VectorStar™ MS464xB Series Microwave Vector Network Analyzer User Interface Reference : Calibration Menus – 4-Port VNAs : Manual Calibration General Dialog Boxes
 
Manual Calibration General Dialog Boxes
These are general information dialog boxes that can be linked to from most manual calibration procedures. A representative typical of dialog box contents are shown below:
AIR EQUIVALENT LENGTH Calculator Dialog Box
MICROSTRIP INFO and USER DEFINED MICROSTRIP Dialog Boxes
STANDARD INFO (OFFSET SHORT) W1-Connector (F) Dialog Box
STANDARD INFO (SOLT/R) V-Connector (M) Dialog Box
STANDARD INFO (Broadband Cal) 0.8 mm Connector (M) Dialog Box
STANDARD INFO (TRIPLE OFFSET SHORT) W1-Connector (M) Dialog Box
THRU INFO Dialog Box – 4-Port VNAs
USER DEFINED MATCH DEVICES Dialog Box – LRL/LRM
WAVEGUIDE INFO and USER DEFINED WAVEGUIDE Dialog Boxes
AIR EQUIVALENT LENGTH Calculator Dialog Box
Use the AIR EQUIVALENT LENGTH calculator dialog box to speed configuration of a through line by entering its length in picoseconds (ps) and its dielectric constant. The calculator returns the air equivalent length in millimeters (mm).
Previous
The AIR EQUIVALENT LENGTH dialog box can be accessed from multiple locations.
THRU INFO Dialog Box – 4-Port VNAs
Navigation
MAIN | Calibration | CALIBRATION | Calibrate | CALIBRATE | AutoCal | AUTOCAL | Modify Cal Setup | MODIFY AUTOCAL SETUP Dialog Box | Thru Info | THRU INFO Dialog Box | Calculator Icon | AIR EQUIVALENT LENGTH Dialog Box ‘
AIR EQUIVALENT LENGTH CONVERSION CALCULATOR Dialog Box
Using the Calculator
1. Use the Enter length in ps (picoseconds) to input a length.
For example, enter a value of 250 ps.
2. Use the Enter constant to change the dielectric constant as required.
For example, change the dielectric constant to 1.2.
3. Click the Calculate Air Equivalent Length button.
4. The required value appears in the Air Equivalent Length in mm field.
Using the examples above, an air equivalent length of 68.465319... appears in the field.
5. Click OK.
6. The Thru Info dialog box reappears with the calculated value in the Length (mm) field.
7. Using the examples above, the Length (mm) field displays 68.4653 mm.
THRU INFO Dialog Box – 4-Port VNAs
8. Click OK on the Thru Info dialog box.
9. The Modify AutoCal Setup dialog box reappears.
MODIFY 4-PORT AUTOCAL SETUP Dialog Box
MODIFY 2-PORT AUTOCAL SETUP Dialog Box – 4-Port VNAs
MODIFY 1-PORT AUTOCAL SETUP Dialog Box – 4-Port VNAs
MICROSTRIP INFO and USER DEFINED MICROSTRIP Dialog Boxes
Dialog Box Name Changes
The exact names of these dialog boxes change depending on the calibration method and connector type selected.
This dialog box can be linked to from a multiple dialog sources. The links below are for dialogs that appear in this chapter.
TWO PORT CAL SETUP (SOLT/R, Microstrip) Dialog Box.
TWO PORT CAL SETUP (LRL/LRM, Coaxial) Dialog Box.
TRANS. FREQ. RESP. CAL SETUP (SSST, Microstrip) Dialog Box.
REFL. FREQ. RESP. CAL SETUP (SSLT, Microstrip) Dialog Box.
Navigation
This dialog box can be linked to from a multiple dialog sources. The links below are for dialogs that appear in this chapter.
