VectorStar™ MS464xB Series Microwave Vector Network Analyzer Programming Manual : SCPI Commands : :SENSe{1-16}:CORRection:COLLect:PORT Subsystem
 
:SENSe{1-16}:CORRection:COLLect:PORT Subsystem
The :SENSe{1-16}:CORRection:COLLect:PORT subsystem commands start an actual instrument calibration.
Calibration Subsystems with Actual Calibrations
Related calibration subsystems that perform actual calibrations are:
:SENSe{1-16}:ABORtcal Subsystem
:SENSe{1-16}:CORRection:COLLect:CALB Subsystem – 4-Port VNAs
:SENSe{1-16}:CORRection:COLLect:FLEXible Subsystem
:SENSe{1-16}:CORRection:COLLect:FULL4 Subsystem – 4-Port VNAs
:SENSe{1-16}:CORRection:COLLect:LRL[:CALa] Subsystem
:SENSe{1-16}:CORRection:COLLect:PORT Subsystem
:SENSe{1-16}:CORRection:COLLect[:CALa]:PORT Subsystem
Calibration Type Abbreviations
The commands in this subsystem use the following abbreviations for different calibration methods:
1P2PF = One-path two-port calibration, forward direction
1P2PR = One path two port calibration, reverse direction
FULL1 = Full one port calibration
FULL2 = Full two port calibration
FULLB = Full one port reflection calibration, both ports
RESP1 = One port response calibration
RESPB = One port response calibration, both ports
TFRB = Transmission frequency response calibration, both directions
TFRF = Transmission frequency response calibration, forward direction
TFRR = Transmission frequency response calibration, reverse direction
:SENSe{1-16}:CORRection:COLLect:PORT{1 | 2 | 3 | 4 | 12 | 13 | 14 | 23 | 24 | 34 | 123 | 124 | 134 | 234 | 1234}:FULL1
Description
The use of Port 3 and/or Port 4 requires a 4-port VNA instrument.
Sets Full 1-Port Calibration as the calibration type on the indicated channel and ports selected. The full 1-port calibration can be performed on any combination of ports by listing them using as PORTw | wx | wxy | wxyz where:
w = the first port
x = the second port
y = the third port
z = the fourth port
For example, to perform a FULL1 calibration on Channel 1 and Port 1, 2, and 4, the command is:
SENS1:CORR:COLL:PORT124:FULL1
No query.
Cmd Parameters
NA
Range
NA
Default
NA
Syntax Example
:SENS1:CORR:COLL:PORT23:FULL1
:SENSe{1-16}:CORRection:COLLect:PORT{1 | 2 | 3 | 4 | 12 | 13 | 14 | 23 | 24 | 34 | 123 | 124 | 134 | 234 | 1234}:RESP1
Description
The use of Port 3 and/or Port 4 requires a 4-port VNA instrument.
Sets 1-Port Response Calibration as the calibration type on the indicated channel and ports selected. The response calibration can be performed on one or more ports by listing them using as PORTw | wx | wxy | wxyz where:
w = the first port
x = the second port
y = the third port
z = the fourth port
For example, to perform a response calibration on Channel 1 and Ports 1, 2, and 3, the command is:
SENS1:CORR:COLL:PORT123:RESP1
No query.
Cmd Parameters
NA
Range
NA
Default
NA
Syntax Example
:SENS1:CORR:COLL:PORT24:RESP1
:SENSe{1-16}:CORRection:COLLect:PORT{12 | 13 | 14 | 23 | 24 | 34}:ISOL
Description
For a 2-port VNA, the available port pair is 12.
Port pairs using Port 3 and/or Port 4 require a 4-port VNA instrument.
Initiates collection of the isolation standard data for the calibration on the indicated channel.
No query.
Cmd Parameters
NA
Range
NA
Default Value
NA
Syntax Example
:SENS1:CORR:COLL:PORT12:ISOL
:SENSe{1-16}:CORRection:COLLect:PORT{12 | 13 | 14 | 23 | 24 | 34}:LSELection <char>
:SENSe{1-16}:CORRection:COLLect:PORT{12 | 13 | 14 | 23 | 24 | 34}:LSELection?
Description
Sets the calibration line selection on the selected port-pair of the indicated channel.
Returns the line selection on the selected port-pair of the indicated channel.
Cmd Parameters
<char1> THRu | RECiprocal | S2PThru
Query Parameters
NA
Query Output
<char1> THR | REC | S2PT
Range
NA
Default Value
THR
Syntax Example
:SENS1:CORR:COLL:PORT12:LSEL REC
:SENS1:CORR:COLL:PORT12:LSEL?
:SENSe{1-16}:CORRection:COLLect:PORT{12 | 13 | 14 | 23 | 24 | 34}:S2PThru:FILe <string>
:SENSe{1-16}:CORRection:COLLect:PORT{12 | 13 | 14 | 23 | 24 | 34}:S2PThru:FILe?
Description
Loads an s2p file for thru and saves the loaded file in the default location:
'C:\AnritsuVNA\Data\S2PThruCal'
Returns the loaded s2p file path.
Cmd Parameters
<string> Filename and path in the form: 'x:\directory\filename.s2p', where
x:\directory\filename.s2p must exist. See Programming the VectorStar Series VNA, Notational Conventions for more information.
Query Parameters
NA
Query Output
<char1> Filename and path in the form: x:\directory\filename.s2p
Range
NA
Default Value
NA
Syntax Example
:SENS:CORR:COLL:PORT12:S2PT:FIL 'C:\AnritsuVNA\Data\S2PThruCal.s2p'
:SENS:CORR:COLL:PORT12:S2PT:FIL?
:SENSe{1-16}:CORRection:COLLect:PORT{12 | 13 | 14 | 23 | 24 | 34}:S2PThru:ADAPter[:STATe] <char>
:SENSe{1-16}:CORRection:COLLect:PORT{12 | 13 | 14 | 23 | 24 | 34}:S2PThru:ADAPter[:STATe]?
Description
Sets the S2P Thru generic adapter flag state to on/off on the selected port-pair of the indicated channel.
Returns the S2P Thru generic adapter flag status of the selected port-pair of the indicated channel.
Cmd Parameters
<char> 1 | 0 | ON | OFF
Query Parameters
NA
Query Output
<char> 1 | 0
Range
NA
Default Value
0
Syntax Example
:SENS:CORR:COLL:PORT12:S2PT:ADAP 1
:SENS:CORR:COLL:PORT12:S2PT:ADAP?
:SENSe{1-16}:CORRection:COLLect:PORT{12 | 13 | 14 | 23 | 24 | 34}:TFRB
Description
The use of Port 3 and/or Port 4 requires a 4-port VNA instrument.
Adds Transmission Frequency Response Calibration Both directions on the indicated Channel and port pair.
The Transmission Frequency Response calibration can be any combination of port pairs, and any combination of Forward only (TFRF), Reverse only (TFRR) or both (TFRB). This command adds a Transmission Frequency Response Calibration for Both directions.
No query.
Cmd Parameters
NA
Range
NA
Default
NA
Syntax Example
:SENS1:CORR:COLL:PORT13:TFRB
:SENSe{1-16}:CORRection:COLLect:PORT{12 | 13 | 14 | 23 | 24 | 34}:TFRF
Description
The use of Port 3 and/or Port 4 requires a 4-port VNA instrument.
Adds Transmission Frequency Response Calibration Forward direction on the indicated Channel and port pair.
The Transmission Frequency Response calibration can be any combination of port pairs, and any combination of Forward only (TFRF), Reverse only (TFRR) or both (TFRB). This command adds a Transmission Frequency Response Calibration Forward direction.
No query.
Cmd Parameters
NA
Range
NA
Default
NA
Syntax Example
:SENS1:CORR:COLL:PORT24:TFRF
:SENSe{1-16}:CORRection:COLLect:PORT{12 | 13 | 14 | 23 | 24 | 34}:TFRR
Description
The use of Port 3 and/or Port 4 requires a 4-port VNA instrument.
Adds Transmission Frequency Response Calibration Reverse direction on the indicated Channel and port pair.
The Transmission Frequency Response calibration can be any combination of port pairs, and any combination of Forward only (TFRF), Reverse only (TFRR) or both (TFRB). This command adds a Transmission Frequency Response Calibration Reverse direction.
No query.
Cmd Parameters
NA
Range
NA
Default
NA
Syntax Example
:SENS1:CORR:COLL:PORT12:TFRR
:SENSe{1-16}:CORRection:COLLect:PORT{12 | 13 | 14 | 23 | 24 | 34}:THRu
Description
The use of Port 3 and/or Port 4 requires a 4-port VNA instrument.
Initiates collection of the through (thru) standard data for the calibration on the indicated channel. For 2-port VNAs, the available port pair is 12.
No query.
Cmd Parameters
NA
Range
NA
Default Value
NA
Syntax Example
:SENS1:CORR:COLL:PORT12:THR
:SENSe{1-16}:CORRection:COLLect:PORT{12 | 13 | 14 | 23 | 24 | 34}:THRu:FREQuency <NRf>
:SENSe{1-16}:CORRection:COLLect:PORT{12 | 13 | 14 | 23 | 24 | 34}:THRu:FREQuency?
Description
The use of Port 3 and/or Port 4 requires a 4-port VNA instrument.
Sets the through-line (thru-line) reference frequency of the loss on the selected port-pair of the indicated channel.
Returns the thru-line reference frequency of the loss on the selected port-pair of the indicated channel.
Cmd Parameters
<NRf> The input parameter is in Hertz.
Query Parameters
<NR3> The output parameter is in Hertz.
Range
MPND
Default Value
See Calibration Component Parameters.
Syntax Example
:SENS1:CORR:COLL:PORT12:THR:FREQ 1.0E6
:SENS1:CORR:COLL:PORT12:THR:FREQ?
:SENSe{1-16}:CORRection:COLLect:PORT{12 | 13 | 14 | 23 | 24 | 34}:THRu:LABEL <string>
:SENSe{1-16}:CORRection:COLLect:PORT{12 | 13 | 14 | 23 | 24 | 34}:THRu:LABEL?
Description
The use of Port 3 and/or Port 4 requires a 4-port VNA instrument.
Sets the through-line (thru-line) label on the selected port-pair of the indicated channel.
On 2-port VNAs, the available port pair is 12.
Returns the thru-line label on the selected port-pair of the indicated channel.
Cmd Parameters
<string> The input parameter is any combination of numbers and letters.
See definition of <string>.
Query Parameters
<char> The output parameter can be any combination of numbers and letters.
See definition of <char>.
Range
NA
Default Value
NA
Syntax Example
:SENS1:CORR:COLL:PORT12:THR:LABEL 'IC7000'
:SENS1:CORR:COLL:PORT12:THR:LABEL?
:SENSe{1-16}:CORRection:COLLect:PORT{12 | 13 | 14 | 23 | 24 | 34}:THRu:LENGth <NRf>
:SENSe{1-16}:CORRection:COLLect:PORT{12 | 13 | 14 | 23 | 24 | 34}:THRu:LENGth?
Description
The use of Port 3 and/or Port 4 requires a 4-port VNA instrument.
Sets the through-line (thru-line) length on the selected port-pair of the indicated channel.
The port-pair for 2-port VNAs is Port 1-2.
Returns the thru-line length on the selected port-pair of the indicated channel.
Cmd Parameters
<NRf> The input parameter is in Meters.
Query Parameters
<NR3> The output parameter is in Meters.
Range
MPND
Default Value
See Calibration Component Parameters.
Syntax Example
:SENS1:CORR:COLL:PORT12:THR:LENG 1.0E0
:SENS1:CORR:COLL:PORT12:THR:LENG?
:SENSe{1-16}:CORRection:COLLect:PORT{12 | 13 | 14 | 23 | 24 | 34}:THRu:LOSS <NRf>
:SENSe{1-16}:CORRection:COLLect:PORT{12 | 13 | 14 | 23 | 24 | 34}:THRu:LOSS?
Description
The use of Port 3 and/or Port 4 requires a 4-port VNA instrument.
Sets the through-line (thru-line) loss on the selected port-pair of the indicated channel.
The port-pair for 2-port VNAs is Port 1-2.
Returns the thru-line loss on the selected port-pair of the indicated channel.
Cmd Parameters
<NRf> The input parameter is in dB/mm.
Query Parameters
<NR3> The output parameter is in dB/mm.
Range
MPND
Default Value
See Calibration Component Parameters.
Syntax Example
:SENS1:CORR:COLL:PORT12:THR:LOSS 3.0E0
:SENS1:CORR:COLL:PORT12:THR:LOSS?
:SENSe{1-16}:CORRection:COLLect:PORT{12 | 13 | 14 | 23 | 24 | 34}:THRu:RECIProcal <char>
:SENSe{1-16}:CORRection:COLLect:PORT{12 | 13 | 14 | 23 | 24 | 34}:THRu:RECIProcal?
Description
The use of Port 3 and/or Port 4 requires a 4-port VNA instrument.
Sets the through-line (thru-line) use reciprocal flag on the selected port-pair of the indicated channel.
The port-pair for 2-port VNAs is Port 1-2.
Returns the thru-line use reciprocal flag on the selected port-pair of the indicated channel.
Cmd Parameters
<char> 1 | 0 | ON | OFF
Query Parameters
<char> 1 | 0
Range
NA>
Default Value
0
Syntax Example
:SENS1:CORR:COLL:PORT12:THR:RECIP ON
:SENS1:CORR:COLL:PORT12:THR:RECIP?
