VectorStar™ MS4640A Series Microwave Vector Network Analyzers : Calibration Menus - 4-Port VNAs : Calibration Utility Functions
 
Calibration Utility Functions
The calibration utility functions and management menus and dialog boxes are:
CAL OPTIONS Menu - 4-Port VNAs
FLEXIBLE CAL SETUP Dialog Box - 4-Port VNAs
MANUAL ADAPTER REMOVAL Dialog Box - 4-Port VNAs
CAL MERGE Dialog Box - 4-Port VNAs
NETWORK EXTRACTION Dialog Box - 4-Port VNAs
CAL KIT/AUTOCAL Menu - 4-Port VNAs
INSTALL (AUTOCAL CHARAC./CAL KIT) Dialog Box - 4-Port VNAs
SAVE (AUTOCAL CHARAC./CAL KIT) Dialog Box - 4-Port VNAs
CAL KIT INFO Dialog Box - 4-Port VNAs
RESTORE DEFAULT COEF. Dialog Box - 4-Port VNAs
 
CAL OPTIONS Menu - 4-Port VNAs
The CAL OPTIONS menu and related sub-menus provide access to advanced calibration options either before or after a calibration procedure has been completed.
Previous
CALIBRATION [TR] Menu - 4-Port VNAs
Navigation
MAIN | Calibration | CALIBRATION | Cal Options | CAL OPTIONS
 
CAL OPTIONS (CALIBRATION OPTIONS) Menu 
 
Flexible Cal Status (Off/On)
Flexible calibration is the ability to select a sub-set of a currently active calibration status. For example, if a Full 2 Port calibration has been completed, flexible calibration allows the calibration to be reduced to a Transmission Frequency Response calibration. Select toggles flexible calibration off and on.
Flexible Cal Setup
Use Flexible Cal Setup to change the parameters above. Select displays the FLEXIBLE CAL SETUP dialog box.
FLEXIBLE CAL SETUP Dialog Box - 4-Port VNAs
Interpolation (Off/On)
Interpolation allows additional interpolated measurement points between calibrated measurement points. This is useful if the user wants to zoom into a specific area without having to recalibrate the instrument. The interpolated points must lie within the calibration frequency points. Select toggles interpolation OFF and ON with a default state of OFF.
Apply Isolation (Off/On)
If this button is unavailable, isolation calibration was not performed. The isolation calibration procedure is started by a button on the calibration type menu.
If this button is available, isolation calibration was performed, and select toggles isolation calibration OFF and ON.
Perform Manual Adapter Removal
Manual adapter removal extracts the behavior of the adapter from the setup after a successful calibration. Select displays the MANUAL ADAPTER REMOVAL dialog box.
MANUAL ADAPTER REMOVAL Dialog Box - 4-Port VNAs
Hybrid
Hybrid calibration takes either up to four 1-port cals and hybridizes them into a 2-port, 3-port, or 4-port cal, or takes two 2-port cals and hybridizes them into 1 4-port cal.
See Calibration Menus - 2-Port VNAs and the section HYBRID CAL Menu - 2-Port VNAs.
Cal Merge
Cal Merge allows merging two calibration files into a single file where the calibrations can use different methods at different frequencies. The calibrations must be of the same calibration type. Select displays the CAL MERGE dialog box.
CAL MERGE Dialog Box - 4-Port VNAs
Network Extraction
Use network extraction to generate an S-Parameter (.s2p) file for a set of networks. The file can ben be embedded or de-embedded as required. Select displays the NETWORK EXTRACTION dialog box.
NETWORK EXTRACTION Dialog Box - 4-Port VNAs
 
 
FLEXIBLE CAL SETUP Dialog Box - 4-Port VNAs
Use the Flexible Cal Setup dialog box to define a calibration sub-set of the currently active calibration.
Previous
CAL OPTIONS Menu - 4-Port VNAs
Navigation
MAIN | Calibration | CALIBRATION | Cal Options | CAL OPTIONS | Flexible Cal Setup | FLEXIBLE CAL SETUP Dialog Box
 
