The :SENSe{1-16}:CORRection:COLLect subsystem commands configure or start an actual calibration and are limited to Port1 and/or Port 2. Both 2-port and 4-port VNA instruments can use these commands.
Calibration Setup Subsystems
These subsystems are used during various phases of calibration configuration setup:
These commands are similar to those in the :SENSe{1-16}:CORRection:COLLect[:CALa]:PORT Subsystem above where any port pair on a 2-port or 4-port VNA can be specified.
Related calibration types use the same command structure of:
:SENSe{1-16}:CORRection:COLLect:XXXXX
where the calibration XXXXX type is one of the following:
• 1P2PF = One-path two-port calibration, forward direction
• 1P2PR = One path two port calibration, reverse direction
• FULL1 = Full one port calibration
• FULL2 = Full two port calibration
• FULLB = Full one port reflection calibration, both ports
• RESP1 = One port response calibration
• RESPB = One port response calibration, both ports
• TFRB = Transmission frequency response calibration, both directions
• TFRF = Transmission frequency response calibration, forward direction
• TFRR = Transmission frequency response calibration, reverse direction
Each calibration type command is described in greater detail in the following sections. Related to the calibration commands are calibration simulation commands in the general form of :SENSe{1-16}:CORRection:COEFFicient:XXXXX that use the same abbreviations above.
To query the state of this calibration command, use:
Sets One-Path 2-Port Reverse calibration (1P2PR) as the calibration type on the indicated channel. After the method is set, the calibration must be performed.
No query. To query the state of this command, use:
Sets the state of the True Source Match flag in an Enhanced Match Calibration on the indicated channel.
Returns the on/off status of the True Source Match flag in an Enhanced Match Calibration on the indicated channel.
Cmd Parameters
<char> 1 | 0 | ON | OFF
Query Parameters
NA
Query Output:
<char> 1 | 0
Range
NA
Default
0
Syntax Example
:SENS1:CORR:COLL:ENHM:MIX:USE:TSM 1
:SENSe{1-16}:CORRection:COLLect:FULL1
Description
Sets full one port reflection calibration as the calibration type on the indicated channel. After the method is set, the calibration must be performed.
No query. To query the state of this command, use:
Sets full one port reflection calibration both ports as the calibration type on the indicated channel. After the method is set, the calibration must be performed.
No query. To query the state of this command, use:
The use of this command requires an instrument equipped with Option 08x.
Sets the greater than 54 GHz receiver setting power value to be used for RF calibration when in mixer receiver configuration for the indicated channel.
Returns the greater than 54 GHz receiver setting power value to be used for RF calibration when in mixer receiver configuration for the indicated channel.
Cmd Parameters
<NRf> The input parameter is in dBm.
Query Parameters
NA
<NR1> The output parameter is in dBm.
Range
-60 to +30 dBm
Default Value
-10 dBm (-18 dBm for MA25400A module selection)
Syntax Example
:SENS1:CORR:MIX:CAL:MODBB:POW 5.0
:SENS1:CORR:MIX:CAL:MODBB:POW?
:SENSe{1-16}:CORRection:COLLect:PORT <char>
:SENSe{1-16}:CORRection:COLLect:PORT?
Description
The use of Port 3 and/or Port 4 requires a 4-port VNA instrument.
Sets the calibration port for a RESP1 or FULL1 calibration on the indicated channel.
Returns the calibration ports or port pairs on the indicated channel.
The command outputs character data for the port combinations. Several examples of various calibration setups are described in the subsections below.
Reflection Frequency Response Calibration
A Reflection Frequency Response Calibration can have up to four reflection frequency response calibrations in one session.
Transmission Frequency Response Calibration
A Transmission Frequency Response Calibration can have up to six transmission frequency response calibrations picked from these types: TFRF, TFRR and TFRB.
One-Port Calibration
A 1-Port Calibration can consist of up to four, FULL1 calibrations in one session.
