VectorStar™ MS4640A Series Microwave Vector Network Analyzers : Application Menus - Broadband/Millimeter-Wave : Broadband/Millimeter-Wave Setup
 
Broadband/Millimeter-Wave Setup
 
3738 SETUP Menu
Previous
APPLICATION Menu
Navigation
The navigation path changes depending on the instrument model and the equipped options. For MS4642A, 44A, and 45A with Option 007 and Option 082/083:
MAIN | Application | APPLICATION | mmWave (3738 Setup) | 3738 SETUP
For MS4647A with Options 007 and 082/083 or Options 007 and 080/081.
MAIN | Application | APPLICATION | BB/mmWave (3738 Setup) | 3738 SETUP
 
3738 SETUP Menu Variants
1. 3738 SETUP menu for MS4642A, 44A, and 45A VNAs equipped with Receiver Offset Option 007 only.
This is six (6) button menu variant.
Option 080/081 is not compatible with 42A, 44A, and 47A VNAs.
2. 3738 SETUP menu for MS4647A VNA.
VNA equipped with Option 007 only.
This is the seven (7) button menu variant as in #3 and #4.
3. 3738A SETUP menu for MS4647A VNA
VNA equipped with Option 007 and
Option 082/083 – 3738A Test Set with Millimeter-Wave Modules.
This is the same seven (7) button menu variant as in #2 and #4.
4. 3738A SETUP menu for MS4647A VNA
VNA equipped with Option 007 and
Option 080/081 3739x Test Set with Broadband/Millimeter-Wave Modules.
This is the same seven (7) button menu variant as in #2 and #3.
 
BB/mmWave
Select toggles broadband/millimeter-wave capability on and off. The broadband state is shown in the button field.
Receiver Configuration Button Group Start
Depending on the instrument model and equipped options, the receiver configuration button group allows only one type to be selected.
Broadband
Select sets the receiver configuration as Broadband.
mm-OML
Select sets the receiver configuration as mm-OML using OML brand millimeter-wave modules.
mm-VDI
Select sets the receiver configuration as mm-VDI using VDI brand millimeter-wave modules.
Receiver Configuration Button Group End
The mm-VDI button is the last button in the receiver configuration button group.
External Module
Select displays the EXTERNAL MODULE SELECTION dialog box.
EXTERNAL MODULE SELECTION Dialog Box
Ext. Source Eqn
Select displays the MODIFY EXTERNAL SOURCE EQUATIONS dialog box.
MODIFY EXTERNAL SOURCE EQUATIONS Dialog Box
Ext. Src Fast Trigger
Select displays the EXTERNAL SOURCE FAST TRIGGER MODE dialog box.
EXTERNAL SOURCE FAST TRIGGER MODE Dialog Box
 
3739 SETUP Menu
Previous
APPLICATION Menu
Navigation
The navigation path changes depending on the instrument model and the equipped options.
For MS4642A, 44A, and 45A with Option 007 and Option 082/083:
MAIN | Application | APPLICATION | mmWave (3739 Setup) | 3739 SETUP
For MS4647A with Options 007 and 082/083 or Options 007 and 080/081:
MAIN | Application | APPLICATION | BB/mmWave (3739 Setup) | 3739 SETUP
 