:SENSe{1-16}:CORRection:COLLect:PORT{12 | 13 | 14 | 23 | 24 | 34}:THRu:SERial <string>
:SENSe{1-16}:CORRection:COLLect:PORT{12 | 13 | 14 | 23 | 24 | 34}:THRu:SERial?
Description
The use of Port 3 and/or Port 4 requires a 4-port VNA instrument.
Sets the through-line (thru-line) serial number on the selected port-pair of the indicated channel.
The port-pair for 2-port VNAs is Port 1-2.
Returns the thru-line serial number on the selected port-pair of the indicated channel.
Cmd Parameters
<string> The input parameter is any combination of numbers and letters. See definition of <string>.
Query Parameters
<char> The output parameter can be any combination of numbers and letters.
Range
NA
Default Value
NA
Syntax Example
:SENS1:CORR:COLL:PORT12:THR:SER '123456'
:SENS1:CORR:COLL:PORT12:THR:SER?
:SENSe{1-16}:CORRection:COLLect:PORT{12 | 13 | 14 | 23 | 24 | 34}:THRu:UPDate:ECAL[:STATe]
:SENSe{1-16}:CORRection:COLLect:PORT{12 | 13 | 14 | 23 | 24 | 34}:THRu:UPDate:ECAL[:STATe]?
Description
Sets whether to use the AutoCal thru for thru update for the indicated port pair of the indicated channel.
Returns whether the AutoCal thru will be used for thru update for the indicated port pair in the given channel.
Cmd Parameters
<char> ON | OFF | 1 | 0
Query Parameters
NA
Query Parameters
1 | 0
Range
1 | 0
Default Value
0
Syntax Example
:SENS1:CORR:COLL:PORT12:THR:UPD:ECAL ON
:SENS1:CORR:COLL:PORT12:THR:UPD:ECAL?
:SENSe{1-16}:CORRection:COLLect:PORT{12 | 13 | 14 | 23 | 24 | 34}:THRu:UPDate
Description
Updates the collection of the through (thru) standard data for the calibration.
No query.
Cmd Parameters
NA
Range
NA
Default Value
NA
Syntax Example
:SENS1:CORR:COLL:PORT12:THR:UPD
:SENSe{1-16}:CORRection:COLLect:PORT{12 | 13 | 14 | 23 | 24 | 34}:THRu:Z0 <NRf>
:SENSe{1-16}:CORRection:COLLect:PORT{12 | 13 | 14 | 23 | 24 | 34}:THRu:Z0?
Description
The use of Port 3 and/or Port 4 requires a 4-port VNA instrument.
Sets the through-line (thru-line) impedance on the selected port-pair of the indicated channel.
The port-pair for 2-port VNAs is Port 1-2.
Returns the thru-line impedance on the selected port-pair of the indicated channel.
Cmd Parameters
<NRf> The input parameter is in Ohms.
Query Parameters
<NR3> The output parameter is in Ohms.
Range
1E-4 to 1E10
Default Value
See Calibration Component Parameters.
Syntax Example
:SENS1:CORR:COLL:PORT12:THR:Z0 7.5E1
:SENS1:CORR:COLL:PORT12:THR:Z0?
:SENSe{1-16}:CORRection:COLLect:PORT{123 | 124 | 134 | 234}:FULL3
Description
This command requires a 4-port VNA instrument.
Sets the calibration type to Full Three Port for the indicated channel and port triplet where:
123 = Port Triplet of Port 1, Port 2, and Port 3
124 = Port Triplet of Port 1, Port 2, and Port 4
134 = Port Triplet of Port 1, Port 3, and Port 4
234 = Port Triplet of Port 2, Port 3, and Port 4
No query.
Cmd Parameters
NA
Range
NA
Default
NA
Syntax Example
:SENS1:CORR:COLL:PORT234:FULL3
:SENSe{1-16}:CORRection:COLLect:PORT{1-4}:CONNector <char>
:SENSe{1-16}:CORRection:COLLect:PORT{1-4}:CONNector?
Description
The use of Port 3 and/or Port 4 requires a 4-port VNA instrument.
Sets the connector type for the indicated port on the indicated channel.
Connector availabilities are based on active cal method. The connector types are:
CM2-V = Anritsu 2.4 mm (male)
CM2 = 2.4 mm (male)
CM3 = GPC 3.5 (male)
CMK = K (male)
CMN = K (male)
CMS = SMA (male)
CMC = TNC (male)
CMV = V (male)
CM1 = 1 mm (male)
CM0.8 = 0.8mm (male)
CM716 = 7/16 (male)
CNG = GPC7 (none)
CMN75 = N 75 Ohm (male)
CMUn = User-defined male connector from 1 to 32, where: CMU1 = User-defined male 1, CMU2 = User-defined male 2, CMU3 = User-defined male 3, etc.
CF2-V = Anritsu 2.4 mm (female)
CF2 = 2.4 mm (female)
CF3 = GPC 3.5 (female)
CFK = K (female)
CFN = N (female)
CFS = SMA (female)
CFC = TNC (female)
CFV = V (female)
CF1 = 1 mm (female)
CF0.8 = 0.8mm (female)
CF716 = 7/16 (female)
CFN75 = N 75 Ohm (female)
CFUn = User-defined female connector from 1 to 32, where: CFU1 = User-defined female 1, CFU2 = User-defined female 2, CFU3 = User-defined female 3, etc.
Returns the connector type for the indicated port on the indicated channel.
See Table: Connector Type Abbreviations and Descriptions for a complete listing of calibration components, connectors, and their Command Parameters.
Cmd Parameters
<char> CM1 | CMV | CMK | CM2 | CM2-V | CM3 | CMS | CMN | CMN75 | CNG | CM716 | CMC | CM0.8 | CMU1 | CMU2 | CMU3 | CMU4 | CMU5 | CMU6 | CMU7 | CMU8 | CMU9 | CMU10 | CMU11 | CMU12 | CMU13 | CMU14 | CMU15 | CMU16 | CMU17 | CMU18 | CMU19 | CMU20 | CMU21 | CMU22 | CMU23 | CMU24 | CMU25 | CMU26 | CMU27 | CMU28 | CMU29 | CMU30 | CMU31 | CMU32 | CF1 | CFV | CFK | CF2 | CF2-V | CF3 | CFS | CFN | CFN75 | CF716 | CFC | CF0.8 | CFU1 | CFU2 | CFU3 | CFU4 | CFU5 | CFU6 | CFU7 | CFU8 | CFU9 | CFU10 | CFU11 | CFU12 | CFU13 | CFU14 | CFU15 | CFU16 | CFU17 | CFU18 | CFU19 | CFU20 | CFU21 | CFU22 | CFU23 | CFU24 | CFU25 | CFU26 | CFU27 | CFU28 | CFU29 | CFU30 | CFU31 | CFU32
Query Parameters
<char> CM1 | CMV | CMK | CM2 | CM2-V | CM3 | CMS | CMN | CMN75 | CNG | CM716 | CMC | CM0.8 | CMU1 | CMU2 | CMU3 | CMU4 | CMU5 | CMU6 | CMU7 | CMU8 | CMU9 | CMU10 | CMU11 | CMU12 | CMU13 | CMU14 | CMU15 | CMU16 | CMU17 | CMU18 | CMU19 | CMU20 | CMU21 | CMU22 | CMU23 | CMU24 | CMU25 | CMU26 | CMU27 | CMU28 | CMU29 | CMU30 | CMU31 | CMU32 | CF1 | CFV | CFK | CF2 | CF2-V | CF3 | CFS | CFN | CFN75 | CF716 | CFC | CF0.8 | CFU1 | CFU2 | CFU3 | CFU4 | CFU5 | CFU6 | CFU7 | CFU8 | CFU9 | CFU10 | CFU11 | CFU12 | CFU13 | CFU14 | CFU15 | CFU16 | CFU17 | CFU18 | CFU19 | CFU20 | CFU21 | CFU22 | CFU23 | CFU24 | CFU25 | CFU26 | CFU27 | CFU28 | CFU29 | CFU30 | CFU31 | CFU32
Range
NA
Default Value
CMV
Syntax Example
:SENS1:CORR:COLL:PORT1:CONN CM2
:SENS1:CORR:COLL:PORT1:CONN?
:SENSe{1-16}:CORRection:COLLect:PORT{1-4}:LOAD
Description
The use of Port 3 and/or Port 4 requires a 4-port VNA instrument.
Initiates collection of the Load Standard data for the calibration on the indicated channel and port.
No query.
Cmd Parameters
NA
Range
NA
Default Value
NA
Syntax Example
:SENS1:CORR:COLL:PORT1:LOAD
:SENSe{1-16}:CORRection:COLLect:PORT{1-4}:LOAD:SELect <char>
:SENSe{1-16}:CORRection:COLLect:PORT{1-4}:LOAD:SELect?
Description
The use of Port 3 and/or Port 4 requires a 4-port VNA instrument.
Sets the load standard of LOAD1 or LOAD2 to be used for calibration on the indicated channel and port.
Returns the selected load standard of LOAD1 or LOAD2 to be used for calibration on the indicated channel and port.
Cmd Parameters
<char> LOAD1 | LOAD2
Query Parameters
<char> LOAD1 | LOAD2
Range
NA
Default Value
NA
Syntax Example
:SENS1:CORR:COLL:PORT1:LOAD:SEL LOAD2
:SENS1:CORR:COLL:PORT1:LOAD:SEL?
:SENSe{1-16}:CORRection:COLLect:PORT{1-4}:LOAD:TYPe <char>
:SENSe{1-16}:CORRection:COLLect:PORT{1-4}:LOAD:TYPe?
Description
The use of Port 3 and/or Port 4 requires a 4-port VNA instrument.
Sets the Broadband Load Type as Fixed or Sliding for calibration on the indicated channel and port.
Returns the Broadband Load Type selection on the indicated channel and port.
Cmd Parameters
<char> FIXed | SLIDing
Query Parameters
<char> FIX | SLID
Range
NA
Default
FIX
Syntax Example
:SENS1:CORR:COLL:PORT1:LOAD:TYP FIX
:SENS1:CORR:COLL:PORT1:LOAD:TYP?
:SENSe{1-16}:CORRection:COLLect:PORT{1-4}:LOAD1:C0 <NRf>
:SENSe{1-16}:CORRection:COLLect:PORT{1-4}:LOAD1:C0?
Description
The use of Port 3 and/or Port 4 requires a 4-port VNA instrument.
Sets the Load1 standard C0 (C zero) capacitance on the selected port of the indicated channel.
The C0 coefficient is measured in Farads.
Returns the Load1 standard C0 capacitance on the selected port of the indicated channel.
See Calibration Component Parameters for a complete listing of calibration components, connectors, and their Command Parameters.
Cmd Parameters
<NRf> The input parameter is in Farads.
Cmd Parameters
<NR3> The output parameter is in Farads.
Range
0 to 1E12
Default Value
0.00000000000E+000
Syntax Example
:SENS1:CORR:COLL:PORT1:LOAD1:C0 3.0E-11
:SENS1:CORR:COLL:PORT1:LOAD1:C0?
:SENSe{1-16}:CORRection:COLLect:PORT{1-4}:LOAD1:C1 <NRf>
:SENSe{1-16}:CORRection:COLLect:PORT{1-4}:LOAD1:C1?
Description
The use of Port 3 and/or Port 4 requires a 4-port VNA instrument.
Sets the Load1 Standard C1 (C one) Coefficient on the selected port of the indicated channel.
The C1 coefficient is measured in Farads/Hertz.
Returns the Load1 Standard C1 Coefficient on the Selected port of the indicated channel.
Cmd Parameters
<NRf> The input parameter is in Farads/Hertz.
Query Parameters
<NR3> The output parameter is in Farads/Hertz.
Range
MPND
Default
0.00000000000E+000
Syntax Example
:SENS1:CORR:COLL:PORT1:LOAD1:C1 2.0E0
:SENS1:CORR:COLL:PORT1:LOAD1:C1?
:SENSe{1-16}:CORRection:COLLect:PORT{1-4}:LOAD1:C2 <NRf>
:SENSe{1-16}:CORRection:COLLect:PORT{1-4}:LOAD1:C2?
Description
The use of Port 3 and/or Port 4 requires a 4-port VNA instrument.
Sets the Load1 Standard C2 Coefficient on the selected port of the indicated channel.
The C2 coefficient is measured in Farads/Hertz^2.
Returns the Load1 Standard C2 Coefficient on the selected port of the indicated channel.
Cmd Parameters
<NRf> The input parameter is in Farads/Hertz^2 (Hertz E2).
Query Parameters
<NR3> The output parameter is in Farads/Hertz^2.
Range
MPND
Default
0.00000000000E+000
Syntax Example
:SENS1:CORR:COLL:PORT1:LOAD1:C2 2.0E0
:SENS1:CORR:COLL:PORT1:LOAD1:C2?
:SENSe{1-16}:CORRection:COLLect:PORT{1-4}:LOAD1:C3 <NRf>
:SENSe{1-16}:CORRection:COLLect:PORT{1-4}:LOAD1:C3?
Description
The use of Port 3 and/or Port 4 requires a 4-port VNA instrument.
Sets the Load1 Standard C3 Coefficient on the selected port of the indicated channel.
The C3 coefficient is measured in Farads/Hertz^3.
