VectorStar™ MS4640A Series Microwave Vector Network Analyzers : Application Menus - Multiple Source : Multiple and External Source Control
 
Multiple and External Source Control
 
MULTIPLE SOURCE Menu
Prerequisites
 
Previous
APPLICATION Menu
Navigation
MAIN | Application | APPLICATION | Multiple Source Setup | MULTIPLE SOURCE
 
MULTIPLE SOURCE Menu 
 
MULTIPLE SOURCE Menu Source Tableau
When the MULTIPLE SOURCE menu is selected, the MULTIPLE SOURCE Tableau Area appears at the bottom of the main display area.
Multiple Source (Off/Off)
Select toggles multiple source capability off or on.
Add Band
Select adds a frequency band to the multiple source tableau at the bottom of the display.
Multiple Source Tableau Dialog
Delete Band
Select a row in the frequency band tableau at the bottom of the display. Click the Delete Band button to permanently delete it.
Multiple Source Tableau Dialog
Clear All Bands
Select deletes all band entries in the tableau.
Multiple Source Tableau Dialog
Done Editing
If no changes have been made to the frequency band settings, this button is unavailable.
Once changes are made, the button becomes available.
When all changes are complete, select the Done Editing button to apply the frequency band changes (additions, edits, and deletions) to the current running measurement and its attached test sets and modules.
After the change have been applied, the button again becomes unavailable.
If the button is not selected, the changes are not applied to the instrument.
Ext. Src Control
Select displays the EXT. SRC CONTROL menu.
EXT. SRC CONTROL Menu
Ext. Src Fast Trigger (Enabled/Disabled)
Select displays the EXT SRC FAST TRIGGER dialog box where the trigger mode can be enabled or disabled. The button shows the current fast trigger mode.
EXTERNAL SOURCE FAST TRIGGER MODE Dialog Box
Ext. Mod. Ctrl.
The button displays the currently broadband setting as either OFF, mmWave [3738], mmWave [3739], Broadband to 125 GHz, E-Band 56 GHz to 96 GHz, or W-Band 65 GHz to 110 GHz.
Select displays the EXTERNAL MODULE CTRL dialog box. The dialog box allows the module type to be set as OFF, mmWave [3738], mmWave [3739], Broadband to 125 GHz, Broadband to 145 GHz, E-Band 56 GHz to 96 GHz, or W-Band 65 GHz to 110 GHz which also depends on the connected equipment and installed options. If any configuration other than OFF is selected, additional configuration parameters are available for each defined band. The dialog box content depends on the broadband mode and the number of bands selected. Sample dialogs for some configurations are shown below.
EXTERNAL MODULE CTRL Dialog Box Overview
EXTERNAL MODULE CTRL Dialog Box – Module Type = OFF
EXTERNAL MODULE CTRL Dialog Box – Module Type = mm-Wave (3738)
EXTERNAL MODULE CTRL Dialog Box – Module Type = mm-Wave (3739)
EXTERNAL MODULE CTRL Dialog Box – Module Type = BB/mm-Wave (3739)
Phase Inversion (Off/On)
Select toggles the phase inversion function off and on.
 
Multiple Source Tableau Dialog
The multiple source tableau is available at the bottom of the screen when the MULTIPLE SOURCE menu is selected. typical example is shown below.
MULTIPLE SOURCE TABLEAU Dialog - Sample
 
Band Tableau Controls
Bands are added by selecting the Add Band button on the MULTIPLE SOURCE menu. Up to 50 Bands can be added. Each band can be defined for any combination of the following sources:
Internal Source
External Source 1
External Source 2
External Source 3
External Source 4
External Source 1
Receiver
Receiver Source
General Operation
The following general operation notes and requirements apply:
Tabular field data can be entered in any sequence.
If invalid data is entered, a warning icon appears in the left margin of the table. Selecting the Done Editing button with invalid data results in a warning dialog.
Read-only tabular elements are distinguished from editable contents.
Table focus remains on the last added/ deleted band.
Data is not applied to the instrument and its attached equipment until the Done Editing button is selected. When successful, the button again becomes unavailable.
CW OFF
Each source is defaulted as CW off. A check box can be selected to enable CW for each source. If CW is not selected, the source equation is:
Where:
M = Multiplier
D = Divisor
F = Frequency in user-defined units of Hz, kHz, MHz, or GHz. This is the current instrument frequency within the band being defined.
OS = Offset frequency in Hertz
CW ON
If CW is selected, the source equation is:
Where:
M = Multiplier
D = Divisor
OS = Offset frequency in Hertz
Band Management
Band management is done by using the Add Band, Delete Band, and Clear All Bands button. After all bands are configured, select Done Editing button to apply the setting to the instrument and its attached equipment.
Saving/Recalling Band Configurations
A multiple source band configuration can be saved by using MENU BAR | File | Save Setup. On the SAVE SETUP dialog box, the setup can be saved as a:
Active Channel Setup and Calibration CHX file
Active Channel Setup STX file
All Channel Setup and Calibration CHA file
All Channel Setup STA file
Recall previously saved setups by using MENU BAR | File | Recall Setup.
File operations are duplicated on the FILE menu at MAIN | File | FILE.
 
