Reference CALIBRATION Menu
:
10 MHz Calibration Menu (Option 3)
10 MHz Calibration Menu (Option 3)
To calibrate the 10 MHz internal reference:
1.
Connect an external 10 MHz reference signal to the MG362x1A rear panel 10 MHz REF IN connector.
2.
Press the START CALIBRATION button to display the dialog shown in
Figure: 10 MHz Calibration Dialog
.
3.
Press CONTINUE to start the calibration. The status bar will display Reference Calibration in Progress. When complete, the CALIBRATION DATE will update.
10 MHz Calibration Dialog