MAIN | Calibration | CALIBRATION | Calibrate | CALIBRATE | Manual Cal | MANUAL CAL | 3-Port Cal | THREE PORT CAL | Modify Cal Setup | CAL SETUP | Cal Method = SSLT | Line Type = Microstrip | Edit Cal Params | THREE PORT CAL SETUP (SSLT, Microstrip) Dialog Box
MAIN | Calibration | CALIBRATION | Calibrate | CALIBRATE | Manual Cal | MANUAL CAL | 2-Port Cal | TWO PORT CAL | Modify Cal Setup | CAL SETUP | Cal Method = SOLT/R | Line Type = Microstrip | Edit Cal Params | TWO PORT CAL SETUP (SOLT/R, Microstrip) Dialog Box
MAIN | Calibration | CALIBRATION | Calibrate | CALIBRATE | Manual Cal | MANUAL CAL | 2-Port Cal | TWO PORT CAL | Modify Cal Setup | CAL SETUP | Cal Method = LRL/LRM | Line Type = Microstrip | Edit Cal Params | TWO PORT CAL SETUP (LRL/LRM, Microstrip) Dialog Box
MAIN | Calibration | CALIBRATION | Calibrate | CALIBRATE | Manual Cal | MANUAL CAL | Trans. Freq. Resp.| TRANS. RESP. | Modify Cal Setup | CAL SETUP | Cal Method = SSST | Line Type = Microstrip | Edit Cal Params | TRANSMISSION REFLECTION RESPONSE SETUP (SSST, Microstrip) Dialog Box
MAIN | Calibration | CALIBRATION | Calibrate | CALIBRATE | Manual Cal | MANUAL CAL | Refl. Freq. Resp. | REFL. RESP. | Modify Cal Setup | CAL SETUP | Cal Method = SSLT | Line Type = Microstrip | Edit Cal Params | REFLECTION RESPONSE SETUP (SSLT, Microstrip) Dialog Box
MICROSTRIP INFO and USER DEFINED MICROSTRIP Dialog Boxes
1. At left MICROSTRIP INFO for 10 Mil Kit.
2. At right USER DEFINED MICROSTRIP Input Dialog Box
STANDARD INFO (OFFSET SHORT) W1-Connector (F) Dialog Box
The figure below shows a typical connector standard information dialog box. The contents depend on the calibration method, the calibration line type, and the calibration connectors and genders used. Anritsu- provided dialogs provide read-only information. User-defined dialogs allow user input of parameters and names.
Previous
This dialog box can be linked to from multiple dialog sources.
FULL FOUR PORT CAL SETUP (SSLT, Coaxial) Dialog Box
FULL FOUR PORT CAL SETUP (SSST, Coaxial) Dialog Box
Navigation
The links below are for dialogs that appear in this chapter.
MAIN | Calibration | CALIBRATION | Calibrate | CALIBRATE | Manual Cal | MANUAL CAL | 4-Port Cal | FOUR PORT CAL | Modify Cal Setup | CAL SETUP | Cal Method = SSLT | Line Type = Coaxial | Edit Cal Params | FULL FOUR PORT CAL SETUP (SSLT, Coaxial) Dialog Box | DUT Connector = W1-Conn (F) | Standard Info | STANDARD INFO (OFFSET SHORT, W1-CONN (F)) Dialog Box
STANDARD INFO (OFFSET SHORT) W-1Conn (F)
Representative Standard Information Dialog Box
STANDARD INFO (SOLT/R) V-Connector (M) Dialog Box
Figure: l (SOLT/R) V-Conn (M) shows a typical connector standard information dialog box. The contents depend on the calibration method, the calibration line type, and the calibration connectors and genders used. Anritsu dialogs provide read-only information. User-defined dialogs allow user input of parameters and names.
Previous
This dialog box can be linked from multiple dialog sources. The links below are for dialogs that appear in this chapter.
FULL FOUR PORT CAL SETUP (SOLT/R, Coaxial) Dialog Box
FULL FOUR PORT CAL SETUP (SSLT, Coaxial) Dialog Box
FULL FOUR PORT CAL SETUP (SSST, Coaxial) Dialog Box
Navigation
This dialog box can be linked to from a multiple dialog sources. The links below are for dialogs that appear in this chapter.