Returns the Load1 Standard C3 Coefficient on the selected port of the indicated channel.
Cmd Parameters
<NRf> The input parameter is in Farads/Hertz^3 (Hertz E3).
Query Parameters
<NR3> The output parameter is in Farads/Hertz^3.
Range
MPND
Default
0.00000000000E+000
Syntax Example
:SENS1:CORR:COLL:PORT1:LOAD1:C3 2.0E0
:SENS1:CORR:COLL:PORT1:LOAD1:C3?
:SENSe{1-16}:CORRection:COLLect:PORT{1-4}:LOAD1:L0 <NRf>
:SENSe{1-16}:CORRection:COLLect:PORT{1-4}:LOAD1:L0?
Description
The use of Port 3 and/or Port 4 requires a 4-port VNA instrument.
Sets the Load1 standard L0 (L zero) inductance on the selected port of the indicated channel.
The L0 coefficient is measured in Henrys.
Returns the Load1 standard L0 inductance on the selected port of the indicated channel.
See Calibration Component Parameters for a complete listing of calibration components, connectors, and their Command Parameters.
Cmd Parameters
<NRf> The input parameter is in Henrys.
Query Parameters
<NR3> The output parameter is in Henrys.
Range
MPND
Default Value
0.00000000000E+000
Syntax Example
:SENS1:CORR:COLL:PORT1:LOAD1:L0 2.0E0
:SENS1:CORR:COLL:PORT1:LOAD1:L0?
:SENSe{1-16}:CORRection:COLLect:PORT{1-4}:LOAD1:L1 <NRf>
:SENSe{1-16}:CORRection:COLLect:PORT{1-4}:LOAD1:L1?
Description
The use of Port 3 and/or Port 4 requires a 4-port VNA instrument.
Sets the Load1 standard L1 (L one) inductance coefficient on the selected port of the indicated channel.
The L1 coefficient is measured in Henrys/Hertz.
Returns the Load1 standard L1 inductance coefficient on the selected port of the indicated channel.
See Calibration Component Parameters for a complete listing of calibration components, connectors, and their Command Parameters.
Cmd Parameters
<NRf> The input parameter is in Henrys/Hertz.
Query Parameters
<NR3> The output parameter is in Henrys/Hertz.
Range
MPND
Default Value
0.00000000000E+000
Syntax Example
:SENS1:CORR:COLL:PORT1:LOAD1:L1 2.0E0
:SENS1:CORR:COLL:PORT1:LOAD1:L1?
:SENSe{1-16}:CORRection:COLLect:PORT{1-4}:LOAD1:L2 <NRf>
:SENSe{1-16}:CORRection:COLLect:PORT{1-4}:LOAD1:L2?
Description
The use of Port 3 and/or Port 4 requires a 4-port VNA instrument.
Sets the Load1 standard L2 inductance coefficient on the selected port of the indicated channel.
The L2 coefficient is measured in Henrys’/Hertz^2.
Returns the Load1 standard L2 inductance coefficient on the selected port of the indicated channel.
See Calibration Component Parameters for a complete listing of calibration components, connectors, and their Command Parameters.
Cmd Parameters
<NRf> The input parameter is in Henrys/Hertz^2.
Query Parameters
<NR3> The output parameter is in Henrys/Hertz^2.
Range
MPND
Default Value
0.00000000000E+000
Syntax Example
:SENS1:CORR:COLL:PORT1:LOAD1:L2 2.0E0
:SENS1:CORR:COLL:PORT1:LOAD1:L2?
:SENSe{1-16}:CORRection:COLLect:PORT{1-4}:LOAD1:L3 <NRf>
:SENSe{1-16}:CORRection:COLLect:PORT{1-4}:LOAD1:L3?
Description
The use of Port 3 and/or Port 4 requires a 4-port VNA instrument.Sets the Load1 standard L3 inductance coefficient on the selected port of the indicated channel.
The L3 coefficient is measured in Henrys/Hertz^3.
Returns the Load1 standard L3 inductance coefficient on the selected port of the indicated channel.
See Calibration Component Parameters for a complete listing of calibration components, connectors, and their Command Parameters.
Cmd Parameters
<NRf> The input parameter is in Henrys/Hertz^3.
Query Parameters
<NR3> The output parameter is in Henrys/Hertz^3.
Range
MPND
Default Value
0.00000000000E+000
Syntax Example
:SENS1:CORR:COLL:PORT1:LOAD1:L3 2.0E0
:SENS1:CORR:COLL:PORT1:LOAD1:L3?
:SENSe{1-16}:CORRection:COLLect:PORT{1-4}:LOAD1:LABEL <string>
:SENSe{1-16}:CORRection:COLLect:PORT{1-4}:LOAD1:LABEL?
Description
The use of Port 3 and/or Port 4 requires a 4-port VNA instrument.
Sets the Load1 standard label on the selected port of the indicated channel.
Returns the Load1 standard label on the selected port of the indicated channel.
See Calibration Component Parameters for a complete listing of calibration components, connectors, and their Command Parameters.
Cmd Parameters
<string> The input parameter is any combination of numbers and letters.
See definition of <string>.
Query Parameters
<char> The output parameter can be any combination of numbers and letters.
Range
NA
Default Value
NA
Syntax Example
:SENS1:CORR:COLL:PORT1:LOAD1:LABEL 'IC7000'
:SENS1:CORR:COLL:PORT1:LOAD1:LABEL?
:SENSe{1-16}:CORRection:COLLect:PORT{1-4}:LOAD1:OFF1 <NRf>
:SENSe{1-16}:CORRection:COLLect:PORT{1-4}:LOAD1:OFF1?
Description
The use of Port 3 and/or Port 4 requires a 4-port VNA instrument.
Sets the Load1 Standard Offset length1 coefficient on the selected port of the indicated channel.
Returns the Load1 Standard Offset length1 coefficient on the Selected port of the indicated channel.
Cmd Parameters
<NRf> The input parameter is in Meters/Hertz.
Query Parameters
<NR3> The output parameter is in Meters/Hertz.
Range
MPND
Default
0.00000000000E+000
Syntax Example
:SENS1:CORR:COLL:PORT1:LOAD1:OFF1 2.0E0
:SENS1:CORR:COLL:PORT1:LOAD1:OFF1?
:SENSe{1-16}:CORRection:COLLect:PORT{1-4}:LOAD1:OFF2 <NRf>
:SENSe{1-16}:CORRection:COLLect:PORT{1-4}:LOAD1:OFF2?
Description
The use of Port 3 and/or Port 4 requires a 4-port VNA instrument.
Sets the Load1 Standard Offset length2 coefficient on the Selected port of the indicated channel.
Returns the Load1 Standard Offset length2 coefficient on the Selected port of the indicated channel.
Cmd Parameters
<NRf> The input parameter is in Meters/Hertz^2.
Query Parameters
<NR3> The output parameter is in Meters/Hertz^2.
Range
MPND
Default
0.00000000000E+000
Syntax Example
:SENS1:CORR:COLL:PORT1:LOAD1:OFF2 2.0E0
:SENS1:CORR:COLL:PORT1:LOAD1:OFF2?
:SENSe{1-16}:CORRection:COLLect:PORT{1-4}:LOAD1:OFF3 <NRf>
:SENSe{1-16}:CORRection:COLLect:PORT{1-4}:LOAD1:OFF3?
Description
The use of Port 3 and/or Port 4 requires a 4-port VNA instrument.
Sets the Load1 Standard Offset length3 coefficient on the Selected port of the indicated channel.
Returns the Load1 Standard Offset length3 coefficient on the Selected port of the indicated channel.
Cmd Parameters
<NRf> The input parameter is in Meters/Hertz^3.
Query Parameters
<NR3> The output parameter is in Meters/Hertz^3.
Range
MPND
Default
0.00000000000E+000
Syntax Example
:SENS1:CORR:COLL:PORT1:LOAD1:OFF3 2.0E0
:SENS1:CORR:COLL:PORT1:LOAD1:OFF3?
:SENSe{1-16}:CORRection:COLLect:PORT{1-4}:LOAD1:OFFS <NRf>
:SENSe{1-16}:CORRection:COLLect:PORT{1-4}:LOAD1:OFFS?
Description
The use of Port 3 and/or Port 4 requires a 4-port VNA instrument.
Sets the Load1 (Load one) standard offset on the selected port of the indicated channel.
Returns the Load1 standard offset on the selected port of the indicated channel.
See Calibration Component Parameters for a complete listing of calibration components, connectors, and their Command Parameters.
Cmd Parameters
<NRf> The input parameter is in Meters.
Query Parameters
<NR3> The output parameter is in Meters.
Range
MPND
Default Value
0.00000000000E+000
Syntax Example
:SENS1:CORR:COLL:PORT1:LOAD1:OFFS 2.5E1
:SENS1:CORR:COLL:PORT1:LOAD1:OFFS?
:SENSe{1-16}:CORRection:COLLect:PORT{1-4}:LOAD1:R <NRf>
:SENSe{1-16}:CORRection:COLLect:PORT{1-4}:LOAD1:R?
Description
The use of Port 3 and/or Port 4 requires a 4-port VNA instrument.
Sets the Load1 (Load one) standard resistance on the selected port of the indicated channel.
Returns the Load1 standard resistance on the selected port of the indicated channel.
Cmd Parameters
<NRf> The input parameter is in Ohms.
Query Parameters
<NR3> The output parameter is in Ohms.
Range
0 to 1E10
Default Value
5.00000000000E+001
Syntax Example
:SENS1:CORR:COLL:PORT1:LOAD1:R 7.5E1
:SENS1:CORR:COLL:PORT1:LOAD1:R?
:SENSe{1-16}:CORRection:COLLect:PORT{1-4}:LOAD1:SERial <string>
:SENSe{1-16}:CORRection:COLLect:PORT{1-4}:LOAD1:SERial?
Description
The use of Port 3 and/or Port 4 requires a 4-port VNA instrument.
Sets the Load1 (Load one) standard serial number on the selected port of the indicated channel.
Returns the Load1 standard serial number on the selected port of the indicated channel.
Cmd Parameters
<string> The input parameter is any combination of numbers and letters.
See definition of <string>.
Query Parameters
<char> The output parameter can be any combination of numbers and letters.
Range
NA
Default Value
NA
Syntax Example
:SENS1:CORR:COLL:PORT1:LOAD1:SER '123456'
:SENS1:CORR:COLL:PORT1:LOAD1:SER?
:SENSe{1-16}:CORRection:COLLect:PORT{1-4}:LOAD1:Z0 <NRf>
:SENSe{1-16}:CORRection:COLLect:PORT{1-4}:LOAD1:Z0?
Description
The use of Port 3 and/or Port 4 requires a 4-port VNA instrument.
Sets the Load1 (Load one) standard Impedance on the selected port of the indicated channel.
Returns the Load1 standard Impedance on the selected port of the indicated channel.
Cmd Parameters
<NRf> The input parameter is in Ohms.
Query Parameters
<NR3> The output parameter is in Ohms.
Range
1E-4 to 1E10
Default Value
5.00000000000E+001
Syntax Example
:SENS1:CORR:COLL:PORT1:LOAD1:Z0 7.5E1
:SENS1:CORR:COLL:PORT1:LOAD1:Z0?
:SENSe{1-16}:CORRection:COLLect:PORT{1-4}:LOAD2:C0 <NRf>
:SENSe{1-16}:CORRection:COLLect:PORT{1-4}:LOAD2:C0?
Description
The use of Port 3 and/or Port 4 requires a 4-port VNA instrument.
Sets the Load2 standard C0 (C zero) capacitance on the selected port of the indicated channel.
The C0 coefficient is measured in Farads.
Returns the Load2 standard C0 capacitance on the selected port of the indicated channel.
Cmd Parameters
<NRf> The input parameter is in Farads.
Query Parameters
<NR3> The output parameter is in Farads.
Range
0 to 1E12
Default Value
0.00000000000E+000
Syntax Example
:SENS1:CORR:COLL:PORT1:LOAD2:C0 2.0E-11
:SENS1:CORR:COLL:PORT1:LOAD2:C0?
:SENSe{1-16}:CORRection:COLLect:PORT{1-4}:LOAD2:C1 <NRf>
:SENSe{1-16}:CORRection:COLLect:PORT{1-4}:LOAD2:C1?
Description
The use of Port 3 and/or Port 4 requires a 4-port VNA instrument.
Sets the Load2 Standard C1 (C one) Coefficient on the Selected port of the indicated channel.
The C1 coefficient is measured in Farads/Hertz.
Returns the Load2 Standard C1 Coefficient on the Selected port of the indicated channel
Cmd Parameters
<NRf> The input parameter is in Farads/Hertz.
Query Parameters
<NR3> The output parameter is in Farads/Hertz.
Range
MPND
Default
0.00000000000E+000
Syntax Example
:SENS1:CORR:COLL:PORT1:LOAD2:C1 2.0E0
:SENS1:CORR:COLL:PORT1:LOAD2:C1?
:SENSe{1-16}:CORRection:COLLect:PORT{1-4}:LOAD2:C2 <NRf>
:SENSe{1-16}:CORRection:COLLect:PORT{1-4}:LOAD2:C2?
Description
The use of Port 3 and/or Port 4 requires a 4-port VNA instrument.
Sets the Load2 Standard C2 Coefficient on the selected port of the indicated channel.
The C2 coefficient is measured in Farads/Hertz^2.
