VectorStar™ MS464xB Series Microwave Vector Network Analyzer User Interface Reference : PulseView™ : PulseView Setup and Configuration
 
PulseView Setup and Configuration
PULSEVIEW SETUP Menu
Previous
Application Menu for PulseView and IF Digitizer
Prerequisites
Option 35 and Option 42 Installed.
2-Port VNA
Navigation
MAIN | Application | APPLICATION | PulseView Setup | PULSEVIEW SETUP
PulseView Setup Menu Variants in Different Pulse Modes
PulseView Setup Menu – Profile Mode
PULSEVIEW SETUP Menu – Profile Mode
PulseView Configuration
PRI
Pulse Repetition Interval – Sets the primary clock sync pulse period in units of time and is inversely proportional to the pulse repetition frequency (PRI = 1 / PRF). PRI can also be set in the PULSEVIEW CONFIGURATION Dialog Box.
PRF
Pulse Repetition Frequency – Sets the primary clock sync pulse repetition rate in units of hertz and is inversely proportional to the pulse repetition interval (PRF = 1 / PRI). PRF can also be set in the PULSEVIEW CONFIGURATION Dialog Box.
Mode
Opens the PULSE MODE Menu to select the pulse mode: Profiling, Point-in-Pulse, Pulse-to-Pulse, Continuous Point-in-Pulse, and Continuous Profiling.
Note: The bottom trace display status bar indicates the PulseView pulse mode.
# of Points
Sets the number of measurement points from the Start to the Stop time settings. The number of points setting only applies to the Pulse Profile and Continuous Profiling measurement modes.
Receiver
Selects a receiver channel to configure.
Measurement Width
Sets the measurement width time (aperture).
Start Time
Enter a time to wait relative to T0 before measuring a pulse. The start time setting only applies to the pulse profile measurement mode.
Stop Time
Enter a time relative to T0 at which the measurement should stop. The stop time setting only applies to the pulse profile measurement mode.
Resolution
Enter a time to allow configuration of the pulse acquisition rate. This allows the user to speed up the measurements by changing the acquisition rate. This is useful when measurements are made at lower rep rates which may not require a high resolution.
Note 
T0 marks the beginning of each primary clock period. All Delay, Width, and the PRI parameter resolutions are limited to increments of 2.5 ns.
PulseView Setup Menu – Point in Pulse Mode
PULSEVIEW SETUP Menu – Point in Pulse Mode
PulseView Configuration
PRI
Pulse Repetition Interval – Sets the primary clock sync pulse period in units of time and is inversely proportional to the pulse repetition frequency (PRI = 1 / PRF). PRI can also be set in the PULSEVIEW CONFIGURATION Dialog Box.
PRF
Pulse Repetition Frequency – Sets the primary clock sync pulse repetition rate in units of hertz and is inversely proportional to the pulse repetition interval (PRF = 1 / PRI). PRF can also be set in the PULSEVIEW CONFIGURATION Dialog Box.
Mode
Opens the PULSE MODE Menu to select the pulse mode: Profiling, Point-in-Pulse, Pulse-to-Pulse, Continuous Point-in-Pulse, and Continuous Profiling.
Note: the bottom trace display status bar indicates the PulseView pulse mode.
Receiver
Selects a receiver channel to configure.
Measurement Width
Sets the measurement width time (aperture).
Delay
Enter a time relative to T0 before starting the measurement. The delay time setting applies to point-in-pulse and pulse-to-pulse measurement modes only.
Resolution
Enter a time to allow configuration of the pulse acquisition rate. This allows the user to speed up the measurements by changing the acquisition rate. This is useful when measurements are made at lower rep rates which may not require a high resolution.
Note 
T0 marks the beginning of each primary clock period. All Delay, Width, and the PRI parameter resolutions are limited to increments of 2.5 ns.
PulseView Setup Menu – Pulse to Pulse Mode
PULSEVIEW SETUP Menu – Pulse to Pulse Mode
PulseView Configuration
PRI
Pulse Repetition Interval – Sets the primary clock sync pulse period in units of time and is inversely proportional to the pulse repetition frequency (PRI = 1 / PRF). PRI can also be set in the PULSEVIEW CONFIGURATION Dialog Box.
PRF
Pulse Repetition Frequency – Sets the primary clock sync pulse repetition rate in units of hertz and is inversely proportional to the pulse repetition interval (PRF = 1 / PRI). PRF can also be set in the PULSEVIEW CONFIGURATION Dialog Box.
