:SENSe{1-16}:CORRection:COLLect[:CALa]:PORT Subsystem
These commands set either a 2-Port Calibration for a 2-port or 4-port VNA, or the first 2-Port Calibration (called the :CALa calibration) for a 4-port VNA. In this case, the second 2-Port Calibration (called the :CALB calibration) is specified by this subsystem
Calibration Subsystems with Actual Calibration
Related calibration subsystems that perform actual calibrations are:
Calibration Abbreviations
The calibration abbreviations and their calibration types are:
• :1P2PF refers to a one-path two-port calibration forward direction
• :1P2PR refers to a one path two port calibration reverse direction
• :FULL1 refers to a full one port calibration
• :FULL2 refers to a full two port calibration
• :FULL3 refers to a full three port calibration
• :FULLB refers to a full one port reflection calibration on both ports
• :FULL4 refers to a full four port calibration
• :RESP1 refers to a one port response calibration
• :RESPB refers to a one port response calibration both ports
• :TFRB refers to a transmission frequency response calibration both directions
• :TFRF refers to a transmission frequency response calibration forward direction
• :TFRR refers to a transmission frequency response calibration reverse direction
Each calibration simulation type command is described in greater detail in the individual command descriptions below.
Most calibration commands of this type do not have a directly related query. To query the state of these commands, use:
The :SENSe{1-16}:CORRection:COLLect:PORT subsystem commands start an actual instrument calibration.
:SENSe{1-16}:CORRection:COLLect[:CALa]:PORT{12 | 13 | 14 | 23 | 24 | 34}:1P2PF
Description
The use of Port 3 and/or Port 4 requires a 4-port VNA instrument.
Sets the first calibration (:CALa ) to One-Path 2-Port Forward Calibration.
If a second 2-Port Calibration (:CALB ) is required, use the following command:
No query.
Cmd Parameters
NA
Range
NA
Default
NA
Syntax Example
:SENS1:CORR:COLL:PORT1:1P2PF
:SENSe{1-16}:CORRection:COLLect[:CALa]:PORT{12 | 13 | 14 | 23 | 24 | 34}:1P2PR
Description
The use of Port 3 and/or Port 4 requires a 4-port VNA instrument.
Sets the first 2-Port calibration to One-Path 2-Port Reverse on the indicated channel and port pair.
No query
Cmd Parameters
NA
Range
NA
Default
NA
Syntax Example
:SENS1:CORR:COLL:PORT1:1P2PR
:SENSe{1-16}:CORRection:COLLect[:CALa]:PORT{12 | 13 | 14 | 23 | 24 | 34}:FULL2
Description
The use of Port 3 and/or Port 4 requires a 4-port VNA instrument.
Sets the first 2-Port calibration to Full 2-Port on the indicated channel and port pair.
No query
Cmd Parameters
NA
Range
NA
Default
NA
Syntax Example
:SENS1:CORR:COLL:PORT1:FULL2