Returns the Load2 Standard C2 Coefficient on the selected port of the indicated channel.
Cmd Parameters
<NRf> The input parameter is in Farads/Hertz^2.
Query Parameters
<NR3> The output parameter is in Farads/Hertz^2.
Range
MPND
Default
0.00000000000E+000
Syntax Example
:SENS1:CORR:COLL:PORT1:LOAD2:C2 2.0E0
:SENS1:CORR:COLL:PORT1:LOAD2:C2?
:SENSe{1-16}:CORRection:COLLect:PORT{1-4}:LOAD2:C3 <NRf>
:SENSe{1-16}:CORRection:COLLect:PORT{1-4}:LOAD2:C3?
Description
The use of Port 3 and/or Port 4 requires a 4-port VNA instrument.
Sets the Load2 Standard C3 Coefficient on the selected port of the indicated channel.
The C3 coefficient is measured in Farads/Hertz^3.
Returns the Load2 Standard C3 Coefficient on the selected port of the indicated channel.
Cmd Parameters
<NRf> The input parameter is in Farads/Hertz^3.
Query Parameters
<NR3> The output parameter is in Farads/Hertz^3.
Range
MPND
Default
0.00000000000E+000
Syntax Example
:SENS1:CORR:COLL:PORT1:LOAD2:C3 2.0E0
:SENS1:CORR:COLL:PORT1:LOAD2:C3?
:SENSe{1-16}:CORRection:COLLect:PORT{1-4}:LOAD2:L0 <NRf>
:SENSe{1-16}:CORRection:COLLect:PORT{1-4}:LOAD2:L0?
Description
The use of Port 3 and/or Port 4 requires a 4-port VNA instrument.
Sets the Load2 standard L0 (L zero) inductance on the selected port of the indicated channel.
The L0 coefficient is measured in Henrys.
Returns the Load2 standard L0 inductance on the selected port of the indicated channel.
Cmd Parameters
<NRf> The input parameter is in Henrys.
Query Parameters
<NR3> The output parameter is in Henrys.
Range
MPND
Default Value
0.00000000000E+000
Syntax Example
:SENS1:CORR:COLL:PORT1:LOAD2:L0 2.0E-6
:SENS1:CORR:COLL:PORT1:LOAD2:L0?
:SENSe{1-16}:CORRection:COLLect:PORT{1-4}:LOAD2:L1 <NRf>
:SENSe{1-16}:CORRection:COLLect:PORT{1-4}:LOAD2:L1?
Description
The use of Port 3 and/or Port 4 requires a 4-port VNA instrument.
Sets the Load2 standard L1 (L one) inductance coefficient on the selected port of the indicated channel.
The L1 coefficient is measured in Henrys/Hertz.
Returns the Load2 standard L1 inductance coefficient on the selected port of the indicated channel.
Cmd Parameters
<NRf> The input parameter is in Henrys/Hertz.
Query Parameters
<NR3> The output parameter is in Henrys/Hertz.
Range
MPND
Default Value
0.00000000000E+000
Syntax Example
:SENS1:CORR:COLL:PORT1:LOAD2:L1 2.0E0
:SENS1:CORR:COLL:PORT1:LOAD2:L1?
:SENSe{1-16}:CORRection:COLLect:PORT{1-4}:LOAD2:L2 <NRf>
:SENSe{1-16}:CORRection:COLLect:PORT{1-4}:LOAD2:L2?
Description
The use of Port 3 and/or Port 4 requires a 4-port VNA instrument.
Sets the Load2 standard L2 inductance coefficient on the selected port of the indicated channel.
The L2 coefficient is measured in Henrys/Hertz^2.
Returns the Load2 standard L2 inductance coefficient on the selected port of the indicated channel.
Cmd Parameters
<NRf> The input parameter is in Henrys/Hertz^2.
Query Parameters
<NR3> The output parameter is in Henrys/Hertz^2.
Range
MPND
Default Value
0.00000000000E+000
Syntax Example
:SENS1:CORR:COLL:PORT1:LOAD2:L2 2.0E0
:SENS1:CORR:COLL:PORT1:LOAD2:L2?
:SENSe{1-16}:CORRection:COLLect:PORT{1-4}:LOAD2:L3 <NRf>
:SENSe{1-16}:CORRection:COLLect:PORT{1-4}:LOAD2:L3?
Description
The use of Port 3 and/or Port 4 requires a 4-port VNA instrument.
Sets the Load2 standard L3 inductance coefficient on the selected port of the indicated channel.
The L3 coefficient is measured in Henrys/Hertz^3.
Returns the Load2 standard L3 inductance coefficient on the selected port of the indicated channel.
Cmd Parameters
<NRf> The input parameter is in Henrys/Hertz^3.
Query Parameters
<NR3> The output parameter is in Henrys/Hertz^3.
Range
MPND
Default Value
0.00000000000E+000
Syntax Example
:SENS1:CORR:COLL:PORT1:LOAD2:L3 2.0E0
:SENS1:CORR:COLL:PORT1:LOAD2:L3?
:SENSe{1-16}:CORRection:COLLect:PORT{1-4}:LOAD2:LABEL <string>
:SENSe{1-16}:CORRection:COLLect:PORT{1-4}:LOAD2:LABEL?
Description
The use of Port 3 and/or Port 4 requires a 4-port VNA instrument.
Sets the Load2 standard label on the selected port of the indicated channel.
Returns the Load2 standard label on the selected port of the indicated channel.
Cmd Parameters
<string> The input parameter is any combination of numbers and letters.
See definition of <string>.
Query Parameters
<char> The output parameter can be any combination of numbers and letters.
Range
NA
Default Value
NA
Syntax Example
:SENS1:CORR:COLL:PORT1:LOAD2:LABEL 'IC7000'
:SENS1:CORR:COLL:PORT1:LOAD2:LABEL?
:SENSe{1-16}:CORRection:COLLect:PORT{1-4}:LOAD2:OFF1 <NRf>
:SENSe{1-16}:CORRection:COLLect:PORT{1-4}:LOAD2:OFF1?
Description
The use of Port 3 and/or Port 4 requires a 4-port VNA instrument.
Sets the Load2 Standard Offset length1 (length one) coefficient on the selected port of the indicated channel.
The OFF1 coefficient is measured in Meters/Hertz.
Returns the Load2 Standard Offset length1 coefficient on the selected port of the indicated channel.
Cmd Parameters
<NRf> The input parameter is in Meters/Hertz.
Query Parameters
<NR3> The output parameter is in Meters/Hertz.
Range
MPND
Default
0.00000000000E+000
Syntax Example
:SENS1:CORR:COLL:PORT1:LOAD2:OFF1 2.0E0
:SENS1:CORR:COLL:PORT1:LOAD2:OFF1?
:SENSe{1-16}:CORRection:COLLect:PORT{1-4}:LOAD2:OFF2 <NRf>
:SENSe{1-16}:CORRection:COLLect:PORT{1-4}:LOAD2:OFF2?
Description
The use of Port 3 and/or Port 4 requires a 4-port VNA instrument.
Sets the Load2 Standard Offset length2 coefficient on the selected port of the indicated channel.
The OFF2 coefficient is measured in Meters/Hertz^2.
Returns the Load2 Standard Offset length2 coefficient on the selected port of the indicated channel.
Cmd Parameters
<NRf> The input parameter is in Meters/Hertz^2.
Query Parameters
<NR3> The output parameter is in Meters/Hertz^2.
Range
MPND
Default
0.00000000000E+000
Syntax Example
:SENS1:CORR:COLL:PORT1:LOAD2:OFF2 2.0E0
:SENS1:CORR:COLL:PORT1:LOAD2:OFF2?
:SENSe{1-16}:CORRection:COLLect:PORT{1-4}:LOAD2:OFF3 <NRf>
:SENSe{1-16}:CORRection:COLLect:PORT{1-4}:LOAD2:OFF3?
Description
The use of Port 3 and/or Port 4 requires a 4-port VNA instrument.
Sets the Load2 Standard Offset length3 coefficient on the selected port of the indicated channel.
The OFF3 coefficient is measured in Meters/Hertz^3.
Returns the Load2 Standard Offset length3 coefficient on the selected port of the indicated channel.
Cmd Parameters
<NRf> The input parameter is in Meters/Hertz^3.
Query Parameters
<NR3> The output parameter is in Meters/Hertz^3.
Range
MPND
Default
0.00000000000E+000
Syntax Example
:SENS1:CORR:COLL:PORT1:LOAD2:OFF3 2.0E0
:SENS1:CORR:COLL:PORT1:LOAD2:OFF3?
:SENSe{1-16}:CORRection:COLLect:PORT{1-4}:LOAD2:OFFS <NRf>
:SENSe{1-16}:CORRection:COLLect:PORT{1-4}:LOAD2:OFFS?
Description
The use of Port 3 and/or Port 4 requires a 4-port VNA instrument.
Sets the Load2 standard offset on the selected port of the indicated channel.
Returns the Load2 standard offset on the selected port of the indicated channel.
Cmd Parameters
<NRf> The input parameter is in Meters.
Query Parameters
<NR3> The output parameter is in Meters.
Range
MPND
Default Value
0.00000000000E+000
Syntax Example
:SENS1:CORR:COLL:PORT1:LOAD2:OFFS 2.5E1
:SENS1:CORR:COLL:PORT1:LOAD2:OFFS?
:SENSe{1-16}:CORRection:COLLect:PORT{1-4}:LOAD2:R <NRf>
:SENSe{1-16}:CORRection:COLLect:PORT{1-4}:LOAD2:R?
Description
The use of Port 3 and/or Port 4 requires a 4-port VNA instrument.
Sets the Load2 standard resistance on the selected port of the indicated channel.
Returns the Load2 standard resistance on the selected port of the indicated channel.
Cmd Parameters
<NRf> The input parameter is in Ohms.
Query Parameters
<NR3> The output parameter is in Ohms.
Range
0 to 1E10
Default Value
5.00000000000E+001
Syntax Example
:SENS1:CORR:COLL:PORT1:LOAD2:R 7.5E1
:SENS1:CORR:COLL:PORT1:LOAD2:R?
:SENSe{1-16}:CORRection:COLLect:PORT{1-4}:LOAD2:SERial <string>
:SENSe{1-16}:CORRection:COLLect:PORT{1-4}:LOAD2:SERial?
Description
The use of Port 3 and/or Port 4 requires a 4-port VNA instrument.
Sets the Load2 standard serial number on the selected port of the indicated channel.
Returns the Load2 standard serial number on the selected port of the indicated channel.
Cmd Parameters
<string> The input parameter is any combination of numbers and letters.
See definition of <string>.
Query Parameters
<char> The output parameter can be any combination of numbers and letters.
Range
NA
Default Value
NA
Syntax Example
:SENS1:CORR:COLL:PORT1:LOAD2:SER '123456'
:SENS1:CORR:COLL:PORT1:LOAD2:SER?
:SENSe{1-16}:CORRection:COLLect:PORT{1-4}:LOAD2:Z0 <NRf>
:SENSe{1-16}:CORRection:COLLect:PORT{1-4}:LOAD2:Z0?
Description
The use of Port 3 and/or Port 4 requires a 4-port VNA instrument.
Sets the Load2 standard Impedance on the selected port of the indicated channel.
Returns the Load2 standard Impedance on the selected port of the indicated channel.
Cmd Parameters
<NRf> The input parameter is in Ohms.
Query Parameters
<NR3> The output parameter is in Ohms.
Range
1E-4 to 1E10
Default Value
5.00000000000E+001
Syntax Example
:SENS1:CORR:COLL:PORT1:LOAD2:Z0 7.5E1
:SENS1:CORR:COLL:PORT1:LOAD2:Z0?
:SENSe{1-16}:CORRection:COLLect:PORT{1-4}:OPEN
Description
The use of Port 3 and/or Port 4 requires a 4-port VNA instrument.
Initiates collection of the open standard data for the calibration on the indicated channel and port.
No query.
Cmd Parameters
NA
Range
NA
Default Value
NA
Syntax Example
:SENS1:CORR:COLL:PORT1:OPEN
:SENSe{1-16}:CORRection:COLLect:PORT{1-4}:OPEN:C0 <NRf>
:SENSe{1-16}:CORRection:COLLect:PORT{1-4}:OPEN:C0?
Description
The use of Port 3 and/or Port 4 requires a 4-port VNA instrument.
Sets the open standard C0 (C zero) capacitance on the selected port of the indicated channel.
The C0 coefficient is measured in Farads.
Returns the open standard C0 capacitance on the selected port of the indicated channel.
Cmd Parameters
<NRf> The input parameter is in Farads.
Query Parameters
<NR3> The output parameter is in Farads.
Range
MPND
Default Value
See Calibration Component Parameters.
Syntax Example
:SENS1:CORR:COLL:PORT1:OPEN:C0 2.0E-11
:SENS1:CORR:COLL:PORT1:OPEN:C0?
:SENSe{1-16}:CORRection:COLLect:PORT{1-4}:OPEN:C1 <NRf>
:SENSe{1-16}:CORRection:COLLect:PORT{1-4}:OPEN:C1?
Description
The use of Port 3 and/or Port 4 requires a 4-port VNA instrument.
Sets the open standard C1 (C one) capacitance coefficient on the selected port of the indicated channel.
The C1 coefficient is measured in Farads/Hertz.
Returns the open standard C1 capacitance coefficient on the selected port of the indicated channel.