3739 SETUP Menu – Variants
1. The 3739 SETUP Menu for MS4642A, 44A, 45A, and 47A VNAs with Option 007 – Receiver Offset and Option 082/083 – Millimeter-Wave Module Connectivity Capability.
The menu has eight (8) buttons.
The first button is mmWave (ON/OFF).
Option 082/083 and Option 081/082 are mutually exclusive and only one can be installed on 47A VNAs.
2. The 3739 SETUP Menu for MS4647A VNAs with Option 007 and Option 080/081Broadband/Millimeter-Wave Modules Connectivity Capability.
The menu has ten (10) buttons.
The first button name is changed to BB/mmWave (ON/OFF).
Note that Option 080/081 is only available for MS4647A VNAs.
Receiver Configuration Button Group
The receiver configuration button group changes depending on the VNA model and the equipped options:
For MS4642A, 44A, 45A, and 47A45A VNAs with Option 082/083, E-Band 56 GHz to 95 GHz, W-Band 65 GHz to 110 GHz, mm-OML, and mm-VDI buttons are members of a button selection group where only one button can be selected.
For MS4647A VNAs with Option 082/083, the Broadband to 125 GHz, E-Band 56 GHz to 95 GHz, W-Band 65 GHz to 110 GHz, mm-OML, and mm-VDI buttons are members of a button selection group where only one button can be selected.
mmWave
Part of the Receiver Configuration button selection group. Only on MS4642A, MS4644A, MS4745A, and MS4647A VNAs equipped with Option 082/083. Toggles the millimeter-wave capabilities on and off.
BB/mmWave
Part of the Receiver Configuration button selection group. Only on MS4647A VNAs equipped with Option 080/081. Toggles the broadband/millimeter-wave capabilities on and off.
Broadband to 145 GHz
Part of the Receiver Configuration button selection group. Only on MS4647A VNAs equipped with Option 080/081. If this is selected, the External Module and Ext. Source Eqn buttons below are not available.
Broadband to 125 GHz
Part of the Receiver Configuration button selection group. Only on MS4647A VNAs equipped with Option 080/081. If this is selected, the External Module and Ext. Source Eqn buttons below are not available.
E-Band 56 GHz to 95 GHz
Part of the Receiver Configuration button selection group. Only on MS4640A Series VNAs equipped with Option 082/083 or 081/082. If this is selected, the External Module and Ext. Source Eqn buttons below are not available.
W-Band 65 GHz to 110 GHz
Part of the Receiver Configuration button selection group. Only on MS4640A Series VNAs equipped with Option 082/083. If this is selected, the External Module and Ext. Source Eqn buttons below are not available.
mm-OML
Part of the Receiver Configuration button selection group. Only on MS4640A Series VNAs equipped with Option 082/083 or 081/082. Select enables the External Module and the Ext. Source Eqn buttons. The External Module button links to a module selection dialog and configures the VNA to use OML millimeter-wave modules. The module selection is done using the External Module button described below to access the EXTERNAL MODULE SELECTION dialog box. The Ext. Source Eqn (External Source Equation) button links to the MODIFY EXTERNAL SOURCE EQUATIONS dialog box.
mm-VDI
Part of the Receiver Configuration button selection group. Only on MS4640A Series VNAs equipped with Option 082/083 or 081/082. Select enables the External Module button which links to a module selection dialog and configures the VNA to use VDI millimeter-wave modules. The module selection is done using the External Module button below to access the EXTERNAL MODULE SELECTION dialog box.
Receiver Configuration Button Selection End
The mm-VDI button is the end of the Receiver Configuration button selection group.
External Module
Select displays the EXTERNAL MODULE SELECTION dialog box where separate control tabs allow configuration of either OML or VDI millimeter-wave modules. Once a module has been selected, the module name appears in the button field.
EXTERNAL MODULE SELECTION Dialog Box.
Ext. Source Eqn
Select displays the MODIFY EXTERNAL SOURCE EQUATIONS dialog box for OML or VDI module selected.
MODIFY EXTERNAL SOURCE EQUATIONS Dialog Box.
mmWave ALC Cal
Select displays the MMWAVE ALC CAL (Millimeter-Wave Automatic Level Control Calibration) menu.
mmWAVE ALC CAL Menu.
MODIFY EXTERNAL SOURCE EQUATIONS Dialog Box
Previous
The previous menu varies depending on the instrument model.
MS4642A or MS4644A VNAs: 3738 SETUP Menu
MS4645A or MS4647A VNAs: 3739 SETUP Menu
Navigation
MAIN | Application | APPLICATION | mmWave (3738 Setup) | 3738 SETUP | Ext. Source Eqn | MODIFY EXTERNAL SOURCE EQUATIONS Dialog Box
MAIN | Application | APPLICATION | BB/mm Wave (3738 Setup | 3738 SETUP | Ext. Source Eqn | MODIFY EXTERNAL SOURCE EQUATIONS Dialog Box
MAIN | Application | APPLICATION | mmWave (3739 Setup) | 3739 SETUP | Ext. Source Eqn | MODIFY EXTERNAL SOURCE EQUATIONS Dialog Box
MAIN | Application | APPLICATION | BB/mm Wave (3739 Setup | 3739 SETUP | Ext. Source Eqn | MODIFY EXTERNAL SOURCE EQUATIONS Dialog Box
MODIFY EXTERNAL SOURCE EQUATIONS Dialog Box
1. 3738 SETUP Menu and setup dialog for Broadband External Source Equations at top left.
Note the minus sign (“–”) in the frequency portion of the equation.
2. 3738 SETUP Menus and setup dialog for Millimeter-Wave External Source Equations at bottom left.
Note the minus sign (“–”) in the frequency portion of the equation.
3. 3739 SETUP Menu and setup dialog for Millimeter-Wave External Source Equations at bottom right.
Note the plus sign (“+”) in the frequency portion of the equation.
Also, there is no corresponding dialog with a breakpoint frequency as shown above left at #1.
Broadband Source Equation
For broadband source equations, enter the following parameters:
Start Frequency (GHz)
Stop Frequency (GHz)
Broadband Break Frequency (GHz)
LO External Source 1 Multiplier (Integer)
LO External Source 1 Divisor (Integer)
RF External Source 2 Multiplier (Integer)
RF External Source 2 Divisor (Integer
Millimeter Wave Source Equation
For the millimeter wave source equations, the same parameters above are entered omitting the Broadband Break Frequency.
mmWAVE ALC CAL Menu
Full Name
Millimeter-Wave Automatic Level Control Calibration Menu
Previous
3739 SETUP Menu
Navigation
There are multiple navigation paths to this menu.
MAIN | Application | APPLICATION | mmWave (3739 Setup) | 3739 SETUP | mmWave ALC Cal | mmWAVE ALC CAL
MAIN | Application | APPLICATION | BB/mmWave (3739 Setup) | 3739 SETUP | mmWave ALC Cal | mmWAVE ALC CAL
 
