:CALCulate{1-16}:REFerence Subsystem
The
:CALCulate{1-16}:REFerence subsystem commands configure various parameters related to the reference plane in line types such as coaxial, microstrip, and waveguides. Note that the per-port and per-trace reference plane extensions are independent. The commands for per-port are in this section while those for per-trace are in the
:CALCulate{1-16}:PARameter{1-16}:REFerence Subsystem . Emulated instruments have different allowed choices (Lightning only uses per-trace reference plane extensions and HP8510 only uses per-port reference plane extensions), so behaviors may change when a non-native language is used.
Calibration Setup Subsystems
These subsystems are used during various phases of calibration configuration setup:
Time Domain, Group Delay, and Reference Plane Subsystems
Related time domain, group delay, and reference place subsystems are:
:CALCulate{1-16}:REFerence:EXTension:COAXial:DIELectric <char>
:CALCulate{1-16}:REFerence:EXTension:COAXial:DIELectric?
Description
Sets the reference plane extension coaxial line dielectric type on the indicated channel.
Returns the reference plane extension coaxial line dielectric type on the indicated channel.
Cmd Parameters
<char> AIR | MICROporous | OTHER | POLYethylene | TEFLON
Query Parameters
<char> AIR | MICRO | OTHER | POLY | TEFLON
Range
NA
Default Value
AIR
Syntax Example
:CALC1:REF:EXT:COAX:DIEL AIR
:CALC1:REF:EXT:COAX:DIEL?
:CALCulate{1-16}:REFerence:EXTension:COAXial:DIELectric:OTHer <NRf>
:CALCulate{1-16}:REFerence:EXTension:COAXial:DIELectric:OTHer?
Description
Sets the reference plane extension coaxial line manual dielectric value on the indicated channel.
Returns the reference plane extension coaxial line manual dielectric value on the indicated channel.
Cmd Parameters
<NRf> The input parameter is a unitless number.
Query Parameters
<NR3> The output parameter is a unitless number.
Range
1 to 9.99E3
Default Value
1.00000000000E+000
Syntax Example
:CALC1:REF:EXT:COAX:DIEL:OTH 1.0E3
:CALC1:REF:EXT:COAX:DIEL:OTH?
:CALCulate{1-16}:REFerence:EXTension:COAXial:DIELectric:VALue?
Description
Query only.
Returns the reference plane extension coaxial line dielectric value on the indicated channel.
Cmd Parameters
NA
Query Parameters
<NR3> The output parameter is a unitless number.
Range
NA
Default Value
1.00064900000E+000
Syntax Example
:CALC1:REF:EXT:COAX:DIEL:VAL?
:CALCulate{1-16}:REFerence:EXTension:LINE <char>
:CALCulate{1-16}:REFerence:EXTension:LINE?
Description
Sets the reference plane extension line type on the indicated channel.
Returns the reference plane extension line type on the indicated channel.
Cmd Parameters
<char> COAXial | MICROstrip | NONDISpersive | WAVEguide
Query Parameters
<char> COAX | MICRO | NONDIS | WAVE
Range
NA
Default Value
COAX
Syntax Example
:CALC1:REF:EXT:LINE COAX
:CALC1:REF:EXT:LINE?
:CALCulate{1-16}:REFerence:EXTension:MICrostrip:DIELectric <NRf>
:CALCulate{1-16}:REFerence:EXTension:MICrostrip:DIELectric?
Description
Sets the reference plane extension microstrip substrate dielectric value on the indicated channel.
Returns the reference plane extension microstrip substrate dielectric value on the indicated channel.
Cmd Parameters
<NRf> The input parameter is a unitless number.
Query Parameters
<NR3> The output parameter is a unitless number.
Range
Command range: 1 to 10 Query range: 1 to 9.99E3
Default Value
9.96000000000E+000
Syntax Example
:CALC1:REF:EXT:MIC:DIEL 2.80
:CALC1:REF:EXT:MIC:DIEL?
:CALCulate{1-16}:REFerence:EXTension:MICrostrip:EFFective <NRf>
:CALCulate{1-16}:REFerence:EXTension:MICrostrip:EFFective?