MAIN | Calibration | CALIBRATION | Calibrate | CALIBRATE | Manual Cal | MANUAL CAL | 4-Port Cal | FOUR PORT CAL | Modify Cal Setup | CAL SETUP | Cal Method = SOLT/SOLR | Line Type = Coaxial | Edit Cal Params | FULL FOUR PORT CAL SETUP (SOLT/R, Coaxial) Dialog Box | DUT Connector = V-Conn (M) | Standard Info | STANDARD INFO (SOLT/R, V-CONN(M))
l (SOLT/R) V-Conn (M)
STANDARD INFO (Broadband Cal) 0.8 mm Connector (M) Dialog Box
Figure: l Standard Info (Broadband Cal) Dialog shows a typical connector standard information dialog box for broadband cal. The contents depend on the calibration line type, and the calibration connectors and genders used. Figure: l USER DEFINED (Broadband Cal) Dialog for User-Defined Connector shows standard information for a user-defined connector dialog box.
Previous
FULL FOUR PORT CAL SETUP Broadband Cal (SOLT/R-SSST/R, Coaxial) Dialog Box.
THREE PORT CAL SETUP Broadband Cal (SOLT/R-SSST/R, Coaxial) Dialog Box
TWO PORT CAL SETUP Broadband Cal (SOLT/R-SSST/R, Coaxial) Dialog Box
ONE PORT CAL SETUP Broadband Cal (SOLT/R-SSST/R, Coaxial) Dialog Box
Navigation
MAIN | Calibration | CALIBRATION | Calibrate | CALIBRATE | Manual Cal | MANUAL CAL | 4-Port Cal | FOUR PORT CAL | Modify Cal Setup | CAL SETUP | Cal Method = Broadband Cal (SOLT/R-SSST/R) | Line Type = Coaxial | Edit Cal Params | FULL FOUR PORT CAL SETUP BROADBAND CAL, MERGED SOLT-SSST/SOLR-SSSR) Dialog Box | DUT Connector = 0.8 mm-Conn (M) | Standard Info | STANDARD INFO (Broadband)
 
l Standard Info (Broadband Cal) Dialog
STANDARD INFO (TRIPLE OFFSET SHORT) W1-Connector (M) Dialog Box
The figure below shows a typical connector standard information dialog box. The box format and contents vary depending on the calibration method, the calibration line type, and the calibration connectors and genders used. Anritsu-provided dialogs provide read-only information. User-defined dialogs allow user input of parameters and names.
Previous
This dialog box can be linked from multiple dialog sources. The links below are for dialogs that appear in this chapter.
Navigation
This dialog box can be linked to from a multiple dialog sources. The links below are for dialogs that appear in this chapter.
MAIN | Calibration | CALIBRATION | Calibrate | CALIBRATE | Manual Cal | MANUAL CAL | 4-Port Cal | FOUR PORT CAL | Modify Cal Setup | CAL SETUP | Cal Method = SSST | Line Type = Coaxial | Edit Cal Params | FULL FOUR PORT CAL SETUP (SSST, Coaxial) Dialog Box | DUT Connector = W1-Conn (M) | Standard Info | STANDARD INFO (TRIPLE OFFSET SHORT,W1-CONN(M))
STANDARD INFO (TRIPLE OFFSET SHORT) W1-Conn (M)
Representative Standard Information Dialog Box
THRU INFO Dialog Box – 4-Port VNAs
Use the THRU INFO dialog to update the thru information for most calibration types. The dialog includes access to the AIR EQUIVALENT LENGTH calculator function dialog box to speed configuration.
The dialog box name includes the port-pair being configured.
