VectorStar™ MS464xB Series Microwave Vector Network Analyzer User Interface Reference : Response Menus – 2-Port VNAs : User-Defined Parameter Menus – 2-Port VNAs
 
User-Defined Parameter Menus – 2-Port VNAs
USER-DEFINED Menu – 2-Port VNAs
The USER DEFINED menu is used to establish various mathematical combinations of incident and reflected power values. See Table: User-Defined Numerator/Denominator Combinations below for all possible combinations and definitions of the more common parameters.
Previous
RESPONSE Menu – 2-Port VNAs
Navigation
MAIN | Response | RESPONSE | User Defined | USER DEFINED
USER DEFINED Menu – 2-Port VNAs
Numerator
Select displays the NUMERATOR menu. Once a numerator value is selected, the menu auto-returns to the USER DEFINED menu.
NUMERATOR Menu – 2-Port VNAs
Denominator
Select displays the DENOMINATOR menu. Once a denominator value is selected, the menu auto-returns to the USER DEFINED menu.
DENOMINATOR Menu – 2-Port VNAs
Driver Port (Port 1/Port 2)
Select toggles the driving port between Port 1 and Port 2.
 
User-Defined Numerator/Denominator Combinations 
Denominator
Numerator
A1
A2
B1
B2
1
 
A1
 
S11 Forward Reflection
S21 Forward Transmission
 
A2
 
S12 Reverse Transmission
S22 Reverse Reflection
 
B1
 
 
B2
 
 
1
 
NUMERATOR Menu – 2-Port VNAs
Previous
USER-DEFINED Menu – 2-Port VNAs
Navigation
MAIN | Response | RESPONSE | User Defined | USER DEFINED | Numerator | NUMERATOR
NUMERATOR Menu – 2-Port VNAs
Numerator Auto-Return Button Selection Group
The buttons of the NUMERATOR menu form an auto-return button selection group, where the selection of any one button de-selects all other buttons and auto-returns to the USER-DEFINED menu.
The possible combinations of the NUMERATOR and DENOMINATOR functions are summarized above in Table: User-Defined Numerator/Denominator Combinations.
A1 (Numerator)
Select specifies that A1, incident power on port 1, will be the numerator value.
A2 (Numerator)
Select specifies that A2 will be the numerator value.
B1 (Numerator)
Select specifies that B1, received power on port 1, will be the numerator value.
B2 (Numerator)
Select specifies that B2 will be the numerator value.
1 (One) (Numerator)
Select specifies that 1 (one) will be the numerator value.
DENOMINATOR Menu – 2-Port VNAs
Previous
USER-DEFINED Menu – 2-Port VNAs
Navigation
MAIN | Response | RESPONSE | User Defined | USER DEFINED | Denominator | DENOMINATOR
DENOMINATOR Menu – 2-Port VNAs
Denominator Auto-Return Button Selection Group
The buttons of the DENOMINATOR menu form an auto-return button selection group, where the selection of any one button de-selects all other buttons and auto-returns to the USER DEFINED menu.
The possible combinations of the NUMERATOR and DENOMINATOR functions are summarized in Table: User-Defined Numerator/Denominator Combinations above.
A1 (Denominator)
Select sets A1 as the denominator value.
A2 (Denominator)
Select sets A2 as the denominator value.
B1 (Denominator)
Select sets B1 as the denominator value.
B2 (Denominator)
Select sets B2 as the denominator value.
1 (One) (Denominator)
Select sets 1 (one) as the denominator value.