:SENSe{1-16}:CORRection:COLLect:MSSS Subsystem
The :SENSe{1-16}:CORRection:COLLect:MSSS subsystem command commands set the parameters for the multiple offset short calibration method.
Calibration Setup Subsystems
These subsystems are used during various phases of calibration configuration setup:
:SENSe{1-16}:CORRection:COLLect:MSSS:PORT{1-4}:SHORt:COUNt <NRf>
:SENSe{1-16}:CORRection:COLLect:MSSS:PORT{1-4}:SHORt:COUNt?
Description
The use of Port 3 and/or Port 4 requires a 4-port VNA instrument.
Sets the number of shorts to use in the mSSS calibration on the indicated channel and port.
Returns the number of shorts to use in the mSSS calibration on the indicated channel and port.
Cmd Parameters
<NRf> The input parameter is an integer.
Query Parameters
NA
Query Output
<NR1> The output parameter is an integer.
Range
3 to 10
Default Value
3
Syntax Example
:SENS1:CORR:COLL:MSSS:PORT1:SHOR:COUN 5
:SENS1:CORR:COLL:MSSS:PORT1:SHOR:COUN?
:SENSe{1-16}:CORRection:COLLect:MSSS:PORT{1-4}:SHORt{1-10}:OFFSet <NRf>
:SENSe{1-16}:CORRection:COLLect:MSSS:PORT{1-4}:SHORt{1-10}:OFFSet?
Description
The use of Port 3 and/or Port 4 requires a 4-port VNA instrument.
Sets the offset length of the indicated short of the mSSS calibration on the indicated channel and port.
Returns the offset length of the indicated short of the mSSS calibration on the indicated channel and port.
Cmd Parameters
<NRf> The input parameter is in Meters.
Query Parameters
NA
Query Output
<NR3> The output parameter is in Meters.
Range
MPND
Default Value
Depends on the kits selected. If any of the shorts have kits other than OTHer , then the value will be populated from the selected kit. If all selected kits are OTHer , then default values are:
Short 1: 0 mm
Short 2: 1 mm
Short 3: 2 mm
Short 4: 3 mm
Short 5: 4 mm
Short 6: 5 mm
Short 7: 6 mm
Short 8: 7 mm
Short 9: 8 mm
Short 10: 9 mm
Syntax Example
:SENS1:CORR:COLL:MSSS:PORT1:SHOR1:OFFS 1.0E0
:SENS1:CORR:COLL:MSSS:PORT1:SHOR1:OFFS?
:SENSe{1-16}:CORRection:COLLect:MSSS:PORT{1-4}:SHORt{1-10}:KIT <char>
:SENSe{1-16}:CORRection:COLLect:MSSS:PORT{1-4}:SHORt{1-10}:KIT?
Description
Only valid for Coaxial and Non-Dispersive Line Types.
The use of Port 3 and/or Port 4 requires a 4-port VNA instrument.
Sets the cal kit selection for the indicated short of the mSSS calibration on the indicated channel and port, where the cal kit selections are:
• CM1 = 1 mm (male).
• CM0.8 = 0.8mm (male).
• CMUn = User-defined male connector from 1 to 32, where: CMU1 = User-defined male 1, CMU2 = User-defined male 2, CMU3 = User-defined male 3, etc.
• CF1 = 1 mm (female).
• CF0.8 = 0.8mm (female).
• CFUn = User-defined female cal kit from 1 to 32, where: CFU1 = User-defined female 1, CFU2 = User-defined female 2, CFU3 = User-defined female 3, etc.
• OTHer = Other, short length to be specified by user
Returns the cal kit selection for the indicated short of the mSSS calibration on the indicated channel and port.