Mode
Opens the PULSE MODE Menu to select the pulse mode: Profiling, Point-in-Pulse, Pulse-to-Pulse, Continuous Point-in-Pulse, and Continuous Profiling.
Note: The bottom trace display status bar indicates the PulseView pulse mode
# of Pulses
Sets the number of pulse measurements to acquire. The number of pulses setting only applies to the pulse-to-pulse measurement mode.
Receiver
Selects a receiver channel to configure.
Measurement Width
Sets the measurement width time (aperture).
Delay
Enter a time relative to T0 before starting the measurement. The delay time setting applies to point-in-pulse and pulse-to-pulse measurement modes only.
Resolution
Enter a time to allow configuration of the pulse acquisition rate. This allows the user to speed up the measurements by changing the acquisition rate. This is useful when measurements are made at lower rep rates which may not require a high resolution.
Note 
T0 marks the beginning of each primary clock period. All Delay, Width, and the PRI parameter resolutions are limited to increments of 2.5 ns.
PulseView Setup Menu – Continuous Point in Pulse Mode
PULSEVIEW SETUP Menu – Continuous Point in Pulse Mode
PulseView Configuration
PRI
Pulse Repetition Interval – Sets the primary clock sync pulse period in units of time and is inversely proportional to the pulse repetition frequency (PRI = 1 / PRF). PRI can also be set in the PULSEVIEW CONFIGURATION Dialog Box.
Note: PRI is limited to limited to a minimum of 100 μs for Synch mode and 1 ms for Time mode Control Types
PRF
Pulse Repetition Frequency – Sets the primary clock sync pulse repetition rate in units of hertz and is inversely proportional to the pulse repetition interval (PRF = 1 / PRI). PRF can also be set in the PULSEVIEW CONFIGURATION Dialog Box.
Mode
Opens the PULSE MODE Menu to select the pulse mode: Profiling, Point-in-Pulse, Pulse-to-Pulse, Continuous Point-in-Pulse, and Continuous Profiling.
Note: The bottom trace display status bar indicates the PulseView pulse mode.
Control Type
Enabled only in CPIP or CProf modes. Choices are Sync and Time. Synch control requires that Trigger Out be connected to Pulse Synch In. This is noted with a tool tip that appears when the pointer is over the Control Type button as well as over the Pulse Configuration dialog button.
Receiver
Selects a receiver channel to configure.
Measurement Width
Sets the measurement width time (aperture).
Delay
Enter a time relative to T0 before starting the measurement. The delay time setting applies to point-in-pulse and pulse-to-pulse measurement modes only.
Note: In CPIP mode, minimum delay is 10 μs
Resolution
Enter a time to allow configuration of the pulse acquisition rate. This allows the user to speed up the measurements by changing the acquisition rate. This is useful when measurements are made at lower rep rates which may not require a high resolution.
Advanced Setup
Allows user to override the calculated capture duration. See ADVANCED SETUP Menu
Note 
T0 marks the beginning of each primary clock period. All Delay, Width, and the PRI parameter resolutions are limited to increments of 2.5 ns.
PulseView Setup Menu – Continuous Profiling Mode
PULSEVIEW SETUP Menu – Continuous Profiling Mode
PulseView Configuration
PRI
Pulse Repetition Interval – Sets the primary clock sync pulse period in units of time and is inversely proportional to the pulse repetition frequency (PRI = 1 / PRF). PRI can also be set in the PULSEVIEW CONFIGURATION Dialog Box.
Note: PRI is limited to limited to a minimum of 100 μs for Synch mode and 1 ms for Time mode Control Types
PRF
Pulse Repetition Frequency – Sets the primary clock sync pulse repetition rate in units of hertz and is inversely proportional to the pulse repetition interval (PRF = 1 / PRI). PRF can also be set in the PULSEVIEW CONFIGURATION Dialog Box.
Mode
Opens the PULSE MODE Menu to select the pulse mode: Profiling, Point-in-Pulse, Pulse-to-Pulse, Continuous Point-in-Pulse, and Continuous Profiling. The bottom trace display status bar indicates the PulseView pulse mode:
Control Type
Appears only in CPIP or CProf mode. Selecting Control Type toggles between Time and Synch. When Synch is toggled, a notification appears saying “Synch ctrl needs Trigger Out connected to Pulse Synch In” See encircled images at left.
Profiling Points
Sets the number of displayed analysis points from the Start to the Stop time settings. This can be greater or less than the # of sweep points. Applies only to CProf mode.
Receiver – Selects a receiver channel to configure.
