Selects Full One-Port Calibration as the calibration type on the indicated channel and ports selected. The Full 1 port calibration can be performed on any combination of ports by listing them using as PORTw | wx | wxy | wxyz where:
• w = the first port
• x = the second port
• y = the third port
• z = the fourth port
The use of Port 3 and/or Port 4 requires a 4-Port VNA instrument.
For example, to perform a FULL1 calibration on Channel 1 and Port 1, 2, and 4, the command is:
Selects One-Port Response Calibration as the calibration type on the indicated channel and ports selected. The response calibration can be performed on one or more ports by listing them using as PORTw | wx | wxy | wxyz where:
• w = the first port
• x = the second port
• y = the third port
• z = the fourth port
For example, to perform a response calibration on Channel 1 and Ports 1, 2, and 3, the command is:
SENS1:CORR:COLL:PORT123:RESP1
The use of Port 3 and/or Port 4 requires a 4-Port VNA instrument.
Initiates collection of the isolation standard data for the calibration on the given channel. For a 2-Port VNA, the available port pair is 12. Port pairs using Port 3 and/or Port 4 require a 4-Port VNA instrument. No query.
Adds Transmission Frequency Response Calibration Both directions on the indicated Channel and port pair. The Transmission Frequency Response calibration can be any combination of port pairs, and any combination of Forward only (TFRF), Reverse only (TFRR) or both (TFRB). This command adds a Transmission Frequency Response Calibration for Both directions. The use of Port 3 and/or Port 4 requires a 4-Port VNA instrument. No query.
Adds Transmission Frequency Response Calibration Forward direction on the indicated Channel and port pair. The Transmission Frequency Response calibration can be any combination of port pairs, and any combination of Forward only (TFRF), Reverse only (TFRR) or both (TFRB). This command adds a Transmission Frequency Response Calibration Forward direction. The use of Port 3 and/or Port 4 requires a 4-Port VNA instrument. No query.
Adds Transmission Frequency Response Calibration Reverse direction on the indicated Channel and port pair. The Transmission Frequency Response calibration can be any combination of port pairs, and any combination of Forward only (TFRF), Reverse only (TFRR) or both (TFRB). This command adds a Transmission Frequency Response Calibration Reverse direction. The use of Port 3 and/or Port 4 requires a 4-Port VNA instrument. No query.
Initiates collection of the through (thru) standard data for the calibration on the given channel. For 2-Port VNAs, the available port pair is 12. The use of Port 3 and/or Port 4 requires a 4-Port VNA instrument. No query.
Sets the through-line (thru-line) reference frequency of the loss on the selected port-pair of the indicated channel. The use of Port 3 and/or Port 4 requires a 4-Port VNA instrument. Outputs the thru-line reference frequency of the loss on the selected port-pair of the indicated channel.
Sets the through-line (thru-line) label on the selected port-pair of the indicated channel. On 3-Port VNAs, the available port pair is 12. The use of Port 3 and/or Port 4 requires a 4-Port VNA instrument. Outputs the thru-line label on the selected port-pair of the indicated channel.
Cmd Parameters
<string> The input parameter is any combination of numbers and letters.
Sets the through-line (thru-line) length on the selected port-pair of the indicated channel. The port-pair for 2-Port VNAs is Port 1-2. The use of Port 3 and/or Port 4 requires a 4-Port VNA instrument. Outputs the thru-line length on the selected port-pair of the indicated channel.
Sets the through-line (thru-line) loss on the selected port-pair of the indicated channel. The port-pair for 2-Port VNAs is Port 1-2. The use of Port 3 and/or Port 4 requires a 4-Port VNA instrument. Outputs the thru-line loss on the selected port-pair of the indicated channel.
Sets the through-line (thru-line) use reciprocal flag on the selected port-pair of the indicated channel. The port-pair for 2-Port VNAs is Port 1-2. The use of Port 3 and/or Port 4 requires a 4-Port VNA instrument. Outputs the thru-line use reciprocal flag on the selected port-pair of the indicated channel.