FLEXIBLE CAL SETUP (FLEXIBLE CALIBRATION SETUP) Dialog Box – 4-Port VNAs 
 
Instructions
The flexible calibration setup feature makes it possible to select a subset of the S-Parameters to be corrected from the available full-term calibration.
Procedure
1. Perform the necessary full-term calibration.
2. Navigate to the FLEXIBLE CAL SETUP dialog box appears:
MAIN | Calibration | CALIBRATION | Cal Options | CAL OPTIONS | Flexible Cal Setup | FLEXIBLE CAL SETUP
3. Select the input method (full term, reflection only, or customize). The default selection is full term calibration with all the ports turned on for the available calibration type. The following radio button options are available in the Input Method area:
Full Term Cal (By Port)
Reflection Only (By Port)
Customize Cal (By S-Param)
4. If Full Term Cal (By Port) is selected above, the area label is Port Selections (Full Term) with available options of:
Port 1 and/or
Port 2
5. If Reflection Only (By Port) is selected above, the area label is Port Selections (Reflection Only) with the available options of:
Port 1 and/or
Port 2
6. If Customize Cal (By S-Param) is selected, the S-Param Selections area appears with S-Parameter selection check boxes. Select any combination of:
S11
S21
S12
S22
7. Select Apply Selection to apply the selections.
8. Select Close to close the dialog without saving the selections.
 
 
MANUAL ADAPTER REMOVAL Dialog Box - 4-Port VNAs
Use manual adapter removal to extract the electrical behavior of an adapter after a successful calibration procedure. This is especially useful when the DUT configuration is not entirely compatible with common calibration procedures such as having different connectors at each end.
Previous
CAL OPTIONS Menu - 4-Port VNAs
Navigation
MAIN | Calibration | CALIBRATION | Cal Options | CAL OPTIONS | Perform Manual Adapter Removal | MANUAL ADAPTER REMOVAL Dialog Box
 
MANUAL ADAPTER REMOVAL Dialog Box 
Instructions
Adapter removal permits accurate measurement of non-insertable devices. The process involves using an adapter of known electrical length and performing two full 12-term calibrations. In the procedure below:
The Y file is the file with the calibration when the adapter connected to Port 1.
The X file is the file with the calibration when the adapter connected to Port 2.
Note 
ONLY AVAILABLE FOR 2-PORT.
For 4-port Adapter Removal, use network extraction for the SNP file of the adapter, and use de-embedding to de-embed it from the calibration.
Procedure
Select the port pair to be used from the following port combinations: 1,2; 1,3; 1,4; 2,3; 2,4; or 3,4.
1. Connect the adapter to Port X where X signifies any port. Perform a full 12-term calibration using Y' and Y as the test ports and store calibration to disk.
2. Connect the adapter to Port Y where Y signifies any port that is not X. Perform a full 12-term calibration using X and X' as the test ports and store calibration to disk.
3. Call up the X and Y files.
4. Input the estimated adapter electrical length in picoseconds (ps).
5. Select Perform Adapter Removal to remove adapter.
 
 
CAL MERGE Dialog Box - 4-Port VNAs
Calibration merge allows merging two calibration files into a single file where the prior calibrations can use different methods (such as SOLT and SSLT) at different frequencies. The calibrations must be the same calibration type (such as Full 1 Port) and be available on the instrument hard drive. Calibration merge has other requirements (described below) as to the number of points, start and stop frequencies, and front panel settings.
Prerequisites
4-Port Mode
Previous
CAL OPTIONS Menu - 4-Port VNAs
Navigation
MAIN | Calibration | CALIBRATION | Cal Options | CAL OPTIONS | Cal Merge | CAL MERGE Dialog Box
 
CAL MERGE Dialog Box 
Instructions
Cal Merge allows merging two RF calibration files into one calibration file. The RF calibrations can be done using different methods and different frequencies.
Requirements and Notes
The two RF calibrations must be of the same calibration type (such as Full 2 Port) and must be stored on the VectorStar VNA hard drive.
The combined frequency list for the two source cal files cannot exceed the maximum allowable instrument points, which is either 25K or 100K points depending on MAX POINTS menu selection.
The frequency lists and matching correction terms are combined. For the frequencies which coincide, the terms from the first cal file are used.
The start and stop frequencies are adjusted to include the entire frequency range provided by the two cal files.
The first cal file provides all other front panel setup configuration information.
Since Calibration Merge can result in non-discrete step size, the cal merge file sweep type is set to Frequency-Based Segmented Sweep.
Procedure
1. 1) Select Browse to select the appropriate cal files. For example, select a file such as calfile1.chx for the first file and calfile2.chx for the second file.
2. Select Perform Cal Merge to perform the merge of the frequency lists and matching correction terms.
3. For best practices it is recommend saving the resultant Cal Merge file, but a save is not required.
 
