VectorStar™ MS464xB Series Microwave Vector Network Analyzer Programming Manual : SCPI Commands : :CALCulate{1-16}:POLar Subsystem
 
:CALCulate{1-16}:POLar Subsystem
The :CALCulate{1-16}:POLar subsystem commands configure the polar chart trace displays on a per-channel basis.d and how they are displayed on each trace.
Trace Subsystems
Related trace subsystems are:
:CALCulate{1-16}:PARameter and :PARameter{1-16} Subsystem
:CALCulate{1-16}:PARameter{1-16}:FSIMulator Subsystem
:CALCulate{1-16}:PARameter{1-16}:SELect Subsystem
:CALCulate{1-16}:PARameter{1-16}:COUPling Subsystem
:CALCulate{1-16}:POLar Subsystem
:CALCulate{1-16}:PROCessing:ORDer Subsystem
:CALCulate{1-16}[:SELected]:CONVersion Subsystem
:CALCulate{1-16}[:SELected]:DATA Subsystem
:CALCulate{1-16}[:SELected]:FORmat Subsystem
:CALCulate{1-16}[:SELected]:MATH Subsystem
:CALCulate{1-16}[:SELected]:MDATA Subsystem
:CALCulate{1-16}[:SELected]:SMITh Subsystem
:CALCulate{1-16}[:SELected]:SMOothing Subsystem
:CALCulate{1-16}[:SELected]:TRACE{1-16}:MARKer{1-12} Subsystem
:CALCulate{1-16}[:SELected]:TDATA Subsystem
:DISPlay Subsystem
:CALCulate{1-16}:POLar:ANGLe:STARt <NRf>
:CALCulate{1-16}:POLar:ANGLe:STARt?
Description
Sets the start angle to use on all polar chart displays on the indicated channel.
Returns the start angle to use on all polar chart displays on the indicated channel.
Cmd Parameters
<NRf> The input parameter is in Degrees.
Query Parameters
<NR3> The output parameter is in Degrees.
Range
-3.6E2 to 3.6E2
Default Value
0.00000000000E+000
Syntax Example
:CALC1:POL:ANGL:STAR 2.10E1
:CALC1:POL:ANGL:STAR?
:CALCulate{1-16}:POLar:ANGLe:STOP <NRf>
:CALCulate{1-16}:POLar:ANGLe:STOP?
Description
Sets the stop angle to use on all polar chart displays on the indicated channel.
Returns the stop angle to use on all polar chart displays on the indicated channel.
Cmd Parameters
<NRf> The input parameter is in Degrees.
Query Parameters
<NR3> The output parameter is in Degrees.
Range
-3.6E2 to 3.6E2
Default Value
3.60000000000E+002
Syntax Example
:CALC1:POL:ANGL:STOP 2.10E1
:CALC1:POL:ANGL:STOP?
:CALCulate{1-16}:POLar:CHARt <char>
:CALCulate{1-16}:POLar:CHARt?
Description
Sets the polar chart mode of all polar chart displays on the indicated channel.
Returns the polar chart mode of all polar chart displays on the indicated channel.
Cmd Parameters
<char> MAGPhase | MAGSweep
Query Parameters
<char> MAGP | MAGS
Range
NA
Default Value
MAGP
Syntax Example
:CALC1:POL:CHAR MAGS
:CALC1:POL:CHAR?