:SENSe{1-16}:CORRection:COEFficient:PORT Subsystem – Simulation
The :SENSe{1-16}:CORRection:COEFficient:PORT subsystem commands are used to simulate a instrument calibration on 2-port or 4-port VNA instrument so that a set of user-defined calibration parameters can be input using the :SENSe{1-16}:CORRection:COEFficient <char>,<block> command. Note that this subsystem does not perform an actual calibration.
Calibration Simulation Subsystems
These subsystems are used to create a calibrated state in the instrument which is followed by adding the required error correction coefficients for the required calibration type. If this approach is used, each error correction coefficient is entered by separate commands. Simulated calibration subsystems are:
Calibration Type Abbreviations
The calibration abbreviations and their calibration types are:
• :1P2PF refers to a one-path two-port calibration forward direction
• :1P2PR refers to a one path two port calibration reverse direction
• :FULL1 refers to a full one port calibration
• :FULL2 refers to a full two port calibration
• :FULL3 refers to a full three port calibration
• :FULLB refers to a full one port reflection calibration on both ports
• :FULL4 refers to a full four port calibration
• :RESP1 refers to a one port response calibration
• :RESPB refers to a one port response calibration both ports
• :TFRB refers to a transmission frequency response calibration both directions
• :TFRF refers to a transmission frequency response calibration forward direction
• :TFRR refers to a transmission frequency response calibration reverse direction
Each calibration simulation type command is described in greater detail in the individual command descriptions below.
Related Query
These commands set the Calibration Type. To query which Calibration Type has been set, use the command:
:SENSe{1-16}:CORRection:COLLect:TYPe?
:SENSe{1-16}:CORRection:COEFficient <char>, <block>
:SENSe{1-16}:CORRection:COEFficient? <char>
Description
Sets the values for a single calibration correction coefficient such as ED1. The command identifies the coefficient name and then the coefficient data for a calibration on the indicated channel. Separate commands are issued for each required correction coefficient.
Returns the values of a correction coefficient of the calibration on the indicated channel.
• There are 12 correction coefficients available for a Full 2-Port Calibration which are: ED1, EP1S, ET11, ET21, EP2L, EX21, ED2, EP2S, ET22, ET12, EP1L, and EX12.
• There are 24 correction coefficients available for a Full 3-Port Calibration that include the 12 coefficients above and the 12 coefficients below: ED3, ET31, ET32, ET13, ET23, ET33, EP3L, EP3S, EX31, EX32, EX13, and EX23.
• There are 40 correction coefficients available for a Full 4-Port Calibration include the 24 coefficients above and the 16 coefficients below: ED4, ET14, ET41, ET24, ET42, ET34, ET43, ET44, EP4L, EP4S, EX14, EX24, EX34, EX41, EX42, and EX43.
Cmd Parameters
<char> ED1 | EP1S | ET11 | ET21 | EP2L | EX21 | ED2 | EP2S | ET22 | ET12 | EP1L | EX12 | ED3 | ET31 | ET32 | ET13 | ET23 | ET33 | EP3L | EP3S | EX31 | EX32 | EX13 | EX23 | ED4 | ET14 | ET41 | ET24 | ET42 | ET34 | ET43 | ET44 | EP4L | EP4S | EX14 | EX24 | EX34 | EX41 | EX42 | EX43
<block> The second input parameter is a block ASCII value. The actual ASCII block must be constructed and sent.
Query Parameters
<char> ED1 | EP1S | ET11 | ET21 | EP2L | EX21 | ED2 | EP2S | ET22 | ET12 | EP1L | EX12 | ED3 | ET31 | ET32 | ET13 | ET23 | ET33 | EP3L | EP3S | EX31 | EX32 | EX13 | EX23 | ED4 | ET14 | ET41 | ET24 | ET42 | ET34 | ET43 | ET44 | EP4L | EP4S | EX14 | EX24 | EX34 | EX41 | EX42 | EX43
Range
NA
Default Value
NA
Syntax Example
:SENS1:CORR:COEF ED1, <block>
:SENS1:CORR:COEF? ED1
:SENSe{1-16}:CORRection:COEFficient:FULL4
Description
Simulates a Full 4-Port Calibration on the indicated channel. This command requires a 4-port VNA instrument.
No query. To query the state of this command use:
Cmd Parameters
NA
Range
NA
Default
NA
Syntax Example
:SENS1:CORR:COEF:FULL4
:SENSe{1-16}:CORRection:COEFficient:PORT{12 | 13 | 14 | 23 | 24 | 34}:1P2PF
Description
The use of Port 3 and/or Port 4 requires a 4-port VNA instrument.
Simulates a One-Path 2-Port Calibration Forward Direction on the indicated port set on the indicated channel.
