: Instrument Operation : Reference CALIBRATION : 100 MHz Calibration (Standard Reference)
 
100 MHz Calibration (Standard Reference)
To calibrate the 100 MHz internal reference:
1. To calibrate the 100 MHz internal reference, connect an external 10 MHz reference signal to the MG362x1A rear panel 10 MHz REF IN connector or a 1 PPS.
2. Press the START CALIBRATION button to display the dialog shown in Figure: 100 MHz Calibration Dialog.
3. Press CONTINUE to start the calibration. The status bar will display Reference Calibration in Progress. When complete, the CALIBRATION DATE will update.
100 MHz Calibration Dialog