 
EXT. SRC CONTROL Menu
Full Name
EXTERNAL SOURCE CONTROL Menu
Previous
MULTIPLE SOURCE Menu
Navigation
MAIN | Application | APPLICATION | Multiple Source Setup | MULTIPLE SOURCE | Ext. Src Control | EXT. SRC CONTROL
 
EXT. SRC CONTROL (EXTERNAL SOURCE CONTROL) Menu 
 
EXT. SRC CONTROL Menu Note
The RF of External Sources is not automatically turned on. See the installation and configuration procedure below this figure.
Ext. Src1 State [Inactive/Active]
For External Source 1, select toggles the State from active to inactive.
If the State is Inactive and Unavailable, either:
No external source is connected, or
The GPIB address set by the VNA instrument for External Source 1 differs from the GPIB address set on the external source instrument.
If the State is Inactive and Available, an external source is connected with a GPIB address that matches that set in the external source instrument.
Select toggles the Ext. Src1 State button between states of Inactive and Active.
When the State is Active, the external source is controlled by the VNA instrument via the GPIB connection.
See the procedure section below this figure a more detailed description.
If in Modular BB mode, each external source is turned on/off manually from this menu.
If in BB/MM Wave mode, and external source is set to ON, Source 1 and Source 2 are automatically set to Active.
Ext. Src2 State [Inactive/Active]
For External Source 2, select toggles the State from active to inactive.
See the description above in the Ext. Src1 State button.
Ext. Src3 State [Inactive/Active]
For External Source 3, select toggles the State from inactive to active.
See the description above in the Ext. Src1 State button.
Ext. Src4 State [Inactive/Active]
For External Source 4, select toggles the State from inactive to active.
See the description above in the Ext. Src1 State button.
External Source State Buttons
This description applies to these buttons described in the figure above:
(Ext. Src1) State [Inactive/Active] button
(Ext. Src2) State [Inactive/Active] button
(Ext. Src3) State [Inactive/Active] button
(Ext. Src4) State [Inactive/Active] button
In this section, these buttons are referred to as the External Source State buttons.
External Source State Buttons Inactive and Unavailable
If the External Source State buttons are Inactive and Unavailable (grayed out), either no external sources are connected to the instrument, or the external sources have different GPIB addresses from those set on the VNA instrument. Once one or more external sources are connected with matching GPIB addresses, the buttons change to Inactive and Available (not grayed out).
Connecting an External Source
Connect from one (1) to four (4) external sources via the rear panel Dedicated GPIB Port D-24 (f) Connector.
This GPIB port is used when the VNA instrument controls external instruments such as power meters, frequency counters, and in this case, external sources.
Up to four (4) external sources can be connected to the instrument.
Note that each external instrument and the VNA must have a unique GPIB address with no duplicates. The factory-default GPIB addresses for the four external sources and other likely external instruments are:
External Source 1 = GPIB Address 4
External Source 2 = GPIB Address 5
External Source 3 = GPIB Address 2
External Source 4 = GPIB Address 3
Power Meter = GPIB Address 13
W-Band Power Meter = GPIB Address 15
Frequency Counter = GPIB Address 7
MN4690B Series Multiport Test Set = GPIB Address 16
Frequency Counter = GPIB Address 7
The VNA Instrument itself = GPIB Address 6
The default addresses for external sources can be changed at the EXT. SRC. ADDR. (EXTERNAL SOURCE ADDRESS) menu
EXT. SRC ADDR. Menu
MAIN | System | SYSTEM | Remote Interface | REMOTE INTER. | Ext. Sources | EXT. SRC. ADDR.
The default addresses for the instrument itself, the external power meter, and the external frequency counter can be changed at the REMOTE INTER. menu.
REMOTE INTER. Menu
MAIN | System | SYSTEM | Remote Interface | REMOTE INTER.
Only Anritsu external sources such as the MG3696A Signal Generator or the MS3695B Signal Generator can be connected.
The external sources must be off during the connection process. The VNA instrument can be running.
Connect the GPIB cables between the instrument and the external sources.
Note 
Best practices recommend no more than two (2) GPIB cable connectors be connected to any one GPIB port. Once connected, do not bend the connectors as this might damage the instrument Dedicated GPIB Port Connector.
Matching GPIB Addresses Between the VNA and the External Source
The GPIB addresses set on the VNA instrument must match those set on each external source.
If the GPIB addresses set on the VNA match those on the external source, the appropriate External Source State button should change from unavailable (grayed out) to available. Once an External Source State button is available, it can be used to toggle the external source between Inactive and Active.
Manual Refresh Required for the EXT. SRC. CONTROL Menu
Note that the EXT. SRC. CONTROL menu does not refresh itself if address changes are made in any of the external sources. If in doubt about the state of connected external sources, click the Back button at the bottom of the display, and then re-enter the menu.
For example, if a single external source with a GPIB address of 4 is connected to the VNA, the (Ext. Src1) State [Inactive/Active] button is available and can toggle the source as Active or Inactive.
The user changes the GPIB address of the external source from 4 to 5.
The current EXT. SRC. CONTROL menu will still show the (Ext. Src1) State [Inactive/Active] button as available.
To verify the currently connected external sources, select Back to exit the EXT. SRC ADDR. menu to the REMOTE INTER. menu and then select again the Ext. Sources button to re-enter the EXT. SRC. CONTROL menu.
In this example, the EXT. SRC. CONTROL menu will now show:
The (Ext. Src1) State [Inactive/Active] button as unavailable
The (Ext. Src2) State [Inactive/Active] button as available
All external source control buttons follow this behavior.
EXTERNAL SOURCE FAST TRIGGER MODE Dialog Box
Previous
This menu can be accessed from multiple locations.
MULTIPLE SOURCE Menu
3738 SETUP Menu
Navigation
MAIN | Application | APPLICATION | Multiple Source Setup | MULTIPLE SOURCE | Ext. Src Fast Trigger | EXTERNAL SOURCE FAST TRIGGER MODE Dialog Box
MAIN | Application | APPLICATION | mmWave (3738 Setup) | 3738 SETUP | Ext. Src Fast Trigger | EXTERNAL SOURCE FAST TRIGGER MODE Dialog Box
MAIN | Application | APPLICATION | BB/mm Wave (3738 Setup) | 3738 SETUP | Ext. Src Fast Trigger | EXTERNAL SOURCE FAST TRIGGER MODE Dialog Box
 
