Site Masterâ„¢ S331P : Classic Mode Operation : Calibration Menu
 
Calibration Menu
Key Sequence: Calibration
Calibration Menu
Start Calibration
Press this key when you want to manually calibrate the instrument. The Calibration Setup window will open. Refer to Calibration Procedure.
Measure: Press this key to perform a measurement at the end of each calibration step, after making the connections as instructed in the Setup window.
Back: If present, press this key to go back to the previous calibration step.
Apply: This key is displayed only in the final calibration step. Press the key to apply the calibration and close the Setup window.
Cal Method: The S331P supports only OSL calibration in Classic mode.
Cal Type (Standard Flex): Press this key to toggle between Standard Cal and Flex Cal. Refer to Calibration Types for details.
Cal Info
Displays the Active Cal Settings and current instrument settings. See Figure: Calibration Information and refer to Cal Info for details.
User Cal (On Off)
Determines whether the Active Cal Settings are applied to the current measurement. With User Cal off, the factory default calibration (1-Port ReadyCal) is applied. Turning User Cal back on restores the Active Cal Settings, including frequency range.
Calibration Information