Measurement Width – Sets the measurement width time (aperture).
Start Time
Enter a time to wait relative to T0 before measuring a pulse. The start time setting only applies to the pulse profile measurement mode.
Stop Time
Enter a time relative to T0 at which the measurement should stop. The stop time setting only applies to the pulse profile measurement mode.
(Note: T0 marks the beginning of each primary clock period. All Delay, Width, and the PRI parameter resolutions are limited to increments of 2.5 ns.)
Resolution
Enter a time to allow configuration of the pulse acquisition rate. This allows the user to speed up the measurements by changing the acquisition rate and is useful when measurements are made at lower rep rates which may not require a high resolution. Increasing the resolution increases the maximum acquisition depth proportionately. Default is 2.5253 ns resolution and 0.5 secs of depth. Minimum resolution is 2.5 ns.
Advanced Setup
Allows user to override the calculated capture duration. See ADVANCED SETUP Menu.
PULSE MODE Menu
Previous
PULSEVIEW SETUP Menu
Prerequisites
Option 35 and Option 042 Installed.
2-Port VNA (PulseView menus are disabled when a 4-port test set is connected).
Navigation
MAIN | Application | APPLICATION | PulseView Setup | PULSEVIEW SETUP | Mode | PULSE MODE
PULSEVIEW PULSE MODE Menu
Pulse Profiling (PROF)
A measurement mode where there is an incoming pulse and it is desired to acquire S-parameters at sub-intervals within that pulse. These values are mapped versus position (time) within the pulse. This can be visualized as a profiling pulse (the sub-interval segment/measurement width) that steps through the DUT output pulse in time increments and makes a measurement at each step. This measurement is often used to check the evolution of DUT behavior over the duration of the pulse due to thermal, trapping, or other effects.
Point-in-Pulse (PIP)
A measurement mode where an S-parameter is only to be acquired at one time point within a pulse (or more precisely, within a small time window/measurement width). This is usually somewhere in the middle of the pulse and is selected when transient effects associated with pulse edges are not a concern, but the DUT must operate in pulsed mode (for example, to minimize device heating). It is desired to see how the behavior at the point within the pulse changes over frequency or power ranges.
Pulse-to-Pulse (P2P)
A measurement mode similar to pulse profiling except the time interval is between multiple pulses rather than within a pulse. Longer time-scale transient effects (such as thermal or memory related effects) are of interest. This is usually a triggered measurement since the DUT often needs to be in a long-time-scale-equilibrium before the excitation starts.
Continuous Point-in-Pulse (CPIP)
A measurement mode similar to Point-in-Pulse which is adequate for many swept pulse measurements. However, there are cases, particularly related to DUT transient response, where more detailed control of the VNA sweep operation relative to the pulses and the measurements is required. This is the purpose of the Continuous Point-in-Pulse mode (CPIP) where the entire sweep of frequency or power is done during one acquisition.
Continuous Profiling
As in CPIP, Continuous Profiling (CProf) measurement also uses a single acquisition for an entire frequency or power sweep but is a generalization in data analysis relative to CPIP. In CProf, one is not restricted to a measurement window associated with each sweep point but one can have up to 25000 measurement windows dispersed over the entire sweep time or located in some subset of that range. The number of displayed points is equal to the number of profiling points.
PULSE RECEIVER Menu
Previous
PULSEVIEW SETUP Menu
Prerequisites
Option 35 and Option 042 Installed.
Navigation
MAIN | Application | APPLICATION | PulseView Setup | PULSEVIEW SETUP | Receiver | RECEIVER
PULSEVIEW RECEIVER Menu (2-Port Systems)
A1 Receiver
Selects the A1 receiver.
A2 Receiver
Selects the A2 receiver.
B1 Receiver
Selects the B1 receiver.
B2 Receiver
Selects the B2 receiver.
PULSEVIEW RECEIVER Menu (4-Port Systems)
A1 Receiver
Selects the A1 receiver.
A2 Receiver
Selects the A2 receiver.
A3 Receiver
Selects the A3 receiver.
A4 Receiver
Selects the A4 receiver.
B1 Receiver
Selects the B1 receiver.
B2 Receiver
Selects the B2 receiver.
B3 Receiver
Selects the B3 receiver.
B4 Receiver
Selects the B4 receiver.
 
ADVANCED SETUP Menu
Previous
PulseView Setup Menu – Continuous Point in Pulse Mode or PulseView Setup Menu – Continuous Profiling Mode
Prerequisites
Option 35 and Option 042 Installed.