Sets the through-line (thru-line) serial number on the selected port-pair of the indicated channel. The port-pair for 2-Port VNAs is Port 1-2. The use of Port 3 and/or Port 4 requires a 4-Port VNA instrument. Outputs the thru-line serial number on the selected port-pair of the indicated channel.
Cmd Parameters
<string> The input parameter is any combination of numbers and letters. See definition of <string>.
Query Parameters
<char1> The output parameter can be any combination of numbers and letters.
Sets the through-line (thru-line) impedance on the selected port-pair of the indicated channel. The port-pair for 2-Port VNAs is Port 1-2. The use of Port 3 and/or Port 4 requires a 4-Port VNA instrument. Outputs the thru-line impedance on the selected port-pair of the indicated channel. The port-pair for 2-Port VNAs is Port 1-2. The use of Port 3 and/or Port 4 requires a 4-Port VNA instrument.
Sets the connector type for the indicated port on the indicated channel. The use of Port 3 and/or Port 4 requires a 4-Port VNA instrument. The connector types are:
• CM2-V = Anritsu 2.4 mm (male)
• CM2 = 2.4 mm (male)
• CM3 = GPC 3.5 (male)
• CMK = K (male)
• CMN = K (male)
• CMS = SMA (male)
• CMC = TNC (male)
• CMV = V (male)
• CM1 = 1 mm (male)
• CM0.8 = 0.8mm (male)
• CM716 = 7/16 (male)
• CNG = GPC7 (none)
• CMN75 = N 75 Ohm (male)
• CMUn = User-defined male connector from 1 to 32, where: CMU1 = User-defined male 1, CMU2 = User-defined male 2, CMU3 = User-defined male 3, etc.
• CF2-V = Anritsu 2.4 mm (female)
• CF2 = 2.4 mm (female)
• CF3 = GPC 3.5 (female)
• CFK = K (female)
• CFN = N (female)
• CFS = SMA (female)
• CFC = TNC (female)
• CFV = V (female)
• CF1 = 1 mm (female)
• CF0.8 = 0.8mm (female)
• CF716 = 7/16 (female)
• CFN75 = N 75 Ohm (female)
• CFUn = User-defined female connector from 1 to 32, where: CFU1 = User-defined female 1, CFU2 = User-defined female 2, CFU3 = User-defined female 3, etc.
Outputs the connector type for the indicated port on the indicated channel.
Initiates collection of the Load Standard data for the calibration on the given channel and port. The use of Port 3 and/or Port 4 requires a 4-Port VNA instrument. No query.
Selects the load standard of LOAD1 or LOAD2 to be used for calibration on the indicated channel and port. Outputs the selected load standard of LOAD1 or LOAD2 to be used for calibration on the indicated channel and port. The use of Port 3 and/or Port 4 requires a 4-Port VNA instrument.
Selects the Broadband Load Type as Fixed or Sliding for calibration on the indicated channel and port. Outputs the Broadband Load Type selection on the indicated channel and port. The use of Port 3 and/or Port 4 requires a 4-Port VNA instrument.
Sets the Load1 standard C0 (C zero) capacitance on the selected port of the indicated channel. The C0 coefficient is measured in Farads. Outputs the Load1 standard C0 capacitance on the selected port of the indicated channel. The use of Port 3 and/or Port 4 requires a 4-Port VNA instrument. See Calibration Component Parameters for a complete listing of calibration components, connectors, and their Command Parameters.
Sets the Load1 Standard C1 (C one) Coefficient on the selected port of the indicated channel. The C1 coefficient is measured in Farads/Hertz. The use of Port 3 and/or Port 4 requires a 4-Port VNA instrument. Outputs the Load1 Standard C1 Coefficient on the Selected port of the indicated channel.
Sets the Load1 Standard C2 Coefficient on the selected port of the indicated channel. The C2 coefficient is measured in Farads/Hertz^2. The use of Port 3 and/or Port 4 requires a 4-Port VNA instrument. Outputs the Load1 Standard C2 Coefficient on the selected port of the indicated channel.
Cmd Parameters
<NRf> The input parameter is in Farads/Hertz^2 (Hertz E2).