mmWAVE ALC CAL Menu – Millimeter-Wave Automatic Level Control Calibration
 
Type of Leveling Button Group Start
The IF, RF, and VNA buttons at the top of the menu form a button selection group where only one type of leveling can be selected.
IF
If the IF type of leveling is selected, it provides the advantages of RF leveling described below plus the benefit of greater power control range and a lower minimum leveled power (at least 20 dB lower typically). As one approaches a 100 MHz difference between source and receiver frequencies, the power control range will start to decrease as will power control accuracy. The use of IF leveling is not recommended for source-receiver frequency differences greater than 100 MHz.
RF
If the RF type of leveling is selected, it provides a leveled RF output power with improved source match, protection against over powering the DUT, and insuring that the DUT is operating in its designated power range.
VNA
If the VNA type of leveling is selected, it turns the RF or IF ALC off and the ALC accuracy is set as that for the standalone VNA instrument.
Type of Leveling Button Group End
The VNA button is the end of the Type of Leveling button group.
Perform Cal
Select starts the calibration type selected above (IF, RF, or VNA) and displays the mmWAVE ALC CAL dialog box. A suffix is added to the dialog name denoting the selected mm-Wave module.
mmWAVE ALC CAL (MM MODULE NAME) Dialog Box
Save IF Leveling Cal
Select displays a SAVE IF LEVELING dialog box. Navigate to the required folder to save the calibration.
Save RF Leveling Cal
Select displays a SAVE RF LEVELING dialog box. Navigate to the required folder to save the calibration.
Recall IF Leveling Cal
Select displays a RECALL IF LEVELING dialog box. Navigate to the required folder to recall the calibration.
Recall RF Leveling Cal
Select displays a RECALL RF LEVELING dialog box. Navigate to the required folder to recall the calibration.
mmWAVE ALC CAL (MM MODULE NAME) Dialog Box
Full Name
Millimeter-Wave Automatic Level Control (Typical module name such as WR-10 Extended) Dialog Box
Previous
mmWAVE ALC CAL Menu.
Navigation
There are multiple navigation paths to this menu.
MAIN | Application | APPLICATION | mmWave (3739 Setup) | 3739 SETUP | mmWave ALC Cal | mmWAVE ALC CAL | mmWAVE ALC CAL | Perform Cal | mmWAVE ALC CAL dialog box
MAIN | Application | APPLICATION | BB/mmWave (3739 Setup) | 3739 SETUP | mmWave ALC Cal | mmWAVE ALC CAL | mmWAVE ALC CAL | Perform Cal | mmWAVE ALC CAL dialog box
 
mmWAVE ALC CAL (MODULE NAME) Dialog Box
1. mmWAVE ALC CAL (MODULE NAME) Dialog Box
2. General instructions. See below.
3. # of Entries (Number of Entries) input field. Up to 25 frequency/power combinations can be entered.
4. Select Port Radio Buttons – Select either Port 1 or Port 2.
5. Select Source Leveling as either IF or RF.
6. Frequency Input. When this has the focus, the Frequency Field Toolbar (not shown here) appears to allow value and units input for each entry row.
7. Power (dB) level for each frequency. When it has the focus, the Power (dB) field toolbar appears to allow value input for each entry row.
8. Power (dB) Field Toolbar
9. Calibration Progress Bar
10. Dialog box control buttons
 
