VectorStar™ MS464xB Series Microwave Vector Network Analyzer Programming Manual : SCPI Commands : :SENSe{1-16}:CORRection:COLLect:WAVeguide Subsystem
 
:SENSe{1-16}:CORRection:COLLect:WAVeguide Subsystem
The :SENSe{1-16}:CORRection:COLLect:WAVeguide subsystem sets the calibration parameters and coefficients for waveguide line types. Use this subsystem to set waveguide parameters before starting a calibration.
Calibration Setup Subsystems
These subsystems are used during various phases of calibration configuration setup:
:CALCulate{1-16}:IMPedance:TRANsformation Subsystem
:CALCulate{1-16}:REFerence Subsystem
:SENSe{1-16}:CORRection:COLLect:METHod Subsystem
:SENSe{1-16}:CORRection:COLLect:MICrostrip Subsystem
:SENSe{1-16}:CORRection:COLLect:MSSS Subsystem
:SENSe{1-16}:CORRection:COLLect:MULTiple Subsystem
:SENSe{1-16}:CORRection:COLLect:MTRL[:CALa] Subsystem
:SENSe{1-16}:CORRection:COLLect:MTRL:CALB Subsystem – 4-Port VNAs
:SENSe{1-16}:CORRection:COLLect:TRL[:CALa] Subsystem
:SENSe{1-16}:CORRection:COLLect:TRL:CALB Subsystem
:SENSe{1-16}:CORRection:COLLect:TRL:SINGleton Subsystem – 4-Port VNAs
:SENSe{1-16}:CORRection:COLLect:TRL:WAVeguide Subsystem
:SENSe{1-16}:CORRection:COLLect Subsystem
:SENSe{1-16}:CORRection:COLLect:WAVeguide Subsystem
:SENSe{1-16}:CORRection:COLLect[:CALa]:PORT Subsystem
:SENSe{1-16}:CORRection:INTerpolation Subsystem
:SENSe{1-16}:CORRection:STATe Subsystem
:SENSe{1-16}:CORRection:COLLect:WAVeguide:DIELectric <NRf>
:SENSe{1-16}:CORRection:COLLect:WAVeguide:DIELectric?
Description
Sets the waveguide calibration kit dielectric value on the indicated channel.
Returns the waveguide calibration kit dielectric value on the indicated channel.
See Calibration Component Parameters for a complete listing of calibration components, connectors, and their Command Parameters.
Cmd Parameters
<NRf> The input parameter is a unitless number.
Query Parameters
<NR3> The output parameter is a unitless number.
Range
MPND
Default Value
1.00000000000E+000
Syntax Example
:SENS1:CORR:COLL:WAV:DIEL 1.2E0
:SENS1:CORR:COLL:WAV:DIEL?
:SENSe{1-16}:CORRection:COLLect:WAVeguide:FREQuency <NRf>
:SENSe{1-16}:CORRection:COLLect:WAVeguide:FREQuency?
Description
Sets the waveguide calibration kit cutoff frequency on the indicated channel.
Returns the waveguide calibration kit cutoff frequency on the indicated channel.
Cmd Parameters
<NRf> The input parameter is in Hertz.
Query Parameters
<NR3> The output parameter is in Hertz.
Range
MPND
Default Value
See Calibration Component Parameters.
Syntax Example
:SENS1:CORR:COLL:WAV:FREQ 5.0E9
:SENS1:CORR:COLL:WAV:FREQ?
:SENSe{1-16}:CORRection:COLLect:WAVeguide:KIT <char>
:SENSe{1-16}:CORRection:COLLect:WAVeguide:KIT?
Description
Sets the waveguide kit type on the indicated channel.
Returns the waveguide kit type on the indicated channel.
See Calibration Component Parameters for a complete listing of calibration components, connectors, and their Command Parameters.