Two-Port Calibration
A 2-Port Calibration can consist of one or two calibrations picked from the following types:
• FULL2, 1P2PF, and 1P2PR.
• The port pair assignments must be independent.
• The Calibration Methods can be Reflection and Thru measurements, AutoCal and LRL/LRM. PORT12 and PORT34 are excluded from LRL/LRM calibrations.
• Both Reflection and Thru measurements and AutoCal can provide FULL2, 1P2PF, and 1P2PR. LRL/LRM calibrations provides FULL2 only.
Three-Port Calibration
A 3-Port Calibration can be accomplished using one of the following techniques:
• Reflection and Thru measurements or two LRL/LRM calibrations.
• PORT12 and PORT34 are excluded from LRL/LRM calibrations.
Four-Port Calibration
A 4-Port Calibration can be accomplished using one of the following techniques:
• Reflection and Thru measurements.
• Two LRL/LRM calibrations
• Two FULL2 AutoCal calibrations.
• PORT12 and PORT34 are excluded from LRL/LRM calibrations.
Sets 1-port Response Calibration on both ports as the calibration type on the indicated channel. After the method is set, the calibration must be performed.
No query. To query the state of this command, use:
Sets transmission frequency response calibration both directions as the calibration type on the indicated channel. After the method is set, the calibration must be performed.
No query. To query the state of this command, use:
Sets transmission frequency response calibration forward direction as the calibration type on the indicated channel. After the method is set, the calibration must be performed.
No query. To query the state of this command, use:
Sets transmission frequency response calibration reverse direction as the calibration type on the indicated channel. After the method is set, the calibration must be performed.
No query. To query the state of this command, use:
Query only. Outputs the calibration types of the indicated channel. Several examples of various calibration setups are described below.
Reflection Frequency Response Calibration
A Reflection Frequency Response Calibration can have up to four reflection frequency response calibrations in one session. The following is a return for the calibration type query for a response calibration on ports 2, 3 and 4:
RESP1,RESP1,RESP1
Transmission Frequency Response Calibration
A Transmission Frequency Response Calibration can have up to six transmission frequency response calibrations picked from these types: TFRF, TFRR and TFRB. The following is the return for the calibration type query for a mix of transmission response calibrations on all possible port pairs:
TFRF,TFRR,TFRB,TFRR,TFRF,TFRB
One-Port Calibration
A 1-Port Calibration can consist of up to four, FULL1 calibrations in one session. The following is a return for the calibration type query for a FULL1 calibration on PORT134:
FULL1,FULL1,FULL1
Two-Port Calibration
A 2-Port Calibration can consist of one or two calibrations picked from the following types: FULL2, 1P2PF, and 1P2PR. The port pair assignments must be independent. The Calibration Methods can be Reflection and Thru measurements, AutoCal and LRL/LRM. PORT12 and PORT34 are excluded from LRL/LRM calibrations. Both Reflection and Thru measurements and AutoCal can provide FULL2, 1P2PF, and 1P2PR. LRL/LRM calibrations provides FULL2 only. The following is a typical return for the calibration type query:
1P2PF,FULL2
Three-Port Calibration
A 3-Port Calibration can be accomplished using either Reflection and Thru measurements or two LRL/LRM calibrations. PORT12 and PORT34 are excluded from LRL/LRM calibrations. The following is a return for the calibration type query for each technique:
FULL3 | FULL3
Four-Port Calibration
A 4-Port Calibration can be accomplished using Reflection and Thru measurements, two LRL/LRM calibrations, or two FULL2 AutoCal calibrations. PORT12 and PORT34 are excluded from LRL/LRM calibrations. The following is a typical return for the calibration type query for each technique:
FULL4 | FULL4
Response Calibration Both and Full Calibration Both
The RESPB and FULLB calibration types return the command arguments shown below:
RESPB returns RESP1,RESP1
FULLB returns FULL1,FULL1
Set of Query Calibration Ports
Use the following commands to set or query the calibration ports: