Share CF & Pwr Offset All Modes Not Shared Press this submenu key to toggle the setting between All Modes and Not Shared. Select All Modes to have the current center frequency setting and power offset setting carried over when changing measurement modes. This function is not applicable to measurements that do not have a center frequency or power offset setting or to measurements in which the current center frequency or power offset setting is outside the range of the new measurement. Set Remote Access Password Press this submenu key to open the Password entry dialog (see Figure: Remote Access Password Dialog). Use the touch screen keyboard to enter the new password. Passwords may contain digits, uppercase and lowercase letters, and the symbols - _ + . (hyphen, underscore, plus, period). Press Enter to save or Esc to cancel. Setting the remote password and rebooting the instrument will lock it from Telnet and FTP connections. This means that connecting to the instrument via Master Software Tools (v2.21.1 or later), Wireless Remote Tools (WRT), or the Web Remote Control browser interface will require the set password. To unlock the instrument, remove the password by setting it to blank. The password is also removed following a Master Reset, Factory Defaults reset, or a firmware update. Power–On Press this submenu key to open the Power-On Menu. Back Press this submenu key to return to the System Options Menu. |
Warning | Do not set a remote access password when using SCPI commands to communicate with the instrument. |
Power Switch Press this submenu key to set the Spectrum Master for normal use of the On/Off button on the instrument front panel. When DC Applied Press this submenu key to set the Spectrum Master to automatically restart when DC power is applied to the External Power connector (see Figure: Test Panel Connectors for MS2720TOptions 709, 713, 720). Refer to External Power On. Back Press this submenu key to return to the System Options 2/2 Menu. |