Cmd Parameters
<char> CM1 | CM0.8 | CMU1 | CMU2 | CMU3 | CMU4 | CMU5 | CMU6 | CMU7 | CMU8 | CMU9 | CMU10 | CMU11 | CMU12 | CMU13 | CMU14 | CMU15 | CMU16 | CMU17 | CMU18 | CMU19 | CMU20 | CMU21 | CMU22 | CMU23 | CMU24 | CMU25 | CMU26 | CMU27 | CMU28 | CMU29 | CMU30 | CMU31 | CMU32 | CF1 | CF0.8 | CFU1 | CFU2 | CFU3 | CFU4 | CFU5 | CFU6 | CFU7 | CFU8 | CFU9 | CFU10 | CFU11 | CFU12 | CFU13 | CFU14 | CFU15 | CFU16 | CFU17 | CFU18 | CFU19 | CFU20 | CFU21 | CFU22 | CFU23 | CFU24 | CFU25 | CFU26 | CFU27 | CFU28 | CFU29 | CFU30 | CFU31 | CFU32 | OTH
Query Parameters
NA
Query Output
<char> CM1 | CM0.8 | CMU1 | CMU2 | CMU3 | CMU4 | CMU5 | CMU6 | CMU7 | CMU8 | CMU9 | CMU10 | CMU11 | CMU12 | CMU13 | CMU14 | CMU15 | CMU16 | CMU17 | CMU18 | CMU19 | CMU20 | CMU21 | CMU22 | CMU23 | CMU24 | CMU25 | CMU26 | CMU27 | CMU28 | CMU29 | CMU30 | CMU31 | CMU32 | CF1 | CF0.8 | CFU1 | CFU2 | CFU3 | CFU4 | CFU5 | CFU6 | CFU7 | CFU8 | CFU9 | CFU10 | CFU11 | CFU12 | CFU13 | CFU14 | CFU15 | CFU16 | CFU17 | CFU18 | CFU19 | CFU20 | CFU21 | CFU22 | CFU23 | CFU24 | CFU25 | CFU26 | CFU27 | CFU28 | CFU29 | CFU30 | CFU31 | CFU32 | OTH
Range
NA
Default Value
OTH
Syntax Example
:SENS1:CORR:COLL:MSSS:PORT1:SHOR1:KIT CM1
:SENS1:CORR:COLL:MSSS:PORT1:SHOR1:KIT?
:SENSe{1-16}:CORRection:COLLect:MSSS:WAVeguide:PORT{1-4}:SHORt{1-10}:KIT <char>
:SENSe{1-16}:CORRection:COLLect:MSSS:WAVeguide:PORT{1-4}:SHORt{1-10}:KIT?
Description
Only valid for Waveguide Line Type.
The use of Port 3 and/or Port 4 requires a 4-port VNA instrument.
Sets the waveguide cal kit selection for the indicated short of the mSSS calibration on the indicated channel and port, where the cal kit selections are:
• USERn = User-defined Waveguide connector from 1 to 32, where: USER1 = User-defined Waveguide Kit 1 USER2 = User-defined Waveguide Kit 2 USER3 = User-defined Waveguide Kit 3 ... USER32 = User-defined Waveguide Kit 32
• OTHer = Other, short length to be specified by user
Returns the waveguide cal kit selection for the indicated short of the mSSS calibration on the indicated channel and port.
Cmd Parameters
<char> USER1 | USER2 | USER3 | USER4 | USER5 | USER6 | USER7 | USER8 | USER9 | USER10 | USER11 | USER12 | USER13 | USER14 | USER15 | USER16 | USER17 | USER18 | USER19 | USER20 | USER21 | USER22 | USER23 | USER24 | USER25 | USER26 | USER27 | USER28 | USER29 | USER30 | USER31 | USER32 | OTH
Query Parameters
NA
Query Output
<char> USER1 | USER2 | USER3 | USER4 | USER5 | USER6 | USER7 | USER8 | USER9 | USER10 | USER11 | USER12 | USER13 | USER14 | USER15 | USER16 | USER17 | USER18 | USER19 | USER20 | USER21 | USER22 | USER23 | USER24 | USER25 | USER26 | USER27 | USER28 | USER29 | USER30 | USER31 | USER32 | OTH
Range
NA
Default Value
OTH
Syntax Example
:SENS1:CORR:COLL:MSSS:WAV:PORT1:SHOR1:KIT USER1
:SENS1:CORR:COLL:MSSS:WAV:PORT1:SHOR1:KIT?
:SENSe{1-16}:CORRection:COLLect:MSSS:MICrostrip:PORT{1-4}:SHORt{1-10}:KIT <char>
:SENSe{1-16}:CORRection:COLLect:MSSS:MICrostrip:PORT{1-4}:SHORt{1-10}:KIT?
Description
Only valid for Microstrip Line Type.
The use of Port 3 and/or Port 4 requires a 4-port VNA instrument.