Instructions
ALC Calibration consists of several sub calibrations that ensure accurate output power and overall system.
Requirements and Notes
The following requirements and notes apply:
Please ensure that the appropriate module is selected and enabled.
Power value entered per frequency should be the maximum power.
Maximum of 25 entries can be entered.
Instructions
1. Input the # of entries (Number of Entries).
2. Select appropriate port.
3. Enter corresponding frequencies and power. Power entered should be maximum power.
4. Select Start Cal to perform the calibration.
5. Select Abort Cal to stop the calibration.
6. Select Close to exit the dialog box to the mmWAVE ALC CAL menu.
EXTERNAL MODULE SELECTION Dialog Box
Previous
There are multiple paths to this dialog box:
3738 SETUP Menu
3739 SETUP Menu
Navigation
There are multiple paths to this dialog box depending on the instrument model number and the equipped options:
MAIN | Application | APPLICATION | mmWave (3738 Setup) | 3738 SETUP | External Module | EXTERNAL MODULE SELECTION dialog box
MAIN | Application | APPLICATION | mmWave (3739 Setup) | 3739 SETUP | External Module | EXTERNAL MODULE SELECTION dialog box
MAIN | Application | APPLICATION | BB/mmWave (3738 Setup) | 3738 SETUP | External Module | EXTERNAL MODULE SELECTION dialog box
MAIN | Application | APPLICATION | BB/mmWave (3739 Setup) | 3739 SETUP | External Module | EXTERNAL MODULE SELECTION dialog box
 
EXTERNAL MODULE SELECTION Dialog Box
1. At top, the OML tab is selected and the available OML Modules are shown. Select a module radio button configure that module.
2. At bottom, the VDI tab is selected and the available VDI Modules are shown. Select a module radio button to configure that module. Note that VDI modules have two basic types at -20 G and -40G. The module type selected here must match the module being used as the two types have slightly different frequency plans.
External Module Summary Notes
WR Numbers
The WR number abbreviation is for “rectangular waveguide” using English measurement units. The interior dimensions of waveguides have an aspect ration of 2:1, where the broadwall is twice the dimension of the narrow wall. The number following is approximately the broadwall length in mils (0.001 in) divided by 10. For example, the WR-62 waveguide has a broadwall dimension of 620 mils (0.620 in) divided by 10 which yields 62. The larger the WR Number, the lower the operating frequency range. Both OML Modules and VDI Modules reference WR and the equivalent WM Numbers.
WM Numbers
WM numbers are an emerging metric specification for equivalent and new waveguides using metric measurements. The interior opening aspect ratio follows the same 2:1 ratio. The number is derived from the broadwall dimension in micrometers (µm or 0.0001 m). For example, the WM-570 has a broadwall dimension of 570 µm. The larger the WM Number, the lower the operating frequency range. Both OML Modules and VDI Modules reference both WR and the equivalent WM Numbers.
Anritsu Numbers
Anritsu has the following five Broadband/Millimeter-Wave module numbers:
MA25300A Millimeter-Wave Module, 54 to 145 GHz – Requires a MS4647A VNA with Broadband/Millimeter-Wave Option 08x and a 3739C Test Set. Two MA25300A modules are required.
3743A Millimeter-Wave Module, 54 to 125 GHz – Requires a MS4647A VNA with Broadband/Millimeter-Wave Option 08x and a 3739x Test Set. Two 3743A modules are required.
3744A-EW/SM6597 Millimeter-Wave Module, W-Band, 65 to 110 GHz – Requires any MS464xA Series VNA equipped with Millimeter-Wave Option 08x and a Millimeter-Wave Test Set.
3744A-EE/SM6499 Millimeter-Wave Module, E-Band,56 to 95 GHz – Requires any MS464xA Series VNA equipped with Millimeter-Wave Option 08x and a Millimeter-Wave Test Set.
3743A-Rx; This module is typically ordered as an accessory to the ME7838 system. When ordering this module, the base VectorStar VNA must have Noise Figure Option 041.
Note 
When using MA25300A BB/mm-Wave modules, the 3739C Test Set is required.