VectorStar™ MS4640A SeriesMicrowave Vector Network Analyzer : SCPI Commands : :CALCulate{1-16}[:SELected]:MARKer Subsystem
 
:CALCulate{1-16}[:SELected]:MARKer Subsystem
The :CALCulate{1-16}:PARameter{1-16}:MARKer subsystem commands control the active marker display, value, and search functions.
Marker Subsystems
Related marker configuration, control, and reporting commands are described multiple subsystems:
:CALCulate{1-16}:DISPlay:MARKer Subsystem
:CALCulate{1-16}:MARKer Subsystem
:CALCulate{1-16}:PARameter{1-16}:MARKer Subsystem
:CALCulate{1-16}:PARameter{1-16}:MSTatistics Subsystem
:CALCulate{1-16}[:SELected]:MARKer Subsystem
:CALCulate{1-16}[:SELected]:MARKer{1-13} Subsystem
:DISPlay Subsystem
 
:CALCulate{1-16}[:SELected]:MARKer:ACTivate?
Description
Query only. On the indicated channel, the query outputs the active marker number of the active trace.
To activate a marker, use the command:
:CALCulate{1-16}[:SELected]:MARKer{1-13}:ACTivate
Markers 1 through 12 are standard measurement markers. Marker 13 is the reference marker.
Query Parameters
<NR1> The output parameter is an integer.
Range
1 to 13
Default Value
0
Syntax Example
:CALC1:MARK:ACT?
 
:CALCulate{1-16}[:SELected]:MARKer:ALL[:STATe] <char1>
Description
The command toggles all markers on/off on the active trace of the given channel.
No query.
Cmd Parameters
<char1> 1 | 0 | ON | OFF
Range
NA
Default Value
0
Syntax Example
:CALC1:MARK:ALL ON
 
:CALCulate{1-16}[:SELected]:MARKer:MOVe:CENTer
Description
The command moves the active marker range value to the stimulus center range on the active trace of the indicated channel.
No query.
Cmd Parameters
NA
Range
NA
Default Value
NA
Syntax Example
:CALC1:MARK:MOV:CENT
 
:CALCulate{1-16}[:SELected]:MARKer:MOVe:REFMarker
Description
The command moves the active marker to the reference marker on the active trace of the indicated channel. The active marker cannot be a reference marker. The reference marker must be on.
No query.
Cmd Parameters
NA
Range
NA
Default Value
NA
Syntax Example
:CALC1:MARK:MOV:REFM
 
:CALCulate{1-16}[:SELected]:MARKer:MOVe:STARt
Description
The command moves the active marker range value to the stimulus start range on the active trace of the indicated channel.
No query.
Cmd Parameters
NA
Range
NA
Default Value
NA
Syntax Example
:CALC1:MARK:MOV:STAR
 
:CALCulate{1-16}[:SELected]:MARKer:MOVe:STOP
Description
The command moves the active marker range value to the stimulus stop range on the active trace of the indicated channel. No query.
Cmd Parameters
NA
Range
NA
Default Value
NA
Syntax Example
:CALC1:MARK:MOV:STOP
 
:CALCulate{1-16}[:SELected]:MARKer:MPSEArch
Description
The command performs a multiple peak search on the active trace of the indicated channel.
No query.
Cmd Parameters
NA
Range
NA
Default Value
NA
Syntax Example
:CALC1:MARK:MPSEA
 
:CALCulate{1-16}[:SELected]:MARKer:MPSEArch:EXCursion <NRf>
:CALCulate{1-16}[:SELected]:MARKer:MPSEArch:EXCursion?
Description
The command sets the marker search excursion value for multiple peak searches on the active trace of the indicated channel. The query outputs the marker search excursion value for multiple peak searches on the active trace of the indicated channel.
Cmd Parameters
<NRf> The input parameter is in dB.
Query Parameters
<NR3> The output parameter depends on the display type.
Range
MPNF
Default Value
0.000000E+000
Syntax Example
:CALC1:MARK:MPSEA:EXC 1.8E1
:CALC1:MARK:MPSEA:EXC?
 
