Returns normal RGB color of the indicated memory trace on the active channel.
Cmd Parameters
<NRf> The input parameters are integers between 0 and 255 representing the Red, Green, and Blue color values.
Query Parameters
<NR1>, <NR1>, <NR1> The output parameters are integers between 0 and 255 representing the Red, Green, and Blue color values.
Range
0 to 255
Default Value
175, 143, 47
Syntax Example
:DISP:COL:NORM:TRAC1:MEM 100, 0, 100
:DISP:COL:NORM:TRAC1:MEM?
:DISPlay:COLor:RESet
Description
Sets all colors and inverted colors to their normal default values.
No query.
Cmd Parameters
NA
Range
NA
Default Value
NA
Syntax Example
:DISP:COL:RES
:DISPlay:COUNt <NRf>
:DISPlay:COUNt?
Description
Sets the number of displayed channels.
When the VNA is in 25,000 point mode, the number of channels can only be 1 (one), 2, 3, 4, 6, 8, 9, 10, 12, or 16 channels. If the channel display is set to a non-listed number (5, 7, 11, 13, 14, 15), the instrument is set to the next higher channel number. If a number of greater than 16 is entered, the instrument is set to 16 channels. If the instrument is set to 100,000 points, any input results in 1 (one) channel.
Returns the number of displayed channels.
The example below assumes that the instrument is set to 25,000 points.
Cmd Parameters
<NRf> The input parameter is a unitless number.
Query Parameters
<NR1> The output parameter is an integer.
Range
For 25,000 point mode, 1 to 16 channels.
For 100,000 point mode, 1 channel.
Default Value
1
Syntax Example
:DISP:COUN 4
:DISP:COUN?
:DISPlay:FONT:SIZe:CTITle <NRf>
:DISPlay:FONT:SIZe:CTITle?
Description
Sets the font size of the channel title.
Returns the font size of the channel title.
Cmd Parameters
<NRf> Font size value in points
Query Output
<NR1> Outputs the set font size value in points
Range
8 to 48
Default Value
9
Syntax Example
:DISP:FONT:SIZ:CTIT 12
:DISP:FONT:SIZ:CTIT?
:DISPlay:FONT:SIZe:LTResult <NRf>
:DISPlay:FONT:SIZe:LTResult?
Description
Sets the font size of the limit test result.
Returns the font size of the limit test result.
Cmd Parameters
<NRf> Font size value in points
Query Output
<NR1> Outputs the set font size value in points
Range
8 to 14
Default Value
9
Syntax Example
:DISP:FONT:SIZ:LTR 14
:DISP:FONT:SIZ:LTR?
:DISPlay:FONT:SIZe:MARKer:READout <NRf>
:DISPlay:FONT:SIZe:MARKer:READout?
Description
Sets the font size of the marker readout values
Returns the font size of the marker readout values.
Cmd Parameters
<NRf> Font size value in points
Query Output
<NR1> Outputs the set font size value in points
Range
8 to 48
Default Value
9
Syntax Example
:DISP:FONT:SIZ:MARK:READ 28
:DISP:FONT:SIZ:MARK:READ?
:DISPlay:FONT:SIZe:MARKer:TABLe <NRf>
:DISPlay:FONT:SIZe:MARKer:TABLe?
Description
Sets the font size of the marker table
Returns the font size of the marker table.
Cmd Parameters
<NRf> Font size value in points
Query Output
<NR1> Outputs the set font size value in points
Range
8 to 48
Default Value
9
Syntax Example
:DISP:FONT:SIZ:MARK:TABL 35
:DISP:FONT:SIZ:MARK:TABL?
:DISPlay:FONT:SIZe:TSCale <NRf>
:DISPlay:FONT:SIZe:TSCale?
Description
Sets the font size of the trace scale values.
Returns the font size of the trace scale values.
Cmd Parameters
<NRf> Font size value in points
Query Output
<NR1> Outputs the set font size value in points
Range
8 to 48
Default Value
9
Syntax Example
:DISP:FONT:SIZ:TSC 18
:DISP:FONT:SIZ:TSC?
:DISPlay:FONT:SIZe:TTITle <NRf>
:DISPlay:FONT:SIZe:TTITle?