HYBRID CAL Menu - 4-Port VNAs
Prerequisites
4-Port Mode
Previous
CAL OPTIONS Menu - 4-Port VNAs
Navigation
MAIN | Calibration | CALIBRATION | Cal Options | CAL OPTIONS | Hybrid Cal | HYBRID CAL
 
CAL OPTIONS (CALIBRATION OPTIONS) Menu 
Hybrid Cal Setup
Use Hybrid Cal Setup to change the hybrid calibration parameters. Select displays the HYBRID CAL SETUP dialog box.
Thru
Read-only display. Shows the thru port pair configured in the HYBRID CAL SETUP dialog box.
Done
The Done button is unavailable until a successful hybrid calibration has been completed. When available, select returns to the CAL OPTIONS menu.
Abort Hybrid Cal
The Abort button is unavailable until a hybrid calibration is started. Selection cancels the hybrid calibration, and returns to the CAL OPTIONS menu.
 
HYBRID CAL SETUP Dialog Box - 4-Port VNAs
Prerequisites
4-Port Mode
Previous
HYBRID CAL Menu - 4-Port VNAs
Navigation
MAIN | Calibration | CALIBRATION | Cal Options | CAL OPTIONS | Hybrid Cal | HYBRID CAL | Hybrid Cal Setup | HYBRID CAL SETUP Dialog Box
 
HYBRID CAL (CALIBRATION) Dialog Box – 4-Port VNAs 
 
NETWORK EXTRACTION Dialog Box - 4-Port VNAs
The network extraction features provides a method of generating an S-Parameter (S2P) file for a set of networks. The S2P file can then be embedded or de-embedded into the VNA’s error coefficient as required. The first NETWORK EXTRACTION dialog box allow selection of one of the seven types of extraction available.
Prerequisites
4-Port Mode
Previous
CAL OPTIONS Menu - 4-Port VNAs
Navigation
MAIN | Calibration | CALIBRATION | Cal Options | CAL OPTIONS | Network Extraction | NETWORK EXTRACTION Dialog Box
 
NETWORK EXTRACTION Dialog Box – 4-Port VNAs 
 
Instructions
The network extraction feature provides the method for generating the S-Parameters for a given set of networks. This generated S2P file can then be embedded or de-embedded into the VNA error coefficient at a later stage. The table below summarizes the types of network extractions and their general attributes. Follow the cross references for additional information about each type of network extraction.
Network Extraction Type Definitions 
Network Extraction
Type
Extract
These
Networks
Extraction
Method
Cross Reference
to Dialog Box
Description
Type A
One 2-Port Networks
Adapter Extraction
Type B
Two Tier Calibration
Type C
Two 2-Port Networks
Inner and Outer Cals Available
Type D
Outer Cal only using Divide-by-Two Method
Type E
Four 2-Port Networks
Inner and Outer Cals Available
Type F
Outer Cal only using Divide-by-Two Method
Type G
Two 4-Port Networks
Outer Cal only using Divide-by-Two Method
 
Requirements and Notes
The outer and inner files must meet the following requirements:
Both files must be from a full term calibration such as a Full 12-Term (SOLT) calibration.
Both files must be the same calibration type.
Both files must have the same frequencies and number of points.
Procedure
1. Select Browse to display the OPEN (SETUP CAL FILE) dialog box and select the appropriate Outer File and Inner File.
2. Select the Perform Ntwk. Extraction button on the dialog box to perform the extraction.
3. If the extraction is successful, follow the prompt to save the appropriate S2P files. For a Full 12 Term Calibration, two (2) S2P files will be generated.
4. If either file is incorrect, a warning message is displayed.
5. Click Close to exit the dialog box.
 
