
The :SENSe{1-16}:SEGMent subsystem command is used to query the segmented sweep type as frequency-based or index based on the active sweep.
Limit Line and Segment Subsystems
Related limit line and segment configuration and control subsystems are:
:CALCulate{1-16}[:SELected]:LIMit Subsystem
:DISPlay Subsystem
:SENSe{1-16}:FSEGMent Subsystem.
:SENSe{1-16}:FSEGMent{1-50} Subsystem.
:SENSe{1-16}:ISEGMent Subsystem.
:SENSe{1-16}:ISEGMent{1-50} Subsystem.
:SENSe{1-16}:ISEGMent{1-50} Subsystem
:SENSe{1-16}:SEGMent Subsystem
:SENSe{1-16}:SEGment:AVERage[:SWEepbysweep][:STATe] <char>
:SENSe{1-16}:SEGment:AVERage[:SWEepbysweep][:STATe]?
Description
Sets the on/off state of sweep-by-sweep averaging for segmented sweep (frequency based and indexed based) for the given channel.
Returns the averaging function on/off status for sweep-by-sweep averaging for segmented sweep (frequency-based and indexed-based) for the given channel.
Cmd Parameters
<char> ON | OFF | 1 | 0
Query Parameters
NA
Query Output
1 | 0
Range
NA
Default Value
0
Syntax Example
:SENS1:SEG:AVER ON
:SENS1:SEG:AVER?
 
:SENSe{1-16}:SEGMent:TYPe?
Description
Query only.
Returns the segmented sweep type for the active segment On the indicated channel as frequency-based or index-based, where:
FREQ = Frequency-based sweep type
INDEX = Index-based sweep type
Query Parameters
<char> FREQ | INDEX
Range
NA
Default Value
NA
Syntax Example
:SENS1:SEGM:TYP?