Sets the microstrip cal kit selection for the indicated short of the mSSS calibration on the indicated channel and port, where the cal kit selections are:
• USERn = User-defined Microstrip connector from 1 to 32, where: USER1 = User-defined Microstrip Kit 1 USER2 = User-defined Microstrip Kit 2 USER3 = User-defined Microstrip Kit 3 ... USER32 = User-defined Waveguide Kit 32
• OTHer = Other, short length to be specified by user
Returns the microstrip cal kit selection for the indicated short of the mSSS calibration on the indicated channel and port.
Cmd Parameters
<char> USER1 | USER2 | USER3 | USER4 | USER5 | USER6 | USER7 | USER8 | USER9 | USER10 | USER11 | USER12 | USER13 | USER14 | USER15 | USER16 | USER17 | USER18 | USER19 | USER20 | USER21 | USER22 | USER23 | USER24 | USER25 | USER26 | USER27 | USER28 | USER29 | USER30 | USER31 | USER32 | OTH
Query Parameters
NA
Query Output
<char> USER1 | USER2 | USER3 | USER4 | USER5 | USER6 | USER7 | USER8 | USER9 | USER10 | USER11 | USER12 | USER13 | USER14 | USER15 | USER16 | USER17 | USER18 | USER19 | USER20 | USER21 | USER22 | USER23 | USER24 | USER25 | USER26 | USER27 | USER28 | USER29 | USER30 | USER31 | USER32 | OTH
Range
NA
Default Value
OTH
Syntax Example
:SENS1:CORR:COLL:MSSS:MIC:PORT1:SHOR1:KIT USER1
:SENS1:CORR:COLL:MSSS:MIC:PORT1:SHOR1:KIT?
:SENSe{1-16}:CORRection:COLLect:MSSS:PORT{1-4}:SHORt{1-10}:LOSS <NRf>
:SENSe{1-16}:CORRection:COLLect:MSSS:PORT{1-4}:SHORt{1-10}:LOSS?
Description
The use of Port 3 and/or Port 4 requires a 4-port VNA instrument.
Sets the loss on the indicated short of the mSSS calibration on the indicated channel and port.
Returns the loss on the indicated short of the mSSS calibration on the indicated channel and port.
Cmd Parameters
<NRf> The input parameter is in dB/mm.
Query Parameters
NA
Query Output
<NR3> The output parameter is in dB/mm.
Range
MPND
Default Value
0.00000000000E+000
Syntax Example
:SENS1:CORR:COLL:MSSS:PORT1:SHOR1:LOSS 1.0E0
:SENS1:CORR:COLL:MSSS:PORT1:SHOR1:LOSS?
:SENSe{1-16}:CORRection:COLLect:MSSS:PORT{1-4}:SHORt{1-10}:FREQuency <NRf>
:SENSe{1-16}:CORRection:COLLect:MSSS:PORT{1-4}:SHORt{1-10}:FREQuency?
Description
The use of Port 3 and/or Port 4 requires a 4-port VNA instrument.
Sets the reference frequency (for loss computation) on the indicated short of the mSSS calibration on the indicated channel and port.
Returns the reference frequency (for loss computation) on the indicated short of the mSSS calibration on the indicated channel and port.
Cmd Parameters
<NRf> The input parameter is in Hertz.
Query Parameters
NA
Query Output
<NR1> The output parameter is in Hertz.
Range
7E4 to 7E10
Default Value
0.00000000000E+000
Syntax Example
:SENS1:CORR:COLL:MSSS:PORT1:SHOR1:FREQ 1.0E7
:SENS1:CORR:COLL:MSSS:PORT1:SHOR1:FREQ?
:SENSe{1-16}:CORRection:COLLect:MSSS:PORT{1-4}:MATCH[:STATe] <char>
:SENSe{1-16}:CORRection:COLLect:MSSS:PORT{1-4}:MATCH[:STATe]?
Description
The use of Port 3 and/or Port 4 requires a 4-port VNA instrument.
Sets the state of whether to include match in the mSSS calibration on the indicated channel and port.
Returns whether match will be included in the mSSS calibration on the indicated channel and port.