 
NETWORK EXTRACTION Dialog Box – Type A – 4-Port VNAs
Full Name
NETWORK EXTRACTION Dialog Box – Type A – One 2-Port Network - Adapter Extraction - 4-Port VNAs
Previous
NETWORK EXTRACTION Dialog Box - 4-Port VNAs
Navigation
MAIN | Calibration | CALIBRATION | Cal Options | CAL OPTIONS | Network Extraction | NETWORK EXTRACTION Dialog Box | Extract One 2-Port Network | Type A - Adapter Extraction | NETWORK EXTRACTION [EXTRACT ONE 2-PORT NETWORK (TYPE A)] Dialog Box
 
NETWORK EXTRACTION - EXTRACT ONE 2-PORT NETWORK - TYPE A - 4-Port
 
 
Instructions
Network Extraction provides the means of generating SnP files of networks. The generated files can than be embedded or de-embedded. Based on the type of extraction chosen, multiple SnP files may be generated, as shown in the graphics for each extraction type. Port Swapping can be performed in the Embedding/De-embedding menus.
All calibration files must be Full cals, of the same Cal type, and over the same exact frequency points.
Procedure
1. Select Browse to select the appropriate cal file or files.
2. Select Perform Network Extraction to perform the extraction.
3. If the extraction is successful, follow the prompt to save the generated SnP file or files.
 
NETWORK EXTRACTION Dialog Box – Type B – 4-Port VNAs
Full Name
NETWORK EXTRACTION Dialog Box – Type B – Extract One 2-Port Network - Two Tier Calibration - 4-Port VNAs
Previous
NETWORK EXTRACTION Dialog Box - 4-Port VNAs
Navigation
MAIN | Calibration | CALIBRATION | Cal Options | CAL OPTIONS | Network Extraction | NETWORK EXTRACTION Dialog Box | Extract One 2-Port Network | Type B - Two Tier Calibration | NETWORK EXTRACTION [EXTRACT ONE 2-PORT NETWORK (TYPE B)] Dialog Box
 
NETWORK EXTRACTION - EXTRACT ONE 2-PORT NETWORK - TYPE B - 4-Port
 
Instructions
Network Extraction provides the means of generating SnP files of networks. The generated files can than be embedded or de-embedded. Based on the type of extraction chosen, multiple SnP files may be generated, as shown in the graphics for each extraction type. Port Swapping can be performed in the Embedding/De-embedding menus.
Cal A and Cal B must share a common test port. Cal A in this extraction type must only be a full 1-port cal, which is ideal if a Thru is not available. Both cals must have the exact same frequency points.
Procedure
1. Select Browse to select the appropriate cal file/s.
2. Select Perform Network Extraction to perform the extraction.
3. If the extraction is successful, follow the prompt to save the generated SnP file or files.
 
 
NETWORK EXTRACTION Dialog Box – Type C – 4-Port VNAs
Full Name
NETWORK EXTRACTION Dialog Box – Type C – Extract Two 2-Port Networks - Inner and Outer Cals Available - 4-Port VNAs
Previous
NETWORK EXTRACTION Dialog Box - 4-Port VNAs
Navigation
MAIN | Calibration | CALIBRATION | Cal Options | CAL OPTIONS | Network Extraction | NETWORK EXTRACTION Dialog Box | Extract Two 2-Port Networks | Type C - Inner and Outer Cals Available | NETWORK EXTRACTION [EXTRACT TWO 2-PORT NETWORKS (TYPE C)] Dialog Box
 
NETWORK EXTRACTION - EXTRACT TWO 2-PORT NETWORKS - TYPE C - 4-Port
 
Instructions
Network Extraction provides the means of generating SnP files of networks. The generated files can than be embedded or de-embedded. Based on the type of extraction chosen, multiple SnP files may be generated, as shown in the graphics for each extraction type. Port Swapping can be performed in the Embedding / De-embedding menus.
All calibration files must be Full cals, of the same Cal type, and over the same exact frequency points.
Procedure
1. Select Browse to select the appropriate cal file/s. Note CalA is the inner file and CalB is the outer file.
2. Select Perform Network Extraction to perform the extraction.
3. If the extraction is successful, follow the prompt to save the generated SnP file or files.
 