Sets the Load1 Standard C3 Coefficient on the selected port of the indicated channel. The C3 coefficient is measured in Farads/Hertz^3. The use of Port 3 and/or Port 4 requires a 4-Port VNA instrument. Outputs the Load1 Standard C3 Coefficient on the selected port of the indicated channel.
Cmd Parameters
<NRf> The input parameter is in Farads/Hertz^3 (Hertz E3).
Sets the Load1 standard L0 (L zero) inductance on the selected port of the indicated channel. The L0 coefficient is measured in Henrys. The use of Port 3 and/or Port 4 requires a 4-Port VNA instrument. Outputs the Load1 standard L0 inductance on the selected port of the indicated channel. See Calibration Component Parameters for a complete listing of calibration components, connectors, and their Command Parameters.
Sets the Load1 standard L1 (L one) inductance coefficient on the selected port of the indicated channel. The L1 coefficient is measured in Henrys/Hertz. The use of Port 3 and/or Port 4 requires a 4-Port VNA instrument. Outputs the Load1 standard L1 inductance coefficient on the selected port of the indicated channel. See Calibration Component Parameters for a complete listing of calibration components, connectors, and their Command Parameters.
Sets the Load1 standard L2 inductance coefficient on the selected port of the indicated channel. The L2 coefficient is measured in Henrys’/Hertz^2. The use of Port 3 and/or Port 4 requires a 4-Port VNA instrument. Outputs the Load1 standard L2 inductance coefficient on the selected port of the indicated channel. See Calibration Component Parameters for a complete listing of calibration components, connectors, and their Command Parameters.
Sets the Load1 standard L3 inductance coefficient on the selected port of the indicated channel. The L3 coefficient is measured in Henrys/Hertz^3. The use of Port 3 and/or Port 4 requires a 4-Port VNA instrument. Outputs the Load1 standard L3 inductance coefficient on the selected port of the indicated channel. See Calibration Component Parameters for a complete listing of calibration components, connectors, and their Command Parameters.
Sets the Load1 standard label on the selected port of the indicated channel. The use of Port 3 and/or Port 4 requires a 4-Port VNA instrument. Outputs the Load1 standard label on the selected port of the indicated channel. See Calibration Component Parameters for a complete listing of calibration components, connectors, and their Command Parameters.
Cmd Parameters
<string> The input parameter is any combination of numbers and letters.
Sets the Load1 Standard Offset length1 coefficient on the selected port of the indicated channel. The use of Port 3 and/or Port 4 requires a 4-Port VNA instrument. Outputs the Load1 Standard Offset length1 coefficient on the Selected port of the indicated channel.
Sets the Load1 Standard Offset length2 coefficient on the Selected port of the indicated channel. The use of Port 3 and/or Port 4 requires a 4-Port VNA instrument. Outputs the Load1 Standard Offset length2 coefficient on the Selected port of the indicated channel.
Sets the Load1 Standard Offset length3 coefficient on the Selected port of the indicated channel. The use of Port 3 and/or Port 4 requires a 4-Port VNA instrument. Outputs the Load1 Standard Offset length3 coefficient on the Selected port of the indicated channel.
Sets the Load1 (Load one) standard offset on the selected port of the indicated channel. The use of Port 3 and/or Port 4 requires a 4-Port VNA instrument. Outputs the Load1 standard offset on the selected port of the indicated channel. See Calibration Component Parameters for a complete listing of calibration components, connectors, and their Command Parameters.
Sets the Load1 (Load one) standard resistance on the selected port of the indicated channel. The use of Port 3 and/or Port 4 requires a 4-Port VNA instrument. Outputs the Load1 standard resistance on the selected port of the indicated channel.
Sets the Load1 (Load one) standard serial number on the selected port of the indicated channel. The use of Port 3 and/or Port 4 requires a 4-Port VNA instrument. Outputs the Load1 standard serial number on the selected port of the indicated channel.
Cmd Parameters
<string> The input parameter is any combination of numbers and letters.
Sets the Load1 (Load one) standard Impedance on the selected port of the indicated channel. The use of Port 3 and/or Port 4 requires a 4-Port VNA instrument. Outputs the Load1 standard Impedance on the selected port of the indicated channel.