:CALCulate{1-16}[:SELected]:MARKer:MPSEArch:POLarity <char1>
:CALCulate{1-16}[:SELected]:MARKer:MPSEArch:POLarity?
Description
The command sets the marker search polarity value for multiple peak searches on the active trace of the indicated channel. The query outputs the marker search polarity value for multiple peak searches on the active trace of the indicated channel.
Cmd Parameters
<char1> POSitive | NEGative | BOTH
Query Parameters
<char1> POS | NEG | BOTH
Range
NA
Default Value
POS
Syntax Example
:CALC1:MARK:MPSEA:POL POS
:CALC1:MARK:MPSEA:POL?
 
:CALCulate{1-16}[:SELected]:MARKer:MPSEArch:THREshold <NRf>
:CALCulate{1-16}[:SELected]:MARKer:MPSEArch:THREshold?
Description
The command sets the marker search threshold value for multiple peak searches on the active trace of the indicated channel. The query outputs the marker search threshold value for multiple peak searches on the active trace of the indicated channel.
Cmd Parameters
<NRf> The input parameter is in dB.
Query Parameters
<NR3> The output parameter depends on the display type.
Range
MPNF
Default Value
0.000000E+000
Syntax Example
:CALC1:MARK:MPSEA:THRE 1.8E1
:CALC1:MARK:MPSEA:THRE?
 
:CALCulate{1-16}[:SELected]:MARKer:MTSEArch
Description
Performs a multiple target search on the active trace of the indicated channel.
No query.
Cmd Parameters
NA
Range
NA
Default Value
NA
Syntax Example
:CALC1:MARK:MTSEA
 
:CALCulate{1-16}[:SELected]:MARKer:MTSEArch:TARget <NRf>
:CALCulate{1-16}[:SELected]:MARKer:MTSEArch:TARget?
Description
The command sets the marker search target value for multiple target searches on the active trace of the indicated channel. The query outputs the marker search target value for multiple target searches on the active trace of the indicated channel.
Cmd Parameters
<NRf> The input parameter depends on the display type.
Query Parameters
<NR3>, <NR3>, <NR3> The output parameter depends on the display type.
Range
MPNI
Default Value
0.000000E+000
Syntax Example
:CALC1:MARK:MTSEA:TAR 1E7
:CALC1:MARK:MTSEA:TAR?
 
:CALCulate{1-16}[:SELected]:MARKer:MTSEArch:TRANsition <char1>
:CALCulate{1-16}[:SELected]:MARKer:MTSEArch:TRANsition?
Description
The command sets the marker search transition value for multiple target searches on the active trace of the indicated channel. The query outputs the marker search transition value for multiple target searches on the active trace of the indicated channel.
Cmd Parameters
<char1> POSitive | NEGative | BOTH
Query Parameters
<char1> POS | NEG | BOTH
Range
NA
Default Value
BOTH
Syntax Example
:CALC1:MARK:MTSEA:TRAN POS
:CALC1:MARK:MTSEA:TRAN?
 
:CALCulate{1-16}[:SELected]:MARKer:OFF
Description
The command turns all markers of the indicated channel off. No query.
Cmd Parameters
NA
Range
NA
Default Value
NA
Syntax Example
:CALC1:MARK:OFF
 
:CALCulate{1-16}[:SELected]:MARKer:PSEArch <char1>
:CALCulate{1-16}[:SELected]:MARKer:PSEArch?
Description
The command performs a peak search on the active trace of the indicated channel. The query outputs the last marker search type for peak searches on the active trace of the indicated channel.
Cmd Parameters
<char1> PEAK | LEFT | RIGHT
Query Parameters
<char1> PEAK | LEFT | RIGHT
Range
NA
Default Value
PEAK
Syntax Example
:CALC1:MARK:PSEA PEAK
:CALC1:MARK:PSEA?
 
:CALCulate{1-16}[:SELected]:MARKer:PSEArch:EXCursion <NRf>
:CALCulate{1-16}[:SELected]:MARKer:PSEArch:EXCursion?
Description
The command sets the marker search excursion value for peak searches on the active trace of the indicated channel. The query outputs the marker search excursion value for peak searches on the active trace of the indicated channel.
Cmd Parameters
<NRf> The input parameter is in dB.
Query Parameters
<NR3> The output parameter depends on the display type.
Range
MPNI
Default Value
0.000000E+000
Syntax Example
:CALC1:MARK:PSEA:EXC 1.8E1
:CALC1:MARK:PSEA:EXC?
 