NETWORK EXTRACTION Dialog Box – Type D – 4-Port VNAs
Full Name
NETWORK EXTRACTION Dialog Box – Type D – Extract Two 2-Port Networks - Outer Cal Only Using Divide-By-Two Method - 4-Port VNAs
Previous
NETWORK EXTRACTION Dialog Box - 4-Port VNAs
Navigation
MAIN | Calibration | CALIBRATION | Cal Options | CAL OPTIONS | Network Extraction | NETWORK EXTRACTION Dialog Box | Extract Two 2-Port Networks | Type D - Outer Cal Only Using Divide-By-Two Method | NETWORK EXTRACTION [EXTRACT TWO 2-PORT NETWORKS (TYPE D)] Dialog Box
 
NETWORK EXTRACTION - EXTRACT TWO 2-PORT NETWORKS - TYPE D - 4-Port
 
Instructions
Network Extraction provides the means of generating SnP files of networks. The generated files can than be embedded or de-embedded. Based on the type of extraction chosen, multiple SnP files may be generated, as shown in the graphics for each extraction type. Port Swapping can be performed in the Embedding/De-embedding menus.
These extraction types are for cases where an inner-cal is not possible. The network measured is assumed symmetrical, and SnP files generated using Divide-By-Two (2) schemes. An option is given to zero-out the match terms instead of fully allocating them to the outer-ports.
Procedure
1. Make sure the appropriate calibration is active.
2. Zero-out the match terms if needed.
3. Connect the network and select Perform Network Extraction to perform the extraction.
4. If the extraction is successful, follow the prompt to save the generated SnP file or files.
 
 
NETWORK EXTRACTION Dialog Box – Type E – 4-Port VNAs
Full Name
NETWORK EXTRACTION Dialog Box – Type E – Extract Four 2-Port Networks - Inner and Outer Cals Available - 4-Port VNAs
Previous
NETWORK EXTRACTION Dialog Box - 4-Port VNAs
Navigation
MAIN | Calibration | CALIBRATION | Cal Options | CAL OPTIONS | Network Extraction | NETWORK EXTRACTION Dialog Box | Extract Four 2-Port Networks | Type E - Inner and Outer Cals Available | NETWORK EXTRACTION [EXTRACT FOUR 2-PORT NETWORKS (TYPE E)] Dialog Box
 
NETWORK EXTRACTION - EXTRACT TWO 2-PORT NETWORKS - TYPE E - 4-Port
 
Instructions
Network Extraction provides the means of generating SnP files of networks. The generated files can than be embedded or de-embedded. Based on the type of extraction chosen, multiple SnP files may be generated, as shown in the graphics for each extraction type. Port Swapping can be performed in the Embedding/De-embedding menus.
All calibration files must be Full cals, of the same Cal type, and over the same exact frequency points.
Procedure
1. Select Browse to select the appropriate cal file/s.
2. Select Perform Network Extraction to perform the extraction.
3. If the extraction is successful, follow the prompt to save the generated SnP file or files.
 
 
NETWORK EXTRACTION Dialog Box – Type F – 4-Port VNAs
Full Name
NETWORK EXTRACTION Dialog Box – Type F – Extract Four 2-Port Networks - Outer Cal Only Using Divide-By-Two Method - 4-Port VNAs
Previous
NETWORK EXTRACTION Dialog Box - 4-Port VNAs
Navigation
MAIN | Calibration | CALIBRATION | Cal Options | CAL OPTIONS | Network Extraction | NETWORK EXTRACTION Dialog Box | Extract Four 2-Port Networks | Type F - Outer Cal Only Using Divide-By-Two Method | NETWORK EXTRACTION [EXTRACT FOUR 2-PORT NETWORKS (TYPE F)] Dialog Box
 
NETWORK EXTRACTION - EXTRACT TWO 2-PORT NETWORKS - TYPE F - 4-Port
 
Instructions
Network Extraction provides the means of generating SnP files of networks. The generated files can than be embedded or de-embedded. Based on the type of extraction chosen, multiple SnP files may be generated, as shown in the graphics for each extraction type. Port Swapping can be performed in the Embedding/De-embedding menus.
These extraction types are for cases where an inner-cal is not possible. The network measured is assumed symmetrical, and SnP files generated using Divide-By-Two (2) schemes. An option is given to zero-out the match terms instead of fully allocating them to the outer-ports.
Procedure
1. Make sure the appropriate calibration is active.
2. Zero-out the match terms if needed.
3. Connect the network and select Perform Network Extraction to perform the extraction.
4. If the extraction is successful, follow the prompt to save the generated SnP file or files.
 
