VectorStar™ MS464xB Series Microwave Vector Network Analyzer Programming Manual
:
SCPI Commands
:
:CALCulate{1-16}:PARameter{1-16}:COUPling Subsystem
:CALCulate{1-16}:PARameter{1-16}:COUPling Subsystem
The
:CALCulate{1-16}:PARameter{TRAC, 1-16}:COUPling
subsystem command sets the active trace coupling mode.
Trace Subsystems
Related trace subsystems are:
•
:CALCulate{1-16}:PARameter and :PARameter{1-16} Subsystem
•
:CALCulate{1-16}:PARameter{1-16}:FSIMulator Subsystem
•
:CALCulate{1-16}:PARameter{1-16}:SELect Subsystem
•
:CALCulate{1-16}:PARameter{1-16}:COUPling Subsystem
•
:CALCulate{1-16}:POLar Subsystem
•
:CALCulate{1-16}:PROCessing:ORDer Subsystem
•
:CALCulate{1-16}[:SELected]:CONVersion Subsystem
•
:CALCulate{1-16}[:SELected]:DATA Subsystem
•
:CALCulate{1-16}[:SELected]:FORmat Subsystem
•
:CALCulate{1-16}[:SELected]:MATH Subsystem
•
:CALCulate{1-16}[:SELected]:MDATA Subsystem
•
:CALCulate{1-16}[:SELected]:SMITh Subsystem
•
:CALCulate{1-16}[:SELected]:SMOothing Subsystem
•
:CALCulate{1-16}[:SELected]:TRACE{1-16}:MARKer{1-12} Subsystem
•
:CALCulate{1-16}[:SELected]:TDATA Subsystem
•
:DISPlay Subsystem
:CALCulate{1-16}:PARameter{1-16}:COUPling <char>
:CALCulate{1-16}:PARameter{1-16}:COUPling?
Description
Sets the graph fo
Returns the graph format group coupling for the given trace in the indicated channel.
The coupling of traces is to facilitate double-y-axis displays. Traces within a coupled group will be forced to the same graph type and the same scale.
Cmd Parameters
<char1> UNCoupled | CPLA | CPLB
Query Parameters
NA
Query Output
<char1> UNC | CPLA | CPLB
Min-Max Range
NA
Default Value
UNC
Syntax Example
:CALC1:PAR1:COUP CPLA
:CALC1:PAR1:COUP?