:CALCulate{1-16}[:SELected]:MARKer:PSEArch:POLarity <char1>
:CALCulate{1-16}[:SELected]:MARKer:PSEArch:POLarity?
Description
The command sets the marker search polarity value for peak searches on the active trace of the indicated channel. The query outputs the marker search polarity value for peak searches on the active trace of the indicated channel.
Cmd Parameters
<char1> POSitive | NEGative | BOTH
Query Parameters
<char1> POS | NEG | BOTH
Range
NA
Default Value
POS
Syntax Example
:CALC1:MARK:PSEA:POL POS
:CALC1:MARK:PSEA:POL?
 
:CALCulate{1-16}[:SELected]:MARKer:PSEArch:THREshold <NRf>
:CALCulate{1-16}[:SELected]:MARKer:PSEArch:THREshold?
Description
The command sets the marker search threshold value for peak searches on the active trace of the indicated channel. The query outputs the marker search threshold value for peak searches on the active trace of the indicated channel.
Cmd Parameters
<NRf> The input parameter is in dB.
Query Parameters
<NR3> The output parameter depends on the display type.
Range
MPNF
Default Value
0.000000E+000
Syntax Example
:CALC1:MARK:PSEA:THRE 1.8E1
:CALC1:MARK:PSEA:THRE?
 
:CALCulate{1-16}[:SELected]:MARKer:SEArch <char1>
:CALCulate{1-16}[:SELected]:MARKer:SEArch?
Description
The command sets the marker search type for the active marker on the active trace of the indicated channel. The query outputs the marker search type for the active marker on the active trace of the indicated channel.
Cmd Parameters
<char1> MAX | MIN | PEAK | TARGet
Query Parameters
<char1> MAX | MIN | PEAK | TARG
Range
NA
Default Value
MAX
Syntax Example
:CALC1:MARK:SEA MAX
:CALC1:MARK:SEA?
 
:CALCulate{1-16}[:SELected]:MARKer:SEArch:BANDwidth:DATA?
Description
Query only. The query outputs the marker search bandwidth data of the active marker on the active trace of the indicated channel.
Query Parameters
<NR3>, <NR3>, <NR3>, <NR3>{,<NR3>} The command outputs four or five <NR3> data elements in the following sequence:
First <NR3> - Bandwidth in Hertz.
Second <NR3> - Center of marker search range in Hertz, Meters, or Seconds.
Third <NR3> - Q which is a unitless number.
Fourth <NR3> - Loss in dB.
Fifth <NR3> - Optional. The Shape Factor which is a unitless number. This parameter is optional depending on if the Shape Function is turned on or off using the Shape Command.
Syntax Example
:CALC1:MARK:SEA:BAND:DATA?
 
:CALCulate{1-16}[:SELected]:MARKer:SEArch:BANDwidth:DEFine <NRf>
:CALCulate{1-16}[:SELected]:MARKer:SEArch:BANDwidth:DEFine?
Description
The command sets the marker search define value for bandwidth calculation of the active marker on the active trace of the indicated channel. The query outputs the marker search define value for bandwidth calculation of the active marker on the active trace of the indicated channel.
Cmd Parameters
<NRf> The input parameter depends on the display type.
Query Parameters
<NR3> The output parameter is in Hertz, Meters, or Seconds.
Range
MPNF
Default Value
0.000000E+000
Syntax Example
:CALC1:MARK:SEA:BAND:DEF 3.0E4
:CALC1:MARK:SEA:BAND:DEF?
 
:CALCulate{1-16}[:SELected]:MARKer:SEArch:BANDwidth:SHAPe:HIGH <NRf>
:CALCulate{1-16}[:SELected]:MARKer:SEArch:BANDwidth:SHAPe:HIGH?
Description
The command sets the marker search high value for bandwidth shape factor calculation of the active marker on the active trace of the indicated channel. The query outputs the marker search high value for bandwidth shape factor calculation of the active marker on the active trace of the indicated channel.
Cmd Parameters
<NRf> The input parameter depends on the display type.
Query Parameters
<NR3> The output parameter is a unitless number.
Range
MPNF
Default Value
0.000000E+000
Syntax Example
:CALC1:MARK:SEA:BAND:SHAP:HIGH 4E0
:CALC1:MARK:SEA:BAND:SHAP:HIGH?
 