Previous
The THRU INFO dialog box can be accessed from multiple locations, including:
MODIFY 4-PORT AUTOCAL SETUP Dialog Box
MODIFY 2-PORT AUTOCAL SETUP Dialog Box – 4-Port VNAs
FULL FOUR PORT CAL SETUP (SOLT/R, Coaxial) Dialog Box
FULL FOUR PORT CAL SETUP (SSLT, Coaxial) Dialog Box
FULL FOUR PORT CAL SETUP (SSST, Coaxial) Dialog Box
FULL FOUR PORT CAL SETUP (SSST, Waveguide) Dialog Box
FULL FOUR PORT CAL SETUP Broadband Cal (SOLT/R-SSST/R, Coaxial) Dialog Box
FULL FOUR PORT CAL SETUP (mSSS, Coaxial) Dialog Box
THREE PORT CAL SETUP (SOLT/R, Coaxial) Dialog Box
THREE PORT CAL SETUP (SSLT, Microstrip) Dialog Box
THREE PORT CAL SETUP (SSST, Coaxial) Dialog Box
THREE PORT CAL SETUP Broadband Cal (SOLT/R-SSST/R, Coaxial) Dialog Box
THREE PORT CAL SETUP (mSSS, Coaxial) Dialog Box
TWO PORT CAL SETUP (SOLT/R, Coaxial) Dialog Box
TWO PORT CAL SETUP (SOLT/R, Microstrip) Dialog Box
TWO PORT CAL SETUP Broadband Cal (SOLT/R-SSST/R, Coaxial) Dialog Box
TWO PORT CAL SETUP (mSSS, Coaxial) Dialog Box
Navigation
MAIN | Calibration | CALIBRATION | Calibrate | CALIBRATE | AutoCal | AUTOCAL | Modify Cal Setup | MODIFY AUTOCAL SETUP Dialog Box | Thru Info | THRU INFO Dialog Box
MAIN | Calibration | CALIBRATION | Calibrate | CALIBRATE | Manual Cal | MANUAL CAL
| 4-Port CAL| Modify Cal Setup | CAL SETUP Menu | Cal Method | SOLT/SOLR |Edit Cal Params | Full Four PORT CAL SETUP (SOLT/R, Coaxial) | Thru x-y Info | THRU INFO Dialog Box
THRU INFO (Thru/Reciprocal Ports 1-2 Setup) Dialog Box (1 of 2)
Thru 2-3 Info Example:
1. Through selected allows user entries for length, line impedance, line loss, and frequency.
2. Reciprocal selected allows user entry for length.
3. S2P Thru selected provides buttons for loading, viewing, and characterization (to generate S2P files).
4. (Only available with K and V connector and when Cal Method is SOLT/R.) Use Generic Anritsu Adapter (checkbox) – Checking this box will automatically load a default .s2p file for a generic Anritsu adapter. Note that selecting the checkbox disables the Load S2P for Thru and Characterize Thru buttons.
Overview
The THRU INFO dialog box is available for all possible through combinations for all 4-port, 3-port, and 2-port calibrations. Each box is labeled with the port pair it represents.
Through Info
Through Selected
Length (mm)
Input line length in mm.
Calculator icon displays the AIR EQUIVALENT LENGTH CONVERSION dialog box.
AIR EQUIVALENT LENGTH Calculator Dialog Box
Line Impedance (Ohms)
Input defaults to be 50 Ohms. Any numeric value accepted.
Line Loss (dB/mm)
Allows input of a line loss in dB per mm at the frequency specified in the field below.
@ Frequency (GHz)
Allows input of a frequency setting for the Line Loss factor input above.
Reciprocal Selected
Length (mm)
Input line length in mm.
Calculator icon displays the AIR EQUIVALENT LENGTH CONVERSION dialog box.
AIR EQUIVALENT LENGTH Calculator Dialog Box
S2P Thru Selected
Load S2P for Thru
Opens navigation window to load S2P file.
View S2P File
Opens window to view contents of S2P file.
Characterize Thru
Allows a thru characterization and generation of an .s2p file.
Use Generic Anritsu Adapter (checkbox)
Checking this box will automatically load a default .s2p file for a generic Anritsu adapter.
Note that selecting the checkbox disables the Load S2P for Thru and Characterize Thru buttons.