Cmd Parameters
<NRf> The input parameter is in Farads/Hertz.
Query Parameters
<NR3> The output parameter is in Farads/Hertz.
Range
MPND
Default Value
See Calibration Component Parameters.
Syntax Example
:SENS1:CORR:COLL:PORT1:OPEN:C1 2.0E0
:SENS1:CORR:COLL:PORT1:OPEN:C1?
:SENSe{1-16}:CORRection:COLLect:PORT{1-4}:OPEN:C2 <NRf>
:SENSe{1-16}:CORRection:COLLect:PORT{1-4}:OPEN:C2?
Description
The use of Port 3 and/or Port 4 requires a 4-port VNA instrument.
Sets the open standard C2 capacitance coefficient on the selected port of the indicated channel.
The C2 coefficient is measured in Farads/Hertz^2.
Returns the open standard C2 capacitance coefficient on the selected port of the indicated channel.
Cmd Parameters
<NRf> The input parameter is in Farads/Hertz^2.
Query Parameters
<NR3> The output parameter is in Farads/Hertz^2.
Range
MPND
Default Value
See Calibration Component Parameters.
Syntax Example
:SENS1:CORR:COLL:PORT1:OPEN:C2 2.0E0
:SENS1:CORR:COLL:PORT1:OPEN:C2?
:SENSe{1-16}:CORRection:COLLect:PORT{1-4}:OPEN:C3 <NRf>
:SENSe{1-16}:CORRection:COLLect:PORT{1-4}:OPEN:C3?
Description
The use of Port 3 and/or Port 4 requires a 4-port VNA instrument.
Sets the open standard C3 capacitance coefficient on the selected port of the indicated channel.
The C3 coefficient is measured in Farads/Hertz^3.
Returns the open standard C3 capacitance coefficient on the selected port of the indicated channel.
Cmd Parameters
<NRf> The input parameter is in Farads/Hertz^3.
Query Parameters
<NR3> The output parameter is in Farads/Hertz^3.
Range
MPND
Default Value
See Calibration Component Parameters.
Syntax Example
:SENS1:CORR:COLL:PORT1:OPEN:C3 2.0E0
:SENS1:CORR:COLL:PORT1:OPEN:C3?
:SENSe{1-16}:CORRection:COLLect:PORT{1-4}:OPEN:FREQuency <NRf>
:SENSe{1-16}:CORRection:COLLect:PORT{1-4}:OPEN:FREQuency?
Description
The use of Port 3 and/or Port 4 requires a 4-port VNA instrument.
Sets the open standard reference frequency on the selected port of the indicated channel.
Returns the open standard reference frequency on the selected port of the indicated channel.
See Calibration Component Parameters for a complete listing of calibration components, connectors, and their Command Parameters.
Cmd Parameters
<NRf> The input parameter is in Hertz.
Query Parameters
<NR3> The output parameter is in Hertz.
Range
MPND
Default Value
See Calibration Component Parameters.
Syntax Example
:SENS1:CORR:COLL:PORT1:OPEN:FREQ 1.0E7
:SENS1:CORR:COLL:PORT1:OPEN:FREQ?
:SENSe{1-16}:CORRection:COLLect:PORT{1-4}:OPEN:LOSS <NRf>
:SENSe{1-16}:CORRection:COLLect:PORT{1-4}:OPEN:LOSS?
Description
The use of Port 3 and/or Port 4 requires a 4-port VNA instrument.
Sets the open standard loss value on the selected port of the indicated channel.
Returns the open standard loss value on the selected port of the indicated channel.
Cmd Parameters
<NRf> The input parameter is in dB/mm.
See definition of <string>.
Query Parameters
<NR3> The output parameter can be in dB/mm.
Range
MPND
Default Value
0
Syntax Example
:SENS1:CORR:COLL:PORT1:OPEN:LOSS
:SENS1:CORR:COLL:PORT1:OPEN:LOSS?
:SENSe{1-16}:CORRection:COLLect:PORT{1-4}:OPEN:LABEL <string>
:SENSe{1-16}:CORRection:COLLect:PORT{1-4}:OPEN:LABEL?
Description
The use of Port 3 and/or Port 4 requires a 4-port VNA instrument.
Sets the open standard label on the selected port of the indicated channel.
Returns the open standard label on the selected port of the indicated channel.
Cmd Parameters
<string> The input parameter is any combination of numbers and letters.
See definition of <string>.
Query Parameters
<char> The output parameter can be any combination of numbers and letters.
Range
NA
Default Value
NA
Syntax Example
:SENS1:CORR:COLL:PORT1:OPEN:LABEL 'IC7000'
:SENS1:CORR:COLL:PORT1:OPEN:LABEL?
:SENSe{1-16}:CORRection:COLLect:PORT{1-4}:OPEN:OFFS <NRf>
:SENSe{1-16}:CORRection:COLLect:PORT{1-4}:OPEN:OFFS?
Description
The use of Port 3 and/or Port 4 requires a 4-port VNA instrument.
Sets the open standard offset on the selected port of the indicated channel.
The OFFS parameter is measured in Meters.
Returns the open standard offset on the selected port of the indicated channel.
See Calibration Component Parameters for a complete listing of calibration components, connectors, and their Command Parameters.
Cmd Parameters
<NRf> The input parameter is in Meters.
Query Parameters
<NR3> The output parameter is in Meters.
Range
MPND
Default Value
See Calibration Component Parameters.
Syntax Example
:SENS1:CORR:COLL:PORT1:OPEN:OFFS 2.5E1
:SENS1:CORR:COLL:PORT1:OPEN:OFFS?
:SENSe{1-16}:CORRection:COLLect:PORT{1-4}:OPEN:S1P:FILe <string>
:SENSe{1-16}:CORRection:COLLect:PORT{1-4}:OPEN:S1P:FILe?
Description
Sets the OPEN Standard S1P file on the selected port of the indicated channel.
Returns the OPEN Standard S1P file on the selected port of the indicated channel.
Cmd Parameters
<string> s1p file path
See definition of <string>.
Query Parameters
NA
Query Output
<char1> s1p file path
Range
NA
Default Value
NA
Syntax Example
:SENS1:CORR:COLL:PORT1:OPEN:S1P:FIL 'C:\Test\Open.s1p'
:SENS1:CORR:COLL:PORT1:OPEN:S1P:FIL?
:SENSe{1-16}:CORRection:COLLect:PORT{1-4}:OPEN:S1P[:STATe] <char>
:SENSe{1-16}:CORRection:COLLect:PORT{1-4}:OPEN:S1P[:STATe]?
Description
Sets the OPEN S1P file settings ON/OFF on the selected port of the indicated channel.
ON means that the S1P file data will be used, OFF means the Circuit Model data will be used.
Returns the OPEN S1P file settings on the selected port of the indicated channel.
Cmd Parameters
<char> 1 | 0 | ON | OFF
See definition of <string>.
Query Parameters
NA
Query Output
<char> 1 | 0
Range
NA
Default Value
0
Syntax Example
:SENS1:CORR:COLL:PORT1:OPEN:S1P 1
:SENS1:CORR:COLL:PORT1:OPEN:S1P?
:SENSe{1-16}:CORRection:COLLect:PORT{1-4}:SHORT:S1P[:STATe] <char>
:SENSe{1-16}:CORRection:COLLect:PORT{1-4}:SHORT:S1P[:STATe]?
Description
Sets the SHORT S1P file settings ON/OFF on the selected port of the indicated channel.
ON means that the S1P file data will be used, OFF means the Circuit Model data will be used.
Returns the SHORT S1P file settings on the selected port of the indicated channel.
Cmd Parameters
<char> 1 | 0 | ON | OFF
See definition of <string>.
Query Parameters
NA
Query Output
<char> 1 | 0
Range
NA
Default Value
0
Syntax Example
:SENS1:CORR:COLL:PORT1:SHORT:S1P 1
:SENS1:CORR:COLL:PORT1:SHORT:S1P?
:SENSe{1-16}:CORRection:COLLect:PORT{1-4}:LOAD1:S1P:FILe <string>
:SENSe{1-16}:CORRection:COLLect:PORT{1-4}:LOAD1:S1P:FILe?
Description
Sets the LOAD1 Standard S1P file on the selected port of the indicated channel.
Returns the LOAD1 Standard S1P file on the selected port of the indicated channel.
Cmd Parameters
<string> s1p file path
See definition of <string>.
Query Parameters
NA
Query Output
<char1> s1p file path
Range
NA
Default Value
NA
Syntax Example
:SENS1:CORR:COLL:PORT1:LOAD1:S1P:FIL 'C:\Test\Load1.s1p'
:SENS1:CORR:COLL:PORT1:LOAD1:S1P:FIL?
:SENSe{1-16}:CORRection:COLLect:PORT{1-4}:LOAD1:S1P[:STATe] <char>
:SENSe{1-16}:CORRection:COLLect:PORT{1-4}:LOAD1:S1P[:STATe]?
Description
Sets the LOAD1 S1P file settings ON/OFF on the selected port of the indicated channel.
ON means that the S1P file data will be used, OFF means the Circuit Model data will be used.
Returns the LOAD1 S1P file settings on the selected port of the indicated channel.
Cmd Parameters
<char> 1 | 0 | ON | OFF
See definition of <string>.
Query Parameters
NA
Query Output
<char> 1 | 0
Range
NA
Default Value
0
Syntax Example
:SENS1:CORR:COLL:PORT1:LOAD1:S1P 1
:SENS1:CORR:COLL:PORT1:LOAD1:S1P?
:SENSe{1-16}:CORRection:COLLect:PORT{1-4}:LOAD2:S1P:FILe <string>
:SENSe{1-16}:CORRection:COLLect:PORT{1-4}:LOAD2:S1P:FILe?
Description
Sets the LOAD2 Standard S1P file on the selected port of the indicated channel.
Returns the LOAD2 Standard S1P file on the selected port of the indicated channel.
Cmd Parameters
<string> s1p file path
See definition of <string>.
Query Parameters
NA
Query Output
<char1> s1p file path
Range
NA
Default Value
NA
Syntax Example
:SENS1:CORR:COLL:PORT1:LOAD2:S1P:FIL 'C:\Test\Load1.s1p'
:SENS1:CORR:COLL:PORT1:LOAD2:S1P:FIL?
:SENSe{1-16}:CORRection:COLLect:PORT{1-4}:LOAD2:S1P[:STATe] <char>
:SENSe{1-16}:CORRection:COLLect:PORT{1-4}:LOAD2:S1P[:STATe]?
Description
Sets the LOAD2 S1P file settings ON/OFF on the selected port of the indicated channel.
ON means that the S1P file data will be used, OFF means the Circuit Model data will be used.
Returns the LOAD2 S1P file settings on the selected port of the indicated channel.
Cmd Parameters
<char> 1 | 0 | ON | OFF
See definition of <string>.
Query Parameters
NA
Query Output
<char> 1 | 0
Range
NA
Default Value
0
Syntax Example
:SENS1:CORR:COLL:PORT1:LOAD2:S1P 1
:SENS1:CORR:COLL:PORT1:LOAD2:S1P?
:SENSe{1-16}:CORRection:COLLect:PORT{1-4}:SHORT:S1P:FILe <string>
:SENSe{1-16}:CORRection:COLLect:PORT{1-4}:SHORT:S1P:FILe?
Description
Sets the SHORT Standard S1P file on the selected port of the indicated channel.
Returns the SHORT Standard S1P file on the selected port of the indicated channel.
Cmd Parameters
<string> s1p file path
See definition of <string>.
Query Parameters
NA
Query Output
<char1> s1p file path
Range
NA
Default Value
NA
Syntax Example
:SENS1:CORR:COLL:PORT1:SHORT:S1P:FIL 'C:\Test\SHORT.s1p'
:SENS1:CORR:COLL:PORT1:SHORT:S1P:FIL?
:SENSe{1-16}:CORRection:COLLect:PORT{1-4}:SHORT1:S1P:FILe <string>
:SENSe{1-16}:CORRection:COLLect:PORT{1-4}:SHORT1:S1P:FILe?
Description
Sets the SHORT1 Standard S1P file on the selected port of the indicated channel.
Returns the SHORT1 Standard S1P file on the selected port of the indicated channel.
Cmd Parameters
<string> s1p file path
See definition of <string>.
Query Parameters
NA
Query Output
<char1> s1p file path
Range
NA
Default Value
NA
Syntax Example
:SENS1:CORR:COLL:PORT1:SHORT1:S1P:FIL 'C:\Test\SHORT1.s1p'
:SENS1:CORR:COLL:PORT1:SHORT1:S1P:FIL?
:SENSe{1-16}:CORRection:COLLect:PORT{1-4}:SHORT1:S1P[:STATe] <char>
:SENSe{1-16}:CORRection:COLLect:PORT{1-4}:SHORT1:S1P[:STATe]?
Description
Sets the SHORT1 S1P file settings ON/OFF on the selected port of the indicated channel.
ON means that the S1P file data will be used, OFF means the Circuit Model data will be used.
Returns the SHORT1 S1P file settings on the selected port of the indicated channel.
Cmd Parameters
<char> 1 | 0 | ON | OFF
Query Parameters
NA
Query Output
<char> 1 | 0
Range
NA
Default Value
0
Syntax Example
:SENS1:CORR:COLL:PORT1:SHORT1:S1P 1
:SENS1:CORR:COLL:PORT1:SHORT1:S1P?
