VectorStar™ MS464xB Series Microwave Vector Network Analyzer User Interface Reference : Calibration Menus – 2-Port VNAs : AutoCal 2-Port Cal Setup – 2-Port VNAs
 
AutoCal 2-Port Cal Setup – 2-Port VNAs
AUTOCAL SETUP Menu – 2-Port Cal – 2-Port VNAs
Instrument Mode
2-Port VNA Mode
Previous
AUTOCAL PORT Menu – 2-Port VNAs
Navigation
MAIN | Calibration | CALIBRATION | Calibrate | CALIBRATE | AutoCal | AUTOCAL PORT | 2-Port Cal | AUTOCAL SETUP
AUTOCAL 2-PORT CAL SETUP Menu – 2-Port VNAs 
Modify Cal Setup
Select displays the MODIFY 2-PORT AUTOCAL SETUP dialog box. The dialog box provides control settings for Auto Sense Module Orientation, Select Cal Type, Through Setup, Adapter Removal Port, and links to the THRU INFO and AIR EQUIVALENT LENGTH CONVERSION dialog boxes.
Port Selection (Read Only)
Displays the Ports selected for the AutoCal procedure. The settings are determined in the MODIFY 2-PORT AUTOCAL SETUP dialog box.
Cal Type (Read Only)
Displays the Cal Type selected for the AutoCal procedure. The settings are determined in the MODIFY 2-PORT AUTOCAL SETUP dialog box.
Thru Type (Read Only)
Displays the Thru Type selected for the AutoCal procedure as Internal Thru, Internal Reciprocal, True Thru, or True Reciprocal. The settings are determined in the MODIFY 2-PORT AUTOCAL SETUP dialog box.
Module Orientation (Read Only)
Displays the left/right VNA Port orientation and assignment for the AutoCal procedure. Options are either:
Left=P1, Right=P2
Left=P2, Right=P1
The settings are determined in the MODIFY 2-PORT AUTOCAL SETUP dialog box.
Begin Cal (AutoCal 2-Port Cal) (2-Port VNAs)
Starts the AutoCal procedure. On-screen dialogs and prompts provide user instructions for the selected calibration procedure. When the calibration is complete, the display returns to the CALIBRATION [TR] menu where the Cal Status button is enabled and set to ON.
MODIFY 2-PORT AUTOCAL SETUP Dialog Box
Use the MODIFY 2-PORT AUTOCAL SETUP dialog box to change the calibration parameters prior to an AutoCal calibration procedure. Options include the calibration and thru types to be used. If required, an adapter removal calibration can be configured. For production installations, the left/right port sense can be automatically or manually configured.
Previous
AUTOCAL SETUP Menu – 2-Port Cal – 2-Port VNAs
Navigation
MAIN | Calibration | CALIBRATION | Calibrate | CALIBRATE | AutoCal | AUTOCAL PORT | 2-Port Cal | AUTOCAL SETUP | Modify Cal Setup | MODIFY AUTOCAL SETUP Dialog Box
 
MODIFY 2-PORT AUTOCAL SETUP Dialog Box – 2-Port VNAs) (1 of 2)
Thru Info Dialog Notes:
Thru selected allows user entries for length, line impedance, line loss, and frequency.
Reciprocal selected allows user entry for length.
S2P Thru selected provides buttons for loading, viewing, and characterization (to generate S2P files).
Auto Sense Module Orientation Check Box
See Callout #1 and Callout #2 in MODIFY 2-PORT AUTOCAL SETUP Dialog Box – 2-Port VNAs) (1 of 2).
If selected, the AutoCal Module determines the Port Left/Right assignments.
If not selected, the Left/Right manual assignment area at the bottom of the dialog box is available. The user can select a radio button for either:
Left = Port 1, Right = Port 2
Right = Port 1, Left = Port 2.
Select Cal Type Area
Select one of the following using the provided radio buttons:
Full 2 Port
When selected, the Thru Select area becomes available.
When selected, the For Adapter Removal Only area is unavailable.
See Callout #2 in Figure: MODIFY 2-PORT AUTOCAL SETUP Dialog Box – 2-Port VNAs) (1 of 2).
Adapter Removal
When selected, the Thru Select area becomes available and, if True Thru is selected, the Thru Info button appears.
Select the Thru Info button to display the THRU INFO dialog box. See Callout #3.
When selected, the For Adapter Removal Only area becomes available.
Select the Calculator icon to display the AIR EQUIVALENT LENGTH CONVERSION dialog box. Use the calculator to convert length in ps to air equivalent length in mm.