Cmd Parameters
<char> WR10 | WR12 | WR15 | USER1 | USER2 | USER3 | USER4 | USER5 | USER6 | USER7 | USER8 | USER9 | USER10 | USER11 | USER12 | USER13 | USER14 | USER15 | USER16 | USER17 | USER18 | USER19 | USER20 | USER21 | USER22 | USER23 | USER24 | USER25 | USER26 | USER27 | USER28 | USER29 | USER30 | USER31 | USER32
Where:
WR10 = Typical WR-10 RF waveguide
WR12 = Typical WR-12 RF waveguide
WR15 = Typical WR-13 RF waveguide
USERx = User defined waveguide x. The device can be renamed by the user but programmatically is always referred to as “USERx”:
Where x is any number between 1 and 32.
USER1 = User defined waveguide 1. The device can be renamed by the user but programmatically is always referred to as “USERx”:
Query Parameters
<char> WR10 | WR12 | WR15 | USER1 | USER2 | USER3 | USER4 | USER5 | USER6 | USER7 | USER8 | USER9 | USER10 | USER11 | USER12 | USER13 | USER14 | USER15 | USER16 | USER17 | USER18 | USER19 | USER20 | USER21 | USER22 | USER23 | USER24 | USER25 | USER26 | USER27 | USER28 | USER29 | USER30 | USER31 | USER32
Range
NA
Default Value
USER1
Syntax Example
:SENS1:CORR:COLL:WAV:KIT WR12
:SENS1:CORR:COLL:WAV:KIT?
:SENSe{1-16}:CORRection:COLLect:WAVeguide:LABel <string>
:SENSe{1-16}:CORRection:COLLect:WAVeguide:LABel?
Description
Sets the waveguide calibration kit label on the indicated channel.
Returns the waveguide calibration kit label on the indicated channel.
See Calibration Component Parameters for a complete listing of calibration components, connectors, and their Command Parameters.
Cmd Parameters
<string> The input parameter is any combination of numbers and letters.
See definition of <string>.
Query Parameters
<char> The output parameter can be any combination of numbers and letters.
Range
NA
Default Value
WR10
Syntax Example
:SENS1:CORR:COLL:WAV:LAB 'wave1'
:SENS1:CORR:COLL:WAV:LAB?
:SENSe{1-16}:CORRection:COLLect:WAVeguide:LOAD:L0 <NRf>
:SENSe{1-16}:CORRection:COLLect:WAVeguide:LOAD:L0?
Description
Sets the waveguide calibration kit L0 (L zero) load inductance on the indicated channel.
Returns the waveguide calibration kit load inductance on the indicated channel.
See Calibration Component Parameters for a complete listing of calibration components, connectors, and their Command Parameters.
Cmd Parameters
<NRf> The input parameter is in Henrys.
Cmd Parameters
<NR3> The output parameter is in Henrys.
Range
MPND
Default Value
0.00000000000E+000
Syntax Example
:SENS1:CORR:COLL:WAV:LOAD:L0 7.5E1
:SENS1:CORR:COLL:WAV:LOAD:L0?
:SENSe{1-16}:CORRection:COLLect:WAVeguide:LOAD:OFFSet <NRf>
:SENSe{1-16}:CORRection:COLLect:WAVeguide:LOAD:OFFSet?
Description
Sets the waveguide calibration kit load offset on the indicated channel.
Returns the waveguide calibration kit load offset on the indicated channel.
See Calibration Component Parameters for a complete listing of calibration components, connectors, and their Command Parameters.
Cmd Parameters
<NRf> The input parameter is in Meters.
Query Parameters
<NR3> The output parameter is in Meters.
Range
MPND
Default Value
0.00000000000E+000
Syntax Example
:SENS1:CORR:COLL:WAV:LOAD:OFFS 2.5E1
:SENS1:CORR:COLL:WAV:LOAD:OFFS?
:SENSe{1-16}:CORRection:COLLect:WAVeguide:LOAD:R <NRf>
:SENSe{1-16}:CORRection:COLLect:WAVeguide:LOAD:R?
Description
Sets the waveguide calibration kit load resistance on the indicated channel.
Returns the waveguide calibration kit load resistance on the indicated channel.
See Calibration Component Parameters for a complete listing of calibration components, connectors, and their Command Parameters.
Cmd Parameters
<NRf> The input parameter is in Ohms.