Sets the Load2 standard C0 (C zero) capacitance on the selected port of the indicated channel. The C0 coefficient is measured in Farads. The use of Port 3 and/or Port 4 requires a 4-Port VNA instrument. Outputs the Load2 standard C0 capacitance on the selected port of the indicated channel.
Sets the Load2 Standard C1 (C one) Coefficient on the Selected port of the indicated channel. The C1 coefficient is measured in Farads/Hertz. The use of Port 3 and/or Port 4 requires a 4-Port VNA instrument. Outputs the Load2 Standard C1 Coefficient on the Selected port of the indicated channel
Sets the Load2 Standard C2 Coefficient on the selected port of the indicated channel. The C2 coefficient is measured in Farads/Hertz^2. The use of Port 3 and/or Port 4 requires a 4-Port VNA instrument. Outputs the Load2 Standard C2 Coefficient on the selected port of the indicated channel.
Sets the Load2 Standard C3 Coefficient on the selected port of the indicated channel. The C3 coefficient is measured in Farads/Hertz^3. The use of Port 3 and/or Port 4 requires a 4-Port VNA instrument. Outputs the Load2 Standard C3 Coefficient on the Selected port of the indicated channel.
Sets the Load2 standard L0 (L zero) inductance on the selected port of the indicated channel. The L0 coefficient is measured in Henrys. The use of Port 3 and/or Port 4 requires a 4-Port VNA instrument. Outputs the Load2 standard L0 inductance on the selected port of the indicated channel.
Sets the Load2 standard L1 (L one) inductance coefficient on the selected port of the indicated channel. The L1 coefficient is measured in Henrys/Hertz. The use of Port 3 and/or Port 4 requires a 4-Port VNA instrument. Outputs the Load2 standard L1 inductance coefficient on the selected port of the indicated channel.
Sets the Load2 standard L2 inductance coefficient on the selected port of the indicated channel. The L2 coefficient is measured in Henrys/Hertz^2. The use of Port 3 and/or Port 4 requires a 4-Port VNA instrument. Outputs the Load2 standard L2 inductance coefficient on the selected port of the indicated channel.
Sets the Load2 standard L3 inductance coefficient on the selected port of the indicated channel. The L3 coefficient is measured in Henrys/Hertz^3. The use of Port 3 and/or Port 4 requires a 4-Port VNA instrument. Outputs the Load2 standard L3 inductance coefficient on the selected port of the indicated channel.
Sets the Load2 standard label on the selected port of the indicated channel. The use of Port 3 and/or Port 4 requires a 4-Port VNA instrument. Outputs the Load2 standard label on the selected port of the indicated channel.
Cmd Parameters
<string> The input parameter is any combination of numbers and letters.
Sets the Load2 Standard Offset length1 (length one) coefficient on the selected port of the indicated channel. The OFF1 coefficient is measured in Meters/Hertz. The use of Port 3 and/or Port 4 requires a 4-Port VNA instrument. Outputs the Load2 Standard Offset length1 coefficient on the selected port of the indicated channel.
Sets the Load2 Standard Offset length2 coefficient on the selected port of the indicated channel. The OFF2 coefficient is measured in Meters/Hertz^2. The use of Port 3 and/or Port 4 requires a 4-Port VNA instrument. Outputs the Load2 Standard Offset length2 coefficient on the selected port of the indicated channel.
Sets the Load2 Standard Offset length3 coefficient on the selected port of the indicated channel. The OFF3 coefficient is measured in Meters/Hertz^3. The use of Port 3 and/or Port 4 requires a 4-Port VNA instrument. Outputs the Load2 Standard Offset length3 coefficient on the selected port of the indicated channel.
Sets the Load2 standard offset on the selected port of the indicated channel. The use of Port 3 and/or Port 4 requires a 4-Port VNA instrument. Outputs the Load2 standard offset on the selected port of the indicated channel.
Sets the Load2 standard resistance on the selected port of the indicated channel. The use of Port 3 and/or Port 4 requires a 4-Port VNA instrument. Outputs the Load2 standard resistance on the selected port of the indicated channel.