Description
Sets the font size of the trace title.
Returns the font size of the trace title.
Cmd Parameters
<NRf> Font size value in points
Query Output
<NR1> Outputs the set font size value in points
Range
8 to 14
Default Value
10
Syntax Example
:DISP:FONT:SIZ:TTIT 13
:DISP:FONT:SIZ:TTIT?
:DISPlay:FSIGn[:STATe] <char>
:DISPlay:FSIGn[:STATe]?
Description
Sets the on/off state indicating a limit failure with a failure sign on the VNA screen.
Returns the on/off status of indicating a limit failure with a failure sign on the VNA screen.
Cmd Parameters
<char> 1 | 0 | ON | OFF
Query Parameters
<char> 1 | 0
Range
NA
Default Value
0
Syntax Example
:DISP:FSIG ON
:DISP:FSIG?
:DISPlay:MARKer:FREQuency:RESolution <NRf>
:DISPlay:MARKer:FREQuency:RESolution?
Description
Sets the marker frequency display resolution for the Marker Table display area.
Returns the marker frequency display resolution.
Cmd Parameters
<NRf> The input parameter is an integer.
Query Parameters
<NR1> The output parameter is an integer.
Range
3, 6, or 9
Default Value
9
Syntax Example
:DISP:MARK:FREQ:RES 3
:DISP:MARK:FREQ:RES?
:DISPlay:SIZe <char>
:DISPlay:SIZe?
Description
Sets the maximum/normal size of the graticule display.
Returns the maximum/normal size of the graticule display.
Cmd Parameters
<char> MAXimum | NORMal
Query Parameters
<char> MAX | NORM
Range
NA
Default
NA
Syntax Example
:DISP:SIZ NORM
:DISP:SIZ?
:DISPlay:SPLit <char>
:DISPlay:SPLit?
Description
Sets the channel display layout in a Row-by-Column format where the following display window arrangements are available:
• R = rows
• C = columns
• R1C1 = One trace on one row and one column
• R1C2 = Two traces, two across
• R2C1 = Two traces, two down
• R1C3 = Three traces, three across
• R3C1 = Three traces, three down
• R2C2C1 = Three traces, two on top, one on bottom
• R2C1C2 = Three traces, one on top, two on bottom
• C2R2R1 = Three traces, two on left, one on right
• C2R1R2 = Three traces, one on left, two on right
• R1C4 = Four traces, four across
• R4C1 = Four traces, four down
• R2C2 = Four traces, two across, two down
• R2C3 = Six traces, three across, two down
• R3C2 = Six traces, three down, two across
• R2C4 = Eight traces, four across, two down
• R4C2 = Eight traces, two across, four down
• R3C3 = Nine traces, three across, three down
• R5C2 = 10 traces, two across, five down
• R2C5 = 10 traces, five across, two down
• R4C3 = 12 traces, four across, three down
• R3C4 = 12 traces, three across, four down
• R4C4 = 16 traces, four across, four down
Returns the channel display layout code as above. For additional descriptions, see the VectorStar Operation Manual – 10410-00317.
Sets the maximum/normal size of the graphic display for the indicated channel.
Returns the size of the graphic display for the indicated channel.
Cmd Parameters
< char> MAX | NORM
Query Parameters
NA
Query Output
<char> MAX | NORM
Range
NA
Default Value
NORM
Syntax Example
:DISP:WIND:SIZ MAX
:DISP:WIND:SIZ?
:DISPlay:WINDow{1-16}:SPLit <char>
:DISPlay:WINDow{1-16}:SPLit?
Description
Sets the trace display layout in a Row-by-Column format for the indicated channel.
