The :SENSe{1-16}:CORRection:COLLect:CALB subsystem commands set the calibration type for a second calibration (termed :CALB) and can only be used on a 4-port VNA instrument. Once the command is issued, the calibration proceeds.
Calibration Subsystems with Actual Calibration
Related calibration subsystems that perform actual calibrations are:
The calibration abbreviations and their calibration types are:
• :1P2PF refers to a one-path two-port calibration forward direction
• :1P2PR refers to a one path two port calibration reverse direction
• :FULL2 refers to a full two port calibration
Related Query
Most calibration commands of this type do not have a directly related query. To query the state of these commands, use: :SENSe{1-16}:CORRection:COLLect:TYPe?
:SENSe{1-16}:CORRection:COLLect:CALB:1P2PF
Description
This command requires a 4-port VNA instrument.
Sets the second 2-port calibration to One-Path 2-Port Forward on the indicated channel and port pair.
No query.
Cmd Parameters
NA
Range
NA
Default
NA
Syntax Example
:SENS1:CORR:COLL:CALB:1P2PF
:SENSe{1-16}:CORRection:COLLect:CALB:1P2PR
Description
This command requires a 4-port VNA instrument.
The command sets the second 2-port calibration to One-Path 2-Port Reverse on the indicated channel and port pair.
No query.
Cmd Parameters
NA
Range
NA
Default
NA
Syntax Example
:SENS1:CORR:COLL:CALB:1P2PR
:SENSe{1-16}:CORRection:COLLect:CALB:FULL2
Description
This command requires a 4-port VNA instrument.
Sets the second 2-port calibration to Full 2-Port on the indicated channel and port pair.