:SENSe{1-16}:CORRection:COLLect:PORT{1-4}:SHORT2:S1P:FILe <string>
:SENSe{1-16}:CORRection:COLLect:PORT{1-4}:SHORT2:S1P:FILe?
Description
Sets the SHORT2 Standard S1P file on the selected port of the indicated channel.
Returns the SHORT2 Standard S1P file on the selected port of the indicated channel.
Cmd Parameters
<string> s1p file path
See definition of <string>.
Query Parameters
NA
Query Output
<char1> s1p file path
Range
NA
Default Value
NA
Syntax Example
:SENS1:CORR:COLL:PORT1:SHORT2:S1P:FIL 'C:\Test\SHORT2.s1p'
:SENS1:CORR:COLL:PORT1:SHORT2:S1P:FIL?
:SENSe{1-16}:CORRection:COLLect:PORT{1-4}:SHORT2:S1P[:STATe] <char>
:SENSe{1-16}:CORRection:COLLect:PORT{1-4}:SHORT2:S1P[:STATe]?
Description
Sets the SHORT2 S1P file settings ON/OFF on the selected port of the indicated channel.
ON means that the S1P file data will be used, OFF means the Circuit Model data will be used.
Returns the SHORT2 S1P file settings on the selected port of the indicated channel.
Cmd Parameters
<char> 1 | 0 | ON | OFF
Query Parameters
NA
Query Output
<char> 1 | 0
Range
NA
Default Value
0
Syntax Example
:SENS1:CORR:COLL:PORT1:SHORT2:S1P 1
:SENS1:CORR:COLL:PORT1:SHORT2:S1P?
:SENSe{1-16}:CORRection:COLLect:PORT{1-4}:SHORT3:S1P:FILe <string>
:SENSe{1-16}:CORRection:COLLect:PORT{1-4}:SHORT3:S1P:FILe?
Description
Sets the SHORT3 Standard S1P file on the selected port of the indicated channel.
Returns the SHORT3 Standard S1P file on the selected port of the indicated channel.
Cmd Parameters
<string> s1p file path
See definition of <string>.
Query Parameters
NA
Query Output
<char1> s1p file path
Range
NA
Default Value
NA
Syntax Example
:SENS1:CORR:COLL:PORT1:SHORT3:S1P:FIL 'C:\Test\SHORT3.s1p'
:SENS1:CORR:COLL:PORT1:SHORT3:S1P:FIL?
:SENSe{1-16}:CORRection:COLLect:PORT{1-4}:OPEN:SERial <string>
:SENSe{1-16}:CORRection:COLLect:PORT{1-4}:OPEN:SERial?
Description
The use of Port 3 and/or Port 4 requires a 4-port VNA instrument.
Sets the open standard serial number on the selected port of the indicated channel.
Returns the open standard serial number on the selected port of the indicated channel.
Cmd Parameters
<string> The input parameter is any combination of numbers and letters. See definition of <string>.
Query Parameters
<char> The output parameter can be any combination of numbers and letters.
Range
NA
Default Value
NA
Syntax Example
:SENS1:CORR:COLL:PORT1:OPEN:SER '123456'
:SENS1:CORR:COLL:PORT1:OPEN:SER?
:SENSe{1-16}:CORRection:COLLect:PORT{1-4}:REFLection:COMPonent <char>
:SENSe{1-16}:CORRection:COLLect:PORT{1-4}:REFLection:COMPonent?
Description
The use of Port 3 and/or Port 4 requires a 4-port VNA instrument.
Sets the reflection standard to use for the reflection calibration on the indicated channel and port where the reflection standard types can be:
OPEN = Open
SHORt = Short
OFSH1 = Offset Short 1
OFSH2 = Offset Short 2
OFSH3 = Offset Short 3
NONE = Only returned by the query if no reflection component has been defined.
Returns the reflect standard to use for the reflection calibration on the indicated channel and port.
Cmd Parameters
<char> OPEN | SHORt | OFSH1 | OFSH2 | OFSH3
Query Parameters
<char> OPEN | SHOR | OFSH1 | OFSH2 | OFSH3 | NONE
Range
NA
Default
OPEN
Syntax Example
:SENS1:CORR:COLL:PORT1:REFL:COMP OPEN
:SENS1:CORR:COLL:PORT1:REFL:COMP?
:SENSe{1-16}:CORRection:COLLect:PORT{1-4}:SHORT
Description
The use of Port 3 and/or Port 4 requires a 4-port VNA instrument.
Initiates collection of the short standard data for the calibration on the indicated channel and port.
See Calibration Component Parameters for a complete listing of calibration components, connectors, and their command parameters.
No query.
Cmd Parameters
NA
Range
NA
Default Value
NA
Syntax Example
:SENS1:CORR:COLL:PORT1:SHORT
:SENSe{1-16}:CORRection:COLLect:PORT{1-4}:SHORT:FREQuency <NRf>
:SENSe{1-16}:CORRection:COLLect:PORT{1-4}:SHORT:FREQuency?
Description
The use of Port 3 and/or Port 4 requires a 4-port VNA instrument.
Sets the short standard reference frequency on the selected port of the indicated channel.
Returns the short standard reference frequency on the selected port of the indicated channel.
See Calibration Component Parameters for a complete listing of calibration components, connectors, and their Command Parameters.
Cmd Parameters
<NRf> The input parameter is in Hertz.
Query Parameters
<NR3> The output parameter is in Hertz.
Range
MPND
Default Value
See Calibration Component Parameters.
Syntax Example
:SENS1:CORR:COLL:PORT1:SHORT:FREQ 1.0E7
:SENS1:CORR:COLL:PORT1:SHORT:FREQ?
:SENSe{1-16}:CORRection:COLLect:PORT{1-4}:SHORT:L0 <NRf>
:SENSe{1-16}:CORRection:COLLect:PORT{1-4}:SHORT:L0?
Description
Sets the short standard L0 (L zero) inductance on the selected port of the indicated channel
The L0 coefficient is measured in Henrys.
Returns the short standard L0 inductance on the selected port of the indicated channel.
See Calibration Component Parameters for a complete listing of calibration components, connectors, and their Command Parameters.
Cmd Parameters
<NRf> The input parameter is in Henrys.
Query Parameters
<NR3> The output parameter is in Henrys.
Range
MPND
Default Value
See Calibration Component Parameters.
Syntax Example
:SENS1:CORR:COLL:PORT1:SHORT:L0 2.0E0
:SENS1:CORR:COLL:PORT1:SHORT:L0?
:SENSe{1-16}:CORRection:COLLect:PORT{1-4}:SHORT:L1 <NRf>
:SENSe{1-16}:CORRection:COLLect:PORT{1-4}:SHORT:L1?
Description
The use of Port 3 and/or Port 4 requires a 4-port VNA instrument. Sets the short standard L1 (L one) inductance coefficient on the selected port of the indicated channel.
The L1 coefficient is measured in Henrys/Hertz.
Returns the short standard L1 inductance coefficient on the selected port of the indicated channel.
See Calibration Component Parameters for a complete listing of calibration components, connectors, and their command parameters.
Cmd Parameters
<NRf> The input parameter is in Henrys/Hertz.
Query Parameters
<NR3> The output parameter is in Henrys/Hertz.
Range
MPND
Default Value
See Calibration Component Parameters.
Syntax Example
:SENS1:CORR:COLL:PORT1:SHORT:L1 2.0E0
:SENS1:CORR:COLL:PORT1:SHORT:L1?
:SENSe{1-16}:CORRection:COLLect:PORT{1-4}:SHORT:L2 <NRf>
:SENSe{1-16}:CORRection:COLLect:PORT{1-4}:SHORT:L2?
Description
The use of Port 3 and/or Port 4 requires a 4-port VNA instrument.
Sets the short standard L2 inductance coefficient on the selected port of the indicated channel.
The L2 coefficient is measured in Henrys/Hertz^2.
Returns the short standard L2 inductance coefficient on the selected port of the indicated channel.
Cmd Parameters
<NRf> The input parameter is in Henrys/Hertz^2.
Query Parameters
<NR3> The output parameter is in Henrys/Hertz^2.
Range
MPND
Default Value
See Calibration Component Parameters.
Syntax Example
:SENS1:CORR:COLL:PORT1:SHORT:L2 2.0E0
:SENS1:CORR:COLL:PORT1:SHORT:L2?
:SENSe{1-16}:CORRection:COLLect:PORT{1-4}:SHORT:L3 <NRf>
:SENSe{1-16}:CORRection:COLLect:PORT{1-4}:SHORT:L3?
Description
The use of Port 3 and/or Port 4 requires a 4-port VNA instrument.
Sets the short standard L3 inductance coefficient on the selected port of the indicated channel.
The L3 coefficient is measured in Henrys/Hertz^3.
Returns the short standard L3 inductance coefficient on the selected port of the indicated channel.
Cmd Parameters
<NRf> The input parameter is in Henrys/Hertz^3.
Query Parameters
<NR3> The output parameter is in Henrys/Hertz^3.
Range
MPND
Default Value
See Calibration Component Parameters.
Syntax Example
:SENS1:CORR:COLL:PORT1:SHORT:L3 2.0E0
:SENS1:CORR:COLL:PORT1:SHORT:L3?
:SENSe{1-16}:CORRection:COLLect:PORT{1-4}:SHORT:LABEL <string>
:SENSe{1-16}:CORRection:COLLect:PORT{1-4}:SHORT:LABEL?
Description
The use of Port 3 and/or Port 4 requires a 4-port VNA instrument.
Sets the short standard label on the selected port of the indicated channel
Returns the short standard label on the selected port of the indicated channel.
Cmd Parameters
<string> The input parameter is any combination of numbers and letters.
See definition of <string>.
Query Parameters
<char> The output parameter can be any combination of numbers and letters.
Range
NA
Default Value
NA
Syntax Example
:SENS1:CORR:COLL:PORT1:SHORT:LABEL 'IC7000'
:SENS1:CORR:COLL:PORT1:SHORT:LABEL?
:SENSe{1-16}:CORRection:COLLect:PORT{1-4}:SHORT:LOSS <NRf>
:SENSe{1-16}:CORRection:COLLect:PORT{1-4}:SHORT:LOSS?
Description
The use of Port 3 and/or Port 4 requires a 4-port VNA instrument.
Sets the short standard loss value on the selected port of the indicated channel.
Returns the short standard loss value on the selected port of the indicated channel.
Cmd Parameters
<NRf> The input parameter is in dB/mm.
Query Parameters
<NR3> The output parameter can be in dB/mm.
Range
MPND
Default Value
0
Syntax Example
:SENS1:CORR:COLL:PORT1:SHORT:LOSS 1.0E0
:SENS1:CORR:COLL:PORT1:SHORT:LOSS?
:SENSe{1-16}:CORRection:COLLect:PORT{1-4}:SHORT:OFFS <NRf>
:SENSe{1-16}:CORRection:COLLect:PORT{1-4}:SHORT:OFFS?
Description
The use of Port 3 and/or Port 4 requires a 4-port VNA instrument.
Sets the short standard offset on the selected port of the indicated channel.
Returns the short standard offset on the selected port of the indicated channel.
Cmd Parameters
<NRf> The input parameter is in Meters.
Query Parameters
<NR3> The output parameter is in Meters.
Range
MPND
Default Value
See Calibration Component Parameters.
Syntax Example
:SENS1:CORR:COLL:PORT1:SHORT:OFFS 2.5E1
:SENS1:CORR:COLL:PORT1:SHORT:OFFS?
:SENSe{1-16}:CORRection:COLLect:PORT{1-4}:SHORT:SERial <string>
:SENSe{1-16}:CORRection:COLLect:PORT{1-4}:SHORT:SERial?
Description
The use of Port 3 and/or Port 4 requires a 4-port VNA instrument.
Sets the short standard serial number on the selected port of the indicated channel.
Returns the short standard serial number on the selected port of the indicated channel.
Cmd Parameters
<string> The input parameter is any combination of numbers and letters.
See definition of <string>.
Query Parameters
<char> The output parameter can be any combination of numbers and letters.
Range
NA
Default Value
NA
Syntax Example
:SENS1:CORR:COLL:PORT1:SHORT:SER '123456'
:SENS1:CORR:COLL:PORT1:SHORT:SER?
:SENSe{1-16}:CORRection:COLLect:PORT{1-4}:SHORT1
Description
The use of Port 3 and/or Port 4 requires a 4-port VNA instrument.
Initiates collection of the Offset Short1 (Short one) standard data for the calibration on the indicated channel and port.
See Calibration Component Parameters. for a complete listing of calibration components, connectors, and their Command Parameters.
No query.
Cmd Parameters
NA
Range
NA
Default Value
NA
Syntax Example
:SENS1:CORR:COLL:PORT1:SHORT1
:SENSe{1-16}:CORRection:COLLect:PORT{1-4}:SHORT1:FREQuency <NRf>
:SENSe{1-16}:CORRection:COLLect:PORT{1-4}:SHORT1:FREQuency?
Description
The use of Port 3 and/or Port 4 requires a 4-port VNA instrument.
Sets the Short1 standard reference frequency on the selected port of the indicated channel.
Returns the Short1 standard reference frequency on the selected port of the indicated channel.
See Calibration Component Parameters for a complete listing of calibration components, connectors, and their Command Parameters.
Cmd Parameters
<NRf> The input parameter is in Hertz.
