Sets the CW sweep mode number of points of the indicated channel.
Returns the CW sweep mode number of points of the indicated channel.
Cmd Parameters
<NRf> The input parameter is a unitless number.
Query Parameters
<NR1> The output parameter is an integer.
Range
The range depends on the maximum points setting:
• When 25,000 points is set, range = 1 to 25,000.
• When 100,000 points is set, range = 1 to 100,000.
Default Value
1
Syntax Example
:SENS1:SWE:CW:POIN 1.01E2
:SENS1:SWE:CW:POIN?
:SENSe{1-16}:SWEep:CW[:STATe] <char>
:SENSe{1-16}:SWEep:CW[:STATe]?
Description
Sets the on/off state of the CW sweep mode of the indicated channel.
Returns the on/off status of the CW sweep mode of the indicated channel
Cmd Parameters
<char> 1 | 0 | ON | OFF
Query Parameters
<char> 1 | 0
Range
NA
Default
0
Syntax Example
:SENS1:SWE:CW 1
:SENS1:SWE:CW?
:SENSe{1-16}:SWEep:DELay <NRf>
:SENSe{1-16}:SWEep:DELay?
Description
Sets the sweep delay time of the indicated channel.
Returns the sweep delay time of the indicated channel.
Cmd Parameters
<NRf> The input parameter is in Seconds.
Query Parameters
<NR3> The output parameter is in Seconds.
Range
0 to 100
Default Value
0.00000000000E+000
Syntax Example
:SENS1:SWE:DEL 50.0E-3
:SENS1:SWE:DEL?
:SENSe{1-16}:SWEep:DELay:TYPe <char>
:SENSe{1-16}:SWEep:DELay:TYPe?
Description
Sets the sweep delay type of the indicated channel, as either Off, Load Pulse, or Phase Locked.
The sweep delay type definitions are:
• LOAD = Load Pulse which sets the sweep so that the sweep starts, and while sweeping, the VNA digital signal processor sends a command to load the programming for the next frequency. At this point, the user-defined delay time period occurs. The sweep-load-delay cycle repeats for every point in the sweep.
• LOCK = Phase Locked which sets the sweep delay so that the sweep starts, the hardware frequency phase locks at the point, and then user-defined delay time period occurs. The sweep continues until the sweep is completed.
• OFF = The instrument sweeps normally without sweep delay.
Returns the sweep delay type of the indicated channel.
Cmd Parameters
<char> LOCK | LOAD | OFF
Query Parameters
<char> LOCK | LOAD | OFF
Range
NA
Default Value
OFF
Syntax Example
:SENS1:SWE:DEL:TYP LOAD
:SENS1:SWE:DEL:TYP?
:SENSe{1-16}:SWEep:POINt <NRf>
:SENSe{1-16}:SWEep:POINt?
Description
Sets the number of measurement points of the indicated channel.
Returns the number of measurement points of the indicated channel.
Cmd Parameters
<NRf> The input parameter is a unitless number.
Query Parameters
<NR1> The output parameter is an integer.
Range
Range depends on the maximum points setting:
• When 25,000 points is set, range = 2 to 25,000 points.
• When 100,000 points is set, range = 2 to 100,000 points.
Default Value
201
Syntax Example
:SENS1:SWE:POIN 1.01E2
:SENS1:SWE:POIN?
:SENSe{1-16}:SWEep:TIMe <NRf>
:SENSe{1-16}:SWEep:TIMe?
Description
Sets the Sweep Time of the indicated channel.
Returns the Sweep Time of the indicated channel.
Cmd Parameters
<NRf> The input parameter is in Seconds.
Query Parameters
<NR3> The output parameter is in Seconds.
Range
MPND
Default
0.00000000000E+000
Syntax Example
:SENS1:SWE:TIM 5.0E1
:SENS1:SWE:TIM?
:SENSe{1-16}:SWEep:TIMe:DISPlay[:STATe] <char>
:SENSe{1-16}:SWEep:TIMe:DISPlay[:STATe]?
Description
Sets the on/off state of the display of Sweep Time on the indicated channel.
Returns the on/off status the display of Sweep Time on the indicated channel.
Cmd Parameters
<char> 1 | 0 | ON | OFF
Query Parameters
<char> 1 | 0
Range
NA
Default
0
Syntax Example
:SENS1:SWE:TIM:DISP 1
:SENS1:SWE:TIM:DISP?
:SENSe{1-16}:SWEep:TIMe:STATe <char>
:SENSe{1-16}:SWEep:TIMe:STATe?
Description
Sets the on/off state of Sweep Time on the indicated channel.
Return s the on/off status of Sweep Time on the indicated channel.
Cmd Parameters
<char> 1 | 0 | ON | OFF
Query Parameters
<char> 1 | 0
Range
NA
Default
0
Syntax Example
:SENS1:SWE:TIM:STAT 1
:SENS1:SWE:TIM:STAT?
:SENSe{1-16}:SWEep:TIMe:TYPe <char>
:SENSe{1-16}:SWEep:TIMe:TYPe?
Description
Sets the Sweep Time Type of the indicated channel, where:
• AUTo = The VNA controls the sweep time automatically based on the sweep range and the number of sweep points.
• MANual = Allows the user to define the sweep time for the indicated channel.
Returns the Sweep Time Type of the indicated channel.
Cmd Parameters
<char> AUTo | MANual
Query Parameters
<char> AUT | MAN
Range
NA
Default
AUT
Syntax Example
:SENS1:SWE:TIM:TYP AUT
:SENS1:SWE:TIM:TYP?
:SENSe{1-16}:SWEep:TYPe <char>
:SENSe{1-16}:SWEep:TYPe?
Description
Sets the sweep type of the indicated channel where the available sweep types are:
• LINear = Frequency-based linear sweep
• LOGarithmic = Frequency-based logarithmic sweep
• FSEGMent = Segment-based sweep with frequency-based segments
• ISEGMent = Index-based sweep with frequency-based segments
• POWer = Power-based sweep with either Power-based sweep with a CW frequency, or Power-based sweep with swept-frequency
• MFGC = Multiple frequency gain compression
Returns the sweep type of the indicated channel.
Cmd Parameters
<char> LINear | LOGarithmic | FSEGMent | ISEGMent | POWer | MFGC