USER DEFINED STANDARD (Broadband Cal) Dialog Box
Figure: l USER DEFINED (Broadband Cal) Dialog for User-Defined Connector shows a typical connector standard information dialog box for broadband cal with user-defined standard. This dialog allows user input of parameters and names or file names.
Previous
FULL FOUR PORT CAL SETUP Broadband Cal (SOLT/R-SSST/R, Coaxial) Dialog Box.
THREE PORT CAL SETUP Broadband Cal (SOLT/R-SSST/R, Coaxial) Dialog Box
TWO PORT CAL SETUP Broadband Cal (SOLT/R-SSST/R, Coaxial) Dialog Box
ONE PORT CAL SETUP Broadband Cal (SOLT/R-SSST/R, Coaxial) Dialog Box.
Navigation
MAIN | Calibration | CALIBRATION | Calibrate | CALIBRATE | Manual Cal | MANUAL CAL | 4-Port Cal | FOUR PORT CAL | Modify Cal Setup | CAL SETUP | Cal Method = Broadband Cal (SOLT/R-SSST/R) | Line Type = Coaxial | Edit Cal Params | FULL FOUR PORT CAL SETUP BROADBAND CAL, MERGED SOLT-SSST/SOLR-SSSR) Dialog Box | DUT Connector = User-Defined (M) | Standard Info | USER DEFINED STANDARD INFO (Broadband Cal)
 
l USER DEFINED (Broadband Cal) Dialog for User-Defined Connector
USER DEFINED MATCH DEVICES Dialog Box – TRL/TRM
Previous
This dialog box can be linked to from multiple dialog sources. The links below are for dialogs that appear in this chapter.
FULL FOUR PORT CAL SETUP (TRL/TRM, Coaxial) Dialog Box.
THREE PORT CAL SETUP (LRL/LRM, Coaxial) Dialog Box.
TWO PORT CAL SETUP (LRL/LRM, Coaxial) Dialog Box.
Navigation
This dialog box can be linked to from multiple dialog sources. The links below are for dialogs that appear in this chapter.
MAIN | Calibration | CALIBRATION | Calibrate | CALIBRATE | Manual Cal | MANUAL CAL | 4-Port Cal | FOUR PORT CAL | Modify Cal Setup | CAL SETUP | Cal Method = TRL/TRM | Line Type = Coaxial | Edit Cal Params | FULL FOUR PORT CAL SETUP (TRL/TRM, Coaxial) | Band X (Match) | Match Info | USER DEFINED MATCH Dialog Box
MAIN | Calibration | CALIBRATION | Calibrate | CALIBRATE | Manual Cal | MANUAL CAL | 3-Port Cal | THREE PORT CAL | Modify Cal Setup | CAL SETUP | Cal Method = TRL/TRM | Line Type = Coaxial | Edit Cal Params | THREE PORT CAL SETUP (TRL/TRM, Coaxial) | Band X (Match) | Match Info | USER DEFINED MATCH Dialog Box
MAIN | Calibration | CALIBRATION | Calibrate | CALIBRATE | Manual Cal | MANUAL CAL | 2-Port Cal | TWO PORT CAL | Modify Cal Setup | CAL SETUP | Cal Method = TRL/TRM | Line Type = Coaxial | Edit Cal Params | TWO PORT CAL SETUP (TRL/TRM, Coaxial) | Band X (Match) | Match Info | USER DEFINED MATCH Dialog Box
 
Example: USER DEFINED MATCH DEVICES Dialog Box – TRL/TRM
USER DEFINED MATCH DEVICES Dialog Box – LRL/LRM
Previous
This dialog box can be linked to from multiple dialog sources. The links below are for dialogs that appear in this chapter.
FULL FOUR PORT CAL SETUP (LRL/LRM, Coaxial) Dialog Box.
THREE PORT CAL SETUP (LRL/LRM, Coaxial) Dialog Box.
TWO PORT CAL SETUP (LRL/LRM, Coaxial) Dialog Box.
Navigation
This dialog box can be linked to from multiple dialog sources. The links below are for dialogs that appear in this chapter.