Sets the Load2 standard serial number on the selected port of the indicated channel. The use of Port 3 and/or Port 4 requires a 4-Port VNA instrument. Outputs the Load2 standard serial number on the selected port of the indicated channel.
Cmd Parameters
<string> The input parameter is any combination of numbers and letters.
Sets the Load2 standard Impedance on the selected port of the indicated channel. The use of Port 3 and/or Port 4 requires a 4-Port VNA instrument. Outputs the Load2 standard Impedance on the selected port of the indicated channel.
Cmd Parameters
<NRf> The input parameter is in Ohms.
Query Parameters
<NR3> The output parameter is in Ohms.
Range
1E-4 to 1E10
Default Value
5.00000000000E+001
Syntax Example
:SENS1:CORR:COLL:PORT1:LOAD2:Z0 7.5E1
:SENS1:CORR:COLL:PORT1:LOAD2:Z0?
:SENSe{1-16}:CORRection:COLLect:PORT{1-4}:OPEN
Description
Initiates collection of the open standard data for the calibration on the given channel and port. The use of Port 3 and/or Port 4 requires a 4-Port VNA instrument.
Sets the open standard C0 (C zero) capacitance on the selected port of the indicated channel. The C0 coefficient is measured in Farads. The use of Port 3 and/or Port 4 requires a 4-Port VNA instrument. Outputs the open standard C0 capacitance on the selected port of the indicated channel.
Sets the open standard C1 (C one) capacitance coefficient on the selected port of the indicated channel. The C1 coefficient is measured in Farads/Hertz. The use of Port 3 and/or Port 4 requires a 4-Port VNA instrument. Outputs the open standard C1 capacitance coefficient on the selected port of the indicated channel.
Sets the open standard C2 capacitance coefficient on the selected port of the indicated channel. The C2 coefficient is measured in Farads/Hertz^2. The use of Port 3 and/or Port 4 requires a 4-Port VNA instrument. Outputs the open standard C2 capacitance coefficient on the selected port of the indicated channel.
Sets the open standard C3 capacitance coefficient on the selected port of the indicated channel. The C3 coefficient is measured in Farads/Hertz^3. The use of Port 3 and/or Port 4 requires a 4-Port VNA instrument. Outputs the open standard C3 capacitance coefficient on the selected port of the indicated channel.
Sets the open standard label on the selected port of the indicated channel. The use of Port 3 and/or Port 4 requires a 4-Port VNA instrument. Outputs the open standard label on the selected port of the indicated channel.
Cmd Parameters
<string> The input parameter is any combination of numbers and letters.
Sets the open standard offset on the selected port of the indicated channel. The OFFS parameter is measured in Meters. The use of Port 3 and/or Port 4 requires a 4-Port VNA instrument. Outputs the open standard offset on the selected port of the indicated channel. See Calibration Component Parameters for a complete listing of calibration components, connectors, and their Command Parameters.
Sets the open standard serial number on the selected port of the indicated channel. The use of Port 3 and/or Port 4 requires a 4-Port VNA instrument. Outputs the open standard serial number on the selected port of the indicated channel.
Cmd Parameters
<string> The input parameter is any combination of numbers and letters. See definition of <string>.
Query Parameters
<char1> The output parameter can be any combination of numbers and letters.
Selects the reflection standard to use for the reflection calibration on the indicated channel and port where the reflection standard types can be:
• OPEN = Open
• SHORt = Short
• OFSH1 = Offset Short 1
• OFSH2 = Offset Short 2
• OFSH3 = Offset Short 3
• NONE = Only returned by the query if not reflection component has been defined.
The use of Port 3 and/or Port 4 requires a 4-Port VNA instrument. Outputs the reflect standard to use for the reflection calibration on the indicated channel and port.
Initiates collection of the short standard data for the calibration on the given channel and port. See Calibration Component Parameters for a complete listing of calibration components, connectors, and their command parameters. The use of Port 3 and/or Port 4 requires a 4-Port VNA instrument.
Sets the short standard L0 (L zero) inductance on the selected port of the indicated channel. The L0 coefficient is measured in Henrys. Outputs the short standard L0 inductance on the selected port of the indicated channel. See Calibration Component Parameters for a complete listing of calibration components, connectors, and their Command Parameters.
