1. MAIN menu 2. CALIBRATION Menu 3. CALIBRATE Menu 4. THRU UPDATE Menu | 5. AUTOCAL PORTS Menu 6. MANUAL CAL Menu 7. CAL KIT/AUTOCAL Menu for characterization file management. CAL OPTIONS Menu with links to calibration utilities menus and dialog boxes. |
Cal Status (Off/On) The Cal Status toggle button displays the calibration status based on the last calibration run. If not calibrated, Cal Status cannot be changed from OFF and a calibration run must be successfully completed before the status can be changed. If calibrated, Cal Status status can be toggled between OFF and ON. If ON, the Channel Status bar at the bottom of the display area shows a status of CORR in green. Calibrate Use the Calibrate button to start the AutoCal or manual calibration process. Options on sub-menus allow for selection of automatic or manual calibration, calibration type, calibration method, line type and other calibration parameters. Select displays the CALIBRATE menu. Thru Update Thru update is a calibration refreshing technique where the user connects a thru line and quickly refreshes the transmission tracking and load match terms without the time and complexity of a full calibration run. The thru update is essentially a one-step refresh calibration for Full 2 Port and 1 Path-2 Port calibrations. Select displays the THRU UPDATE menu. If the Thru Update button is not available, a Full 4-Port, Full 3-Port, Full 2-Port, or 1 Path-2 Port calibration is not active. Complete one of the required calibrations to activate the Thru Update button. Cal Kit/AutoCal Characterization Use this menu to save, load, and recall characterization files for AutoCal and manual calibration kits. Select displays the CAL KIT/AUTOCAL menu. Cal Options Select displays the CAL OPTIONS menu. |
Existing Cal Setup Restores the setup parameters from the last successful calibration procedure, whether an AutoCal or manual calibration. All menu and dialog box settings are returned to their prior settings and the operator can proceed with the calibration procedure as soon as the necessary external device connections are complete. AutoCal Select displays the AutoCal menu. Manual Cal Select displays the Manual Calibration menu. |
1. Starting THRU UPDATE calibration menu for 4-Port VNA system. The Done button is unavailable. The number of thrus and the thru pairs presented depends on the user configuration before the calibration and thru update. | 2. At right, the completed THRU UPDATE calibration menu for 4-port VNA system on right with completion check marks and Done button available. |