Query Parameters
<NR3> The output parameter is in Hertz.
Range
MPND
Default Value
See Calibration Component Parameters.
Syntax Example
:SENS1:CORR:COLL:PORT1:SHORT1:FREQ 1.0E7
:SENS1:CORR:COLL:PORT1:SHORT1:FREQ?
:SENSe{1-16}:CORRection:COLLect:PORT{1-4}:SHORT1:L0 <NRf>
:SENSe{1-16}:CORRection:COLLect:PORT{1-4}:SHORT1:L0?
Description
The use of Port 3 and/or Port 4 requires a 4-port VNA instrument.
Sets the Short1 standard L0 (L zero) inductance on the selected port of the indicated channel.
The L0 parameter is measured in Henrys.
Returns the Short1 standard L0 inductance on the selected port of the indicated channel.
See Calibration Component Parameters for a complete listing of calibration components, connectors, and their Command Parameters.
Cmd Parameters
<NRf> The input parameter is in Henrys.
Query Parameters
<NR3> The output parameter is in Henrys.
Range
MPND
Default Value
See Calibration Component Parameters.
Syntax Example
:SENS1:CORR:COLL:PORT1:SHORT1:L0 2.0E0
:SENS1:CORR:COLL:PORT1:SHORT1:L0?
:SENSe{1-16}:CORRection:COLLect:PORT{1-4}:SHORT1:L1 <NRf>
:SENSe{1-16}:CORRection:COLLect:PORT{1-4}:SHORT1:L1?
Description
The use of Port 3 and/or Port 4 requires a 4-port VNA instrument.
Sets the Short1 standard L1 (L one) inductance coefficient on the selected port of the indicated channel.
The L1 coefficient is measured in Henrys/Hertz.
Returns the Short1 standard L1 inductance coefficient on the selected port of the indicated channel.
Cmd Parameters
<NRf> The input parameter is in Henrys/Hertz.
Query Parameters
<NR3> The output parameter is in Henrys/Hertz.
Range
MPND
Default Value
See Calibration Component Parameters.
Syntax Example
:SENS1:CORR:COLL:PORT1:SHORT1:L1 2.0E0
:SENS1:CORR:COLL:PORT1:SHORT1:L1?
:SENSe{1-16}:CORRection:COLLect:PORT{1-4}:SHORT1:L2 <NRf>
:SENSe{1-16}:CORRection:COLLect:PORT{1-4}:SHORT1:L2?
Description
The use of Port 3 and/or Port 4 requires a 4-port VNA instrument.
Sets the Short1 standard L2 inductance coefficient on the selected port of the indicated channel.
The L2 coefficient is measured in Henrys/Hertz^2.
Returns the Short1 standard L2 inductance coefficient on the selected port of the indicated channel.
Cmd Parameters
<NRf> The input parameter is in Henrys/Hertz^2.
Query Parameters
<NR3> The output parameter is in Henrys/Hertz^2.
Range
MPND
Default Value
See Calibration Component Parameters.
Syntax Example
:SENS1:CORR:COLL:PORT1:SHORT1:L2 2.0E0
:SENS1:CORR:COLL:PORT1:SHORT1:L2?
:SENSe{1-16}:CORRection:COLLect:PORT{1-4}:SHORT1:L3 <NRf>
:SENSe{1-16}:CORRection:COLLect:PORT{1-4}:SHORT1:L3?
Description
The use of Port 3 and/or Port 4 requires a 4-port VNA instrument.
Sets the Short1 standard L3 inductance coefficient on the selected port of the indicated channel.
The L3 coefficient is measured in Henrys/Hertz^3.
Returns the Short1 standard L3 inductance coefficient on the selected port of the indicated channel.
Cmd Parameters
<NRf> The input parameter is in Henrys/Hertz^3.
Query Parameters
<NR3> The output parameter is in Henrys/Hertz^3.
Range
MPND
Default Value
See Calibration Component Parameters.
Syntax Example
:SENS1:CORR:COLL:PORT1:SHORT1:L3 2.0E0
:SENS1:CORR:COLL:PORT1:SHORT1:L3?
:SENSe{1-16}:CORRection:COLLect:PORT{1-4}:SHORT1:LABEL <string>
:SENSe{1-16}:CORRection:COLLect:PORT{1-4}:SHORT1:LABEL?
Description
The use of Port 3 and/or Port 4 requires a 4-port VNA instrument.
Sets the Short1 (Short one) standard label on the selected port of the indicated channel.
Returns the Short1 standard label on the selected port of the indicated channel.
See Calibration Component Parameters for a complete listing of calibration components, connectors, and their Command Parameters.
Cmd Parameters
<string> The input parameter is any combination of numbers and letters.
See definition of <string>.
Query Parameters
<char> The output parameter can be any combination of numbers and letters.
Range
NA
Default Value
NA
Syntax Example
:SENS1:CORR:COLL:PORT1:SHORT1:LABEL 'IC7000'
:SENS1:CORR:COLL:PORT1:SHORT1:LABEL?
:SENSe{1-16}:CORRection:COLLect:PORT{1-4}:SHORT1:OFFS <NRf>
:SENSe{1-16}:CORRection:COLLect:PORT{1-4}:SHORT1:OFFS?
Description
The use of Port 3 and/or Port 4 requires a 4-port VNA instrument.
Sets the Short1 (Short 1) standard offset on the selected port of the indicated channel.
Returns the Short1 standard offset on the selected port of the indicated channel.
Cmd Parameters
<NRf> The input parameter is in Meters.
Query Parameters
<NR3> The output parameter is in Meters.
Range
MPND
Default Value
See Calibration Component Parameters.
Syntax Example
:SENS1:CORR:COLL:PORT1:SHORT1:OFFS 2.5E1
:SENS1:CORR:COLL:PORT1:SHORT1:OFFS?
:SENSe{1-16}:CORRection:COLLect:PORT{1-4}:SHORT1:LOSS <NRf>
:SENSe{1-16}:CORRection:COLLect:PORT{1-4}:SHORT1:LOSS?
Description
The use of Port 3 and/or Port 4 requires a 4-port VNA instrument.
Sets the Short1 standard loss value on the selected port of the indicated channel.
Returns the Short1 standard loss value on the selected port of the indicated channel.
Cmd Parameters
<NRf> The input parameter is in dB/mm.
Query Parameters
<NR3> The output parameter can be in dB/mm.
Range
MPND
Default Value
0
Syntax Example
:SENS1:CORR:COLL:PORT1:SHORT1:LOSS 1.0E0
:SENS1:CORR:COLL:PORT1:SHORT1:LOSS?
:SENSe{1-16}:CORRection:COLLect:PORT{1-4}:SHORT1:SERial <string>
:SENSe{1-16}:CORRection:COLLect:PORT{1-4}:SHORT1:SERial?
Description
The use of Port 3 and/or Port 4 requires a 4-port VNA instrument.
Sets the Short1 standard serial number on the selected port of the indicated channel.
Returns the Short1 standard serial number on the selected port of the indicated channel.
Cmd Parameters
<string> The input parameter is any combination of numbers and letters. See definition of <string>.
Query Parameters
<char> The output parameter can be any combination of numbers and letters.
Range
NA
Default Value
NA
Syntax Example
:SENS1:CORR:COLL:PORT1:SHORT1:SER '123456'
:SENS1:CORR:COLL:PORT1:SHORT1:SER?
:SENSe{1-16}:CORRection:COLLect:PORT{1-4}:SHORT2
Description
The use of Port 3 and/or Port 4 requires a 4-port VNA instrument.
Initiates collection of the Offset Short2 standard data for the calibration on the indicated channel and port.
No query.
Cmd Parameters
NA
Range
NA
Default Value
NA
Syntax Example
:SENS1:CORR:COLL:PORT1:SHORT2
:SENSe{1-16}:CORRection:COLLect:PORT{1-4}:SHORT2:FREQuency <NRf>
:SENSe{1-16}:CORRection:COLLect:PORT{1-4}:SHORT2:FREQuency?
Description
The use of Port 3 and/or Port 4 requires a 4-port VNA instrument.
Sets the Short2 standard reference frequency on the selected port of the indicated channel.
Returns the Short2 standard reference frequency on the selected port of the indicated channel.
See Calibration Component Parameters for a complete listing of calibration components, connectors, and their Command Parameters.
Cmd Parameters
<NRf> The input parameter is in Hertz.
Query Parameters
<NR3> The output parameter is in Hertz.
Range
MPND
Default Value
See Calibration Component Parameters.
Syntax Example
:SENS1:CORR:COLL:PORT1:SHORT2:FREQ 1.0E7
:SENS1:CORR:COLL:PORT1:SHORT2:FREQ?
:SENSe{1-16}:CORRection:COLLect:PORT{1-4}:SHORT2:L0 <NRf>
:SENSe{1-16}:CORRection:COLLect:PORT{1-4}:SHORT2:L0?
Description
The use of Port 3 and/or Port 4 requires a 4-port VNA instrument.
Sets the Short2 standard L0 (L zero) inductance on the selected port of the indicated channel.
The L0 (zero) coefficient is measured in Henrys.
Returns the Short2 standard L0 inductance on the selected port of the indicated channel.
Cmd Parameters
<NRf> The input parameter is in Henrys.
Query Parameters
<NR3> The output parameter is in Henrys.
Range
MPND
Default Value
See Calibration Component Parameters.
Syntax Example
:SENS1:CORR:COLL:PORT1:SHORT2:L0 2.0E0
:SENS1:CORR:COLL:PORT1:SHORT2:L0?
:SENSe{1-16}:CORRection:COLLect:PORT{1-4}:SHORT2:L1 <NRf>
:SENSe{1-16}:CORRection:COLLect:PORT{1-4}:SHORT2:L1?
Description
The use of Port 3 and/or Port 4 requires a 4-port VNA instrument.
Sets the Short2 standard L1 (L one) inductance coefficient on the selected port of the indicated channel.
The L1 coefficient is measured in Henrys/Hertz.
Returns the Short2 standard L1 inductance coefficient on the selected port of the indicated channel.
Cmd Parameters
<NRf> The input parameter is in Henrys/Hertz.
Query Parameters
<NR3> The output parameter is in Henrys/Hertz.
Range
MPND
Default Value
See Calibration Component Parameters.
Syntax Example
:SENS1:CORR:COLL:PORT1:SHORT2:L1 2.0E0
:SENS1:CORR:COLL:PORT1:SHORT2:L1?
:SENSe{1-16}:CORRection:COLLect:PORT{1-4}:SHORT2:L2 <NRf>
:SENSe{1-16}:CORRection:COLLect:PORT{1-4}:SHORT2:L2?
Description
The use of Port 3 and/or Port 4 requires a 4-port VNA instrument.
Sets the Short2 standard L2 inductance coefficient on the selected port of the indicated channel.
The L2 coefficient is measured in Henrys/Hertz^2.
Returns the Short2 standard L2 inductance coefficient on the selected port of the indicated channel.
Cmd Parameters
<NRf> The input parameter is in Henrys/Hertz^2.
Query Parameters
<NR3> The output parameter is in Henrys/Hertz^2.
Range
MPND
Default Value
See Calibration Component Parameters.
Syntax Example
:SENS1:CORR:COLL:PORT1:SHORT2:L2 2.0E0
:SENS1:CORR:COLL:PORT1:SHORT2:L2?
:SENSe{1-16}:CORRection:COLLect:PORT{1-4}:SHORT2:L3 <NRf>
:SENSe{1-16}:CORRection:COLLect:PORT{1-4}:SHORT2:L3?
Description
The use of Port 3 and/or Port 4 requires a 4-port VNA instrument.
Sets the Short2 standard L3 inductance coefficient on the selected port of the indicated channel.
The L3 coefficient is measured in Henrys/Hertz^3.
Returns the Short2 standard L3 inductance coefficient on the selected port of the indicated channel.
Cmd Parameters
<NRf> The input parameter is in Henrys/Hertz^3.
Query Parameters
<NR3> The output parameter is in Henrys/Hertz^3.
Range
MPND
Default Value
See Calibration Component Parameters.
Syntax Example
:SENS1:CORR:COLL:PORT1:SHORT2:L3 2.0E0
:SENS1:CORR:COLL:PORT1:SHORT2:L3?
:SENSe{1-16}:CORRection:COLLect:PORT{1-4}:SHORT2:LABEL <string>
:SENSe{1-16}:CORRection:COLLect:PORT{1-4}:SHORT2:LABEL?
Description
The use of Port 3 and/or Port 4 requires a 4-port VNA instrument.
Sets the Short2 standard label on the selected port of the indicated channel.
Returns the Short2 standard label on the selected port of the indicated channel.
Cmd Parameters
<string> The input parameter is any combination of numbers and letters.
See definition of <string>.
Query Parameters
<char> The output parameter can be any combination of numbers and letters.
Range
NA
Default Value
NA
Syntax Example
:SENS1:CORR:COLL:PORT1:SHORT2:LABEL 'IC7000'
:SENS1:CORR:COLL:PORT1:SHORT2:LABEL?
:SENSe{1-16}:CORRection:COLLect:PORT{1-4}:SHORT2:LOSS <NRf>
:SENSe{1-16}:CORRection:COLLect:PORT{1-4}:SHORT2:LOSS?
Description
The use of Port 3 and/or Port 4 requires a 4-port VNA instrument.
Sets the Short2 standard loss value on the selected port of the indicated channel.