Query Parameters
<NR3> The output parameter is in Ohms.
Range
MPND
Default Value
5.00000000000E+001
Syntax Example
:SENS1:CORR:COLL:WAV:LOAD:R 7.5E1
:SENS1:CORR:COLL:WAV:LOAD:R?
:SENSe{1-16}:CORRection:COLLect:WAVeguide:OPEN:C0 <NRf>
:SENSe{1-16}:CORRection:COLLect:WAVeguide:OPEN:C0?
Description
Sets the waveguide calibration kit open capacitance C0 (C zero) value on the indicated channel.
The C0 coefficient is measured in Farads.
Returns the waveguide calibration kit open capacitance C0 value on the indicated channel.
See Calibration Component Parameters for a complete listing of calibration components, connectors, and their Command Parameters.
Cmd Parameters
<NRf> The input parameter is in Farads.
Query Parameters
<NR3> The output parameter is in Farads.
Range
MPND
Default Value
0.00000000000E+000
Syntax Example
:SENS1:CORR:COLL:WAV:OPEN:C0 3.0E-12
:SENS1:CORR:COLL:WAV:OPEN:C0?
:SENSe{1-16}:CORRection:COLLect:WAVeguide:OPEN:C1 <NRf>
:SENSe{1-16}:CORRection:COLLect:WAVeguide:OPEN:C1?
Description
Sets the waveguide calibration kit open capacitance C1 (C one) term on the indicated channel.
The C1 coefficient is measured in Farads/Hertz.
Returns the waveguide calibration kit open capacitance1 term on the indicated channel.
See Calibration Component Parameters for a complete listing of calibration components, connectors, and their Command Parameters.
Cmd Parameters
<NRf> The input parameter is in Farads/Hertz.
Query Parameters
<NR3> The output parameter is in Farads/Hertz.
Range
MPND
Default Value
0.00000000000E+000
Syntax Example
:SENS1:CORR:COLL:WAV:OPEN:C1 3.0E-12
:SENS1:CORR:COLL:WAV:OPEN:C1?
:SENSe{1-16}:CORRection:COLLect:WAVeguide:OPEN:C2 <NRf>
:SENSe{1-16}:CORRection:COLLect:WAVeguide:OPEN:C2?
Description
Sets the waveguide calibration kit open capacitance C2 term on the indicated channel.
The C2 coefficient is measured in Farads/Hertz^2.
Returns the waveguide calibration kit open capacitance C2 term on the indicated channel.
See Calibration Component Parameters for a complete listing of calibration components, connectors, and their Command Parameters.
Cmd Parameters
<NRf> The input parameter is in Farads/Hertz^2.
Query Parameters
<NR3> The output parameter is in Farads/Hertz^2.
Range
MPND
Default Value
0.00000000000E+000
Syntax Example
:SENS1:CORR:COLL:WAV:OPEN:C2 3.0E-12
:SENS1:CORR:COLL:WAV:OPEN:C2?
:SENSe{1-16}:CORRection:COLLect:WAVeguide:OPEN:C3 <NRf>
:SENSe{1-16}:CORRection:COLLect:WAVeguide:OPEN:C3?
Description
Sets the waveguide calibration kit open capacitance C3 term on the indicated channel.
The C3 coefficient is measured in Farads/Hertz^3.
Returns the waveguide calibration kit open capacitance C3 term on the indicated channel.
See Calibration Component Parameters for a complete listing of calibration components, connectors, and their Command Parameters.
Cmd Parameters
<NRf> The input parameter is in Farads/Hertz^3.
Query Parameters
<NR3> The output parameter is in Farads/Hertz^3.
Range
MPND
Default Value
0.00000000000E+000
Syntax Example
:SENS1:CORR:COLL:WAV:OPEN:C3 3.0E-12
:SENS1:CORR:COLL:WAV:OPEN:C3?
:SENSe{1-16}:CORRection:COLLect:WAVeguide:OPEN:FREQuency <NRf>
:SENSe{1-16}:CORRection:COLLect:WAVeguide:OPEN:FREQuency?
