S820E Microwave Site Masterâ„¢ User Guide : Cable and Antenna Measurements : Sweep Menu
 
Sweep Menu
Key Sequence: Sweep (3)
Sweep Menu
Sweep Setup
Data Points
Sets the number of data points: 130, 259, 517, 1033, or 2065.
Run/Hold
Run Hold
Toggles between Run and Hold. When in Hold mode, pressing this key starts the sweeping and provides a trigger. When in the Run mode, pressing this key pauses the sweep.
Sweep Trigger
This displays the Sweep Trigger Menu. The mode is displayed on the submenu key. Continuous is the default.
RF Immunity
High Low
The instrument defaults to RF Immunity Low. Refer to RF Immunity for details.
RF Pwr In Hold
On Off
Sets the RF Output power to be left On or turned Off when Run/Hold is toggled to Hold. Refer to RF Pwr In Hold for details.
Source Power
High Low
This toggles the source power between High and Low.
IFBW
Press this submenu key to display the IFBW selection box. Highlight one of the 4 choices: 
100 kHz, the maximum sweep speed
1 kHz, default
100 Hz
10 Hz, the maximum dynamic range
Smoothing
Press this submenu key to set the smoothing in percent (0 % to 20 %). Use the number keypad, the Up/Down Arrow keys, or the rotary knob.
Sweep Averaging
Press this submenu key to set the number of sweeps to average at each sweep point (1 to 1000). Use the number keypad, the Up/Down Arrow keys, or the rotary knob, then press Enter.