:CALCulate{1-16}[:SELected]:MARKer:SEArch:BANDwidth:SHAPe:LOW <NRf>
:CALCulate{1-16}[:SELected]:MARKer:SEArch:BANDwidth:SHAPe:LOW?
Description
The command sets the marker search low value for bandwidth shape factor calculation of the active marker on the active trace of the indicated channel. The query outputs the marker search low value for bandwidth shape factor calculation of the active marker on the active trace of the indicated channel.
Cmd Parameters
<NRf> The input parameter depends on the display type.
Query Parameters
<NR3> The output parameter is a unitless number.
Range
MPNF
Default Value
0.000000E+000
Syntax Example
:CALC1:MARK:SEA:BAND:SHAP:LOW 4E0
:CALC1:MARK:SEA:BAND:SHAP:LOW?
 
:CALCulate{1-16}[:SELected]:MARKer:SEArch:BANDwidth:SHAPe[:STATe] <char1>
:CALCulate{1-16}[:SELected]:MARKer:SEArch:BANDwidth:SHAPe[:STATe]?
Description
The command toggles on/off the marker search bandwidth shape factor calculation of the active marker on the active trace of the indicated channel. The query outputs the on/off status of marker search bandwidth shape factor of the active marker calculation on the active trace of the indicated channel.
Cmd Parameters
<char1> 1 | 0 | ON | OFF
Query Parameters
<char1> 1 | 0
Range
NA
Default Value
0
Syntax Example
:CALC1:MARK:SEA:BAND:SHAP 1
:CALC1:MARK:SEA:BAND:SHAP?
 
:CALCulate{1-16}[:SELected]:MARKer:SEArch:BANDwidth[:STATe] <char1>
:CALCulate{1-16}[:SELected]:MARKer:SEArch:BANDwidth[:STATe]?
Description
The command toggles on/off the marker search bandwidth calculation of the active marker on the active trace of the indicated channel. The query outputs the on/off status of marker search bandwidth calculation of the active marker on the active trace of the indicated channel.
Cmd Parameters
<char1> 1 | 0 | ON | OFF
Query Parameters
<char1> 1 | 0
Range
NA
Default Value
0
Syntax Example
:CALC1:MARK:SEA:BAND ON
:CALC1:MARK:SEA:BAND?
 
:CALCulate{1-16}[:SELected]:MARKer:SEArch:NOTCh:DATA?
Description
Query only. The query outputs the marker search notch bandwidth data of the active marker on the active trace of the indicated channel.
Query Parameters
<NR3>, <NR3>, <NR3>, <NR3>{, <NR3>} The output parameter is in Hertz, Meters, or Seconds.
Range
Outputs multiple NR3 data
Default Value
NA
Syntax Example
:CALC1:MARK:SEA:NOTC:DATA?
 
:CALCulate{1-16}[:SELected]:MARKer:SEArch:NOTCh:DEFine <NRf>
:CALCulate{1-16}[:SELected]:MARKer:SEArch:NOTCh:DEFine?
Description
The command sets the marker search define value for notch bandwidth calculation of the active marker on the active trace of the indicated channel. The query outputs the marker search define value for notch bandwidth calculation of the active marker on the active trace of the indicated channel.
Cmd Parameters
<NRf> The input parameter depends on the display type.
Query Parameters
<NR3> The output parameter is in Hertz, Meters, or Seconds.
Range
MPNF
Default Value
0.000000E+000
Syntax Example
:CALC1:MARK:SEA:NOTC:DEF 1.0E4
:CALC1:MARK:SEA:NOTC:DEF?
 
:CALCulate{1-16}[:SELected]:MARKer:SEArch:NOTCh:SHAPe:HIGH <NRf>
:CALCulate{1-16}[:SELected]:MARKer:SEArch:NOTCh:SHAPe:HIGH?
Description
The command sets the marker search high value for notch shape factor calculation of the active marker on the active trace of the indicated channel. The query outputs the marker search high value for notch shape factor calculation of the active marker on the active trace of the indicated channel.
Cmd Parameters
<NRf> The input parameter depends on the display type.
Query Parameters
<NR3> The output parameter is in Hertz, Meters, or Seconds.
Range
MPNF
Default Value
0.000000E+000
Syntax Example
:CALC1:MARK:SEA:NOTC:SHAP:HIGH 4E0
:CALC1:MARK:SEA:NOTC:SHAP:HIGH?
 