Description
Sets the reference plane extension microstrip effective dielectric value on the indicated channel.
Returns the reference plan extension microstrip effective dielectric value on the indicated channel.
Cmd Parameters
<NRf> The input parameter is a unitless number.
Query Parameters
<NR3> The output parameter is a unitless number.
Range
MPND
Default Value
6.69000000000E+000
Syntax Example
:CALC1:REF:EXT:MIC:EFF 2.80
:CALC1:REF:EXT:MIC:EFF?
:CALCulate{1-16}:REFerence:EXTension:MICrostrip:THICKness <NRf>
:CALCulate{1-16}:REFerence:EXTension:MICrostrip:THICKness?
Description
Sets the reference plane extension microstrip substrate thickness on the indicated channel.
Returns the reference plane extension microstrip substrate thickness on the indicated channel.
Cmd Parameters
<NRf> The input parameter is in Meters.
Query Parameters
<NR3> The output parameter is in Meters.
Range
MPND
Default Value
2.54000000000E-004
Syntax Example
:CALC1:REF:EXT:MIC:THICK 3.0E-4
:CALC1:REF:EXT:MIC:THICK?
:CALCulate{1-16}:REFerence:EXTension:MICrostrip:WIDth <NRf>
:CALCulate{1-16}:REFerence:EXTension:MICrostrip:WIDth?
Description
Sets the reference plane extension microstrip width on the indicated channel.
Returns the reference plane extension microstrip width on the indicated channel.
Cmd Parameters
<NRf> The input parameter is in Meters.
Query Parameters
<NR3> The output parameter is in Meters.
Range
MPND
Default Value
2.38760000000E-004
Syntax Example
:CALC1:REF:EXT:MIC:WID 3.0E-4
:CALC1:REF:EXT:MIC:WID?
:CALCulate{1-16}:REFerence:EXTension:MICrostrip:Z0 <NRf>
:CALCulate{1-16}:REFerence:EXTension:MICrostrip:Z0?
Description
Sets the reference plane extension microstrip impedance on the indicated channel.
Returns the reference plane extension microstrip impedance on the indicated channel.
Cmd Parameters
<NRf> The input parameter is in Ohms.
Query Parameters
<NR3> The output parameter is in Ohms.
Range
MPND
Default Value
5.00000000000E+001
Syntax Example
:CALC1:REF:EXT:MIC:Z0 7.5E1
:CALC1:REF:EXT:MIC:Z0?
:CALCulate{1-16}:REFerence:EXTension:PARameter <char>
:CALCulate{1-16}:REFerence:EXTension:PARameter?
Description
Sets the reference extension parameter type Port or Trace.
Returns the reference extension parameter type Port or Trace.
Cmd Parameters
<char> PORT | TRACe
Query Parameters
NA
Output
<char> PORT | TRAC
Range
NA
Default
PORT
Syntax Example
:CALC1:REF:EXT:PAR <char>
:CALC1:REF:EXT:PAR?
:CALCulate{1-16}:REFerence:EXTension:PORT{1-4}:AUTOmatic
Description
The use of Port 3 or Port 4 requires a 4-port VNA instrument.
Calculates and applies the reference plane extension delay for the indicated port on the indicated channel.
No query.
Cmd Parameters
NA
Query Parameters
NA
Range
NA
Default Value
NA
Syntax Example
:CALC1:REF:EXT:PORT1:AUTO
:CALCulate{1-16}:REFerence:EXTension:PORT{1-4}:DISTance <NRf>
:CALCulate{1-16}:REFerence:EXTension:PORT{1-4}:DISTance?
Description
The use of Port 3 or Port 4 requires a 4-port VNA instrument.
Sets the reference plane extension distance in meters for the indicated port on the indicated channel.
Returns the reference plane extension distance in meters for the indicated port on the indicated channel.
Cmd Parameters
<NRf> The input parameter is in Meters.
Query Parameters
<NR3> The output parameter is in Meters.
Range
MPND
Default Value
0
Syntax Example
:CALC1:REF:EXT:PORT1:DIST 5E-4
:CALC1:REF:EXT:PORT1:DIST?