Sets the short standard L1 (L one) inductance coefficient on the selected port of the indicated channel. The L1 coefficient is measured in Henrys/Hertz. The use of Port 3 and/or Port 4 requires a 4-Port VNA instrument. Outputs the short standard L1 inductance coefficient on the selected port of the indicated channel. See Calibration Component Parameters for a complete listing of calibration components, connectors, and their command parameters.
Sets the short standard L2 inductance coefficient on the selected port of the indicated channel. The L2 coefficient is measured in Henrys/Hertz^2. The use of Port 3 and/or Port 4 requires a 4-Port VNA instrument. Outputs the short standard L2 inductance coefficient on the selected port of the indicated channel.
Sets the short standard L3 inductance coefficient on the selected port of the indicated channel. The L3 coefficient is measured in Henrys/Hertz^3. The use of Port 3 and/or Port 4 requires a 4-Port VNA instrument. Outputs the short standard L3 inductance coefficient on the selected port of the indicated channel.
Sets the short standard label on the selected port of the indicated channel. The use of Port 3 and/or Port 4 requires a 4-Port VNA instrument. Outputs the short standard label on the selected port of the indicated channel.
Cmd Parameters
<string> The input parameter is any combination of numbers and letters.
Sets the short standard offset on the selected port of the indicated channel. The use of Port 3 and/or Port 4 requires a 4-Port VNA instrument. Outputs the short standard offset on the selected port of the indicated channel.
Sets the short standard serial number on the selected port of the indicated channel. The use of Port 3 and/or Port 4 requires a 4-Port VNA instrument. Outputs the short standard serial number on the selected port of the indicated channel.
Cmd Parameters
<string> The input parameter is any combination of numbers and letters.
<char1> The output parameter can be any combination of numbers and letters.
Range
NA
Default Value
NA
Syntax Example
:SENS1:CORR:COLL:PORT1:SHORT:SER ’123456’
:SENS1:CORR:COLL:PORT1:SHORT:SER?
123456
:SENSe{1-16}:CORRection:COLLect:PORT{1-4}:SHORT1
Description
Initiates collection of the Offset Short1 (Short one) standard data for the calibration on the given channel and port. See Calibration Component Parameters. for a complete listing of calibration components, connectors, and their Command Parameters. The use of Port 3 and/or Port 4 requires a 4-Port VNA instrument.
Sets the Short1 standard L0 (L zero) inductance on the selected port of the indicated channel. The L0 parameter is measured in Henrys. The use of Port 3 and/or Port 4 requires a 4-Port VNA instrument. Outputs the Short1 standard L0 inductance on the selected port of the indicated channel. See Calibration Component Parameters for a complete listing of calibration components, connectors, and their Command Parameters.
Sets the Short1 standard L1 (L one) inductance coefficient on the selected port of the indicated channel. The L1 coefficient is measured in Henrys/Hertz. The use of Port 3 and/or Port 4 requires a 4-Port VNA instrument. Outputs the Short1 standard L1 inductance coefficient on the selected port of the indicated channel.
Sets the Short1 standard L2 inductance coefficient on the selected port of the indicated channel. The L2 coefficient is measured in Henrys/Hertz^2. The use of Port 3 and/or Port 4 requires a 4-Port VNA instrument. Outputs the Short1 standard L2 inductance coefficient on the selected port of the indicated channel.
Sets the Short1 standard L3 inductance coefficient on the selected port of the indicated channel. The L3 coefficient is measured in Henrys/Hertz^3. The use of Port 3 and/or Port 4 requires a 4-Port VNA instrument. Outputs the Short1 standard L3 inductance coefficient on the selected port of the indicated channel.
Sets the Short1 (Short one) standard label on the selected port of the indicated channel. The use of Port 3 and/or Port 4 requires a 4-Port VNA instrument. Outputs the Short1 standard label on the selected port of the indicated channel. See Calibration Component Parameters for a complete listing of calibration components, connectors, and their Command Parameters.
Cmd Parameters
<string> The input parameter is any combination of numbers and letters.