:CALCulate{1-16}[:SELected]:MARKer:SEArch:NOTCh:SHAPe:LOW <NRf>
:CALCulate{1-16}[:SELected]:MARKer:SEArch:NOTCh:SHAPe:LOW?
Description
The command sets the marker search low value for notch shape factor calculation of the active marker on the active trace of the indicated channel. The query outputs the marker search low value for notch shape factor calculation of the active marker on the active trace of the indicated channel.
Cmd Parameters
<NRf> The input parameter depends on the display type.
Query Parameters
<NR3> The output parameter is a unitless number.
Range
MPNF
Default Value
0.000000E+000
Syntax Example
:CALC1:MARK:SEA:NOTC:SHAP:LOW 4E0
:CALC1:MARK:SEA:NOTC:SHAP:LOW?
 
:CALCulate{1-16}[:SELected]:MARKer:SEArch:NOTCh:SHAPe[:STATe] <char1>
:CALCulate{1-16}[:SELected]:MARKer:SEArch:NOTCh:SHAPe[:STATe]?
Description
The command toggles on/off the marker search notch shape factor calculation of the active marker on the active trace of the indicated channel. Outputs on/off status of marker search notch shape factor calculation of the active marker on the active trace of the indicated channel.
Cmd Parameters
<char1> 1 | 0 | ON | OFF
Query Parameters
<char1> 1 | 0
Range
NA
Default Value
0
Syntax Example
:CALC1:MARK:SEA:NOTC:SHAP ON
:CALC1:MARK:SEA:NOTC:SHAP?
 
:CALCulate{1-16}[:SELected]:MARKer:SEArch:NOTCh[:STATe] <char1>
:CALCulate{1-16}[:SELected]:MARKer:SEArch:NOTCh[:STATe]?
Description
The command toggles on/off the marker search notch calculation of the active marker on the active trace of the indicated channel. The query outputs the on/off status of marker search notch calculation of the active marker on the active trace of the indicated channel.
Cmd Parameters
<char1> 1 | 0 | ON | OFF
Query Parameters
<char1> 1 | 0
Range
NA
Default Value
0
Syntax Example
:CALC1:MARK:SEA:NOTC ON
:CALC1:MARK:SEA:NOTC?
 
:CALCulate{1-16}[:SELected]:MARKer:SEArch:RANGe:ALL[:STATe] <char1>
:CALCulate{1-16}[:SELected]:MARKer:SEArch:RANGe:ALL[:STATe]?
Description
The command toggles on/off applying the marker search range to all traces of the indicated channel. The query outputs the on/off status of applying the marker search range to all traces of the indicated channel.
Cmd Parameters
<char1> 1 | 0 | ON | OFF
Query Parameters
<char1> 1 | 0
Range
NA
Default Value
0
Syntax Example
:CALC1:MARK:SEA:RANG:ALL ON
:CALC1:MARK:SEA:RANG:ALL?
 
:CALCulate{1-16}[:SELected]:MARKer:SEArch:RANGe:STARt:X <NRf>
:CALCulate{1-16}[:SELected]:MARKer:SEArch:RANGe:STARt:X?
Description
The command sets the marker search range start range value on the active trace of the indicated channel. The query outputs the marker search range start range value on the active trace of the indicated channel.
Cmd Parameters
<NRf> The input parameter is in Hertz, Meters, or Seconds.
Query Parameters
<NR3> The output parameter is in Hertz, Meters, or Seconds.
Range
MPND
Default Value
0.00000000000E+000
Syntax Example
:CALC1:MARK:SEA:RANG:STAR:X 2.5E9
:CALC1:MARK:SEA:RANG:STAR:X?
 
:CALCulate{1-16}[:SELected]:MARKer:SEArch:RANGe:STOP:X <NRf>
:CALCulate{1-16}[:SELected]:MARKer:SEArch:RANGe:STOP:X?
Description
The command sets the marker search range stop range value on the active trace of the indicated channel. The query outputs the marker search range stop range value on the active trace of the indicated channel.
Cmd Parameters
<NRf> The input parameter is in Hertz, Meters, or Seconds.
Query Parameters
<NR3> The output parameter is in Hertz, Meters, or Seconds.
]Range
MPND
Default Value
0.00000000000E+000
Syntax Example
:CALC1:MARK:SEA:RANG:STOP:X 2.5E9
:CALC1:MARK:SEA:RANG:STOP:X?
 