If more traces are set than the trace display contains, the higher numbered trace display windows are blank. If the trace display layout is less than the number of traces set, some traces will have overlapped displays. The following trace display window arrangements are available:
• R = rows
• C = columns
• R1C1 = One trace on one row and one column
• R1C2 = Two traces, two across
• R2C1 = Two traces, two down
• R1C3 = Three traces, three across
• R3C1 = Three traces, three down
• R2C2C1 = Three traces, two on top, one on bottom
• R2C1C2 = Three traces, one on top, two on bottom
• C2R2R1 = Three traces, two on left, one on right
• C2R1R2 = Three traces, one on left, two on right
• R1C4 = Four traces, four across
• R4C1 = Four traces, four down
• R2C2 = Four traces, two across, two down
• R2C3 = Six traces, three across, two down
• R3C2 = Six traces, three down, two across
• R2C4 = Eight traces, four across, two down
• R4C2 = Eight traces, two across, four down
• R3C3 = Nine traces, three across, three down
• R5C2 = 10 traces, two across, five down
• R2C5 = 10 traces, five across, two down
• R4C3 = 12 traces, four across, three down
• R3C4 = 12 traces, three across, four down
• R4C4 = 16 traces, four across, four down
Returns the trace display layout. For additional descriptions, see the VectorStar MS4640B VNA User Interface Reference Manual – 10410-00319.
Working with Dual-Display Rectangular Trace Formats This command is primarily designed to be used to set the per-division (PDIV) scale value on one of the VectorStar VNA dual-display trace formats. There are four dual-display trace formats:
• Linear Magnitude and Phase (LINPHase)
• Log Magnitude and Phase (LOGPHase)
• Real and Imaginary (REIMaginary)
• Impedance Real and Imaginary (ZCOMPlex)
Working with Single-Display Rectangular or Polar Trace Formats If the trace is a single-trace non-Smith chart display, this command can also set the per-division scale value on single trace displays including rectangular and polar graph trace formats.
Working with Smith Chart Trace Displays
If the trace type is a Smith Impedance chart or a Smith Admittance chart, the available scale values are limited to the values of +3dB, 0dB, -10dB, -20dB, -30dB. These are the only values permitted and other entered values result in an execution error.
Cmd Parameters
For Smith charts: <NRf> 3 | 0 | -10 | -20 | -30
Where:
• 3 = +3dB compressed Smith chart
• 0 = 0 dB standard Smith chart
• -10 = -10 dB expanded Smith chart
• -20 = -20 dB expanded Smith chart
• -30 = -30 dB expanded Smith chart
For all other displays: <NRf>
Query Parameters
<NR3> The output parameter units varies depending on the trace display type.
For Smith charts, the output parameter is based on the outside radius of the outer circle. For all other trace displays, the output is in one of the following:
• dB per division
• Hertz per division
• Meters per division
• Seconds per division.
Range
The range varies depending on display type:
• For Log Magnitude display types: 1E-3 to 1E3
• For all other non-Smith display types: 1E-5 to 1E9
• For all Smith chart display types, discrete values only: 3 | 0 | -10 | -20 | -30.
Default Value
1.000000E+001
Syntax Example
:DISP:WIND1:TRAC1:Y:PDIV 1E2
:DISP:WIND1:TRAC1:Y:PDIV?
:DISPlay:WINDow{1-16}:TRACe{1-16}:Y:PDIV2 <NRf>
:DISPlay:WINDow{1-16}:TRACe{1-16}:Y:PDIV2?
Description
Sets the per-division scale value of the bottom display of the indicated channel and trace.
It does not create an error if the command is invoked using a one-trace display, but, when a two-trace display is re-invoked, all changes made to the “hidden” trace are discarded and the trace reverts to its prior visible settings.
Returns the per-division scale value of the bottom display of the indicated channel and trace.
<NRf> The input parameter depends on the display type.
Query Parameters
<NR3> The output parameter depends on the display type.
Range
Minimum Reference Line Position = 0
Maximum Reference Line Position = Maximum Number of Divisions set by ::DISPlay:Y:NDIVisions.
Default Value
5
Syntax Example
:DISP:WIND1:TRAC1:Y:RPOS 10
:DISP:WIND1:TRAC1:Y:RPOS?
:DISPlay:WINDow{1-16}:TRACe{1-16}:Y:RPOS2 <NRf>
:DISPlay:WINDow{1-16}:TRACe{1-16}:Y:RPOS2?
Description
For dual-trace displays only.
Sets the reference line position of the bottom display of the indicated channel and trace. If the trace display is a single-trace display, this command results in an execution error.
Returns the reference line position of the bottom display of the indicated channel and trace.