Sets the Short1 (Short 1) standard offset on the selected port of the indicated channel. The use of Port 3 and/or Port 4 requires a 4-Port VNA instrument. Outputs the Short1 standard offset on the selected port of the indicated channel.
Sets the Short1 (Short 1) standard serial number on the selected port of the indicated channel. The use of Port 3 and/or Port 4 requires a 4-Port VNA instrument. Outputs the Short1 standard serial number on the selected port of the indicated channel.
Cmd Parameters
<string> The input parameter is any combination of numbers and letters. See definition of <string>.
Query Parameters
<char1> The output parameter can be any combination of numbers and letters.
Range
NA
Default Value
NA
Syntax Example
:SENS1:CORR:COLL:PORT1:SHORT1:SER ’123456’
:SENS1:CORR:COLL:PORT1:SHORT1:SER?
123456
:SENSe{1-16}:CORRection:COLLect:PORT{1-4}:SHORT2
Description
Initiates collection of the Offset Short2 standard data for the calibration on the given channel and port.
The use of Port 3 and/or Port 4 requires a 4-Port VNA instrument.
Sets the Short2 standard L0 (L zero) inductance on the selected port of the indicated channel. The L0 (zero) coefficient is measured in Henrys. The use of Port 3 and/or Port 4 requires a 4-Port VNA instrument. Outputs the Short2 standard L0 inductance on the selected port of the indicated channel.
Sets the Short2 standard L1 (L one) inductance coefficient on the selected port of the indicated channel. The L1 coefficient is measured in Henrys/Hertz. The use of Port 3 and/or Port 4 requires a 4-Port VNA instrument. Outputs the Short2 standard L1 inductance coefficient on the selected port of the indicated channel.
Sets the Short2 standard L2 inductance coefficient on the selected port of the indicated channel. The L2 coefficient is measured in Henrys/Hertz^2. The use of Port 3 and/or Port 4 requires a 4-Port VNA instrument. Outputs the Short2 standard L2 inductance coefficient on the selected port of the indicated channel.
Sets the Short2 standard L3 inductance coefficient on the selected port of the indicated channel. The L3 coefficient is measured in Henrys/Hertz^3. The use of Port 3 and/or Port 4 requires a 4-Port VNA instrument. Outputs the Short2 standard L3 inductance coefficient on the selected port of the indicated channel.
Sets the Short2 standard label on the selected port of the indicated channel. The use of Port 3 and/or Port 4 requires a 4-Port VNA instrument. Outputs the Short2 standard label on the selected port of the indicated channel.
Cmd Parameters
<string> The input parameter is any combination of numbers and letters.
Sets the Short2 standard offset on the selected port of the indicated channel. The use of Port 3 and/or Port 4 requires a 4-Port VNA instrument. Outputs the Short2 standard offset on the selected port of the indicated channel.
Sets the Short2 standard serial number on the selected port of the indicated channel. The use of Port 3 and/or Port 4 requires a 4-Port VNA instrument. Outputs the Short2 standard serial number on the selected port of the indicated channel.
Cmd Parameters
<string> The input parameter is any combination of numbers and letters. See definition of <string>.
Query Parameters
<char1> The output parameter can be any combination of numbers and letters.
Range
NA
Default Value
NA
Syntax Example
:SENS1:CORR:COLL:PORT1:SHORT2:SER ’123456’
:SENS1:CORR:COLL:PORT1:SHORT2:SER?
123456
:SENSe{1-16}:CORRection:COLLect:PORT{1-4}:SHORT3
Description
Initiates collection of the Offset Short3 standard data for the calibration on the given channel and port. The use of Port 3 and/or Port 4 requires a 4-Port VNA instrument. No query.
Sets the Short3 standard L0 (L zero) inductance on the selected port of the indicated channel. The L0 coefficient is measured in Henrys. The use of Port 3 and/or Port 4 requires a 4-Port VNA instrument. Outputs the Short3 standard L0 inductance on the selected port of the indicated channel.
Sets the Short3 standard L1 (L one) inductance coefficient on the selected port of the indicated channel. The L1 coefficient is measured in Henrys/Hertz. The use of Port 3 and/or Port 4 requires a 4-Port VNA instrument. Outputs the Short3 standard L1 inductance coefficient on the selected port of the indicated channel.