Description
Sets the waveguide calibration kit open reference frequency on the indicated channel.
Returns the waveguide calibration kit open reference frequency on the indicated channel.
See Calibration Component Parameters for a complete listing of calibration components, connectors, and their Command Parameters.
Cmd Parameters
<NRf> The input parameter is in Hertz.
Query Parameters
<NR3> The output parameter is in Hertz.
Range
MPND
Default Value
See Calibration Component Parameters.
Syntax Example
:SENS1:CORR:COLL:WAV:OPEN:FREQ 1.0E7
:SENS1:CORR:COLL:WAV:OPEN:FREQ?
:SENSe{1-16}:CORRection:COLLect:WAVeguide:OPEN:LOSS <NRf>
:SENSe{1-16}:CORRection:COLLect:WAVeguide:OPEN:LOSS?
Description
Sets the waveguide calibration kit open loss value on the indicated channel.
Returns the waveguide calibration kit open loss value on the indicated channel.
Cmd Parameters
<NRf> The input parameter is in dB/mm.
Query Parameters
<NR3> The output parameter can be in dB/mm.
Range
MPND
Default Value
0
Syntax Example
:SENS1:CORR:COLL:WAV:OPEN:LOSS 1.0E0
:SENS1:CORR:COLL:WAV:OPEN:LOSS?
:SENSe{1-16}:CORRection:COLLect:WAVeguide:OPEN:OFFSet <NRf>
:SENSe{1-16}:CORRection:COLLect:WAVeguide:OPEN:OFFSet?
Description
Sets the waveguide calibration kit open offset on the indicated channel.
Returns the waveguide calibration kit open offset on the indicated channel.
See Calibration Component Parameters for a complete listing of calibration components, connectors, and their Command Parameters.
Cmd Parameters
<NRf> The input parameter is in Meters.
Query Parameters
<NR3> The output parameter is in Meters.
Range
MPND
Default Value
0.00000000000E+000
Syntax Example
:SENS1:CORR:COLL:WAV:OPEN:OFFS 1.2E-4
:SENS1:CORR:COLL:WAV:OPEN:OFFS?
:SENSe{1-16}:CORRection:COLLect:WAVeguide:OPEN:S1P:FILe <string>
:SENSe{1-16}:CORRection:COLLect:WAVeguide:OPEN:S1P:FILe?
Description
Sets the waveguide calibration kit OPEN s1p file on the indicated channel.
Returns the waveguide calibration kit OPEN s1p file on the indicated channel.
Cmd Parameters
<string> s1p file path
See definition of <string>.
Query Parameters
NA
Query Output
<char1> s1p file path
Range
NA
Default Value
NA
Syntax Example
:SENS1:CORR:COLL:WAV:OPEN:S1P:FIL 'C:\Test\WG_Open.s1p'
:SENS1:CORR:COLL:WAV:OPEN:S1P:FIL?
:SENSe{1-16}:CORRection:COLLect:WAVeguide:OPEN:S1P[:STATe] <char>
:SENSe{1-16}:CORRection:COLLect:WAVeguide:OPEN:S1P[:STATe]?
Description
Sets the waveguide calibration kit OPEN S1P file settings ON/OFF on the selected port of the indicated channel.
ON means that the S1P file data will be used, OFF means the Circuit Model data will be used.
Returns the waveguide calibration kit OPEN S1P file settings on the selected port of the indicated channel.
Cmd Parameters
<char> 1 | 0 | ON | OFF
Query Parameters
NA
Query Output
<char> 1 | 0
Range
NA
Default Value
0
Syntax Example
:SENS1:CORR:COLL:WAV:OPEN:S1P 1
:SENS1:CORR:COLL:WAV:OPEN:S1P?
:SENSe{1-16}:CORRection:COLLect:WAVeguide:SERial <string>
:SENSe{1-16}:CORRection:COLLect:WAVeguide:SERial?
Description
Sets the waveguide calibration kit serial number on the indicated channel.
Returns the waveguide calibration kit serial number on the indicated channel.
Cmd Parameters
<string> The input parameter is any combination of numbers and letters. See definition of <string>.
