S820E Microwave Site Master™ User Guide : Calibration, VNA
 
Calibration, VNA
 
Introduction
Calibration Considerations
Calibration Data and Indications
Calibration Setup
Cal Type
Cal Line and Cal Method
Port 1 or Port 2 DUT and Cal Kit
Thru Device
Calibration Procedures
Calibration Procedure
Save and Recall Calibration Coefficients
Calibration Menu
Calibration Dialog Box
Start Calibration Menu
Introduction
This chapter provides details and procedures for calibrating the Vector Network Analyzer modes of the Site Master S820E.
Note 
For accurate results, the instrument must be calibrated before making any measurements.
The Site Master is a high‑precision instrument. When making 1‑port or 2‑port measurements, the instrument must be calibrated in order to remove residual errors due to measurement setup conditions. Anritsu recommends performing the calibration under the same conditions as the measurement: temperature, frequency, number of points, source power, and IFBW. The calibration must be conducted using the appropriate standards at the open end of any test port cables and adapters that are connected to Port 1 or Port 2 of the instrument. This will ensure that the match, phase length, and loss of these cables and adapters are all accounted for. For optimal performance, high quality phase‑stable cables and precision adapters must be used.
Note 
Screen images in this User Guide are illustrations of typical instrument features. Some images may include instruments other than the Site Master S820E. Traces and other display features may differ from the screen displays of your instrument.