2-Port VNA (PulseView menus are disabled when a 4-port test set is connected).
PulseView pulse mode must be in Continuous Point in Pulse or Continuous Profiling mode.
Navigation
MAIN | Application | APPLICATION | PulseView Setup | PULSEVIEW SETUP | Advanced Setup| ADVANCED SETUP
PULSEVIEW MODE Menu
When Advanced Setup is selected this menu is displayed allowing the user to override the calculated capture duration.
Calculated Capt. Dur.
Not active. Simply displays current capture duration.
Override Capt. Dur?
Select toggles override on and off
Desired Capt. Dur.
The desired capture duration maximum is dependent on the PULSEVIEW menu Resolution field entry, which can have a maximum of 14 seconds.
PULSEVIEW CONFIGURATION Dialog Box
When the PULSEVIEW SETUP menu is active, the PULSEVIEW CONFIGURATION dialog is available.
Previous
PULSEVIEW SETUP Menu
Prerequisites
Option 35 and Option 042 Installed.
2-Port VNA (PulseView menus are disabled when a 4-port test set is connected).
Navigation
MAIN | Application | APPLICATION | PulseView Setup | PULSEVIEW SETUP | PulseView Configuration | PULSEVIEW CONFIGURATION
The pulse configuration dialog box configures individual pulse channels, measurement mode, and other options.
PULSEVIEW CONFIGURATION Dialog Box
1. Items in the Pulse Control area are explained in Pulse Controls.
2. Items in the Pulse Display area are detailed in Pulse Display.
3. Items in the Pulse Generator, Measurement Configuration, and Pulse Power Calibration tabs are detailed in Pulse Generator Configuration, Pulse Measurement Setup, and Pulse Power Calibration Setup.
Pulse Controls
PULSE Control Area (1 of 2)
Marker
The pulse Marker feature allows up to two markers to be set on the pulse display. The marker window can then zoomed in and back out. The feature is described further in Zoom Marker
Resolution
Opens a field to change the pulse acquisition rate. This allows the user to speed up the measurements by changing the acquisition rate. This is useful when measurements are made at lower rep rates which may not require a high resolution. As resolution changes, the maximum values of most other PulseView settings changes (maximum acquisition length increases proportionately with resolution; default is 2.5253 ns resolution and 0.5 s maximum acquisition length).
Synch
Toggles the pulse Synch source:
INTERNAL synchronizes the primary clock to the leading edge of an internally provided synch pulse.
CONTINUOUS provides continuous, unsynchronized full-speed data acquisition.
EXT. RISING/EXT FALLING synchronizes the primary clock to the leading edge of an externally provided synch pulse. The synch event can be triggered to either the rising or falling edge of the synch pulse (EXT. RISING or EXT. FALLING).
When the Synch type selected is EXT. RISING or EXT. FALLING, an “Ext. Synch Marking” check box is made available on the PulseView Configuration | Measurement tab. This feature allows the use of the external Pulse Synch In signal to mark the data stream in lieu of using the internally generated T0 pulsed signal. When enabled, the internal pulse generators are disabled
PRI
Pulse Repetition Interval – Sets the primary clock synch pulse period in units of time and is inversely proportional to the pulse repetition frequency (PRI = 1 / PRF). When a PRI value is entered, the PRF is computed automatically. PRI can also be set in the PULSEVIEW SETUP Menu.
PRF
Pulse Repetition Frequency – Sets the primary clock synch pulse repetition rate in units of hertz and is inversely proportional to the pulse repetition interval (PRF = 1 / PRI). When a PRF value is entered, the PRI is computed automatically. PRF can also be set in the PULSEVIEW SETUP Menu.
Control Type
Enabled only in CPIP or CProf modes. Choices are Sync and Time. Synch control requires that Trigger Out be connected to Pulse Synch In. This is noted with a tool tip that appears when a pointer is over the button, as well as on the PULSEVIEW SETUP menu.
 
Cancel
Cancels all settings and closes the PULSEVIEW CONFIGURATION dialog box.
Done
Applies all settings and closes the PULSEVIEW CONFIGURATION dialog box. Done must be clicked to apply any setting changes made.
Pulse Display
The pulse display shows a profile of each pulse channel, the synch signal, and the receiver measurement acquisition profile on a relative time scale. Pulse generators (PG1 through PG4) can be renamed by clicking the channel name in the pulse generator configuration tab.
