:SENSe{1-16}:CORRection:COLLect:TRL:WAVeguide Subsystem
The :SENSe{1-16}:CORRection:COLLect:TRL:WAVeguide subsystem sets the calibration parameters and coefficients for waveguide line types. Use this subsystem to set waveguide parameters before starting a calibration.
Calibration Setup Subsystems
These subsystems are used during various phases of calibration configuration setup:
:SENSe{1-16}:CORRection:COLLect:TRL:WAVeguide:DIELectric <NRf>
:SENSe{1-16}:CORRection:COLLect:TRL:WAVeguide:DIELectric?
Description
Sets the TRL waveguide calibration kit dielectric value on the indicated channel.
Returns the TRL waveguide calibration kit dielectric value on the indicated channel.
Cmd Parameters
<NRf> The input parameter is a unitless number.
Query Parameters
NA
Query Output
<NR3> The output parameter is a unitless number.
Range
MPND
Default Value
1.00000000000E+000
Syntax Example
:SENS1:CORR:COLL:TRL:WAV:DIEL 1.2E0
:SENS1:CORR:COLL:TRL:WAV:DIEL?
:SENSe{1-16}:CORRection:COLLect:TRL:WAVeguide:FREQuency <NRf>
:SENSe{1-16}:CORRection:COLLect:TRL:WAVeguide:FREQuency?
Description
Sets the TRL waveguide calibration kit cutoff frequency on the indicated channel.
Returns the TRL waveguide calibration kit cutoff frequency on the indicated channel.
Cmd Parameters
<NRf> The input parameter is in Hertz.
Query Parameters
NA
Query Output
<NR3> The output parameter is in Hertz.
Range
MPND
Default Value
Syntax Example
:SENS1:CORR:COLL:TRL:WAV:FREQ 5.0E9
:SENS1:CORR:COLL:TRL:WAV:FREQ?
1-800-ANRITSU