- | AUTO RANGE The instrument will dynamically update the attenuation and preamp settings for the measurement. Auto range helps ensure that harmonics and spurs are not introduced into the measurements. OFFSET REF LEVEL The reference level is the top graticule line on the measurement display. If the reference level offset is not zero, the offset reference level is displayed at this menu location. If auto range is enabled, the reference level is graphical only and will have no impact on attenuation or preamp. AUTO REF LEVEL Auto reference level will change the reference value so as to place the highest signal amplitude at about two graticule lines from the top. This feature is not available when auto range is enabled. SCALE/DIV The scale can be set from 1 dB per division to 15 dB per division. The default setting is 10 dB. Y AXIS UNIT Selects the y-axis amplitude units of dBm, dBW, or dBµV, etc. AUTO ATTEN Input attenuation can be either tied to the reference level (on) or manually selected (off). When input attenuation is tied to the reference level, attenuation is increased as higher reference levels are selected to make sure the instrument input circuits are not saturated by large signals that are likely to be present when high reference levels are required. Auto attenuation is not available when auto range is enabled. ATTEN LEVEL Manually sets the attenuation level. This feature is not available when auto range or auto attenuation is enabled. REF LEVEL OFFSET Reference level offset compensates for the presence of external input attenuation or gain. The default offset value is 0 dB. The reference level on the Y-axis will reflect the new offset value. For example, if the reference level was 0 dBm and an offset value of –10 dB is applied, the offset reference level will be 10 dBm. PRE AMP Turns the low-noise front-end preamplifier on or off. To ensure accurate measurement results, the largest signal into the instrument input when the preamplifier is turned on should be less than –40 dBm. The preamplifier cannot be turned on if auto attenuation is on and reference level is above –40 dBm. GESTURES Opens the GESTURES Menu. Not shown in WCDMA Modulation measurements. |