PulseView Configuration Dialog – Pulse Display (with example settings)
1. Pulse Measurement:
 Width = 5 µs
 Start = 3.5 µs
 Stop = 8.5 µs
2. Pulse Synchronization Signal:
 Set to Internal
3. Pulse Generator 1 (PG1):
 Singlet Mode
 6 µs Delay,  1 µs Width
4. Pulse Generator 2 (PG2):
 Doublet Mode
 4 µs Delay 1,  1 µs Width 1
 6 µs Delay 2,  1 µs Width 2
5. Pulse Generator 3 (PG3):
 Quadruplet Mode
 2 µs Delay,  1 µs Width
 4 µs Delay,  1 µs Width
 6 µs Delay,  1 µs Width
 8 µs Delay,  1 µs Width
6. Pulse Generator 4 (PG4):
 Burst Mode
 # of Bursts: 10
 Period: 400 ns
 Delay: 4 µs
 Width: 50 ns
Refer to the following sections for more information on configuring the individual pulse generators, measurement acquisition and pulse power calibration setup:
Pulse Generator Configuration
Pulse Measurement Setup
Pulse Power Calibration Setup
Zoom Marker
The Zoom Marker feature allows the visible resolution of the pulse display to be increased. Click Zoom Marker ON to enable the feature.
Pulse Markers
1. With the Zoom Marker toggled ON, click and drag the left edge of the display to reveal a marker.
2. Click and drag the left edge of the display to reveal a second marker.
3. Marker values change dynamically when the marker is dragged along the time axis.
To increase the visible resolution of a pulse, click Zoom In. The time scale changes to the range between the markers. The above zoom area is shown below.
Pulse Markers – Zoomed
Click Zoom Out to return to the original time scale.
Pulse Generator Configuration
To configure the pulse generators, open the PULSEVIEW CONFIGURATION dialog box and enter the information into the tabbed pulse generator configuration areas (see Figure: Pulse Configuration Tab through Figure: Burst Pulse Profile).
Each pulse generator tab allows full configuration of an MS464xB Series pulse generator channel, including:
Pulse Generator Mode (Singlet, Doublet, Triplet, Quadruplet, or Burst) and Enable/Disable Toggle
Pulse Delay
Pulse Width
Pulse Polarity (note that rear panel signals are always active whether or not the PG is enabled)
To define a pulse width, establish a start and stop time. The minimum pulse width is often twice the minimum settable resolution; for example, with a minimum measurement resolution of 2.5 ns, the minimum measurement pulse width is 5 ns. The usefulness of a given width setting is dependent on the rise/fall times of the pulse generator and the bandwidth of the DUT.
Example of singlet pulse setup:
Pulse Configuration Tab
1. Select the pulse Mode and Enable the Pulse Generator.
2. Select the Delay and Width field to edit.
3. Enter the value in the field toolbar and terminate with the desired units.
4. Click the PG1 Label field and enter the desired pulse generator name (optional).
5. Duty Cycle (display only, appears with Singlet pulse mode).
Singlet Pulse Profile
Note 
Delay + Width PRI
Example of quadruplet pulse setup:
Pulse Configuration Tab
1. Invalid settings are indicated with a warning symbol as shown above.
Quadruplet Pulse Profile (with Delay 4 corrected to 8 μs)
Note 
Delay 2 > Delay 1 + Width 1
Delay 3 > Delay 2 + Width 2
Delay 4 > Delay 3 + Width 3
Delay 4 + Width 4 PRI
When Burst pulse type is selected, the Burst Setup area allows configuration of:
# of Bursts
Burst Period
Burst Delay
Burst Width
Example of burst pulse setup:
Pulse Configuration Tab
Burst Pulse Profile
Note 
Delay + (# of Bursts – 1) x Period + Width PRI
Width Period
Pulse Measurement Setup
To configure the pulse measurement acquisition, open the PulseView Configuration dialog box and enter the information into the tabbed measurement configuration areas (see Figure: PulseView Configuration Measurement Tab vs. Measurement Modes through Figure: Continuous Profile Measurement).
The measurement mode is a per-channel global setting that applies to all receivers. Each receiver can be set for independent acquisition parameters or coupled to the same parameters.
Note 
All receivers are set to the same mode, but the pulse acquisition parameters can be set differently between each of the receivers.
