1. The status panel provides quick access to common spectrum analyzer settings. See STATUS PANEL (SEM). 2. SEM Mask 3. Spectrum Emission 4. SEM Mask 5. All SEM measurement parameters are set via the SETUP menu. See SETUP Menu (SEM). |
AUTO MAX PWR Toggle auto maximum power to set the maximum power automatically. MANUAL MAX PWR Enables when Auto Max Power is toggled. REF LEVEL Sets the reference level of the top graticule line in the selected units. If the reference level offset is not zero, OFFSET REF LEVEL is displayed at this location. Refer to Setting Amplitude Parameters. SCALE/DIV Sets the graticule scale/division for log-based units. This setting does not apply to linear units. REF LEVEL OFFSET Sets the reference level offset in dB units. This setting can compensate for the presence of external input attenuation or gain. FREQ OFFSET Accounts for frequency conversions outside of the analyzer. Refer to Setting Frequency Parameters. ATTEN LEVEL When auto attenuation is off, sets input attenuation. PRE AMP Toggles the low-noise front-end preamplifier on or off. Refer to Setting Amplitude Parameters. TRACES/CURSORS Displays the current status of up to six traces or cursors in a quick-view summary. When the measurement view is set to RF Spectrum, trace information is displayed in this area. When the measurement view is set to Spectrogram, cursor information is displayed in this area. Cursors are available in the Spectrum measurement with the Spectrogram view selected. The summary information includes the trace or cursor number, type, mode, and detector type. The active trace will show a highlighted background with the mode and detector type restated under the table. In Spectrogram, a reference trace (T0) will show you the settings of the trace used to fill the spectrogram. The reference trace settings are applied to all traces and cursors while in Spectrogram view. Selecting a trace or cursor in the summary panel activates the pressed trace or cursor and opens the TRACE menu, allowing you to select and set up an individual trace or cursor as desired. Refer to Setting Trace Parameters. SWEEP Toggles the current sweep setting between continuously or sweep once. Refer to Setting Sweep Parameters. FREQ REFERENCE Indicates the current frequency reference source of Internal High Accuracy (used after GPS has lost sync, but while the internal clock still has good GPS reference), Internal Standard Accuracy, External, or GNSS (GPS) Hi Accy (requires GPS). The instrument automatically selects the frequency reference in the following order of priority: external, GPS, then the internal time base. |
SELECT MASK Selects the desired upper or lower emission mask. Choose the desired mask from the list. IMPORT MASK Used to import a custom user mask. EXPORT MASK Exports an example user mask as a comma separated value file (see Example Custom Signal Emission Mask) or exports the currently selected custom mask. REF CH BW Sets the reference channel bandwidth for the measurement. AUTO MAX PWR Toggles automatic max power. Some segments in the mask are dependent on the main channel power. Enabling this automatically calculates the reference channel power by measuring the channel power of the center channel bandwidth. When disabled, the reference power must be entered manually below. MANUAL MAX PWR Used to manually enter the reference channel power for masks that have limits dependent on main channel power. PRESET SETUP Presets all values on the SETUP menu to default values. |