Cmd Parameters
<char> 1 | 0 | ON | OFF
Query Parameters
NA
Query Output
<char> 1 | 0
Range
NA
Default Value
0
Syntax Example
:SENS1:CORR:COLL:MSSS:PORT1:MATCH 1
:SENS1:CORR:COLL:MSSS:PORT1:MATCH?
:SENSe{1-16}:CORRection:COLLect:MSSS:PORT{1-4}:OPTLength[:STATe] <char>
:SENSe{1-16}:CORRection:COLLect:MSSS:PORT{1-4}:OPTLength[:STATe]?
Description
The use of Port 3 and/or Port 4 requires a 4-port VNA instrument.
Sets the state of whether to use the Optimal Length Super-Weighting in the mSSS calibration on the indicated channel and port.
Returns whether Optimal Length Super-Weighting will be used in the mSSS calibration on the indicated channel and port.
Cmd Parameters
<char> 1 | 0 | ON | OFF
Query Parameters
NA
Query Output
<char> 1 | 0
Range
NA
Default Value
0
Syntax Example
:SENS1:CORR:COLL:MSSS:PORT1:OPTL 1
:SENS1:CORR:COLL:MSSS:PORT1:OPTL?
:SENSe{1-16}:CORRection:COLLect:MSSS:PORT{1-4}:MATCh:C{0-3} <NRf>
:SENSe{1-16}:CORRection:COLLect:MSSS:PORT{1-4}:MATCh:C{0-3}?
Description
The use of Port 3 and/or Port 4 requires a 4-port VNA instrument.
Sets the indicated capacitance coefficient of match on the indicated port of the mSSS calibration on the indicated channel and port.
Returns the value of indicated capacitance coefficient of match on the indicated port of the mSSS calibration on the indicated channel and port.
Cmd Parameters
<NRf> The input parameter is in Farads for C0 coefficient.
<NRf> The input parameter is in Farads/Hertz for C1 coefficient.
<NRf> The input parameter is in Farads/Hertz^2 for C2 coefficient.
<NRf> The input parameter is in Farads/Hertz^3 for C3 coefficient.
Query Parameters
NA
Query Output
<NR3> The input parameter is in Farads for C0 coefficient.
<NR3> The input parameter is in Farads/Hertz for C1 coefficient.
<NR3> The input parameter is in Farads/Hertz^2 for C2 coefficient.
<NR3> The input parameter is in Farads/Hertz^3 for C3 coefficient.
Range
MPND
Default Value
0.00000000000E+000
Syntax Example
:SENS1:CORR:COLL:MSSS:PORT1:MATC:C0 3.01E-12
:SENS1:CORR:COLL:MSSS:PORT1:MATC:C0?
:SENSe{1-16}:CORRection:COLLect:MSSS:PORT{1-4}:MATCh:L{0-3} <NRf>
:SENSe{1-16}:CORRection:COLLect:MSSS:PORT{1-4}:MATCh:L{0-3}?
Description
The use of Port 3 and/or Port 4 requires a 4-port VNA instrument.
Sets the indicated inductance coefficient of match on the indicated port of the mSSS calibration on the indicated channel and port.
Returns the value of indicated inductance coefficient of match on the indicated port of the mSSS calibration on the indicated channel and port.
Cmd Parameters
<NRf> The input parameter is in Henrys for L0 coefficient.
<NRf> The input parameter is in Henrys/Hertz for L1 coefficient.
<NRf> The input parameter is in Henrys/Hertz^2 for L2 coefficient.
<NRf> The input parameter is in Henrys/Hertz^3 for L3 coefficient.
Query Parameters
NA
Query Output
<NR3> The input parameter is in Henrys for L0 coefficient.
<NR3> The input parameter is in Henrys/Hertz for L1 coefficient.
<NR3> The input parameter is in Henrys/Hertz^2 for L2 coefficient.
<NR3> The input parameter is in Henrys/Hertz^3 for L3 coefficient.
Range
MPND
Default Value
0.00000000000E+000
Syntax Example
:SENS1:CORR:COLL:MSSS:PORT1:MATC:L0 2.01E-6
:SENS1:CORR:COLL:MSSS:PORT1:MATC:L0?
:SENSe{1-16}:CORRection:COLLect:MSSS:PORT{1-4}:MATCh:OFFS{0-3} <NRf>
:SENSe{1-16}:CORRection:COLLect:MSSS:PORT{1-4}:MATCh:OFFS{0-3}?