AIR EQUIVALENT LENGTH CONVERSION Dialog Box – 2-Port VNAs
See Callout #5, and Callout #6 in Figure: MODIFY 2-PORT AUTOCAL SETUP Dialog Box – 2-Port VNAs) (1 of 2).
1 Path 2 Port (1→2)
When selected, the Thru Select area becomes available.
When selected, the For Adapter Removal Only area is unavailable.
See Callout #4 in Figure: MODIFY 2-PORT AUTOCAL SETUP Dialog Box – 2-Port VNAs) (1 of 2).
1 Path 2 Port (2→1)
When selected, the Thru Select area becomes available.
When selected, the For Adapter Removal Only area is unavailable.
Through Setup Area
This area is available for all AutoCal Types. The button selection options are:
Internal Thru
If selected, the AutoCal module uses internal circuitry to determine the through values.
The Thru Info button is unavailable.
See Callout #4 in Figure: MODIFY 2-PORT AUTOCAL SETUP Dialog Box – 2-Port VNAs) (1 of 2).
Internal Reciprocal
With this choice, the internal transmissive path within the AutoCal module is employed, but it is treated as a reciprocal (i.e., the S-parameters from the characterization file are not used).
See Callout #7 in Figure: MODIFY 2-PORT AUTOCAL SETUP Dialog Box – 2-Port VNAs) (1 of 2).
True Thru
If selected, the AutoCal module will prompt the user to remove the module and connect the ports with a through line.
If selected, the Thru Info button becomes available. When clicked, the THRU INFO dialog box appears and allows input of values for Thru Length, Thru Line Impedance, Thru Line Loss, and Thru Frequency Setting.
In the Thru Line Length field, enter the line length in mm. For example, enter a value of 30 mm.
In the Line Impedance field, enter the impedance in ohms. For example, use the default value of 50 ohms.
In the Line Loss field, enter the loss as dB per mm. For example, enter a value of 0.1 dB/mm.
In the @ Frequency field, enter the working frequency in GHz. For example, enter a value of 45 GHz.
When all entries are complete, click OK to return to the Modify 2-Port AutoCal Setup dialog box.
See Callout #3 in Figure: MODIFY 2-PORT AUTOCAL SETUP Dialog Box – 2-Port VNAs) (1 of 2).
THRU INFO Dialog Box – 2-Port VNAs
For Adapter Removal Only Area
This area is only available if Adapter Removal button was selected in the Select Cal Type area above. The adapter removal controls are:
Adapter Port Select Radio Buttons
Left
Right.
Adapter Length (mm) Field
Available as either a direct entry field or using the dialog box below to calculate the parameters.
To use the calculator dialog, click the Calculator icon.
The AIR EQUIVALENT LENGTH CONVERSION dialog box appears. Entries can be typed in or incremented by clicking the field up/down arrows, or by pressing the keyboard up/down arrow keys.
Enter the adapter length in ps. For example, enter a length of 100 ps.
Enter the adapter dielectric constant. For example, enter a dielectric constant for polyethylene of 2.26.
Click the Calculate Air Equivalent Length button.
The air equivalent length in mm is: 9.9778515... or 9.9779.
Click OK and the result appears in the For Adapter Removal Only Length field.
See Callout #7 in Figure: MODIFY 2-PORT AUTOCAL SETUP Dialog Box – 2-Port VNAs) (1 of 2).
AIR EQUIVALENT LENGTH CONVERSION Dialog Box – 2-Port VNAs
Manual Port Orientation Area
This area is only available if the Auto Sense Module Orientation check box at the top of the dialog box is deselected.
If the Auto Sense check box is selected, the Left/Right buttons at the bottom of the dialog box are unavailable. The instrument identifies the left/right ports, and directs the user to the appropriate port.
See Callout #1 in Figure: MODIFY 2-PORT AUTOCAL SETUP Dialog Box – 2-Port VNAs) (1 of 2).
If the Auto Sense check box is deselected, the Left/Right buttons at the bottom of the dialog box are available. The user defines which port is “Left” and which port is “Right”. This is especially useful if the instrument is oriented differently from the work environment. Options are:
Left = Port 1, Right = Port 2
Right = Port 1, Left = Port 2
See Callout #2 in Figure: MODIFY 2-PORT AUTOCAL SETUP Dialog Box – 2-Port VNAs) (1 of 2).
Completing AutoCal Setup
When all AutoCal Setup functions are completed, click OK to return to the AutoCal menu. Click Cancel to make no changes and close the dialog box.
AUTOCAL SETUP Menu – 2-Port Cal – 2-Port VNAs