NETWORK EXTRACTION Dialog Box – Type G – 4-Port VNAs
Full Name
NETWORK EXTRACTION Dialog Box – Type G – Extract Two 4-Port Networks - Outer Cal Only Using Divide-By-Two Method - 4-Port VNAs
Previous
NETWORK EXTRACTION Dialog Box - 4-Port VNAs
Navigation
MAIN | Calibration | CALIBRATION | Cal Options | CAL OPTIONS | Network Extraction | NETWORK EXTRACTION Dialog Box | Extract Two 4-Port Networks | Type G - Outer Cal Only Using Divide-By-Two Method | NETWORK EXTRACTION [EXTRACT TWO 4-PORT NETWORKS (TYPE G)] Dialog Box
 
NETWORK EXTRACTION - EXTRACT FOUR 2-PORT NETWORKS - TYPE G - 4-Port
 
Instructions
Network Extraction provides the means of generating SnP files of networks. The generated files can than be embedded or de-embedded. Based on the type of extraction chosen, multiple SnP files may be generated, as shown in the graphics for each extraction type. Port Swapping can be performed in the Embedding/De-embedding menus.
These extraction types are for cases where an inner-cal is not possible. The network measured is assumed symmetrical, and SnP files generated using Divide by 2 schemes. An option is given to zero-out the match terms instead of fully allocating them to the outer-ports.
Procedure
1. Make sure the appropriate calibration is active.
2. Zero-out the match terms if needed.
3. Connect the network and select Perform Network Extraction to perform the extraction.
4. If the extraction is successful, follow the prompt to save the generated SnP file or files.
 
 
CAL KIT/AUTOCAL Menu - 4-Port VNAs
Use the CAL KIT/AUTOCAL menu to install, save, and restore calibration kit characterization files between an external memory device, the instrument firmware, and a hard drive on the instrument or on a network.
Full Name
FULL MANUAL CALIBRATION KIT / AUTOMATIC CALIBRATOR (AUTOCAL) Menu
Previous
CALIBRATION Menu - 2-Port VNAs
Navigation
MAIN | Calibration | CALIBRATION | Cal Kit/AutoCal Characterization | CAL KIT/AUTOCAL
 
CAL KIT/AUTOCAL Menu - 4-Port VNAs
Install Kit/Charac.
Select loads the Calibration Kit file or AutoCal Characterization file from the hard drive or external memory device into the VNA firmware through the INSTALL (AutoCal Characterization/Cal Kit File) dialog box.
Save Kit/Charac.
Select saves the Cal Kit or AutoCal Characterization file from the firmware to the location of choice (typically the instrument hard drive) for later use through the SAVE (AutoCal Characterization/Cal Kit) File dialog box.
Display/Edit Info
Select displays the CAL KIT INFO dialog box which shows parametric information about the calibration kit and allows user edits of the values.
Restore Cal Kit Default Coef.
Select displays the RESTORE DEFAULT COEF dialog box.
 
INSTALL (AUTOCAL CHARAC./CAL KIT) Dialog Box - 4-Port VNAs
Use the INSTALL (AUTOCAL CHARACTERIZATION/CAL KIT) dialog box to install the calibration kit coefficients file or AutoCal Characterization files in the instrument firmware for subsequent use. A recommended best practice is to keep the cal kit serial number as part of the file name.
Full Name
INSTALL (AUTOCAL CHARACTERIZATION/CAL KIT PARAMETERS FILES) Dialog Box
Previous
CAL KIT/AUTOCAL Menu - 4-Port VNAs
Navigation
MAIN | Calibration | CALIBRATION | Cal Kit/AutoCal Characterization | CAL KIT/AUTOCAL | Install Kit/Charac | INSTALL (AUTOCAL CHARACTERIZATION/CAL KIT) Dialog Box
 
INSTALL (AUTOCAL CHARACTERIZATION/CAL KIT) Dialog Box 
Instructions
1. In the Select File Type area, select either the AutoCal Characterization or the Cal Kit radio button.
2. In the Open field, either enter the file name or click Browse to navigate manually.
3. If AutoCal Characterization was selected, the file type will be an AutoCal Characterization ACD file.
4. If Cal Kit was selected, the file type will be a Cal Kit Coefficient CCF file.
5. Click Open to load the file or Cancel to return to the menu.
 