MAIN | Calibration | CALIBRATION | Calibrate | CALIBRATE | Manual Cal | MANUAL CAL | 4-Port Cal | FOUR PORT CAL | Modify Cal Setup | CAL SETUP | Cal Method = LRL/LRM | Line Type = Coaxial | Edit Cal Params | FULL FOUR PORT CAL SETUP (LRL/LRM, Coaxial) | Band 1, Device 2 (Match) | Match Info | USER DEFINED MATCH Dialog Box
MAIN | Calibration | CALIBRATION | Calibrate | CALIBRATE | Manual Cal | MANUAL CAL | 3-Port Cal | THREE PORT CAL | Modify Cal Setup | CAL SETUP | Cal Method = LRL/LRM | Line Type = Coaxial | Edit Cal Params | THREE PORT CAL SETUP (LRL/LRM, Coaxial) | Band 1, Device 2 (Match) | Match Info | USER DEFINED MATCH Dialog Box
MAIN | Calibration | CALIBRATION | Calibrate | CALIBRATE | Manual Cal | MANUAL CAL | 2-Port Cal | TWO PORT CAL | Modify Cal Setup | CAL SETUP | Cal Method = LRL/LRM | Line Type = Coaxial | Edit Cal Params | TWO PORT CAL SETUP (LRL/LRM, Coaxial) | Band 1, Device 2 (Match) | Match Info | USER DEFINED MATCH Dialog Box
Example: USER DEFINED MATCH DEVICES Dialog Box – LRL/LRM
USER DEFINED MATCH DEVICES Dialog Box – LRL/LRM Load/View S1P from File (1 of 2)
1. USER DEFINED MATCH DEVICES Dialog
2. S1PFILE CALKIT EXISTS Dialog
3. LOAD SxP FILE Dialog
4. VIEW FILE – DEVICE4MATCH Dialog
USER DEFINED MATCH DEVICES Dialog Box – mTRL
Previous
This dialog box can be linked to from multiple dialog sources. The links below are for dialogs that appear in this chapter.
FULL FOUR PORT CAL SETUP (mTRL, Coaxial) Dialog Box.
TWO PORT CAL SETUP (mTRL, Coaxial) Dialog Box
Navigation
This dialog box can be linked to from multiple dialog sources. The links below are for dialogs that appear in this chapter.
MAIN | Calibration | CALIBRATION | Calibrate | CALIBRATE | Manual Cal | MANUAL CAL | 4-Port Cal | FOUR PORT CAL | Modify Cal Setup | CAL SETUP | Cal Method = mTRL | Line Type = Coaxial | Edit Cal Params | FULL FOUR PORT CAL SETUP (mTRL, Coaxial) | Match Info | USER DEFINED MATCH Dialog Box
MAIN | Calibration | CALIBRATION | Calibrate | CALIBRATE | Manual Cal | MANUAL CAL | 2-Port Cal | TWO PORT CAL | Modify Cal Setup | CAL SETUP | Cal Method = mTRL | Line Type = Coaxial | Edit Cal Params | TWO PORT CAL SETUP (mTRL, Coaxial) | Match Info | USER DEFINED MATCH Dialog Box
Example: USER DEFINED MATCH DEVICES Dialog Box – mTRL
USER DEFINED MATCH DEVICES Dialog Box – mTRL Load/View S1P from File (1 of 2)
1. USER DEFINED MATCH DEVICES Dialog
2. S1PFILE CALKIT EXISTS Dialog
3. LOAD SxP FILE Dialog
4. VIEW FILE – DEVICE4MATCH Dialog
USER DEFINED MATCH Dialog Box – mSSS
Previous
This dialog box can be linked to from multiple dialog sources. The links below are for dialogs that appear in this chapter.
FULL FOUR PORT CAL SETUP (mSSS, Coaxial) Dialog Box.
THREE PORT CAL SETUP (mSSS, Coaxial) Dialog Box
TWO PORT CAL SETUP (mSSS, Coaxial) Dialog Box
ONE PORT CAL SETUP (mSSS, Coaxial) Dialog Box
Navigation
This dialog box can be linked to from multiple dialog sources. The links below are for dialogs that appear in this chapter.