:CALCulate{1-16}:REFerence:EXTension:PORT{1-4}:FDEPendent:AUTOmatic
Description
Calculates and applies the Reference Plane extension for length and loss for the indicated port on the indicated channel
No query.
Cmd Parameters
NA
Query Parameters
NA
Range
NA
Default Value
NA
Syntax Example
:CALC1:REF:EXT:PORT1:FDEP:AUTO
:CALCulate{1-16}:REFerence:EXTension:PORT{1-4}:FDEPendent:EXPonent <NRf>
:CALCulate{1-16}:REFerence:EXTension:PORT{1-4}:FDEPendent:EXPonent?
Description
Sets the frequency-dependent exponent (the power of the frequency dependence of the loss that will be assumed) for the indicated channel and port. Values are usually between 0.01 to 10)
Returns the frequency-dependent exponent for the indicated channel and port.
Cmd Parameters
<NRf> The input parameter is a unitless number.
Query Parameters
NA
Query Output
<NR3> The output parameter is a unitless number.
Range
0.1 to 10
Default Value
0.5
Syntax Example
:CALC1:REF:EXT:PORT1:FDEP:EXP 5
:CALC1:REF:EXT:PORT1:FDEP:EXP?
:CALCulate{1-16}:REFerence:EXTension:PORT{1-4}:FDEPendent:FREQuency <NRf>
:CALCulate{1-16}:REFerence:EXTension:PORT{1-4}:FDEPendent:FREQuency?
Description
Sets the frequency-dependent reference frequency for the indicated channel and port. This is the frequency at which loss occurs; 0 denotes constant loss.
Returns the frequency-dependent reference frequency for the indicated channel and port.
Cmd Parameters
<NRf> The input parameter is in Hertz.
Query Parameters
NA
Query Output
<NR3> The output parameter is in Hertz.
Range
0 to 99 THz
Default Value
0
Syntax Example
:CALC1:REF:EXT:PORT1:FDEP:FREQ 5E9
:CALC1:REF:EXT:PORT1:FDEP:FREQ?
:CALCulate{1-16}:REFerence:EXTension:PORT{1-4}:FDEPendent:LOSS <NRf>
:CALCulate{1-16}:REFerence:EXTension:PORT{1-4}:FDEPendent:LOSS?
Description
Sets the frequency-dependent reference loss for the indicated channel and port. Usually greater than 0.
Returns the frequency-dependent reference loss for the indicated channel and port.
Cmd Parameters
<NRf> The input parameter is in dB.
Query Parameters
<NR3> The output parameter is in dB.
Range
–1000 to 1000
Default Value
0
Syntax Example
:CALC1:REF:EXT:PORT1:FDEP:LOSS 5
:CALC1:REF:EXT:PORT1:FDEP:LOSS?
:CALCulate{1-16}:REFerence:EXTension:PORT{1-4}:FDEPendent:MSUPpression <char>
:CALCulate{1-16}:REFerence:EXTension:PORT{1-4}:FDEPendent:MSUPpression?
Description
Sets on/off state of the mismatch suppression for reference plane for the indicated channel and port. When on, ensures that even on mismatch ripple peaks, the result will not show gain.
Returns the on/off state of the mismatch suppression for reference plane for the indicated channel and port.
Cmd Parameters
<char> 1 | 0 | ON | OFF
Query Parameters
NA
Query Output
<char> 1 | 0
Range
NA
Default Value
0
Syntax Example
:CALC1:REF:EXT:PORT1:FDEP:MSUP OFF
:CALC1:REF:EXT:PORT1:FDEP:MSUP?
:CALCulate{1-16}:REFerence:EXTension:PORT{1-4}:LOSS <NRf>
:CALCulate{1-16}:REFerence:EXTension:PORT{1-4}:LOSS?
Description
The use of Port 3 or Port 4 requires a 4-port VNA instrument.
Sets the reference plane extension loss in dB for the indicated port on the indicated channel.
Returns the reference plane extension loss in dB for the indicated port on the indicated channel.
Cmd Parameters
<NRf> The input parameter is in dB.