EXTERNAL SOURCE FAST TRIGGER MODE Dialog Box
 
 
Instructions
This Fast External Source Trigger mode takes advantage of faster TTL hand-shaking instead of traditional GPIB. It is only available on sources such as the Anritsu MG37020x Series Signal Generators that have this capability. All external sources must use the same triggering means, either all GPIB or all TTL hand-shaking.
Note 
Ensure that the BNC cable is connected properly before enabling this mode to prevent source lock errors. The length of the BNC cables can vary, while the BNC T-Connectors should all be the same type from the same manufacturer. Select OK when done.
In the Fast Trigger Mode area, select Enabled or Disabled, and then select OK. The general configuration is shown in the figure below.
External Source Fast Trigger BNC Cable Connections
1. VectorStar MS4640A Series VNA.
2. VNA Trigger Out – Rear Panel External Trigger Output (Ext Trig Out) Port.
3. BNC T-Connectors – For this application, the best practice recommendation is to use the same model number BNC T-connectors from the same manufacturer. The BNC cables do not need to be identical length nor do they need to be from the same manufacturer.
4. Anritsu MG3702xX Signal Generators – Up to four (4) signal generators can be connected.
5. MS37020x Rear Panel External Trigger (External Trigger In) Port.
6. MS37020x Rear Panel Lock Status (Lock Status Output) Port.
7. VNA Lock Status – Rear Panel Lock Status (Lock Status In) Port.