Returns the Short2 standard loss value on the selected port of the indicated channel.
Cmd Parameters
<NRf> The input parameter is in dB/mm.
Query Parameters
<NR3> The output parameter can be in dB/mm.
Range
MPND
Default Value
0
Syntax Example
:SENS1:CORR:COLL:PORT1:SHORT2:LOSS 1.0E0
:SENS1:CORR:COLL:PORT1:SHORT2:LOSS?
:SENSe{1-16}:CORRection:COLLect:PORT{1-4}:SHORT2:OFFS <NRf>
:SENSe{1-16}:CORRection:COLLect:PORT{1-4}:SHORT2:OFFS?
Description
The use of Port 3 and/or Port 4 requires a 4-port VNA instrument.
Sets the Short2 standard offset on the selected port of the indicated channel.
Returns the Short2 standard offset on the selected port of the indicated channel.
Cmd Parameters
<NRf> The input parameter is in Meters.
Query Parameters
<NR3> The output parameter is in Meters.
Range
MPND
Default Value
See Calibration Component Parameters.
Syntax Example
:SENS1:CORR:COLL:PORT1:SHORT2:OFFS 2.5E1
:SENS1:CORR:COLL:PORT1:SHORT2:OFFS?
:SENSe{1-16}:CORRection:COLLect:PORT{1-4}:SHORT2:SERial <string>
:SENSe{1-16}:CORRection:COLLect:PORT{1-4}:SHORT2:SERial?
Description
The use of Port 3 and/or Port 4 requires a 4-port VNA instrument.
Sets the Short2 standard serial number on the selected port of the indicated channel.
Returns the Short2 standard serial number on the selected port of the indicated channel.
Cmd Parameters
<string> The input parameter is any combination of numbers and letters. See definition of <string>.
Query Parameters
<char> The output parameter can be any combination of numbers and letters.
Range
NA
Default Value
NA
Syntax Example
:SENS1:CORR:COLL:PORT1:SHORT2:SER '123456'
:SENS1:CORR:COLL:PORT1:SHORT2:SER?
:SENSe{1-16}:CORRection:COLLect:PORT{1-4}:SHORT3
Description
The use of Port 3 and/or Port 4 requires a 4-port VNA instrument.
Initiates collection of the Offset Short3 standard data for the calibration on the indicated channel and port.
No query.
Cmd Parameters
NA
Range
NA
Default Value
NA
Syntax Example
:SENS1:CORR:COLL:PORT1:SHORT3
:SENSe{1-16}:CORRection:COLLect:PORT{1-4}:SHORT3:FREQuency <NRf>
:SENSe{1-16}:CORRection:COLLect:PORT{1-4}:SHORT3:FREQuency?
Description
The use of Port 3 and/or Port 4 requires a 4-port VNA instrument.
Sets the Short3 standard reference frequency on the selected port of the indicated channel.
Returns the Short3 standard reference frequency on the selected port of the indicated channel.
See Calibration Component Parameters for a complete listing of calibration components, connectors, and their Command Parameters.
Cmd Parameters
<NRf> The input parameter is in Hertz.
Query Parameters
<NR3> The output parameter is in Hertz.
Range
MPND
Default Value
See Calibration Component Parameters.
Syntax Example
:SENS1:CORR:COLL:PORT1:SHORT3:FREQ 1.0E7
:SENS1:CORR:COLL:PORT1:SHORT3:FREQ?
:SENSe{1-16}:CORRection:COLLect:PORT{1-4}:SHORT3:L0 <NRf>
:SENSe{1-16}:CORRection:COLLect:PORT{1-4}:SHORT3:L0?
Description
The use of Port 3 and/or Port 4 requires a 4-port VNA instrument.
Sets the Short3 standard L0 (L zero) inductance on the selected port of the indicated channel.
The L0 coefficient is measured in Henrys.
Returns the Short3 standard L0 inductance on the selected port of the indicated channel.
Cmd Parameters
<NRf> The input parameter is in Henrys.
Query Parameters
<NR3> The output parameter is in Henrys.
Range
MPND
Default Value
See Calibration Component Parameters.
Syntax Example
:SENS1:CORR:COLL:PORT1:SHORT3:L0 2.0E0
:SENS1:CORR:COLL:PORT1:SHORT3:L0?
:SENSe{1-16}:CORRection:COLLect:PORT{1-4}:SHORT3:L1 <NRf>
:SENSe{1-16}:CORRection:COLLect:PORT{1-4}:SHORT3:L1?
Description
The use of Port 3 and/or Port 4 requires a 4-port VNA instrument.
Sets the Short3 standard L1 (L one) inductance coefficient on the selected port of the indicated channel.
The L1 coefficient is measured in Henrys/Hertz.
Returns the Short3 standard L1 inductance coefficient on the selected port of the indicated channel.
Cmd Parameters
<NRf> The input parameter is in Henrys/Hertz.
Query Parameters
<NR3> The output parameter is in Henrys/Hertz.
Range
MPND
Default Value
See Calibration Component Parameters.
Syntax Example
:SENS1:CORR:COLL:PORT1:SHORT3:L1 2.0E0
:SENS1:CORR:COLL:PORT1:SHORT3:L1?
:SENSe{1-16}:CORRection:COLLect:PORT{1-4}:SHORT3:L2 <NRf>
:SENSe{1-16}:CORRection:COLLect:PORT{1-4}:SHORT3:L2?
Description
The use of Port 3 and/or Port 4 requires a 4-port VNA instrument.
Sets the Short3 standard L2 inductance coefficient on the selected port of the indicated channel.
The L2 coefficient is measured in Henrys/Hertz^2.
Returns the Short3 standard L2 inductance coefficient on the selected port of the indicated channel.
Cmd Parameters
<NRf> The input parameter is in Henrys/Hertz^2.
Query Parameters
<NR3> The output parameter is in Henrys/Hertz^2.
Range
MPND
Default Value
See Calibration Component Parameters.
Syntax Example
:SENS1:CORR:COLL:PORT1:SHORT3:L2 2.0E0
:SENS1:CORR:COLL:PORT1:SHORT3:L2?
:SENSe{1-16}:CORRection:COLLect:PORT{1-4}:SHORT3:L3 <NRf>
:SENSe{1-16}:CORRection:COLLect:PORT{1-4}:SHORT3:L3?
Description
The use of Port 3 and/or Port 4 requires a 4-port VNA instrument.
Sets the Short3 standard L3 inductance coefficient on the selected port of the indicated channel.
The L3 coefficient is measured in Henrys/Hertz^3.
Returns the Short3 standard L3 inductance coefficient on the selected port of the indicated channel.
Cmd Parameters
<NRf> The input parameter is in Henrys/Hertz^3.
Query Parameters
<NR3> The output parameter is in Henrys/Hertz^3.
Range
MPND
Default Value
See Calibration Component Parameters.
Syntax Example
:SENS1:CORR:COLL:PORT1:SHORT3:L3 2.0E0
:SENS1:CORR:COLL:PORT1:SHORT3:L3?
:SENSe{1-16}:CORRection:COLLect:PORT{1-4}:SHORT3:LABEL <string>
:SENSe{1-16}:CORRection:COLLect:PORT{1-4}:SHORT3:LABEL?
Description
The use of Port 3 and/or Port 4 requires a 4-port VNA instrument.
Sets the Short3 standard label on the selected port of the indicated channel.
Returns the Short1 standard label on the selected port of the indicated channel.
Cmd Parameters
<string> The input parameter is any combination of numbers and letters.
See definition of <string>.
Query Parameters
<char> The output parameter can be any combination of numbers and letters.
Range
NA
Default Value
NA
Syntax Example
:SENS1:CORR:COLL:PORT1:SHORT3:LABEL 'IC7000'
:SENS1:CORR:COLL:PORT1:SHORT3:LABEL?
:SENSe{1-16}:CORRection:COLLect:PORT{1-4}:SHORT3:LOSS <NRf>
:SENSe{1-16}:CORRection:COLLect:PORT{1-4}:SHORT3:LOSS?
Description
The use of Port 3 and/or Port 4 requires a 4-port VNA instrument.
Sets the Short3 standard loss value on the selected port of the indicated channel.
Returns the Short3 standard loss value on the selected port of the indicated channel.
Cmd Parameters
<NRf> The input parameter is in dB/mm.
Query Parameters
<NR3> The output parameter can be in dB/mm.
Range
MPND
Default Value
0
Syntax Example
:SENS1:CORR:COLL:PORT1:SHORT3:LOSS 1.0E0
:SENS1:CORR:COLL:PORT1:SHORT3:LOSS?
:SENSe{1-16}:CORRection:COLLect:PORT{1-4}:SHORT3:OFFS <NRf>
:SENSe{1-16}:CORRection:COLLect:PORT{1-4}:SHORT3:OFFS?
Description
The use of Port 3 and/or Port 4 requires a 4-port VNA instrument.
Sets the Short3 standard offset on the selected port of the indicated channel.
Returns the Short3 standard offset on the selected port of the indicated channel.
Cmd Parameters
<NRf> The input parameter is in Meters.
Query Parameters
<NR3> The output parameter is in Meters.
Range
MPND
Default Value
See Calibration Component Parameters.
Syntax Example
:SENS1:CORR:COLL:PORT1:SHORT3:OFFS 2.5E1
:SENS1:CORR:COLL:PORT1:SHORT3:OFFS?
:SENSe{1-16}:CORRection:COLLect:PORT{1-4}:SHORT3:SERial <string>
:SENSe{1-16}:CORRection:COLLect:PORT{1-4}:SHORT3:SERial?
Description
The use of Port 3 and/or Port 4 requires a 4-port VNA instrument.
Sets the Short3 standard serial number on the selected port of the indicated channel.
Returns the Short1 standard serial number on the selected port of the indicated channel.
Cmd Parameters
<string> The input parameter is any combination of numbers and letters. See definition of <string>.
Query Parameters
<char> The output parameter can be any combination of numbers and letters.
Range
NA
Default Value
NA
Syntax Example
:SENS1:CORR:COLL:PORT1:SHORT3:SER '123456'
:SENS1:CORR:COLL:PORT1:SHORT3:SER?
:SENSe{1-16}:CORRection:COLLect:PORT{1-4}:SHORT3:S1P[:STATe] <char>
:SENSe{1-16}:CORRection:COLLect:PORT{1-4}:SHORT3:S1P[:STATe]?
Description
Sets the SHORT3 S1P file settings ON/OFF on the selected port of the indicated channel.
ON means that the S1P file data will be used, OFF means the Circuit Model data will be used.
Returns the SHORT3 S1P file settings on the selected port of the indicated channel.
Cmd Parameters
<char> 1 | 0 | ON | OFF
See definition of <string>.
Query Parameters
NA
Query Output
<char> 1 | 0
Range
NA
Default Value
0
Syntax Example
:SENS1:CORR:COLL:PORT1:SHORT3:S1P 1
:SENS1:CORR:COLL:PORT1:SHORT3:S1P?
:SENSe{1-16}:CORRection:COLLect:PORT{1-4}:SLOAD1
Description
The use of Port 3 and/or Port 4 requires a 4-port VNA instrument.
Initiates collection of the Sliding Load1 standard data for the calibration on the indicated channel and port.
No query.
Cmd Parameters
NA
Range
NA
Default Value
NA
Syntax Example
:SENS1:CORR:COLL:PORT1:SLOAD1
:SENSe{1-16}:CORRection:COLLect:PORT{1-4}:SLOAD2
Description
The use of Port 3 and/or Port 4 requires a 4-port VNA instrument.
Initiates collection of the Sliding Load2 standard data for the calibration on the indicated channel and port.
No query.
Cmd Parameters
NA
Range
NA
Default Value
NA
Syntax Example
:SENS1:CORR:COLL:PORT1:SLOAD2
:SENSe{1-16}:CORRection:COLLect:PORT{1-4}:SLOAD3
Description
The use of Port 3 and/or Port 4 requires a 4-port VNA instrument.
Initiates collection of the Sliding Load3 standard data for the calibration on the indicated channel and port.
No query.
Cmd Parameters
NA
Range
NA
Default Value
NA
Syntax Example
:SENS1:CORR:COLL:PORT1:SLOAD3
:SENSe{1-16}:CORRection:COLLect:PORT{1-4}:SLOAD4
Description
The use of Port 3 and/or Port 4 requires a 4-port VNA instrument.
Initiates collection of the Sliding Load4 standard data for the calibration on the indicated channel and port.
No query.
Cmd Parameters
NA
Range
NA
Default Value
NA
Syntax Example
:SENS1:CORR:COLL:PORT1:SLOAD4
:SENSe{1-16}:CORRection:COLLect:PORT{1-4}:SLOAD5
Description
The use of Port 3 and/or Port 4 requires a 4-port VNA instrument.
Initiates collection of the Sliding Load5 standard data for the calibration on the indicated channel and port.
No query.
Cmd Parameters
NA
Range
NA
Default Value
NA
Syntax Example
:SENS1:CORR:COLL:PORT1:SLOAD5
:SENSe{1-16}:CORRection:COLLect:PORT{1-4}:SLOAD6
Description
The use of Port 3 and/or Port 4 requires a 4-port VNA instrument.
Initiates collection of the Sliding Load6 standard data for the calibration on the indicated channel and port.
No query.
Cmd Parameters
NA
Range
NA
Default Value
NA
Syntax Example
:SENS1:CORR:COLL:PORT1:SLOAD6