Query Parameters
<NR3> The output parameter is in dB.
Range
-1E3 to +1E3
Default Value
0.00000000000E+000
Syntax Example
:CALC1:REF:EXT:PORT1:LOSS 3E0
:CALC1:REF:EXT:PORT1:LOSS?
:CALCulate{1-16}:REFerence:EXTension:PORT{1-4}:PHAse <NRf>
:CALCulate{1-16}:REFerence:EXTension:PORT{1-4}:PHAse?
Description
The use of Port 3 or Port 4 requires a 4-port VNA instrument.
Sets the reference plane extension phase offset in degrees for the indicated port on the indicated channel.
The allowable range is -360 degrees to +360 degrees. If the -360 value is exceeded, the instrument truncates the input to -360; if the +360 value is exceeded, the instrument truncates the input to +360.
Returns the reference plane extension phase offset in degrees for the indicated port on the indicated channel.
Cmd Parameters
<NRf> The input parameter is in Degrees.
Query Parameters
<NR3> The output parameter is in Degrees.
Range
-360 to +360
Default Value
0.000000E+000
Syntax Example
:CALC1:REF:EXT:PORT1:PHA 1.5E1
:CALC1:REF:EXT:PORT1:PHA?
:CALCulate{1-16}:REFerence:EXTension:PORT{1-4}:TERMinator <char>
:CALCulate{1-16}:REFerence:EXTension:PORT{1-4}:TERMinator?
Description
Sets the terminator selection for reference plane extension for the indicated channel and port.
Returns the terminator selection for reference plane extension for the indicated channel and port.
Cmd Parameters
<char> GENeral | OPEN | SHORt
Query Parameters
NA
Query Output
<NR3> GEN | OPEN | SHOR
Range
NA
Default Value
GEN
Syntax Example
:CALC1:REF:EXT:PORT1:TERM SHOR
:CALC1:REF:EXT:PORT1:TERM?
:CALCulate{1-16}:REFerence:EXTension:PORT{1-4}:TIMe <NRf>
:CALCulate{1-16}:REFerence:EXTension:PORT{1-4}:TIMe?
Description
The use of Port 3 or Port 4 requires a 4-port VNA instrument.
Sets the reference plane extension in time in seconds for the indicated port on the indicated channel.
Returns the reference plane extension in time in seconds for the indicated port on the indicated channel.
Cmd Parameters
<NRf> The input parameter is in Seconds.
Query Parameters
<NR3> The output parameter is in Seconds.
Range
MPND
Default Value
0.00000000000E+000
Syntax Example
:CALC1:REF:EXT:PORT1:TIM 5.0E-2
:CALC1:REF:EXT:PORT1:TIM?
:CALCulate{1-16}:REFerence:EXTension:WAVeguide:DIELectric <NRf>
:CALCulate{1-16}:REFerence:EXTension:WAVeguide:DIELectric?
Description
Sets the reference plane extension waveguide dielectric value on the indicated channel.
Returns the reference plane extension waveguide dielectric value on the indicated channel.
Cmd Parameters
<NRf> The input parameter is a unitless number.
Query Parameters
<NR3> The output parameter is a unitless number.
Range
MPND
Default Value
1.00000000000E+000
Syntax Example
:CALC1:REF:EXT:WAV:DIEL 2.5E0
:CALC1:REF:EXT:WAV:DIEL?
:CALCulate{1-16}:REFerence:EXTension:WAVeguide:FREQuency <NRf>
:CALCulate{1-16}:REFerence:EXTension:WAVeguide:FREQuency?
Description
Sets the reference plane extension waveguide cutoff frequency on the indicated channel.
Returns the reference plane extension waveguide cutoff frequency on the indicated channel.
Cmd Parameters
<NRf> The input parameter is in Hertz.
Query Parameters
<NR3> The output parameter is in Hertz.
Range
MPND
Default Value
0.00000000000E+000
Syntax Example
:CALC1:REF:EXT:WAV:FREQ 1E10
:CALC1:REF:EXT:WAV:FREQ?
1-800-ANRITSU
10450-00040 Rev. AD