 
SAVE (AUTOCAL CHARAC./CAL KIT) Dialog Box - 4-Port VNAs
Use the SAVE (AUTOCAL CHARACTERIZATION/CAL KIT) dialog to save a Calibration Kit Coefficients file or an AutoCal Characterization file from the VNA firmware to external location such as the instrument hard drive, a network drive, or an external memory device. This is useful for storing in the instrument hard drive multiple files from available cal kits or AutoCal modules.
An alternate method is to use a Windows program such as File Manager to copy the file from the supplied USB memory device onto the hard drive. In that case, we recommend using the default locations mentioned later in this paragraph.
Previous
CAL KIT/AUTOCAL Menu - 4-Port VNAs
Navigation
MAIN | Calibration | CALIBRATION | Cal Kit/AutoCal Characterization | CAL KIT/AUTOCAL | Save Kit/Charac | SAVE (AUTOCAL CHARACTERIZATION/CAL KIT) Dialog Box
 
SAVE (AUTOCAL CHARACTERIZATION/CAL KIT FILE Dialog Box
1. AutoCal Characterization radio button selected.
2. Cal Kit radio buttons selected.
Save Dialog Button and Field Availability
The available buttons and fields on the dialog depend on the radio button selection in the Select File Type area.
If AutoCal Characterization is selected, no other buttons/fields are available and instructions are immediately below.
If Cal Kit is selected, the For Cal Kit Only area becomes available with fields for Line Type, Cal Method, and Cal Kit Name. Instructions are in the next sub-section following.
Instructions for AutoCal Characterization File Type
1. Click OK to save the AutoCal Characterization file.
Click Cancel to stop the save.
2. The SAVE AS (AUTOCAL CHARACTERIZATION ACD FILE) dialog box appears.
3. Navigate to the required storage location:
C:\AnritsuVNA\AutoCal is recommended.
4. Click Save. The system auto-returns to the Cal Kit/AutoCal menu.
Instructions for Cal Kit File Types
If Cal Kit is selected, the fields of Line Type, Cal Method, and Cal Kit Name are available (shown at right in Figure: SAVE (AUTOCAL CHARACTERIZATION/CAL KIT FILE Dialog Box above) with the values in drop-down menus.
1. In the Line Type field, select a type value:
Coaxial
Non-Dispersive
Waveguide
2. In the Cal Method field, select a method value:
SOLT/SOLR
SSLT
SSST
3. In the Cal Kit Name field, select a calibration kit characterization file:
0.8 mm Conn
W1-Conn
V-Conn
K-Conn
GPC-3.5
SMA
N-Conn
N-Conn (75)
GPC-7
7/16
TNC
User-Defined1 through User-Defined32
4. Click OK to save the AutoCal Characterization file.
Click Cancel to stop the save.
5. The SAVE AS (CAL KIT COEFFICIENT CCF FILE) dialog box appears.
6. Navigate to the required storage location:
C:\AnritsuVNA\Data is recommended.
7. Click Save. The system auto-returns to the CAL KIT/AUTOCAL menu.
Click Cancel to abort the save and return to the CAL KIT/AUTOCAL menu.
 
CAL KIT INFO Dialog Box - 4-Port VNAs
Use the CAL KIT INFO dialog box to review the available instrument calibration kit information.
Previous
CAL KIT/AUTOCAL Menu - 4-Port VNAs
Navigation
MAIN | Calibration | CALIBRATION | Cal Kit/AutoCal Characterization | CAL KIT/AUTOCAL | Display/Edit Info | CAL KIT INFO Dialog Box
 
CAL KIT INFO Dialog Box
1. At left, controls for AutoCal Characterization files.
2. At right, controls for Cal Kit files.
 