Query Parameters
<char> The output parameter can be any combination of numbers and letters.
Range
NA
Default Value
NA
Syntax Example
:SENS1:CORR:COLL:WAV:SER '12012008'
:SENS1:CORR:COLL:WAV:SER?
:SENSe{1-16}:CORRection:COLLect:WAVeguide:SHORT1:FREQuency <NRf>
:SENSe{1-16}:CORRection:COLLect:WAVeguide:SHORT1:FREQuency?
Description
Sets the waveguide calibration kit Short1 reference frequency on the indicated channel.
Returns the waveguide calibration kit Short1 reference frequency on the indicated channel.
See Calibration Component Parameters for a complete listing of calibration components, connectors, and their Command Parameters.
Cmd Parameters
<NRf> The input parameter is in Hertz.
Query Parameters
<NR3> The output parameter is in Hertz.
Range
MPND
Default Value
See Calibration Component Parameters.
Syntax Example
:SENS1:CORR:COLL:WAV:SHORT1:FREQ 1.0E7
:SENS1:CORR:COLL:WAV:SHORT1:FREQ?
:SENSe{1-16}:CORRection:COLLect:WAVeguide:SHORT1:LOSS <NRf>
:SENSe{1-16}:CORRection:COLLect:WAVeguide:SHORT1:LOSS?
Description
Sets the waveguide calibration kit Short1 loss value on the indicated channel.
Returns the waveguide calibration kit Short1 loss value on the indicated channel.
Cmd Parameters
<NRf> The input parameter is in dB/mm.
Query Parameters
<NR3> The output parameter can be in dB/mm.
Range
MPND
Default Value
0
Syntax Example
:SENS1:CORR:COLL:WAV:SHORT1:LOSS 1.0E0
:SENS1:CORR:COLL:WAV:SHORT1:LOSS?
:SENSe{1-16}:CORRection:COLLect:WAVeguide:SHORT1:OFFSet <NRf>
:SENSe{1-16}:CORRection:COLLect:WAVeguide:SHORT1:OFFSet?
Description
Sets the waveguide calibration kit short1 offset on the indicated channel.
Returns the waveguide calibration kit short1 offset on the indicated channel.
Cmd Parameters
<NRf> The input parameter is in Meters.
Query Parameters
<NR3> The output parameter is in Meters.
Range
MPND
Default Value
See Calibration Component Parameters.
Syntax Example
:SENS1:CORR:COLL:WAV:SHORT1:OFFS 1.2E-4
:SENS1:CORR:COLL:WAV:SHORT1:OFFS?
:SENSe{1-16}:CORRection:COLLect:WAVeguide:SHORT2:FREQuency <NRf>
:SENSe{1-16}:CORRection:COLLect:WAVeguide:SHORT2:FREQuency?
Description
Sets the waveguide calibration kit Short2 reference frequency on the indicated channel.
Returns the waveguide calibration kit Short2 reference frequency on the indicated channel.
See Calibration Component Parameters for a complete listing of calibration components, connectors, and their Command Parameters.
Cmd Parameters
<NRf> The input parameter is in Hertz.
Query Parameters
<NR3> The output parameter is in Hertz.
Range
MPND
Default Value
See Calibration Component Parameters.
Syntax Example
:SENS1:CORR:COLL:WAV:SHORT2:FREQ 1.0E7
:SENS1:CORR:COLL:WAV:SHORT2:FREQ?
:SENSe{1-16}:CORRection:COLLect:WAVeguide:SHORT2:LOSS <NRf>
:SENSe{1-16}:CORRection:COLLect:WAVeguide:SHORT2:LOSS?
Description
Sets the waveguide calibration kit Short2 loss value on the indicated channel.
Returns the waveguide calibration kit Short2 loss value on the indicated channel.
Cmd Parameters
<NRf> The input parameter is in dB/mm.
Query Parameters
<NR3> The output parameter can be in dB/mm.
Range
MPND
Default Value
0
Syntax Example
:SENS1:CORR:COLL:WAV:SHORT2:LOSS 1.0E0
:SENS1:CORR:COLL:WAV:SHORT2:LOSS?