Description
The use of Port 3 and/or Port 4 requires a 4-port VNA instrument.
Sets the indicated offset length coefficient of match of the mSSS calibration on the indicated channel and port.
Returns the value of indicated offset length coefficient of match of the mSSS calibration on the indicated channel and port.
Cmd Parameters
<NRf> The input parameter is in Meters for offset length coefficient.
<NRf> The input parameter is in Meters/Hertz for offset length1 coefficient.
<NRf> The input parameter is in Meters/Hertz^2 for offset length2 coefficient.
<NRf> The input parameter is in Meters/Hertz^3 for offset length3 coefficient.
Query Parameters
NA
Query Output
<NR3> The input parameter is in Meters for offset length coefficient.
<NR3> The input parameter is in Meters/Hertz for offset length1 coefficient.
<NR3> The input parameter is in Meters/Hertz^2 for offset length2 coefficient.
<NR3> The input parameter is in Meters/Hertz^3 for offset length3 coefficient.
Range
MPND
Default Value
0.00000000000E+000
Syntax Example
:SENS1:CORR:COLL:MSSS:PORT1:MATC:OFFS 1.0E0
:SENS1:CORR:COLL:MSSS:PORT1:MATC:OFFS?
:SENSe{1-16}:CORRection:COLLect:MSSS:PORT{1-4}:MATCh:R <NRf>
:SENSe{1-16}:CORRection:COLLect:MSSS:PORT{1-4}:MATCh:R?
Description
The use of Port 3 and/or Port 4 requires a 4-port VNA instrument.
Sets the resistance of match of the mSSS calibration on the indicated channel and port.
Returns the resistance of match of the mSSS calibration on the indicated channel and port.
Cmd Parameters
<NRf> The input parameter is in Ohms.
Query Parameters
NA
Query Output
<NR3> The output parameter is in Ohms.
Range
MPND
Default Value
5.00000000000E+001
Syntax Example
:SENS1:CORR:COLL:MSSS:PORT1:MATC:R 7.5E1
:SENS1:CORR:COLL:MSSS:PORT1:MATC:R?
:SENSe{1-16}:CORRection:COLLect:MSSS:PORT{1-4}:MATCh:Z0 <NRf>
:SENSe{1-16}:CORRection:COLLect:MSSS:PORT{1-4}:MATCh:Z0?
Description
The use of Port 3 and/or Port 4 requires a 4-port VNA instrument.
Sets the impedance of match of the mSSS calibration on the indicated channel and port.
Returns the impedance of match of the mSSS calibration on the indicated channel and port.
Cmd Parameters
<NRf> The input parameter is in Ohms.
Query Parameters
NA
Query Output
<NR3> The output parameter is in Ohms.
Range
MPND
Default Value
5.00000000000E+001
Syntax Example
:SENS1:CORR:COLL:MSSS:PORT1:MATC:Z0 7.5E1
:SENS1:CORR:COLL:MSSS:PORT1:MATC:Z0?
:SENSe{1-16}:CORRection:COLLect:MSSS:PORT{1-4}:MATCh:S1P:FILE <string>
:SENSe{1-16}:CORRection:COLLect:MSSS:PORT{1-4}:MATCh:S1P:FILE?
Description
The use of Port 3 and/or Port 4 requires a 4-port VNA instrument.
Sets the match S1P file on the indicated port of the mSSS calibration on the indicated channel.
Returns the match S1P file on the indicated port of the mSSS calibration on the indicated channel.
Cmd Parameters
<string> Filename and path to the file in the form: x:\directory\filename.s1p , where x:\directory\filename.s1p must exist.
Query Parameters
NA
Query Output
<string> Filename and path in the form: 'x:\directory\filename.s1p'
Range
NA
Default Value
NA
Syntax Example
:SENS1:CORR:COLL:MSSS:PORT1:MATC:S1P:FILE 'c:\filename.s1p'
:SENS1:CORR:COLL:MSSS:PORT1:MATC:S1P:FILE?
:SENSe{1-16}:CORRection:COLLect:MSSS:PORT{1-4}:MATCh:S1P[:STATe] <char>
:SENSe{1-16}:CORRection:COLLect:MSSS:PORT{1-4}:MATCh:S1P[:STATe]?
Description
The use of Port 3 and/or Port 4 requires a 4-port VNA instrument.