Button and Field Availability
The available buttons and fields on the dialog depend on the radio button selection in the Select File Type area.
If AutoCal Characterization is selected, no other buttons/fields are available.
If Cal Kit is selected, the For Cal Kit Only area becomes available with fields for Line Type, Cal Method, and Select Cal Standard.
Instructions
1. Select AutoCal Characterization or Cal Kit as required.
2. If Cal Kit was selected above, in the For Cal Kit Only area, select a Line Type value:
Coaxial
Non-Dispersive
Waveguide
3. Select a Cal Method value:
SOLT/SOLR
SSLT
SSST
4. Select a Cal Standard value:
0.8 mm Conn (M)
0.8 mm Conn (F)
W1-Conn (M)
W1-Conn (F)
V-Conn (M)
V-Conn (F)
K-Conn (M)
K-Conn (F)
2.4 mm (M)
2.4 mm (F)
GPC-3.5 (M)
GPC-3.5 (F)
SMA (M)
SMA (F)
N-Conn (M)
N-Conn (F)
N-Conn (75) (M)
N-Conn (75) (F)
GPC-7
7/16 (M)
7/16 (F)
TNC (M)
TNC (F)
User-Defined1(M) through User-Defined32 (M)
User-Defined1(F) through User-Defined32 (F)
5. Click Display/Edit.
The STANDARD INFO read-only dialog box appears.
Note that the dialog box title and content fields are dependent on the selections made in the steps above for Cal Method, Line Type, and Connector Type.
For example, if:
Calibration type is set to Full 2 Port (12 Term Cal)
Calibration method is set to SOLT/SOLR
Line type is set to Coaxial
Connector type is set to V-Conn (M).
The read-only dialog box is titled:
STANDARD INFO (SOLT/R) STANDARD LABEL (V-Conn (M)) (shown below).
STANDARD INFO (SOLT/R) Standard Label (V-Conn (M)) Dialog Box
 
6. Click OK to close the read-only dialog box.
 
RESTORE DEFAULT COEF. Dialog Box - 4-Port VNAs
Use the RESTORE DEFAULT COEF. dialog box to restore firmware-stored Cal Kit Coefficients fields back to their default coefficients. For best performance, either install the cal kit coefficients file supplied with your cal kit, or enter your user-defined coefficients before starting this procedure. The restore function is not available to AutoCal kits as they do not have restorable characterization data.
Full Name
FULL RESTORE DEFAULT COEFFICIENTS Dialog Box
Previous
CAL KIT/AUTOCAL Menu - 4-Port VNAs
Navigation
MAIN | Calibration | CALIBRATION | Cal Kit/AutoCal Characterization | CAL KIT/AUTOCAL | Restore Cal Kit Default Coef | RESTORE DEFAULT COEF Dialog Box
 
RESTORE DEFAULT COEF. (RESTORE DEFAULT COEFFICIENTS) Dialog Box 
Instructions Restore Default Calibration Coefficients
Use this dialog to restore factory coefficients to available calibration kits.
1. Select the Line Type as required.
2. Select the Calibration Kits as required to be restored.
3. Click OK.
 
 
Available Selections
The table below shows the available calibration kits in the Select Cal Kit field of the RESTORE DEFAULT COEFFICIENTS dialog box. The available kits depend on the input combination selected for Line Type Media and Cal Method
Calibration Kit Availability in the RESTORE DEFAULT COEF. Dialog Box 
LINE TYPE Media Setting
CAL METHOD Setting
Available Calibration Kits
Coaxial
SOLT/SOLR
0.8 mm Conn, W1-Conn, V-Conn, K-Conn, 2.4 mm, 2.4 mm V, GPC-3.5, SMA, N-Conn, N-Conn (75), GPC-7, 7/16, TNC
SSLT
0.8 mm Conn, W1-Conn
SSST
0.8 mm Conn, W1-Conn
Non-Dispersive
SOLT/SOLR
0.8 mm Conn, W1-Conn, V-Conn, K-Conn, 2.4 mm, GPC-3.5, SMA, N-Conn, N-Conn (75), GPC-7, 7/16, TNC
SSLT
0.8 mm Conn, W1-Conn
SSST
0.8 mm Conn, W1-Conn
Microstrip
SOLT/SOLR
10 Mil Kit, 15 Mil Kit, 25 Mil Kit
SSLT
10 Mil Kit, 15 Mil Kit, 25 Mil Kit
SSST
10 Mil Kit, 15 Mil Kit, 25 Mil Kit
Waveguide
SOLT/SOLR
No selections available
SSLT
WR10, WR12, WR15
SSST
No selections available