No query. To query the state of this command use:
To perform an actual calibration, use:
or
Parameters
NA
Range
NA
Default
NA
Syntax Example
:SENS1:CORR:COEF:PORT12:1P2PF
:SENSe{1-16}:CORRection:COEFficient:PORT{12 | 13 | 14 | 23 | 24 | 34}:1P2PR
Description
The use of Port 3 and/or Port 4 requires a 4-port VNA instrument.
Simulates a One-Path 2-Port Calibration Reverse Direction on the indicated port set on the indicated channel.
No query. To query the state of this command use:
Cmd Parameters
NA
Range
NA
Default
NA
Syntax Example
:SENS1:CORR:COEF:PORT12:1P2PR
:SENSe{1-16}:CORRection:COEFficient:PORT{12 | 13 | 14 | 23 | 24 | 34}:FULL2
Description
The use of Port 3 and/or Port 4 requires a 4-port VNA instrument.
Simulates a Full 2-Port Calibration on the indicated port set on the indicated channel.
No query. To query the state of this command use:
Cmd Parameters
NA
Range
NA
Default
NA
Syntax Example
:SENS1:CORR:COEF:PORT12:FULL2
:SENSe{1-16}:CORRection:COEFficient:PORT{12 | 13 | 14 | 23 | 24 | 34}:FULLB
Description
The use of Port 3 and/or Port 4 requires a 4-port VNA instrument.
Simulates a Full 1-port Reflection Calibration on both ports on the indicated channel.
No query. To query the state of this command use:
Cmd Parameters
NA
Range
NA
Default
NA
Syntax Example
:SENS1:CORR:COEF:PORT12:FULLB
:SENSe{1-16}:CORRection:COEFficient:PORT{12 | 13 | 14 | 23 | 24 | 34}:RESPB
Description
The use of Port 3 and/or Port 4 requires a 4-port VNA instrument.
Simulates a 1-port Response Calibration on both ports on the indicated port set on the indicated channel.
No query. To query the state of this command use:
Parameters
NA
Range
NA
Default
NA
Syntax Example
:SENS1:CORR:COEF:PORT12:RESPB
:SENSe{1-16}:CORRection:COEFficient:PORT{12 | 13 | 14 | 23 | 24 | 34}:TFRB
Description
The use of Port 3 and/or Port 4 requires a 4-port VNA instrument.
Simulates a Transmission Frequency Response Calibration in both directions on the indicated port set on the indicated channel.
No query. To query the state of this command use:
Cmd Parameters
NA
Range
NA
Default
NA
Syntax Example
:SENS1:CORR:COEF:PORT12:TFRB
:SENSe{1-16}:CORRection:COEFficient:PORT{12 | 13 | 14 | 23 | 24 | 34}:TFRF
Description
The use of Port 3 and/or Port 4 requires a 4-port VNA instrument.
Simulates a Transmission Frequency Response Calibration in the forward direction on the indicated port set on the indicated channel.
No query. To query the state of this command use:
Parameters
NA
Range
NA
Default
NA
Syntax Example
:SENS1:CORR:COEF:PORT12:TFRF
:SENSe{1-16}:CORRection:COEFficient:PORT{12 | 13 | 14 | 23 | 24 | 34}:TFRR
Description
The use of Port 3 and/or Port 4 requires a 4-port VNA instrument.
Simulates a Transmission Frequency Response Calibration in the reverse direction on the indicated port set on the indicated channel.
No query. To query the state of this command use:
No query.
Cmd Parameters
NA
Range
NA
Default
NA
Syntax Example
:SENS1:CORR:COEF:PORT12:TFRR
:SENSe{1-16}:CORRection:COEFficient:PORT{123 | 124 | 134 | 234}:FULL3
Description
The use of Port 3 and/or Port 4 requires a 4-port VNA instrument.
Simulates a Full Three Port Calibration on the indicated port set on the indicated channel.
No query. To query the state of this command use:
No query.
Cmd Parameters
NA
Range
NA
Default
NA
Syntax Example
:SENS1:CORR:COEF:PORT123:FULL3
:SENSe{1-16}:CORRection:COEFficient:PORT{1-4}:FULL1
Description
The use of Port 3 and/or Port 4 requires a 4-port VNA instrument.
Simulates a Full 1-port Reflection Calibration on the indicated port on the indicated channel.
No query. To query the state of this command use:
Cmd Parameters
NA
Range
NA
Default
NA
Syntax Example
:SENS1:CORR:COEF:PORT1:FULL1
:SENSe{1-16}:CORRection:COEFficient:PORT{1-4}:RESP1
Description
The use of Port 3 and/or Port 4 requires a 4-port VNA instrument. No query.
Simulates a 1-port Response Calibration on the indicated port on the indicated channel.
No query. To query the state of this command use:
Cmd Parameters
NA
Range
NA
Default
NA
Syntax Example
:SENS1:CORR:COEF:PORT1:RESP1