Sets the MATCH S1P file settings ON/OFF on the selected port for mSSS calibration for the indicated channel.
ON means that the S1P file data will be used, OFF means the Circuit Model data will be used.
Outputs the MATCH S1P file settings on the selected port for mSSS calibration for the indicated channel.
Cmd Parameters
<char> 1 | 0 | ON | OFF
Query Parameters
NA
Query Output
<char> 1 | 0
Range
NA
Default Value
0
Syntax Example
:SENS1:CORR:COLL:MSSS:PORT1:MATC:S1P 1
:SENS1:CORR:COLL:MSSS:PORT1:MATC:S1P?
:SENSe{1-16}:CORRection:COLLect:MSSS:PORT{1-4}:MATCh:MEASure <NRf>
Description
The use of Port 3 or Port 4 requires a 4-port VNA instrument.
Initiates collection of match standard data for the mSSS calibration on the indicated channel and port.
No query.
Cmd Parameters
NA
Query Parameters
NA
Query Output
NA
Range
NA
Default Value
NA
Syntax Example
:SENS1:CORR:COLL:MSSS:PORT1:MATC:MEAS
:SENSe{1-16}:CORRection:COLLect:MSSS:PORT{1-4}:SHORt{1-10}:MEASure <NRf>
Description
The use of Port 3 and/or Port 4 requires a 4-port VNA instrument.
Initiates collection of the indicated Offset Short standard data for the mSSS calibration on the indicated channel and port.
No query.
Cmd Parameters
NA
Query Parameters
NA
Query Output
NA
Range
NA
Default Value
NA
Syntax Example
:SENS1:CORR:COLL:MSSS:PORT1:SHOR1:MEAS
:SENSe{1-16}:CORRection:COLLect:MSSS:WAVeguide:DIELectric <NRf>
:SENSe{1-16}:CORRection:COLLect:MSSS:WAVeguide:DIELectric?
Description
Sets the mSSS waveguide calibration kit dielectric value on the indicated channel.
Returns the mSSS waveguide calibration kit dielectric value on the indicated channel.
Cmd Parameters
<NRf> The input parameter is a unitless number.
Query Parameters
NA
Query Output
<NR3> The output parameter is a unitless number.
Range
MPND
Default Value
1.00000000000E+001
Syntax Example
:SENS1:CORR:COLL:MSSS:WAV:DIEL 1.2E0
:SENS1:CORR:COLL:MSSS:WAV:DIEL?
:SENSe{1-16}:CORRection:COLLect:MSSS:WAVeguide:Frequency <NRf>
:SENSe{1-16}:CORRection:COLLect:MSSS:WAVeguide:Frequency?
Description
Sets the mSSS waveguide calibration kit cutoff frequency on the indicated channel.
Returns the mSSS waveguide calibration kit cutoff frequency on the indicated channel.
Cmd Parameters
<NRf> The input parameter is in Hertz.
Query Parameters
NA
Query Output
<NR3> The output parameter is in Hertz.
Range
MPND
Default Value
0.00000000000E+001
Syntax Example
:SENS1:CORR:COLL:MSSS:WAV:FREQ 5.0E9
:SENS1:CORR:COLL:MSSS:WAV:FREQ?
:SENSe{1-16}:CORRection:COLLect:MSSS:PORT{1-4}:SHORt{1-10}:SHORt:SELection <char>
:SENSe{1-16}:CORRection:COLLect:MSSS:PORT{1-4}:SHORt{1-10}:SHORt:SELection?
Description
The use of Port 3 and/or Port 4 requires a 4-port VNA instrument.
Sets the short selection for the indicated short of the mSSS calibration on the indicated channel and port, where:
• S1, S2, S3 are the short selections from the selected cal kit for the indicated short. Not valid if the selected cal kit for the indicated short is OTHer .
Returns the short selection for the indicated short of the mSSS calibration on the indicated channel and port.
Cmd Parameters
<char> S1 | S2 | S3
Query Parameters
NA
Query Output
<char> S1 | S2 | S3
Range
NA
Default Value
OTH
Syntax Example
:SENS1:CORR:COLL:MSSS:PORT1:SHOR1:SHOR:SEL S2
:SENS1:CORR:COLL:MSSS:PORT1:SHOR1:SHOR:SEL?