Vector Network Analyzer (VNA) Measurement Guide : VNA Calibration and Measurements : Setting Up Limit Lines
 
Setting Up Limit Lines
Limit lines allow you to monitor when trace data crosses a defined line. Two types of limit lines can be specified: lower limit lines and upper limit lines. Limit lines are used for visual reference or for pass/fail criteria using the limit alarm and upper/lower limits. Selecting the LIMIT on the main menu displays the LIMIT menu.
Overview of limit lines:
Each measurement has a unique limit line.
The color of the limit line changes to red when a measurement trace exceeds a limit and when LIMIT TEST is enabled, the limit pass/fail result is displayed in the upper right of the measurement data.
Limits set beyond the current amplitude range are displayed at either the top or bottom of the graticule.
The limit line amplitude is stored when a limit line is turned off.
Limit lines must either be all single limit lines or all segmented limit lines. They cannot be mixed, but you can have a single segment as upper or lower limit lines.
PRESET LIMITS will turn off the limit line display, limit alarm and Pass/Fail message, and reset upper and lower limits to their default values.
Single Limit Lines
To enable a simple upper or lower limit lines:
1. Select LIMIT on the main menu.
2. Toggle UPPER LIMIT or LOWER LIMIT on or off.
3. Select UPPER LEVEL or LOWER LEVEL to edit the limit line amplitude, or use the touch screen to drag the limit line up or down.
A single limit line extends over the entire displayed range of the sweep, independent of the start/stop settings of the sweep. For a single limit line, the amplitude for the start/stop points is the same. Upper limit lines are labeled with a “UL”, and lower limit lines are labeled with an “LL”. Limit lines are displayed in green so long as the limits are not reached or exceeded. When a limit is exceeded (upper or lower), the limit line or segment turns red. Any portion of the measurement trace touching or exceeding a limit also turns red, while portions of the trace within limits remain in the default yellow color.
Figure: Simple Limit Line shows single upper and lower limit lines with the lower limit failing the limit test.
Each limit line can consist of a single segment, or as many as 40 segments across the entire frequency span of the instrument. These limit segments are retained regardless of the current frequency span of the instrument, which allows the configuring of specific limit envelopes at various frequencies of interest without having to re-configure them each time the frequency is changed. Limit line parameters are set using the Segmented Limit Lines.
Simple Limit Line
1. Limit points are shown as gray circles. The active point is filled in gray. Points can be dragged into position or set discretely using the frequency and amplitude settings in the.
2. The limit line shown here is a simple upper limit line. The limit line color is green when the trace does not cross the limit line, and the limit line color turns red when the trace crosses it.
3. The limit test pass/fail status is also shown in green or red color at the top of the display. The limit test is applied to the active trace, indicated here by T1.
Segmented Limit Lines
Segmented limit lines need not be continuous or connected as a single limit line. If any of the segments fail the limit test, the entire upper or lower limit fails the test. Additionally, segments can be added in any order and are not necessarily sequential from left to right or as all upper or all lower in sequence (i.e., segment 1 and 10 can be upper limits.
To enable segmented limit lines:
1. Select LIMIT on the main menu.
2. Select MODE and select Segmented. The segmented LIMIT menu will be shown (use the MODE selection to change between single or segmented limits).
3. Select SEGMENT TYPE and select Upper or Lower.
4. Toggle UPPER LIMIT or LOWER LIMIT on or off.
A single limit line segment is placed over the entire displayed range of the sweep, independent of the start/stop settings of the sweep. A segmented limit line can be divided into connected or disconnected segments with different start/stop frequency values (X1/X2) and corresponding start/stop amplitude values (Y1/Y2).
5. Select SEGMENT and select a segment number and make it the active segment. The active segment is shown in blue color.
6. Select X1 to set the start frequency of the active segment.
7. Select Y1 to set the start amplitude of the active segment.
8. Select X2 to set the stop frequency of the active segment.
9. Select Y2 to set the stop amplitude of the active segment.
Note 
Limit lines segments cannot be moved by using the touch screen.
Figure: Single Segment Limit Line shows a single segment limit line.
Single Segment Limit Line
10. Select ADD SEGMENT to add a new segment, then press SEGMENT TYPE to set it as an Upper or Lower segment.
A new segment is added after the currently set active segment and existing segment numbers are incremented by 1. The new segment starts from the stop point of the previous segment and ends at the start point of the next segment of the same type (upper or lower), or to the end of the sweep range if there are no following segments. Figure: Single Segment Limit Line shows three upper segments where the active segment 2 was added by making segment 1 active, then pressing ADD SEGMENT. The new segment was added and connects the end point of segment 1 and start point of segment 3.
Single Segment Limit Line
Segments are added, set as upper and lower, and their start and stop points are edited to create a complex pass/fail limit test. Figure: Single Segment Limit Line shows several upper and lower segmented limit lines with the lower limit failing.
Complex Segmented Limit Lines
Limit Table
A limit table can be displayed for segmented limit lines by toggling LIMIT TABLE on or off. The limit table is useful when creating very complex limit lines as you can directly select the limit segment to make it active and edit any of its parameters (start and stop points and its type) right from the table. Figure: Limit Table shows segmented limit lines with the limit table enabled. The lower segment is failing the limit test.
Limit Table
Limit Alarm
Select the LIMIT > ALARM to toggle the audible limit test alarm on or off.
LIMIT Menu (Single)
Two types of limit lines can be specified, lower limit lines and upper limit lines. Limit lines can be used for visual reference only, or for pass/fail criteria using the limit alarm. Limit alarm failures are reported whenever a signal is above the upper limit line or below the lower limit line. Limit lines cannot be used with Smith Chart or Polar graphs. Each limit line can consist of a single segment, or as many as 40 segments across the entire frequency span of the instrument. These limit segments are retained regardless of the current frequency span of the instrument. Limit segments allow the configuring of specific limit envelopes at various frequencies of interest without having to reconfigure them each time that the frequency is changed.
 
LIMIT Menu (Single)
 
LIMIT TEST: Selects UPPER or LOWER limit line for editing.
MODE: Sets the limit mode to either Single or Segmented. If the mode is set to Segmented you can add a limit segment by using custom values for X and Y axes.
UPPER LIMIT: Activates the upper limit when turned on.
UPPER LEVEL: Sets the upper limit value of the data point of active trace.
LOWER LIMIT: Activates the lower limit when turned on.
LOWER LEVEL: Sets the lower limit line value of the data point of the active trace.
ALARM: This setting is for toggling the alarm function on or off for the currently active limit line. When on, an alarm beep will occur when a data point exceeds the limit.
 
LIMIT Menu (Segmented)
If the limit MODE is set to Segmented you can add a segmented limit by using custom values for X (frequency) and Y (amplitude) axes.
LIMIT Menu
 
LIMIT TEST: When toggled on, a PASS or FAIL message is displayed at the top of the scale to indicate whether the trace touches or exceeds the limit line (FAIL).
MODE: Sets the limit line to single or segmented. Limit line segments are independent of each other and are defined by a start point and an end point. A maximum of 42 limit line segments can be defined. Both the upper and lower limit line segments for a measurement trace must be of the same type: either both are single, or both are segmented; however, you can create just one limit line segment as upper or lower to cover the entire measurement. If using single limit lines, see LIMIT Menu (Single).
SEGMENT: Selects one of the limit line segments and makes it active.
LIMIT TABLE: Toggle on or off the limit table displayed below the screen. The limit table is only available when segmented limit lines are used. The limit table provides an easy to use interface for editing limit line segments. Refer to Limit Table for more info.
ADD SEGMENT
Adds a limit line segment after the current active segment. All segment numbers after the current active segment will be incremented.
DELETE SEGMENT
Deletes the currently active segment. All segment numbers after the current active segment will be decremented.
CLEAR ALL
Removes all segments and does not store any segment information.
X1
Edits the start X value (frequency or distance) of the active limit line segment.
X2
Edits the stop X value (frequency or distance) of the active limit line segment.
Y1
Edits the start Y value (amplitude or phase) of the active limit line segment.
Y2
Edits the stop Y value (amplitude or phase) of the active limit line segment.
SEGMENT TYPE: Toggles the active limit line segment to upper or lower limit.
Y OFFSET
Applies an amplitude offset to all upper limit line segments. The offset value must not offset any upper limit line segments outside of the full scale amplitude for the measurement.
UPPER LIMIT: Toggles all of the upper limit line segments on or off.
LOWER LIMIT: Toggles all of the lower limit line segments on or off.
ALARM: When toggled on, an audible alarm sounds a repeating beep when the trace touches the limit line.