:CALCulate{1-16}[:SELected]:MARKer:SEArch:RANGe[:STATe] <char1>
:CALCulate{1-16}[:SELected]:MARKer:SEArch:RANGe[:STATe]?
Description
The command toggles on/off the marker search range on the active trace of the indicated channel. The query outputs the on/off status of marker search range on the active trace of the indicated channel.
Cmd Parameters
<char1> 1 | 0 | ON | OFF
Query Parameters
<char1> 1 | 0
Range
NA
Default Value
0
Syntax Example
:CALC1:MARK:SEA:RANG ON
:CALC1:MARK:SEA:RANG?
 
:CALCulate{1-16}[:SELected]:MARKer:SEArch:TRACKing[:STATe] <char1>
:CALCulate{1-16}[:SELected]:MARKer:SEArch:TRACKing[:STATe]?
Description
The command toggles on/off marker search tracking on the active trace of the indicated channel. The query outputs the on/off status of marker search tracking on the active trace of the indicated channel.
Cmd Parameters
<char1> 1 | 0 | ON | OFF
Query Parameters
<char1> 1 | 0
Range
NA
Default Value
0
Syntax Example
:CALC1:MARK:SEA:TRACK ON
:CALC1:MARK:SEA:TRACK?
 
:CALCulate{1-16}[:SELected]:MARKer:SET:CENTer
Description
The command sets the stimulus center range to the active marker range value on the active trace of the indicated channel. No query.
Cmd Parameters
NA
Range
NA
Default Value
NA
Syntax Example
:CALC1:MARK:SET:CENT
 
:CALCulate{1-16}[:SELected]:MARKer:SET:REFLevel
Description
The command sets the display reference level to the active marker response value on the active trace of the indicated channel. No query.
Cmd Parameters
NA
Range
NA
Default Value
NA
Syntax Example
:CALC1:MARK:SET:REFL
 
:CALCulate{1-16}[:SELected]:MARKer:SET:STARt
Description
The command sets the stimulus start range to the active marker range value on the active trace of the indicated channel. No query.
Cmd Parameters
NA
Range
NA
Default Value
NA
Syntax Example
:CALC1:MARK:SET:STAR
 
:CALCulate{1-16}[:SELected]:MARKer:SET:STOP
Description
The command sets the stimulus stop range to the active marker range value on the active trace of the indicated channel. No query.
Cmd Parameters
NA
Range
NA
Default Value
NA
Syntax Example
:CALC1:MARK:SET:STOP
 
:CALCulate{1-16}[:SELected]:MARKer:TSEArch <char1>
:CALCulate{1-16}[:SELected]:MARKer:TSEArch?
Description
The command performs a target search on the active trace of the indicated channel. The query outputs the marker search type for target searches on the active trace of the indicated channel.
Cmd Parameters
<char1> TARGet | LEFT | RIGHT
Query Parameters
<char1> TARG | LEFT | RIGHT
Range
NA
Default Value
TARG
Syntax Example
:CALC1:MARK:TSEA TARG
:CALC1:MARK:TSEA?
 
:CALCulate{1-16}[:SELected]:MARKer:TSEArch:TARget <NRf>
:CALCulate{1-16}[:SELected]:MARKer:TSEArch:TARget?
Description
The command sets the marker search target value for target searches on the active trace of the indicated channel. The query outputs the marker search target value for target searches on the active trace of the indicated channel.
See Table: Trace Parameters and Coefficients for a complete listing of trace graph types, default settings, and available ranges.
Cmd Parameters
<NRf> The input parameter depends on the display type.
Query Parameters
<NR3>, <NR3>, <NR3> The output parameter depends on the display type.
Range
MPNF
Default Value
0.000000E+000
Syntax Example
:CALC1:MARK:TSEA:TAR 1E7
:CALC1:MARK:TSEA:TAR?
 
:CALCulate{1-16}[:SELected]:MARKer:TSEArch:TRANsition <char1>
:CALCulate{1-16}[:SELected]:MARKer:TSEArch:TRANsition?
Description
The command sets the marker search transition value for target searches on the active trace of the indicated channel. The query outputs the marker search transition value for target searches on the active trace of the indicated channel.
Cmd Parameters
<char1> POSitive | NEGative | BOTH
Query Parameters
<char1> POS | NEG | BOTH
Range
NA
Default Value
BOTH
Syntax Example
:CALC1:MARK:TSEA:TRAN POS