Acquisition parameters include:
Pulse Measurement Mode (Pulse Profile, Point-in-Pulse, Pulse-to-Pulse, Continuous Point-in-Pulse, and Continuous Profiling)
Number of measurement Points (Pulse Profile mode only)
Number of measurement Pulses (Pulse-to-Pulse mode only)
Pulse generators on only during measurement selection (Pulse Profile, Continuous Point-in-Pulse, and Continuous Profiling modes only)
Receiver channel (A1, A2, B1, B2 – 2-Port Systems)
Receiver channel (A1, A2, A3, A4, B1, B2, B3, B4 – 4-Port Systems)
Couple Receiver Parameters
Measurement Width
Pulse Start and Stop time (Pulse Profile and Continuous Profiling modes only)
Delay time (Point-in-Pulse, Pulse-to-Pulse and Continuous Point-in-Pulse modes only)
Note 
In profile mode, overlapping measurements are permitted when the measurement width is set larger than the duration between measurement points, determined by:
(Stop – Start) / # of Points
 
PulseView Configuration Measurement Tab vs. Measurement Modes
1. Set the measurement mode (the measurement mode is a per-channel setting that applies to all receivers).
2. Select the receiver and check if the measurement parameters are to be applied equally to all receivers.
3. Set the measurement parameters on a per-receiver or coupled basis.
4. When the Measurement tab is selected the Pulse generators on only during measurement toggle selection is only visible when measurement mode is Pulse Profile or Continuous Profile. Selecting this feature allows pulse generators to start up only when a measurement is taken.
5. In Continuous Profiling mode there is a concept of profiling points which is the number of displayed analysis points. This can be greater or less than the # of sweep points.
Pulse Generators On Only During Measurement
Sometimes a user may want the pulse generators to start up only when a measurement is taken, thus avoiding some early device stimulation (the user may be looking at long thermal time constant effects) or some potential overshoot conditions (from level dip release). Since some users rely on a quasi-static measurement, this function can be toggled on or off.
Example of Pulse Profile Measurement:
Pulse Profile Measurement Setup
Pulse Profile Measurement
Note 
Stop – Start PRI
# of points instrument number of points (25k or 100k)
The measurement width can overlap other measurement widths.
Example of Point-in-Pulse Measurement:
Point-in-Pulse Measurement Setup
Point-in-Pulse Measurement
Note 
Delay + Measurement Width PRI
Example of Pulse-to-Pulse Measurement:
Pulse-to-Pulse Measurement Setup
Pulse-to-Pulse Measurement
Note 
Delay + Measurement Width PRI
Example of Continuous Point-in-Pulse Measurement:
Continuous Point-in-Pulse Measurement Setup
Continuous Point-in-Pulse Measurement
Note 
Delay + Measurement Width PRI
Example of Continuous Profile Measurement:
Continuous Profile Measurement Setup
Continuous Profile Measurement
Note 
Delay + Measurement Width PRI
Pulse Power Calibration Setup
To configure the pulse power calibration, open the PULSEVIEW CONFIGURATION dialog box and enter the information into the tabbed Power Cal and Leveling Mode configuration areas (see Figure: Pulse Power Calibration Setup).
When enabled, power calibrations will use a pulsed power sensor to only analyze power in a specific time gate specified in the dialog. This allows one to calibrate power in the middle of pulse, on the overshoot section, or any other specific interval on the pulsed envelope. The pulsed power meter trigger must be connected to the VNA Synch Out port for the time alignment to be established.
Leveling Mode allows the user to more precisely control power based on the signal envelope. Allows selection of:
Real Time CW leveling mode – Level adjustments occur based upon detection of the CW non-pulsed signal. This will result in a leveling of average power, but occurs on a continuous (sub-microsecond) basis and has no impact on sweep time.
Per Sweep leveling mode – Level adjustments occur on a per sweep basis targeting the specific portion of the pulse described by the reference measurement width and delay. The leveling adjustments only occur at the end of the sweep, so fast drift (by a pre-amplifier for example) will not be corrected. This approach has minimal impact on sweep time.
Per Point leveling mode – Level adjustments occur on a per point basis targeting the specific portion of the pulse described by the reference measurement width and delay. The leveling adjustment is made at every point before the main measurement is made, so even fast drifts will be corrected. This approach has more impact on sweep time, but the power level at a specific point in the pulse is more precisely controlled.
Leveling Mode (Figure: Pulse Power Calibration – Leveling Mode Setups) can be set in this tab, or in POWER SETUP | Leveling Mode | LEVELING MODE (LEVELING MODE Menu – 2-Port VNAs or LEVELING MODE Menu – 4-Port VNAs). Tolerance (for Per Point Pulsed) can only be set in this dialog.
 
Pulse Power Calibration Setup
 
Pulse Power Calibration – Leveling Mode Setups