:SENSe{1-16}:CORRection:COLLect:WAVeguide:SHORT2:OFFSet <NRf>
:SENSe{1-16}:CORRection:COLLect:WAVeguide:SHORT2:OFFSet?
Description
Sets the waveguide calibration kit short2 offset on the indicated channel.
Returns the waveguide calibration kit label short2 offset on the indicated channel.
See Calibration Component Parameters for a complete listing of calibration components, connectors, and their Command Parameters.
Cmd Parameters
<NRf> The input parameter is in Meters.
Query Parameters
<NR3> The output parameter is in Meters.
Range
MPND
Default Value
See Calibration Component Parameters.
Syntax Example
:SENS1:CORR:COLL:WAV:SHORT2:OFFS 5.0E-9
:SENS1:CORR:COLL:WAV:SHORT2:OFFS?
:SENSe{1-16}:CORRection:COLLect:WAVeguide:SHORT3:FREQuency <NRf>
:SENSe{1-16}:CORRection:COLLect:WAVeguide:SHORT3:FREQuency?
Description
Sets the waveguide calibration kit Short3 reference frequency on the indicated channel.
Returns the waveguide calibration kit Short3 reference frequency on the indicated channel.
See Calibration Component Parameters for a complete listing of calibration components, connectors, and their Command Parameters.
Cmd Parameters
<NRf> The input parameter is in Hertz.
Query Parameters
<NR3> The output parameter is in Hertz.
Range
MPND
Default Value
See Calibration Component Parameters.
Syntax Example
:SENS1:CORR:COLL:WAV:SHORT3:FREQ 1.0E7
:SENS1:CORR:COLL:WAV:SHORT3:FREQ?
:SENSe{1-16}:CORRection:COLLect:WAVeguide:SHORT3:LOSS <NRf>
:SENSe{1-16}:CORRection:COLLect:WAVeguide:SHORT3:LOSS?
Description
Sets the waveguide calibration kit Short3 loss value on the indicated channel.
Returns the waveguide calibration kit Short3 loss value on the indicated channel.
Cmd Parameters
<NRf> The input parameter is in dB/mm.
Query Parameters
<NR3> The output parameter can be in dB/mm.
Range
MPND
Default Value
0
Syntax Example
:SENS1:CORR:COLL:WAV:SHORT3:LOSS 1.0E0
:SENS1:CORR:COLL:WAV:SHORT3:LOSS?
:SENSe{1-16}:CORRection:COLLect:WAVeguide:SHORT3:OFFSet <NRf>
:SENSe{1-16}:CORRection:COLLect:WAVeguide:SHORT3:OFFSet?
Description
Sets the waveguide calibration kit short3 offset length on the indicated channel.
Returns the waveguide calibration kit short3 offset length on the indicated channel.
See Calibration Component Parameters for a complete listing of calibration components, connectors, and their Command Parameters.
Cmd Parameters
<NRf> The input parameter is in Meters.
Query Parameters
<NR3> The output parameter is in Meters.
Range
MPND
Default Value
See Calibration Component Parameters.
Syntax Example
:SENS1:CORR:COLL:WAV:SHORT3:OFFS 5.0E9
:SENS1:CORR:COLL:WAV:SHORT3:OFFS?
:SENSe{1-16}:CORRection:COLLect:WAVeguide:SLOAD:MINF <NRf>
:SENSe{1-16}:CORRection:COLLect:WAVeguide:SLOAD:MINF?
Description
Sets the waveguide calibration kit sliding load minimum frequency on the indicated channel.
Returns the waveguide calibration kit sliding load minimum frequency on the indicated channel.
See Calibration Component Parameters for a complete listing of calibration components, connectors, and their Command Parameters.
Cmd Parameters
<NRf> The input parameter is in Hertz.
Query Parameters
<NR3> The output parameter is in Hertz.
Range
MPND
Default Value
2.00000000000E+009
Syntax Example
:SENS1:CORR:COLL:WAV:SLOAD:MINF 10.0E7
:SENS1:CORR:COLL:WAV:SLOAD:MINF?