Sets the Short3 standard L2 inductance coefficient on the selected port of the indicated channel. The L2 coefficient is measured in Henrys/Hertz^2. The use of Port 3 and/or Port 4 requires a 4-Port VNA instrument. Outputs the Short3 standard L2 inductance coefficient on the selected port of the indicated channel.
Sets the Short3 standard L3 inductance coefficient on the selected port of the indicated channel. The L3 coefficient is measured in Henrys/Hertz^3. The use of Port 3 and/or Port 4 requires a 4-Port VNA instrument. Outputs the Short3 standard L3 inductance coefficient on the selected port of the indicated channel.
Sets the Short3 standard label on the selected port of the indicated channel. The use of Port 3 and/or Port 4 requires a 4-Port VNA instrument. Outputs the Short1 standard label on the selected port of the indicated channel.
Cmd Parameters
<string> The input parameter is any combination of numbers and letters.
Sets the Short3 standard offset on the selected port of the indicated channel. The use of Port 3 and/or Port 4 requires a 4-Port VNA instrument. Outputs the Short3 standard offset on the selected port of the indicated channel.
Sets the Short3 standard serial number on the selected port of the indicated channel. The use of Port 3 and/or Port 4 requires a 4-Port VNA instrument. Outputs the Short1 standard serial number on the selected port of the indicated channel.
Cmd Parameters
<string> The input parameter is any combination of numbers and letters. See definition of <string>.
Query Parameters
<char1> The output parameter can be any combination of numbers and letters.
Range
NA
Default Value
NA
Syntax Example
:SENS1:CORR:COLL:PORT1:SHORT3:SER ’123456’
:SENS1:CORR:COLL:PORT1:SHORT3:SER?
:SENSe{1-16}:CORRection:COLLect:PORT{1-4}:SLOAD1
Description
Initiates collection of the Sliding Load1 standard data for the calibration on the given channel and port. The use of Port 3 and/or Port 4 requires a 4-Port VNA instrument. No query.
Cmd Parameters
NA
Range
NA
Default Value
NA
Syntax Example
:SENS1:CORR:COLL:PORT1:SLOAD1
:SENSe{1-16}:CORRection:COLLect:PORT{1-4}:SLOAD2
Description
Initiates collection of the Sliding Load2 standard data for the calibration on the given channel and port. The use of Port 3 and/or Port 4 requires a 4-Port VNA instrument. No query.
Cmd Parameters
NA
Range
NA
Default Value
NA
Syntax Example
:SENS1:CORR:COLL:PORT1:SLOAD2
:SENSe{1-16}:CORRection:COLLect:PORT{1-4}:SLOAD3
Description
Initiates collection of the Sliding Load3 standard data for the calibration on the given channel and port. The use of Port 3 and/or Port 4 requires a 4-Port VNA instrument. No query.
Cmd Parameters
NA
Range
NA
Default Value
NA
Syntax Example
:SENS1:CORR:COLL:PORT1:SLOAD3
:SENSe{1-16}:CORRection:COLLect:PORT{1-4}:SLOAD4
Description
Initiates collection of the Sliding Load4 standard data for the calibration on the given channel and port. The use of Port 3 and/or Port 4 requires a 4-Port VNA instrument. No query.
Cmd Parameters
NA
Range
NA
Default Value
NA
Syntax Example
:SENS1:CORR:COLL:PORT1:SLOAD4
:SENSe{1-16}:CORRection:COLLect:PORT{1-4}:SLOAD5
Description
Initiates collection of the Sliding Load5 standard data for the calibration on the given channel and port. The use of Port 3 and/or Port 4 requires a 4-Port VNA instrument. No query.
Cmd Parameters
NA
Range
NA
Default Value
NA
Syntax Example
:SENS1:CORR:COLL:PORT1:SLOAD5
:SENSe{1-16}:CORRection:COLLect:PORT{1-4}:SLOAD6
Description
Initiates collection of the Sliding Load6 standard data for the calibration on the given channel and port. The use of Port 3 and/or Port 4 requires a 4-Port VNA instrument. No query.