MAIN | Calibration | CALIBRATION | Calibrate | CALIBRATE | Manual Cal | MANUAL CAL | 4-Port Cal | FOUR PORT CAL | Modify Cal Setup | CAL SETUP | Cal Method = mSSS | Line Type = Coaxial | Edit Cal Params | FULL FOUR PORT CAL SETUP (mSSS, Coaxial) | Match Info | USER DEFINED MATCH Dialog Box
MAIN | Calibration | CALIBRATION | Calibrate | CALIBRATE | Manual Cal | MANUAL CAL | 2-Port Cal | THREE PORT CAL | Modify Cal Setup | CAL SETUP | Cal Method = mSSS | Line Type = Coaxial | Edit Cal Params | THREE PORT CAL SETUP (mSSS, Coaxial) | Match Info | USER DEFINED MATCH Dialog Box
MAIN | Calibration | CALIBRATION | Calibrate | CALIBRATE | Manual Cal | MANUAL CAL | 2-Port Cal | TWO PORT CAL | Modify Cal Setup | CAL SETUP | Cal Method = mSSS | Line Type = Coaxial | Edit Cal Params | TWO PORT CAL SETUP (mSSS, Coaxial) | Match Info | USER DEFINED MATCH Dialog Box
MAIN | Calibration | CALIBRATION | Calibrate | CALIBRATE | Manual Cal | MANUAL CAL | 2-Port Cal | ONE PORT CAL | Modify Cal Setup | CAL SETUP | Cal Method = mSSS | Line Type = Coaxial | Edit Cal Params | ONE PORT CAL SETUP (mSSS, Coaxial) | Match Info | USER DEFINED MATCH Dialog Box
Example: USER DEFINED MATCH Dialog Box – mSSS
USER DEFINED MATCH DEVICES Dialog Box – mSSS Load/View S1P from File 
 
1. USER DEFINED MATCH Dialog
2. EDIT MATCH POLYNOMIAL TERMS Dialog
WAVEGUIDE INFO and USER DEFINED WAVEGUIDE Dialog Boxes
Dialog Box Name Changes
The exact title and content of these dialog boxes varies depending on the selected calibration method and waveguide kit.
Previous
This dialog box can be linked to from a multiple dialog sources. The links below are for dialogs that appear in this chapter.
FULL FOUR PORT CAL SETUP (SSST, Waveguide) Dialog Box
ONE PORT CAL SETUP (SSLT, Waveguide) Dialog Box
Navigation
This dialog box can be linked to from a multiple dialog sources. The links below are for dialogs that appear in this chapter.
MAIN | Calibration | CALIBRATION | Calibrate | CALIBRATE | Manual Cal | MANUAL CAL | 4-Port Cal | FOUR PORT CAL | Modify Cal Setup | CAL SETUP | Cal Method = SSST | Line Type = Waveguide | Edit Cal Params | FULL FOUR PORT CAL SETUP (SSST, Waveguide) Dialog Box
MAIN | Calibration | CALIBRATION | Calibrate | CALIBRATE | Manual Cal | MANUAL CAL | 1-Port Cal | ONE PORT CAL | Modify Cal Setup | CAL SETUP | Cal Method = SSLT | Line Type = Waveguide | Edit Cal Params | FULL ONE PORT CAL (SSLT, Waveguide) | Waveguide Info Button | WAVEGUIDE INFO (SSLT) Dialog Box
WAVEGUIDE INFO and USER DEFINED WAVEGUIDE Dialog Boxes
1. At left, USER DEFINED WAVEGUIDE Input Dialog Box.
2. At right WAVEGUIDE INFO for WR10 Connector.
For mSSS, the Cutoff Frequency and Dielectric are defined globally in the FULL [FOUR, THREE, TWO, ONE] PORT CAL SETUP (mSSS, Waveguide) Dialog Box.